22
Platinum Resistance Temperature Transmit ter
• Proven 100 Ω platinum resistance sensor provides reliability, stability and
unbeatable performance
• Economical price
• Standard temperature ranges from -40 °F to 120 °F through 0 °F to 1,000 °F
• Custom temperature ranges available from -300 °F to 1,100 °F (-200 °C to 600 °C)
• Wide variety of temperature ranges and connections
• Quick response time
• Optional USB programmable output available
• 316 stainless steel housing
• Burnout protection from 3.3 mA to 23 mA
• Also available with our 1800 Series attachable loop indicator
• CE compliant to suppress RFI, EMI, and ESD
Electronic Temperature Measurement
Water systems
Storage tanks
Industrial machinery and machine tools
HVAC systems
Refrigeration systems
APPLICATIONS
800 SERIES
SPECIFICATIONS
Output signals 4 mA to 20 mA 2-wire, 0 Vdc to 5 Vdc 3-wire
0 Vdc to 10 Vdc 3-wire, 1 Vdc to 5 Vdc 3-wire
Temperature ranges Standard ranges from -40 °F to 1,000 °F
Accuracy
Measuring element Class B per EN 60751 (IEC 751)
Output ±[0.30 +0.005*ltl] °C
±0.5% full scale
Sensor protection Burnout protected from 3.3 mA to 23 mA
Power requirement* 10 Vdc to 30 Vdc (4 mA to 20 mA, 2-wire)
10 Vdc to 30 Vdc (0 Vdc to 5 Vdc, 3-wire)
10 Vdc to 30 Vdc (1 Vdc to 5 Vdc, 3-wire)
14 Vdc to 30 Vdc (0 Vdc to 10 Vdc, 3-wire)
Load limitations ≤ (VPower -10)/0.020 Amp for 4 mA to 20 mA output
≤ 5,000 Ω for 1 Vdc to 5 Vdc output
≤ 10,000 Ω for 0 Vdc to 10 Vdc output
≤ 4,500 Ω for 0.5 Vdc to 4.5 Vdc output
Wetted materials 316 stainless steel
Housing material 316 stainless steel
Ambient temperature -40 °F to 185 °F (-40 °C to 85 °C)
Storage temperature -40 °F to 185 °F (-40 °C to 85 °C)
Environmental rating IP65 according to EN 60529/IEC 529
Electromagnetic rating CE compliant to EMC norm EN 61326:1997/A1:1998
RFI, EMI and ESD protection
Electrical protection Reverse polarity, over-voltage and short circuit protection
Weight Approximately 4 oz.
Also available with our 1800 Series
attachable loop indicator.
See www.noshok.com for details.
Optional USB programmable output
Thermowells are recommended for pressure, corrosive fl uids and high velocity applications, see pages 36-37.
* Unregulated