C25 80PK-26 80PK-25 FVF-SC1 TPAK
50 Series II Thermometers
Laboratory accuracy. Wherever you go.
Additional features for the 53 and
54 Series I I:
• Data Logging up to 500 points of data
with user adjustable recording interval
• Real time clock captures the exact time of
day when events occur
• Recall function allows logged data to be
easily reviewed on the meter display
• IR communication port allows data to be
exported to optional FlukeView®
Temperature PC software
Features
Specifications
Recommended Accessories
Battery life: 1000 hours typical, AA
Size (HxWxD): 17 3 x 86 x 38 mm Weight: 0.4 kg
Three Year Warranty
11080-eng Rev. 01
Included Accessories
Impact absorbing holster
Two bead probe thermocouples 80PK-1
(54+52)
One bead probe thermocouple 80PK-1
(51+53)
Ordering Information
Fluke 51 II Thermometer
Fluke 52 II Thermometer
Fluke 53 II Thermometer
Fluke 54 II Thermometer
FVF-SC1 FlukeView Forms
Software including
interface cable
●● ● ●
●● ● ●
●● ● ●
●●
●●
●●
●●
51 I I 52 II 53 II 54 II
Thermocouple types J,K,T,E J,K,T,E J,K,T,E,N,R,S J,K,T,E,N,R,S
Number of inputs Single Dual Single Dual
Time stamp Relative Time Relative Time Time of Day Time of Day
Splash/Dust resistant
Dual display with backlight
Min/Max/Avg recording
(T1-T2) True differential
Data logging up to 500 points
IR data port for interface
to PC
Compatible with optional
FlukeView Software
Temperature range:
J-type Thermocouples -210°C to 1200°C (-346°F to 2192°F)
K-type Thermocouples -200°C to 1372°C (-328°F to 2501°F)
T-type Thermocouples -250°C to 400°C (-418°F to 752°F)
E-type Thermocouples -150°C to 1000°C (-238°F to 1832°F)
N-type** Thermocouples -200°C to 1300°C (-328°F to 2372°F)
R** and S-type** Thermocouples 0°C to 1767°C (32°F to 3212°F)
Temperature accuracy
Above –100°C (-148°F):
J, K, T, E, and N-type** ± [ 0.05% + 0.3°C (0.5°F) ]
R** and S-type** ± [ 0.05% + 0.4°C (0.7°F) ]
Below –100°C (-148° F):
J, K, E, and N-types ± [ 0.20% + 0.3°C (0.5°F) ]
T-type ± [ 0.50% + 0.3°C (0.5°F) ]
**Only the Fluke Models 53 and 54 Series II thermometers are capable of measuring N, R, and S-type thermocouples.
The Fluke 50 Series II contact
thermometers offer fast response and
laboratory accuracy (0.05% + 0.3°C) in
a rugged handheld test tool.
• Large backlit dual display shows any
combination of T1, T2(52 and 54 only),
T1-T2(52 and 54 only) plus MIN, MAX,
or AVG
• Relative time clock on MIN, MAX, and
AVG provides a time reference for major
events
• Electronic Offset function allows
compensation of thermocouple errors to
maximize overall accuracy
• Readout in °C, °F, or Kelvin (K)
• Sleep mode increases battery life
• Battery door allows easy battery
replacement without breaking the
calibration seal
Fluke 51 II Fluke 52 II Fluke 53 II