Mixed Signal Oscilloscopes
MSO4000B, DPO4000B Series Datasheet
With the MSO/DPO4000B Mixed Signal Oscilloscope Series, you can
analyze up to 20 analog and digital signals with a single instrument to
quickly find and diagnose problems in complex designs. Bandwidths up to
1 GHz and up to 5X oversampling on all channels ensure you have the
performance you need to see fast-changing signal details. To capture long
windows of signal activity while maintaining fine timing resolution, the MSO/
DPO4000B Series offers deep record length of up to 20M points standard
on all channels. And with Wave Inspector® controls for rapid waveform
navigation, automated serial and parallel bus analysis, limit and mask
testing, and automated power analysis – your Tektronix oscilloscope
provides the feature-rich tools you need to simplify and speed debug of
your complex design.
Key performance specifications
1 GHz, 500 MHz, 350 MHz, and 100 MHz bandwidth models
2 and 4 analog channel models
Up to 5 GS/s sample rate on all channels
Up to 20 M record length on all channels
>340,000 wfm/s maximum waveform capture rate
Standard passive voltage probes with 3.9 pF capacitive loading and
500 MHz or 1 GHz analog bandwidth
Suite of advanced triggers
Key features
Wave Inspector® controls provide easy navigation and automated
search of waveform data
42 automated measurements, and FFT analysis for simplified
waveform analysis
16 digital channels (MSO series)
Mixed signal design and analysis (MSO series)
Automated triggering, decode, and search on parallel buses
Per-channel threshold settings
Multichannel setup and hold triggering
MagniVu High-speed acquisition provides 60.6 ps fine timing
resolution on digital channels
Optional serial triggering and analysis - automated serial triggering,
decode, and search options for I2C, SPI, USB, Ethernet, CAN, LIN,
FlexRay, RS-232/422/485/UART, MIL-STD-1553, and I2S/LJ/RJ/TDM
TekVPI® probe interface supports active, differential, and current
probes for automatic scaling and units
10.4 in. (264 mm) bright XGA color display
Small footprint and lightweight – Only 5.8 in. (147 mm) deep and 11 lb.
(5 kg)
Connectivity
Two USB 2.0 host ports on the front panel and two on the rear panel
for quick and easy data storage and connecting a USB keyboard
USB 2.0 device port on the rear panel for easy connection to a PC or
direct printing to a PictBridge®-compatible printer
Integrated 10/100/1000BASE-T Ethernet port for network connection
and video out port to export the oscilloscope display to a monitor or
projector
Optional application support
Power analysis
Limit and mask testing
HDTV and custom video analysis
www.tektronix.com 1
Need more performance?
Looking for an MSO at 100 MHz, 350 MHz, 500 MHz or 1 GHz?
Need an integrated spectrum analyzer?
Need to analyze analog, digital, and RF simultaneously?
Consider the MDO4000B Series oscilloscopes, the world's first Mixed Domain
Oscilloscope. www.tektronix.com/MDO4000
▪ Includes a 16 channel MSO
▪ 3 GHz and 6 GHz integrated spectrum analyzers
▪ Simultaneous, time-correlated acquisition of analog, digital and RF
▪ 20 M record length
Comprehensive features speed every stage
of debug
These oscilloscopes offer a robust set of features to speed every stage of
debugging your design – from quickly discovering an anomaly and
capturing it, to searching your waveform record for the event and analyzing
its characteristics and your device’s behavior.
Discover
To debug a design problem, first you must know it exists. Every design
engineer spends time looking for problems in their design, a time-
consuming and frustrating task without the right debug tools.
The industry’s most complete visualization of signals provides fast insight
into the real operation of your device. A fast waveform capture rate –
greater than 340,000 waveforms per second with FastAcq – enables you
to see glitches and other infrequent transients within seconds, revealing the
true nature of device faults. A digital phosphor display with color- and
intensity-grading shows the history of a signal’s activity by using color or
intensity in areas of the signal that occur more frequently, providing a visual
display of just how often anomalies occur.
Discover Digital phosphor technology with FastAcq enables greater than
340,000 wfms/s waveform capture rate and real-time color-intensity grading.
Capture
Discovering a device fault is only the first step. Next, you must capture the
event of interest to identify root cause.
Accurately capturing any signal of interest begins with proper probing. Low-
capacitance probes are included with the oscilloscope, one for each analog
channel. These industry-first high-impedance passive voltage probes have
less than 4 pF of capacitive loading to minimize the effect of the probe on
your circuit's operation, offering the performance of an active probe with the
flexibility of a passive probe.
A complete set of triggers - including runt, timeout, logic, pulse width/glitch,
setup/hold violation, serial packet, and parallel data - help you quickly find
your event. With up to a 20M point record length, you can capture many
events of interest, even thousands of serial packets, in a single acquisition
for further analysis while maintaining high resolution to zoom in on fine
signal details.
From triggering on specific packet content to automatic decode in multiple
data formats, the oscilloscope provides integrated support for the industry's
broadest range of serial buses - I2C, SPI, USB, Ethernet, CAN, LIN,
FlexRay, RS-232/422/485/UART, MIL-STD-1553, and I2S/LJ/RJ/TDM. The
ability to decode up to four serial and/or parallel buses simultaneously
means you gain insight into system-level problems quickly.
To further help troubleshoot system-level interactions in complex
embedded systems, the oscilloscope offers 16 digital channels in addition
to its analog channels. Since the digital channels are fully integrated into
the oscilloscope, you can trigger across all input channels, automatically
time correlating all analog, digital, and serial signals. The MagniVu high-
speed acquisition on these channels enables you to acquire fine signal
detail (up to 60.6 ps resolution) around the trigger point for precision timing
measurements. MagniVu is essential for making accurate timing
measurements for setup and hold, clock delay, signal skew, and glitch
characterization.
Datasheet
2 www.tektronix.com
Capture - Triggering on a specific transmit data packet going across an RS-232 bus. A
complete set of triggers, including triggers for specific serial packet content, ensures you
quickly capture your event of interest.
Search
Finding your event of interest in a long waveform record can be time
consuming without the right search tools. With today’s record lengths
pushing beyond a million data points, locating your event can mean
scrolling through thousands of screens of signal activity.
The innovative Wave Inspector® controls give you the industry’s most
comprehensive search and waveform navigation capability. These controls
speed panning and zooming through your record. With a unique force-
feedback system, you can move from one end of your record to the other in
just seconds. User marks allow you to mark any location that you may want
to reference later for further investigation. Or, automatically search your
record for criteria you define. Wave Inspector will instantly search your
entire record, including analog, digital, and serial bus data. Along the way it
will automatically mark every occurrence of your defined event so you can
quickly move between events.
Search I2C decode showing results from a Wave Inspector search for Address value
50. Wave Inspector controls provide unprecedented efficiency in viewing and navigating
waveform data.
Analyze
Verifying that your prototype’s performance matches simulations and meets
the project’s design goals requires analyzing its behavior. Tasks can range
from simple checks of rise times and pulse widths to sophisticated power
loss analysis and investigation of noise sources.
The oscilloscope offers a comprehensive set of integrated analysis tools
including waveform- and screen-based cursors, automated measurements,
advanced waveform math including arbitrary equation editing, FFT
analysis, and trend plots for visually determining how a measurement is
changing over time. Specialized application support for serial bus analysis,
power supply design, and video design and development is also available.
Analyze – Waveform histogram of a falling edge showing the distribution of edge position
(jitter) over time. Included are numeric measurements made on the waveform histogram
data. A comprehensive set of integrated analysis tools speeds verification of your
design's performance.
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 3
Wave Inspector® navigation and search
With long record lengths, a single acquisition can include thousands of
screens of waveform data. Wave Inspector®, the industry’s best tool for
navigation and search, enables you to find events of interest in seconds.
Wave Inspector controls provide unprecedented efficiency in viewing, navigating, and
analyzing waveform data. Zip through your long record by turning the outer pan control
(1). Get details from the beginning to end in seconds. See something of interest and want
to see more details? Just turn the inner zoom control (2).
Zoom and pan
A dedicated, two-tier front-panel control provides intuitive control of both
zooming and panning. The inner control adjusts the zoom factor (or zoom
scale); turning it clockwise activates zoom and goes to progressively higher
zoom factors, while turning it counterclockwise results in lower zoom
factors and eventually turning zoom off. No longer do you need to navigate
through multiple menus to adjust your zoom view. The outer control pans
the zoom box across the waveform to quickly get to the portion of waveform
you are interested in. The outer control also utilizes force-feedback to
determine how fast to pan on the waveform. The farther you turn the outer
control, the faster the zoom box moves. Pan direction is changed by simply
turning the control the other way.
Play/Pause
A dedicated Play/Pause front-panel button scrolls the waveform across the
display automatically while you look for anomalies or an event of interest.
Playback speed and direction are controlled using the intuitive pan control.
Once again, turning the control further makes the waveform scroll faster
and changing direction is as simple as turning the control the other way.
User marks
Press the Set Mark front-panel button to place one or more marks on the
waveform. Navigating between marks is as simple as pressing the
Previous (←) and Next (→) buttons on the front panel.
Search marks
The Search button allows you to automatically search through your long
acquisition looking for user-defined events. All occurrences of the event are
highlighted with search marks and are easily navigated to, using the front-
panel Previous (←) and Next (→) buttons. Search types include edge,
pulse width/glitch, timeout, runt, logic, setup and hold, rise/fall time, parallel
bus, and I2C, SPI, USB, Ethernet, CAN, LIN, FlexRay, RS-232/422/485/
UART, MIL-STD-1553, and I2S/LJ/RJ/TDM packet content. A search mark
table provides a tabular view of the events found during the automated
search. Each event is shown with a time stamp, making timing
measurements between events easy.
Search step 1: You define what you would like to find.
Search step 2: Wave Inspector automatically searches through the record and marks
each event with a hollow white triangle. You can then use the Previous and Next buttons
to jump from one event to the next.
Datasheet
4 www.tektronix.com
Search step 3: The Search Mark table provides a tabular view of each of the events
found by the automated search. Each event is shown with a time stamp making timing
measurements between events easy.
Digital phosphor technology with FastAcq
Digital phosphor technology with FastAcq provides you with fast insight into
the real operation of your device. Its fast waveform capture rate – greater
than 340,000 wfms/s – gives you a high probability of quickly seeing the
infrequent problems common in digital systems: runt pulses, glitches, timing
issues, and more.
To further enhance the visibility of rarely occurring events, intensity grading
is used to indicate how often rare transients are occurring relative to normal
signal characteristics. There are four waveform palettes available in
FastAcq acquisition mode.
The Temperature palette uses color-grading to indicate frequency of
occurrence with hot colors like red/yellow indicating frequently
occurring events and colder colors like blue/green indicating rarely
occurring events.
The Spectral palette uses color-grading to indicate frequency of
occurrence with colder colors like blue indicating frequently occurring
events and hot colors like red indicating rarely occurring events.
The Normal palette uses the default channel color (like yellow for
channel one) along with gray-scale to indicate frequency of occurrence
where frequently occurring events are bright.
The Inverted palette uses the default channel color along with gray-
scale to indicate frequency of occurrence where rarely occurring events
are bright.
These color palettes quickly highlight the events that over time occur more
often or, in the case of infrequent anomalies, occur less often.
Infinite or variable persistence choices determine how long waveforms stay
on the display, helping you to determine how often an anomaly is occurring.
Digital phosphor technology enables a greater than 340,000 wfm/s waveform capture
rate and real-time intensity grading.
Accurate high-speed probing
The TPP Series probes, included standard with every MSO/DPO4000B
Series oscilloscope, provide up to 1 GHz of analog bandwidth, and 3.9 pF
of capacitive loading. The extremely low capacitive loading minimizes
adverse effects on your circuits and is more forgiving of longer ground
leads. And, since the probe bandwidth matches or exceeds your
oscilloscope bandwidth, you can see the high-frequency components in
your signal which is critical for high-speed applications. The TPP Series
passive voltage probes offer all the benefits of general-purpose probes like
high dynamic range, flexible connection options, and robust mechanical
design, while providing the performance of active probes. In addition, a low-
attenuation, 2X version of the TPP probes is available for measuring low
voltages. Unlike other low-attenuation passive probes, the TPP0502 has
high bandwidth (500 MHz) as well as low capacitive loading (12.7 pF).
Mixed signal design and analysis (MSO
Series)
The MSO models provide 16 digital channels which are tightly integrated
into the oscilloscope's user interface. This simplifies operation and makes it
possible to solve mixed-signal issues easily.
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 5
The MSO Series provide 16 integrated digital channels enabling you to view and analyze
time-correlated analog and digital signals.
Color-coded digital waveform display
This oscilloscope has redefined the way you view digital waveforms. One
common problem shared by both logic analyzers and mixed-signal
oscilloscopes is determining if data is a one or a zero when zoomed in far
enough that the digital trace stays flat all the way across the display. Color-
coded digital traces display ones in green and zeros in blue.
The multiple transition detection hardware shows you a white edge on the
display when the system detects multiple transitions. White edges indicate
that more information is available by zooming in or acquiring at faster
sampling rates. In most cases zooming in will reveal the pulse that was not
viewable with the previous settings. If the white edge is still present after
zooming in as far as possible, this indicates that increasing the sample rate
on the next acquisition will reveal higher frequency information than the
previous settings could acquire.
You can group digital waveforms and enter waveform labels by using a
USB keyboard. By simply placing digital waveforms next to each other, they
form a group.
With color-coded digital waveform display, groups are created by simply placing digital
channels together on the screen, allowing digital channels to be moved as a group. You
can set threshold values for each channel, enabling support for up to 16 different logic
families.
Once a group is formed, you can position all the channels contained in that
group collectively. This greatly reduces the normal setup time associated
with positioning channels individually.
MagniVu® high-speed acquisition
The main digital acquisition mode on the MSO4000B Series will capture up
to 20M points at 500 MS/s (2 ns resolution). In addition to the main record,
the oscilloscope provides an ultra high-resolution record called MagniVu
which acquires 10,000 points at up to 16.5 GS/s (60.6 ps resolution). Both
main and MagniVu waveforms are acquired on every trigger and can be
switched between in the display at any time, running or stopped. MagniVu
provides significantly finer timing resolution than comparable MSOs on the
market, instilling confidence when making critical timing measurements on
digital waveforms.
The MagniVu high-resolution record provides 60.6 ps timing resolution, enabling you to
make critical timing measurements on your digital waveforms.
Datasheet
6 www.tektronix.com
P6616 MSO probe
This unique probe design offers two eight-channel pods. Each channel
ends with a probe tip featuring a recessed ground for simplified connection
to the device under test. The coax on the first channel of each pod is
colored blue making it easy to identify. The common ground uses an
automotive-style connector making it easy to create custom grounds for
connecting to the device under test. When connecting to square pins, the
P6616 has an adapter that attaches to the probe head extending the probe
ground flush with the probe tip so you can attach to a header. The P6616
offers outstanding electrical characteristics, having only 3 pF of capacitive
loading, a 100 kΩ input resistance, and is capable of acquiring toggle rates
>500 MHz and pulses as short as 1 ns in duration.
The P6616 MSO probe offers two eight-channel pods to simplify connecting to your
device.
Serial triggering and analysis (optional)
On a serial bus, a single signal often includes address, control, data, and
clock information. This can make isolating events of interest difficult.
Automatic trigger, decode, and search on bus events and conditions gives
you a robust set of tools for debugging serial buses. The optional serial
protocol triggering and analysis functionality is offered free for a 30-day trial
period. This free trial period starts automatically when the instrument is
powered on for the first time.
Triggering on a specific OUT Token packet on a USB full-speed serial bus. The yellow
waveform is the D+ and the blue waveform is the D-. A bus waveform provides decoded
packet content including Start, Sync, PID, Address, End Point, CRC, Data values, and
Stop.
Serial triggering
Trigger on packet content such as start of packet, specific addresses,
specific data content, unique identifiers, etc. on popular serial interfaces
such as I2C, SPI, USB, Ethernet, CAN, LIN, FlexRay, RS-232/422/485/
UART, MIL-STD-1553, and I2S/LJ/RJ/TDM.
Bus display
Provides a higher-level, combined view of the individual signals (clock,
data, chip enable, etc.) that make up your bus, making it easy to identify
where packets begin and end and identifying sub-packet components such
as address, data, identifier, CRC, etc.
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 7
Bus decoding
Tired of having to visually inspect the waveform to count clocks, determine
if each bit is a 1 or a 0, combine bits into bytes, and determine the hex
value? Let the oscilloscope do it for you! Once you've set up a bus, the
MSO/DPO4000B Series will decode each packet on the bus, and display
the value in hex, binary, decimal (USB, Ethernet, MIL-STD-1553, LIN, and
FlexRay only), signed decimal (I2S/LJ/RJ/TDM only), or ASCII (USB,
Ethernet, and RS-232/422/485/UART only) in the bus waveform.
Event table
In addition to seeing decoded packet data on the bus waveform itself, you
can view all captured packets in a tabular view much like you would see in
a software listing. Packets are time stamped and listed consecutively with
columns for each component (Address, Data, etc.). You can save the event
table data in .csv format.
Event table showing decoded identifier, DLC, DATA, and CRC for every CAN packet in a
long acquisition.
Technology Trigger, Decode,
Search
Order product
Embedded I2C Yes DPO4EMBD
SPI Yes DPO4EMBD
Computer RS232/422/485,
UART
Yes DPO4COMP
USB USB LS, FS, HS Yes DPO4USB
Ethernet 10Base-T,
100Base-TX
Yes DPO4ENET
Automotive CAN Yes DPO4AUTO or
DPO4AUTOMAX
LIN Yes DPO4AUTO or
DPO4AUTOMAX
FlexRay Yes DPO4AUTOMAX
Military and
Aerospace
MIL-STD-1553 Yes DPO4AERO
Audio I2S Yes DPO4AUDIO
LJ, RJ Yes DPO4AUDIO
TDM Yes DPO4AUDIO
Search (serial triggering)
Serial triggering is very useful for isolating the event of interest, but once
you’ve captured it and need to analyze the surrounding data, what do you
do? In the past, users had to manually scroll through the waveform
counting and converting bits and looking for what caused the event. You
can have the oscilloscope automatically search through the acquired data
for user-defined criteria including serial packet content. Each occurrence is
highlighted by a search mark. Rapid navigation between marks is as simple
as pressing the Previous (←) and Next (→) buttons on the front panel.
Power analysis (optional)
Ever increasing consumer demand for longer battery-life devices and for
green solutions that consume less power require power-supply designers to
characterize and minimize switching losses to improve efficiency. In
addition, the supply’s power levels, output purity, and harmonic feedback
into the power line must be characterized to comply with national and
regional power quality standards. Historically, making these and many
other power measurements on an oscilloscope has been a long, manual,
and tedious process. The optional power analysis tools greatly simplify
these tasks, enabling quick and accurate analysis of power quality,
switching loss, harmonics, safe operating area (SOA), modulation, ripple,
and slew rate (di/dt, dv/dt). Completely integrated into the oscilloscope, the
power analysis tools provide automated, repeatable power measurements
with a touch of a button; no external PC or complex software setup is
required. The optional power analysis functionality is offered free for a 30-
day trial period. This free trial period starts automatically when the
instrument is powered on for the first time.
Datasheet
8 www.tektronix.com
Safe operating area measurement. Automated power measurements enable quick and
accurate analysis of common power parameters.
Limit/Mask testing (optional)
A common task during the development process is characterizing the
behavior of certain signals in a system. One method, called limit testing, is
to compare a tested signal to a known good or "golden" version of the
same signal with user-defined vertical and horizontal tolerances. Another
common method, called mask testing, is to compare a tested signal to a
mask, looking for where a signal under test violates the mask. The MSO/
DPO4000B Series offers both limit and mask testing capability useful for
long-term signal monitoring, characterizing signals during design, or testing
on a production line. A robust set of telecommunications and computer
standards are provided to test for compliance to a standard. Additionally,
custom masks can be created and used for characterizing signals. Tailor a
test to your specific requirements by defining test duration in number of
waveforms or time, a violation threshold that must be met before
considering a test a failure, counting hits along with statistical information,
and actions upon violations, test failure, and test complete. Whether
specifying a mask from a known good signal or from a custom or standard
mask, conducting pass/fail tests in search of waveform anomalies such as
glitches has never been easier. The optional limit/mask test functionality is
offered free for a 30-day trial period. This free trial period starts
automatically when the instrument is powered on for the first time.
Limit Test showing a mask created from a golden waveform and compared against a live
signal. Results showing statistical information about the test are displayed.
Video design and development
Many video engineers have remained loyal to analog oscilloscopes,
believing the intensity gradations on an analog display are the only way to
see certain video waveform details. The fast waveform capture rate,
coupled with its intensity-graded view of the signal, provides the same
information-rich display as an analog oscilloscope, but with much more
detail and all the benefits of digital scopes.
Standard features such as IRE and mV graticules, holdoff by fields, video
polarity, and an Autoset smart enough to detect video signals, make these
the easiest to use oscilloscopes on the market for video applications. And
with high bandwidth and four analog inputs, the oscilloscope provides
ample performance for analog and digital video use.
The video functionality is further extended with an optional video application
module, which provides the industry's most complete suite of HDTV and
custom (nonstandard) video triggers, as well as a video picture mode
enabling you to see the picture of the video signal you are viewing - for
NTSC and PAL signals. The optional video analysis functionality is offered
free for a 30- day trial period. This free trial period starts automatically when
the instrument is powered on for the first time.
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 9
Viewing an NTSC video image. Video picture mode contains automatic contrast and
brightness settings as well as manual controls.
Designed to make your work easier
The MSO/DPO4000B Series is designed to make your work easier. The large, high-
resolution display shows intricate signal details. Dedicated front-panel controls simplify
operation. Two USB host ports on the front panel allow you to easily transfer screen
shots, instrument settings, and waveform data to a USB mass storage device.
Large, high-resolution display
The MSO/DPO4000B Series features a 10.4 in. (264 mm) bright, LED
backlit XGA color display for seeing intricate signal details.
Connectivity
The MSO/DPO4000B contains a number of ports which can be used to
connect the instrument to a network, directly to a PC, or other test
equipment.
Two USB host ports on the front and two USB host ports on the rear
enable easy transfer of screen shots, instrument settings, and
waveform data to a USB mass storage device. A USB keyboard can
also be attached to a USB host port for data entry.
Rear USB device port is useful for controlling the oscilloscope remotely
from a PC or for printing directly to a PictBridge®-compatible printer.
The standard 10/100/1000BASE-T Ethernet port on the rear of the
instrument enables easy connection to networks, provides network and
e-mail printing, and provides LXI Class-C compatibility. The instrument
can also mount network drives for easy storage of screen images,
setup files, or data files.
A video out port on the rear of the instrument allows the display to be
exported to an external monitor or projector.
Compact form factor
With the compact, portable form factor, you can easily move the
oscilloscope between labs. And with a depth of just 5.8 inches (147 mm), it
saves you valuable space on your test bench.
TekVPI® probe interface
The TekVPI probe interface sets the standard for ease of use in probing. In
addition to the secure, reliable connection that the interface provides,
TekVPI probes feature status indicators and controls, as well as a probe
menu button right on the comp box itself. This button brings up a probe
menu on the oscilloscope display with all relevant settings and controls for
the probe. The TekVPI interface enables direct attachment of current
probes without requiring a separate power supply. TekVPI probes can be
controlled remotely through USB, GPIB, or LAN, enabling more versatile
solutions in ATE environments. The instrument provides up to 50 W of
power to the front panel TekVPI connectors from the internal power supply,
giving you the confidence that it can meet any demanding probing
requirements you have.
Datasheet
10 www.tektronix.com
TekVPI probe interface simplifies connecting your probes to the oscilloscope.
Remote connectivity and instrument control
Exporting data and measurements is as simple as connecting a USB cable
from the oscilloscope to your PC. Key software applications –
OpenChoice® Desktop, and Microsoft Excel and Word toolbars – are
included standard with each oscilloscope to enable fast and easy direct
communication with your Windows PC.
The included OpenChoice Desktop enables fast and easy communication
between the oscilloscope and your PC through USB or LAN for transferring
settings, waveforms, and screen images.
OpenChoice Desktop software enables seamless connection between the oscilloscope
and your PC.
The embedded e*Scope® capability enables fast control of the oscilloscope
over a network connection through a standard web browser. Simply enter
the IP address or network name of the oscilloscope and a web page will be
served to the browser. Transfer and save settings, waveforms,
measurements, and screen images or make live control changes to settings
on the oscilloscope directly from the web browser.
The MSO/DPO4000B Series can also be connected to your network using
the LAN port. The included LXI web interface provides information about
the current configuration of your MSO/DPO4000B Series oscilloscope,
including network configuration. You can make changes to the network
configuration, control instrument settings, save screen images and
instrument data, and save/load instrument setups of your oscilloscope
directly from the web interface through a password-protected web page.
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 11
Specifications
All specifications apply to all models unless noted otherwise.
Model overview
DPO4014B 1DPO4034B 1 DPO4054B 1 DPO4102B-L,
MSO4102B-L
DPO4102B,
MSO4102B
DPO4104B-L,
MSO4104B-L
DPO4104B,
MSO4104B 1
Analog channels 4 4 4 2 2 4 4
Bandwidth 100 MHz 350 MHz 500 MHz 1 GHz 1 GHz 1 GHz 1 GHz
Rise time 3.5 ns 1 ns 700 ps 350 ps 350 ps 350 ps 350 ps
Sample rate (1 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s 5 GS/s 5 GS/s 5 GS/s
Sample rate (2 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s 5 GS/s 5 GS/s
Sample rate (4 ch) 2.5 GS/s 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s
Record length (1 ch) 20M 20M 20M 5M 20M 5M 20M
Record length (2 ch) 20M 20M 20M 5M 20M 5M 20M
Record length (4 ch) 20M 20M 20M 5M 20M
Duration at highest sample rate 8 ms 8 ms 8 ms 1 ms 4 ms 1 ms 4 ms
Digital channels MSO models add 16 digital channels to the corresponding DPO model
Vertical system analog channels
Hardware bandwidth limits
≥350 MHz models 20 MHz or 250 MHz
100 MHz models 20 MHz
Input coupling AC, DC
Input impedance 1 MΩ ±1%, 50 Ω ±1%
Input sensitivity range
1 MΩ 1 mV/div to 10 V/div
50 Ω 1 mV/div to 1 V/div
Vertical resolution 8 bits (11 bits with Hi Res)
Maximum input voltage
1 MΩ 300 VRMS CAT II with peaks ≤ ±425 V
50 Ω 5 VRMS with peaks ≤ ±20 V (DF ≤ 6.25%)
DC gain accuracy ±1.5%, derated at 0.10%/°C above 30 °C
Channel-to-channel isolation Any two channels at equal vertical scale ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth
1Looking for the MSO4000B Series at 100 MHz, 350 MHz, 500 MHz or 1 GHz? You will find their replacements in the MDO4000B Series. These oscilloscopes have a built-in spectrum analyzer at no additional
cost. Learn more about these products today, please visit www.tektronix.com/mdo4000.
Datasheet
12 www.tektronix.com
Offset range Volts/div setting Offset range
1 MΩ input 50 Ω
1 mV/div to 50 mV/div ±1 V ±1 V
50.5 mV/div to 99.5 mV/div ±0.5 V ±0.5 V
100 mV/div to 500 mV/div ±10 V ±10 V
505 mV/div to 995 mV/div ±5 V ±5 V
1 V/div to 5 V/div ±100 V ±5 V
5.05 V/div to 10 V/div ±50 V NA
Vertical system digital channels
Input channels 16 digital (D15 to D0)
Thresholds Per-channel thresholds
Threshold selections TTL, CMOS, ECL, PECL, User-defined
User-defined threshold range ±40 V
Threshold accuracy ±[100 mV + 3% of threshold setting]
Maximum input voltage ±42 Vpeak
Input dynamic range 30 Vp-p ≤200 MHz
10 Vp-p >200 MHz
Minimum voltage swing 400 mV
Probe loading 100 kΩ in parallel with 3 pF
Vertical resolution 1 bit
Horizontal system analog channels
Time base range
1 GHz models 400 ps to 1000 s
≤ 500 MHz models 1 ns to 1000 s
Time-base delay time range -10 divisions to 5000 s
Channel-to-channel deskew range ±125 ns
Time base accuracy ±5 ppm over any ≥1 ms interval
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
Vertical system analog channels
www.tektronix.com 13
Horizontal system digital channels
Maximum sample rate (Main) 500 MS/s (2 ns resolution)
Maximum record length (Main) 20M points (5M points on -L models)
Maximum sample rate (MagniVu) 16.5 GS/s (60.6 ps resolution)
Maximum record length (MagniVu) 10k points centered around the trigger
Minimum detectable pulse width
(typical)
1 ns
Channel-to-channel skew (typical) 200 ps
Maximum input toggle rate 500 MHz (Maximum frequency sine wave that can accurately be reproduced as a logic square wave. Requires the use of a short
ground extender on each channel. This is the maximum frequency at the minimum swing amplitude. Higher toggle rates can be
achieved with higher amplitudes.)
Trigger system
Trigger modes Auto, Normal, and Single
Trigger coupling DC, AC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity)
Trigger holdoff range 20 ns to 8 s
Trigger sensitivity
Internal DC coupled Trigger source Sensitivity
1 MΩ path (all models) For 1 mV/div to 4.98 mV/div; 0.75 div from DC to 50 MHz,
increasing to 1.3 div at rated bandwidth
50 Ω path (≤500 MHz models) For ≥5 mV/div; 0.4 div from DC to 50 MHz, increasing to 1 div
at rated bandwidth
50 Ω path (1 GHz models) 0.4 div from DC to 50 MHz, increasing to 1 div at rated
bandwidth
External Trigger source Sensitivity
Auxiliary Input 200 mV from DC to 50 MHz, increasing to 500 mV at rated
bandwidth
Trigger level ranges
Any input channel ±8 divisions from center of screen, ±8 divisions from 0 V when vertical LF reject trigger coupling is selected
Aux Input (external trigger) ±8 V
Line The line trigger level is fixed at about 50% of the line voltage.
Trigger frequency readout Provides 6-digit frequency readout of triggerable events.
Trigger types
Edge Positive, negative, or either slope on any channel or front-panel auxiliary input. Coupling includes DC, AC, HF reject, LF reject, and
noise reject.
Sequence (B-trigger) Trigger Delay by Time: 4 ns to 8 s. Or Trigger Delay by Events: 1 to 4,000,000 events.
Pulse Width Trigger on width of positive or negative pulses that are >, <, =, ≠, or inside/outside a specified period of time.
Timeout Trigger on an event which remains high, low, or either, for a specified time period (4 ns to 8 s).
Runt Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again.
Datasheet
14 www.tektronix.com
Logic Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can be used as a clock
to look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for all input channels defined as High, Low, or
Don’t Care.
Setup and Hold Trigger on violations of both setup time and hold time between clock and data present on any of the analog and digital input
channels.
Rise/Fall Time Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative, or either.
Video Trigger on all lines, odd, even, or all fields on NTSC, PAL, and SECAM video signals.
Extended Video (optional) Trigger on 480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60, 1080i/50, 1080i/60, 1080p/24, 1080p/24sF, 1080p/25, 1080p/
30, 1080p/50, 1080p/60, and custom bi-level and tri-level sync video standards.
I2C (optional) Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address and Data on I2C buses up to 10 Mb/s.
SPI (optional) Trigger on SS active, Start of Frame, MOSI, MISO, or MOSI and MISO on SPI buses up to 50.0 Mb/s.
RS-232/422/485/UART
(optional)
Trigger on Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, Rx Data, Tx Parity Error, and Rx Parity Error up
to 10 Mb/s.
USB: Low speed (optional) Trigger on Sync Active, Start of Frame, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet,
Handshake Packet, Special Packet, Error.
Token packet trigger - Any token type, SOF, OUT, IN, SETUP; Address can be specified for Any Token, OUT, IN, and SETUP
token types. Address can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular value, or inside or outside of a range. Frame
number can be specified for SOF token using binary, hex, unsigned decimal and don't care digits.
Data packet trigger - Any data type, DATA0, DATA1; Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular data
value, or inside or outside of a range.
Handshake packet trigger - Any handshake type, ACK, NAK, STALL.
Special packet trigger - Any special type, Reserved
Error trigger - PID Check, CRC5 or CRC16, Bit Stuffing.
USB: Full speed (optional) Trigger on Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet, Handshake Packet, Special
Packet, Error.
Token packet trigger - Any token type, SOF, OUT, IN, SETUP; Address can be specified for Any Token, OUT, IN, and SETUP
token types. Address can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular value, or inside or outside of a range. Frame
number can be specified for SOF token using binary, hex, unsigned decimal and don't care digits.
Data packet trigger - Any data type, DATA0, DATA1; Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular data
value, or inside or outside of a range.
Handshake packet trigger - Any handshake type, ACK, NAK, STALL.
Special packet trigger - Any special type, PRE, Reserved.
Error trigger - PID Check, CRC5 or CRC16, Bit Stuffing.
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
Trigger system
www.tektronix.com 15
USB: High speed (optional) 2 Trigger on Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet, Handshake Packet, Special
Packet, Error.
Token packet trigger - Any token type, SOF, OUT, IN, SETUP; Address can be specified for Any Token, OUT, IN, and SETUP
token types. Address can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular value, or inside or outside of a range. Frame
number can be specified for SOF token using binary, hex, unsigned decimal and don't care digits.
Data packet trigger - Any data type, DATA0, DATA1, DATA2, MDATA; Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a
particular data value, or inside or outside of a range.
Handshake packet trigger - Any handshake type, ACK, NAK, STALL, NYET.
Special packet trigger - Any special type, ERR, SPLIT, PING, Reserved. SPLIT packet components that can be specified include:
▪ Hub Address
▪ Start/Complete - Don't Care, Start (SSPLIT), Complete (CSPLIT)
▪ Port Address
▪ Start and End bits - Don't Care, Control/Bulk/Interrupt (Full-speed Device, Low-speed Device), Isochronous (Data is Middle, Data
is End, Data is Start, Data is All)
▪ Endpoint Type - Don't Care, Control, Isochronous, Bulk, Interrupt
Error trigger - PID Check, CRC5 or CRC16.
Ethernet (optional) 3 10BASE-T and 100BASE-TX: Trigger on Start Frame Delimiter, MAC Addresses, MAC Q-Tag Control Information, MAC Length/
Type, IP Header, TCP Header, TCP/IPv4/MAC Client Data, End of Packet, and FCS (CRC) Error.
100BASE-TX: Idle.
MAC Addresses - Trigger on Source and Destination 48-bit address values.
MAC Q-Tag Control Information - Trigger on Q-Tag 32-bit value.
MAC Length/Type - Trigger on ≤, <, =, >, ≥, ≠ a particular 16-bit value, or inside or outside of a range.
IP Header - Trigger on IP Protocol 8-bit value, Source Address, Destination Address.
TCP Header - Trigger on Source Port, Destination Port, Sequence Number, and Ack Number.
TCP/IPv4/MAC Client Data - Trigger on ≤, <, =, >, ≥, ≠ a particular data value, or inside or outside of a range. Selectable number
of bytes to trigger on from 1-16. Byte offset options of Don't Care, 0-1499.
CAN (optional) Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data, Identifier and Data,
End of Frame, Missing ACK, or Bit Stuffing Error on CAN signals up to 1 Mb/s. Data can be further specified to trigger on ≤, <, =,
>, ≥, or ≠ a specific data value. User-adjustable sample point is set to 50% by default.
LIN (optional) Trigger on Sync, Identifier, Data, Identifier and Data, Wakeup Frame, Sleep Frame, Errors such as Sync, Parity, or Checksum
Errors up to 100 kb/s (by LIN definition, 20 kb/s).
FlexRay (optional) Trigger on Start of Frame, Type of Frame (Normal, Payload, Null, Sync, Startup), Identifier, Cycle Count, Complete Header Field,
Data, Identifier and Data, End of Frame or Errors such as Header CRC, Trailer CRC, Null Frame, Sync Frame, or Startup Frame
Errors up to 100 Mb/s.
MIL-STD-1553 (optional) Trigger on Sync, Word Type 4 (Command, Status, Data), Command Word (set RT Address, T/R, Sub-address/Mode, Data Word
Count/Mode Code, and Parity individually), Status Word (set RT Address, Message Error, Instrumentation, Service Request Bit,
Broadcast Command Received, Busy, Subsystem Flag, Dynamic Bus Control Acceptance (DBCA), Terminal Flag, and Parity
individually), Data Word (user-specified 16-bit data value), Error (Sync, Parity, Manchester, Non-contiguous data), Idle Time
(minimum time selectable from 2 µs to 100 µs; maximum time selectable from 2 µs to 100 µs; trigger on < minimum, > maximum,
inside range, outside range). RT Address can be further specified to trigger on =, ≠, <, >, ≤, ≥ a particular value, or inside or
outside of a range.
2High-speed support only available on models with 1 GHz analog channel bandwidth.
3≥350 MHz bandwidth models are recommended for 100BASE-TX
4Trigger selection of Command Word will trigger on Command and ambiguous Command/Status words. Trigger selection of Status Word will trigger on Status and ambiguous Command/Status words.
Datasheet
Trigger system
16 www.tektronix.com
I2S/LJ/RJ/TDM (optional) Trigger on Word Select, Frame Sync, or Data. Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a specific data value, or
inside or outside of a range. Maximum data rate for I2S/LJ/RJ is 12.5 Mb/s. Maximum data rate for TDM is 25 Mb/s.
Parallel (available on MSO
models only)
Trigger on a parallel bus data value. Parallel bus can be from 1 to 16 bits (from the digital channels) plus 2 or 4 bits (from the
analog channels) in size. Binary and Hex radices are supported.
Acquisition system
Acquisition Modes
Sample Acquire sampled values.
Peak Detect Captures glitches as narrow as 800 ps (1 GHz models) or 1.6 ns (≤500 MHz models) at all sweep speeds
Averaging From 2 to 512 waveforms included in average.
Envelope Min-max envelope reflecting Peak Detect data over multiple acquisitions.
Hi Res Real-time boxcar averaging reduces random noise and increases vertical resolution.
Roll Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div.
FastAcqFastAcq optimizes the instrument for analysis of dynamic signals and capture of infrequent events, capturing >340,000 wfms/s on
1 GHz models and >270,000 wfms/s on 100 MHz - 500 MHz models.
Waveform measurements
Cursors Waveform and Screen.
Automatic measurements
(time domain)
30, of which up to eight can be displayed on-screen at any one time. Measurements include: Period, Frequency, Delay, Rise Time,
Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase, Positive
Overshoot, Negative Overshoot, Total Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle
RMS, Positive Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Area and Cycle Area.
Measurement statistics Mean, Min, Max, Standard Deviation.
Reference levels User-definable reference levels for automatic measurements can be specified in either percent or units.
Gating Isolate the specific occurrence within an acquisition to take measurements on, using either the screen, or waveform cursors.
Waveform histogram A waveform histogram provides an array of data values representing the total number of hits inside of a user-defined region of the
display. A waveform histogram is both a visual graph of the hit distribution as well as a numeric array of values that can be
measured.
Sources - Channel 1, Channel 2, Channel 3, Channel 4, Ref 1, Ref 2, Ref 3, Ref 4, Math
Types - Vertical, Horizontal
Waveform histogram
measurements
Waveform Count, Hits in Box, Peak Hits, Median, Max, Min, Peak-to-Peak, Mean, Standard Deviation, Sigma 1, Sigma 2, Sigma 3
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
Trigger system
www.tektronix.com 17
Waveform math
Arithmetic Add, subtract, multiply, and divide waveforms.
Math functions Integrate, Differentiate, FFT.
FFT Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or
Blackman-Harris.
Advanced math Define extensive algebraic expressions including waveforms, reference waveforms, math functions (FFT, Intg, Diff, Log, Exp, Sqrt,
Abs, Sine, Cosine, Tangent, Rad, Deg), scalars, up to two user-adjustable variables and results of parametric measurements
(Period, Freq, Delay, Rise, Fall, PosWidth, NegWidth, BurstWidth, Phase, PosDutyCycle, NegDutyCycle, PosOverShoot,
NegOverShoot, PeakPeak, Amplitude, RMS, CycleRMS, High, Low, Max, Min, Mean, CycleMean, Area, CycleArea, and trend
plots), e.g.,(Intg(Ch1 - Mean(Ch1)) × 1.414 × VAR1).
Act on Event
Events None, when a trigger occurs, or when a defined number of acquisitions complete (1 to 1,000,000)
Actions Stop acquisition, save waveform to file, save screen image, print, AUX OUT Pulse, remote interface SRQ, e-mail notification, and
visual notification.
Repeat Repeat the act on event process (1 to 1,000,000 and infinity)
Video Picture Mode (optional)
Sources Channel 1, Channel 2, Channel 3, Channel 4
Video standards NTSC, PAL
Contrast and brightness Manual and automatic
Field selection Odd, Even, Interlaced
Picture location on screen Selectable X and Y location, width and height adjustment, start line and pixel and line-to-line offset control.
Power measurements (optional)
Power Quality Measurements VRMS, VCrest Factor, Frequency, IRMS, ICrest Factor, True Power, Apparent Power, Reactive Power, Power Factor, Phase Angle.
Switching loss measurements
Power loss Ton, Toff, Conduction, Total.
Energy loss Ton, Toff, Conduction, Total.
Harmonics THD-F, THD-R, RMS measurements. Graphical and table displays of harmonics. Test to IEC61000-3-2 Class A and MIL-
STD-1399, Section 300A.
Ripple measurements VRipple and IRipple.
Modulation Analysis Graphical display of +Pulse Width, –Pulse Width, Period, Frequency, +Duty Cycle, and –Duty Cycle modulation types.
Safe operating area Graphical display and mask testing of switching device safe operating area measurements.
dV/dt and dI/dt measurements Cursor measurements of slew rate.
Datasheet
18 www.tektronix.com
Limit/Mask testing (optional)
Included standard masks 5 ITU-T, ANSI T1.102, USB
Test source Limit test: Any Ch1 - Ch4 or any R1 - R4
Mask test: Any Ch1 - Ch4
Mask creation Limit test vertical tolerance from 0 to 1 division in 1 m division increments; Limit test horizontal tolerance from 0 to 500 m division in
1 m division increments
Load standard mask from internal memory
Load custom mask from text file with up to 8 segments
Mask scaling Lock to Source ON (mask automatically re-scales with source-channel settings changes)
Lock to Source OFF (mask does not re-scale with source-channel settings changes)
Test criteria run until Minimum number of waveforms (from 1 to 1,000,000; Infinity)
Minimum elapsed time (from 1 second to 48 hours; Infinity)
Violation threshold From 1 to 1,000,000
Actions on test failure Stop acquisition, save screen image to file, save waveform to file, print screen image, trigger out pulse, set remote interface SRQ
Actions on test complete Trigger out pulse, set remote interface SRQ
Results display Test status, total waveforms, number of violations, violation rate, total tests, failed tests, test failure rate, elapsed time, total hits for
each mask segment
Software
OpenChoice® Desktop Enables fast and easy communication between a Windows PC and your oscilloscope using USB or LAN. Transfer and save
settings, waveforms, measurements, and screen images. Included Word and Excel toolbars automate the transfer of acquisition
data and screen images from the oscilloscope into Word and Excel for quick reporting or further analysis.
IVI driver Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI,
Microsoft .NET, and MATLAB.
e*Scope® Web-based remote
control
Enables control of the oscilloscope over a network connection through a standard web browser. Simply enter the IP address or
network name of the oscilloscope and a web page will be served to the browser. Transfer and save settings, waveforms,
measurements, and screen images or make live control changes to settings on the oscilloscope directly from the web browser.
LXI Class C Web interface Connect to the oscilloscope through a standard Web browser by simply entering the oscilloscope's IP address or network name in
the address bar of the browser. The Web interface enables viewing of instrument status and configuration, status and modification
of network settings, and instrument control through the e*Scope Web-based remote control. All Web interaction conforms to LXI
Class C specification, version 1.3.
5≥350 MHz bandwidth models are recommended for mask testing on telecomm standards >55 Mb/s. 1 GHz bandwidth models are recommended for mask testing on high-speed (HS) USB.
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 19
Display system
Display type 10.4 in. (264 mm) liquid-crystal TFT color display
Display resolution 1,024 horizontal × 768 vertical pixels (XGA)
Interpolation Sin(x)/x
Waveform styles Vectors, Dots, Variable Persistence, Infinite Persistence.
Graticules Full, Grid, Cross Hair, Frame, IRE and mV.
Format YT and simultaneous XY/YT
Maximum waveform capture rate >340,000 wfms/s in FastAcq acquisition mode on 1 GHz models
>270,000 wfms/s in FastAcq acquisition mode on 100 MHz - 500 MHz models
>50,000 wfms/s in DPO acquisition mode on all models.
Input/output ports
USB 2.0 high-speed host port Supports USB mass storage devices and keyboard. Two ports on front and two ports on rear of instrument.
USB 2.0 device port Rear-panel connector allows for communication/control of oscilloscope through USBTMC or GPIB (with a TEK-USB-488), and
direct printing to all PictBridge-compatible printers.
Printing Print to network printer, PictBridge printer, or to a printer that supports e-mail printing. Note: This product includes software
developed by the OpenSSL Project for use in the OpenSSL Toolkit. (http://www.openssl.org/)
LAN port RJ-45 connector, supports 10/100/1000 Mb/s
Video out port DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector. XGA resolution.
Auxiliary input Front-panel BNC connector. Input Impedance 1 MΩ. Max input 300 VRMS CAT II with peaks ≤ ±425 V.
Probe compensator output voltage
and frequency
Front-panel pins
Amplitude 0 to 2.5 V
Frequency 1 kHz
Auxiliary out Rear-panel BNC connector
VOUT (Hi): ≥2.5 V open circuit, ≥1.0 V 50 Ω to ground
VOUT (Lo): ≤0.7 V into a load of ≤4 mA; ≤0.25 V 50 Ω to ground
Output can be configured to provide a pulse out signal when the oscilloscope triggers, the internal oscilloscope reference clock
out, or an event out for limit/mask testing.
External reference input Time-base system can phase lock to an external 10 MHz reference (10 MHz ±1%)
Kensington-style lock Rear-panel security slot connects to standard Kensington-style lock.
VESA mount Standard (MIS-D 100) 100 mm VESA mounting points on rear of instrument.
Datasheet
20 www.tektronix.com
LAN eXtensions for Instrumentation (LXI)
Class LXI Class C
Version V1.3
Power source
Power source voltage 100 to 240 V ±10%
Power source frequency 50 to 60 Hz ±10% at 100 to 240 V ±10%
400 Hz ±10% at 115 V ±13%
Power consumption 225 W maximum
Physical characteristics
Dimensions mm in.
Height 229 9.0
Width 439 17.3
Depth 147 5.8
Weight kg lb.
Net 5 11
Shipping 10.7 23.6
Rackmount configuration 5U
Cooling clearance 2 in. (51 mm) required on left side and rear of instrument
EMC, environment, and safety
Temperature
Operating 0 ºC to +50 ºC (+32 ºF to 122 ºF)
Nonoperating -20 ºC to +60 ºC (-4 ºF to 140 ºF)
Humidity
Operating High: 40 ºC to 50 ºC, 10% to 60% relative humidity, Low: 0 ºC to 40 ºC, 10% to 90% relative humidity
Nonoperating High: 40 ºC to 60 ºC, 5%to 60% relative humidity, Low: 0 ºC to 40 ºC, 5% to 90% relative humidity
Altitude
Operating 3,000 meters (9,843 feet)
Nonoperating 9,144 meters (30,000 feet)
Regulatory
Electromagnetic compatibility EC Council Directive 2004/108/EC
Safety UL61010-1:2004, CAN/CSA-C22.2 No. 61010.1: 2004, Low Voltage Directive 2006/95/EC and EN61010-1:2001,
IEC 61010-1:2001, ANSI 61010-1-2004, ISA 82.02.01
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 21
Ordering information
MSO/DPO4000B family 6
DPO4014B 100 MHz, 2.5/2.5/2.5 GS/s on 1/2/4 channels, 20M record length, 4-channel digital phosphor oscilloscope
DPO4034B 350 MHz, 2.5/2.5/2.5 GS/s on 1/2/4 channels, 20M record length, 4-channel digital phosphor oscilloscope
DPO4054B 500 MHz, 2.5/2.5/2.5 GS/s on 1/2/4 channels, 20M record length, 4-channel digital phosphor oscilloscope
DPO4102B-L 1 GHz, 5/2.5 GS/s on 1/2 channels, 5M record length, 2-channel digital phosphor oscilloscope
DPO4102B 1 GHz, 5/5 GS/s on 1/2 channels, 20M record length, 2-channel digital phosphor oscilloscope
DPO4104B-L 1 GHz, 5/5/2.5 GS/s on 1/2/4 channels, 5M record length, 4-channel digital phosphor oscilloscope
DPO4104B 1 GHz, 5/5/5 GS/s on 1/2/4 channels, 20M record length, 4-channel digital phosphor oscilloscope
MSO4102B-L 1 GHz, 5/2.5 GS/s on 1/2 channels, 5M record length, 2+16 channel mixed signal oscilloscope
MSO4102B 1 GHz, 5/5 GS/s on 1/2 channels, 20M record length, 2+16 channel mixed signal oscilloscope
MSO4104B-L 1 GHz, 5/5/2.5 GS/s on 1/2/4 channels, 5M record length, 4+16 channel mixed signal oscilloscope
MSO4104B 1 GHz, 5/5/5 GS/s on 1/2/4 channels, 20M record length, 4+16 channel mixed signal oscilloscope
Standard accessories
Probes
≤500 MHz models TPP0500B, 500 MHz bandwidth, 10X, 3.9 pF. One passive voltage probe per analog channel.
1 GHz models TPP1000, 1 GHz bandwidth, 10X, 3.9 pF. One passive voltage probe per analog channel.
MSO models also include One P6616 16-channel logic probe and a logic probe accessory kit (020-2662-xx).
Accessories
200-5130-xx Front cover
063-4300-xx Documentation CD
016-2030-xx Accessory bag
User manual
Power cord
OpenChoice® Desktop Software
NI LabVIEW SignalExpress Tektronix Edition Software
Calibration certificate documenting traceability to National Metrology Institute(s) and ISO9001 quality system registration
Warranty
Three-year warranty covering all parts and labor, excluding probes.
6Looking for the MSO4000B Series at 100 MHz, 350 MHz, 500 MHz or 1 GHz? You will find their replacements in the MDO4000B Series. These oscilloscopes have a built-in spectrum analyzer at no additional
cost. Learn more about these products today, please visit www.tektronix.com/mdo4000.
Datasheet
22 www.tektronix.com
Application Modules
Application modules have licenses which can be transferred between an application module and an oscilloscope. The license may be contained in the module; allowing the
module to be moved from one instrument to another. Or, the license can be contained in the oscilloscope; allowing the module to be removed and stored for safekeeping.
Transferring the license to an oscilloscope and removing the module permits the use of more than 4 applications simultaneously.
DPO4AERO Aerospace Serial Triggering and Analysis Module. Enables triggering on packet-level information on MIL-STD-1553 buses as well
as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with
time-stamp information.
Signal Inputs - Any Ch1 - Ch4, Math, Ref1 - Ref4
Recommended Probing - Differential or single ended (only one single-ended signal required)
DPO4AUDIO Audio Serial Triggering and Analysis Module. Enables triggering on packet-level information on I2S, LJ, RJ, and TDM audio buses
as well as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables
with time-stamp information. 7
Signal Inputs - Any Ch1 - Ch4 (and any D0 - D15 on MSO models)
Recommended Probing - Single ended
DPO4AUTO Automotive Serial Triggering and Analysis Module. Enables triggering on packet-level information on CAN and LIN buses as well
as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with
time-stamp information.
Signal Inputs - LIN: Any Ch1 - Ch4 (and any D0 - D15 on MSO models); CAN: Any Ch1 - Ch4 (and any D0 - D15 on MSO models)
Recommended Probing - LIN: Single ended; CAN: Single ended or differential
DPO4AUTOMAX Extended Automotive Serial Triggering and Analysis Module. Enables triggering on packet-level information on CAN, LIN, and
FlexRay buses as well as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, packet
decode tables with time-stamp information, and eye diagram analysis software.
Signal Inputs - LIN: Any Ch1 - Ch4 (and any D0 - D15 on MSO models); CAN: Any Ch1 - Ch4 (and any D0 - D15 on MSO
models); FlexRay: Any Ch1 - Ch4 (and any D0 - D15 on MSO models)
Recommended Probing - LIN: Single ended; CAN, FlexRay: Single ended or differential
DPO4COMP Computer Serial Triggering and Analysis Module. Enables triggering on packet-level information on RS-232/422/485/UART buses
as well as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables
with time-stamp information.
Signal Inputs - Any Ch1 - Ch4 (and any D0 - D15 on MSO models)
Recommended Probing - RS-232/UART: Single ended; RS-422/485: Differential
DPO4EMBD Embedded Serial Triggering and Analysis Module. Enables triggering on packet-level information on I2C and SPI buses as well as
analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with time-
stamp information. 8
Signal Inputs - I2C: Any Ch1 - Ch4 (and any D0 - D15 on MSO models); SPI: Any Ch1 - Ch4 (and any D0 - D15 on MSO models)
Recommended Probing - Single ended
DPO4ENET Ethernet Serial Triggering and Analysis Module. Enables triggering on packet-level information on 10BASE-T and 100BASE-TX 9
buses as well as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode
tables with time-stamp information.
Signal Inputs - Any Ch1 - Ch4, Math, Ref1 - Ref4
Recommended Probing - 10BASE-T: Single ended or differential; 100BASE-TX: Differential
7Not available on DPO4102B or DPO4102B-L models.
8Only 2-wire SPI is supported on DPO4102B and DPO4102B-L models.
9≥350 MHz bandwidth models are recommended for 100BASE-TX
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 23
DPO4USB USB Serial Triggering and Analysis Module. Enables triggering on packet-level content for low-speed, full-speed, and high-speed
USB serial buses. Also enables analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and
packet decode tables with time-stamp information for low-speed, full-speed, and high-speed USB serial buses. 10
Signal Inputs - Low-speed and Full-speed: Any Ch1 - Ch4 (and any D0 - D15 on MSO models); Low-speed, Full-speed, and High-
speed: Any Ch1 - Ch4, Math, Ref1 - Ref4
Recommended Probing - Low-speed and Full-speed: Single ended or differential; High-speed: Differential
DPO4PWR Power Analysis Application Module. Enables quick and accurate analysis of power quality, switching loss, harmonics, safe
operating area (SOA), modulation, ripple, and slew rate (dI/dt, dV/dt).
DPO4LMT Limit and Mask Testing Application Module. Enables testing against limit templates generated from "golden" waveforms and mask
testing using custom or standard telecommunications or computer masks. 11
DPO4VID HDTV and Custom (nonstandard) Video Triggering and Video Picture Module.
Instrument options
Power cord and plug options
Opt. A0 North America power plug (115 V, 60 Hz)
Opt. A1 Universal Euro power plug (220 V, 50 Hz)
Opt. A2 United Kingdom power plug (240 V, 50 Hz)
Opt. A3 Australia power plug (240 V, 50 Hz)
Opt. A5 Switzerland power plug (220 V, 50 Hz)
Opt. A6 Japan power plug (100 V, 50/60 Hz)
Opt. A10 China power plug (50 Hz)
Opt. A11 India power plug (50 Hz)
Opt. A12 Brazil power plug (60 Hz)
Opt. A99 No power cord
Language options
Opt. L0 English manual
Opt. L1 French manual
Opt. L2 Italian manual
Opt. L3 German manual
Opt. L4 Spanish manual
Opt. L5 Japanese manual
Opt. L6 Portuguese manual
Opt. L7 Simplified Chinese manual
Opt. L8 Traditional Chinese manual
Opt. L9 Korean manual
10 USB high-speed supported only on models with 1 GHz analog channel bandwidth.
11 ≥350 MHz bandwidth models are recommended for mask testing on telecomm standards >55 Mb/s. 1 GHz bandwidth models are recommended for mask testing on high-speed (HS) USB.
Datasheet
24 www.tektronix.com
Opt. L10 Russian manual
Opt. L99 No manual
Language options include translated front-panel overlay for the selected language(s).
Service options
Opt. C3 Calibration Service 3 Years
Opt. C5 Calibration Service 5 Years
Opt. D1 Calibration Data Report
Opt. D3 Calibration Data Report 3 Years (with Opt. C3)
Opt. D5 Calibration Data Report 5 Years (with Opt. C5)
Opt. R5 Repair Service 5 Years (including warranty)
Opt. SILV600 Standard warranty extended to 5 years
Probes and accessories are not covered by the oscilloscope warranty and service offerings. Refer to the datasheet of each probe and accessory model for its unique warranty
and calibration terms.
Recommended accessories
Probes
Tektronix offers over 100 different probes to meet your application needs. For a comprehensive listing of available probes, please visit www.tektronix.com/probes.
TPP0500B 500 MHz, 10X TekVPI® passive voltage probe with 3.9 pF input capacitance
TPP0502 500 MHz, 2X TekVPI® passive voltage probe with 12.7 pF input capacitance
TPP0850 2.5 kV, 800 MHz, 50X TekVPI® passive high-voltage probe
TPP1000 1 GHz, 10X TekVPI® passive voltage probe with 3.9 pF input capacitance
TAP1500 1.5 GHz TekVPI® active single-ended voltage probe
TCP0030 120 MHz TekVPI® 30 Ampere AC/DC current probe
TCP0150 20 MHz TekVPI® 150 Ampere AC/DC current probe
TDP0500 500 MHz TekVPI® differential voltage probe with ±42 V differential input voltage
TDP1000 1 GHz TekVPI® differential voltage probe with ±42 V differential input voltage
TDP1500 1.5 GHz TekVPI® differential voltage probe with ±8.5 V differential input voltage
THDP0200 ±1.5 kV, 200 MHz TekVPI® high-voltage differential probe
THDP0100 ±6 kV, 100 MHz TekVPI® high-voltage differential probe
TMDP0200 ±750 V, 200 MHz TekVPI® high-voltage differential probe
P5100A 2.5 kV, 500 MHz, 100X high-voltage passive probe
P5200A 1.3 kV, 50 MHz high-voltage differential probe
Accessories
077-0512-xx Service manual (English only)
TPA-BNC TekVPI® to TekProbe BNC adapter
TEK-DPG TekVPI Deskew pulse generator signal source
Mixed Signal Oscilloscopes — MSO4000B, DPO4000B Series
www.tektronix.com 25
067-1686-xx Power measurement deskew and calibration fixture
TEK-USB-488 GPIB-to-USB adapter
ACD4000B Soft transit case
HCTEK54 Hard transit case (requires ACD4000B)
RMD5000 Rackmount kit
Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar.
Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats.
Datasheet
ASEAN / Australasia (65) 6356 3900 Austria 00800 2255 4835* Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Belgium 00800 2255 4835* Brazil +55 (11) 3759 7627 Canada 1 800 833 9200
Central East Europe and the Baltics +41 52 675 3777 Central Europe & Greece +41 52 675 3777 Denmark +45 80 88 1401
Finland +41 52 675 3777 France 00800 2255 4835* Germany 00800 2255 4835*
Hong Kong 400 820 5835 India 000 800 650 1835 Italy 00800 2255 4835*
Japan 81 (3) 6714 3010 Luxembourg +41 52 675 3777 Mexico, Central/South America & Caribbean 52 (55) 56 04 50 90
Middle East, Asia, and North Africa +41 52 675 3777 The Netherlands 00800 2255 4835* Norway 800 16098
People's Republic of China 400 820 5835 Poland +41 52 675 3777 Portugal 80 08 12370
Republic of Korea 001 800 8255 2835 Russia & CIS +7 (495) 6647564 South Africa +41 52 675 3777
Spain 00800 2255 4835* Sweden 00800 2255 4835* Switzerland 00800 2255 4835*
Taiwan 886 (2) 2722 9622 United Kingdom & Ireland 00800 2255 4835* USA 1 800 833 9200
* European toll-free number. If not accessible, call: +41 52 675 3777 Updated 10 April 2013
For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com.
Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and
price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies.
30 Jul 2014 3GW-20156-22
www.tektronix.com