Change #1, 51120
On page 27, add the following section under the Entity Parameters:
Pressure Module Output Circuit:
Vo, Uo Io Po Co Lo
IIC IIB IIA IIC IIB IIA
7.14 V 102 mA 182 mW 13.5 μF 240 μF 1000 μF 3.4 mH 10.2 mH 27.2 mH
Change #2, 52975
On page 9, under Certification Information, change the following:
From: P ( II 1 G EEx ia IIC T4
0344
To: P ( II 1 G Ex ia IIC T4
0344
On page 27, under Product Compliance Markings, change the following:
From: P ( II 1 G EEx ia IIC T4
0344 KEMA 04ATEX1601 X
To: P ( II 1 G Ex ia IIC T4
0344 KEMA 04ATEX1601 X
718Ex 30G/100G Users Manual Supplement
2 5/11
Change #3, 54529
On page 26, under General Specifications add the following:
Operating altitude: 3000 meters maximum
On page 27, under Product Compliance Markings, replace the fourth line with the following:
Manufactured by Martel Electronics, Inc.
3 Corporate Park Drive
Derry, NH USA
Change #4, 58415
On page 25, add the following to the end of the Pressure Sensor Input Table:
Where in a 3V/m radiated EM field 350 MHz, Pressure Accuracy is 1 % of range.
On page 26, add the following to the end of the DC mA Input Table:
Where in a 3V/m radiated EM field 350 MHz, add 0.1 % to mA input accuracy.
Manual Supplement 718Ex 30G/100G Users
1/12 3
Change #5, 60644
On page 25, replace the Pressure Sensor Input Table and add the Note:
Pressure Sensor Input
Model Range Accuracy
Max Non-
destructive
Pressure
30G -12 to 30 psi
(-83 to 207 kPa)
±0.025 %
of range
(6 month)
±0.035 %
of range
(1 year)
60 psi
(413 kPa)
100G -12 to 100 psi
(-83 to 690 kPa)
200 psi
(1.4 mPa)
300G -12 to 300 psi
(-83 to 2068 kPa)
±0.05 %
of range
(1 year)
375 psi
(2.6 mPa)
Temperature coefficient: 0.01 % of range per
°
C for
temperature ranges -10
°
C to 18
°
C and 28
°
C to 55
°
C
Note
Specifications apply to the 718Ex Pressure Calibrators with firmware
version 2.0 or greater. To verify the firmware version, press and hold
and