cui.com
date 06/20/2017 │ page 2 of 5CUI Inc │ SERIES: ETMA 200W U │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
INPUT
parameter conditions/description min typ max units
voltage 90 264 Vac
frequency 47 63 Hz
current at 90 Vac, 60 Hz 3 A
inrush current at 230 Vac, cold start 150 A
no load power consumption 0.5 W
OUTPUT
parameter conditions/description min typ max units
line regulation ±1 %
load regulation ±5 %
hold-up time at 115 Vac 16 ms
temperature coefcient ±0.05 %/°C
PROTECTIONS
parameter conditions/description
over voltage protection
12 V output
16, 20 V outputs
19 V output
24, 36 V outputs
48 V output
110
110
111
110
110
133
130
137
125
120
%
%
%
%
%
overload protection shutdown and auto restart 110 160 %
short circuit protection continuous, auto restart
SAFETY & COMPLIANCE
parameter conditions/description min typ max units
isolation voltage input to output for 2 seconds
input to frame ground for 2 seconds
5,656
2,121
Vdc
Vdc
insulation resistance input to output, at 500 Vdc
input to frame ground, at 500 Vdc
20
20
MΩ
MΩ
safety approvals medical UL/cUL (UL 60601-1), EN 60601-1, IEC 60601-1 4th edition
EMI/EMC FCC 47 CFR Part 18, CE, EN 61000-3-(2, 3), EN 55011, IEC 61000-4-(2, 3, 4, 5, 6, 8, 11)
leakage current earth, at 264 Vac
enclosure, at 264 Vac
190
10
300
100
µA
µA
MTBF at 25°C, max. load 140,000 hours
RoHS compliant yes
ENVIRONMENTAL
parameter conditions/description min typ max units
operating temperature 0 50 °C
storage temperature -10 70 °C
humidity non-condensing 20 90 %