Continuous, Automatic recovery
Output Current Limitation
(depending on model)
3.7 - 5 VDC (3.3 VDC model)
5.6 - 7 VDC (5 VDC model)
13.5 - 16 VDC (12 VDC model)
18.3 - 22 VDC (15 VDC model)
29.1 - 34.5 VDC (24 VDC model)
5.6 - 7 VDC (±5 VDC model)
13.5 - 18.2 VDC (±12 VDC model)
17 - 22 VDC (±15 VDC model)
250 µs typ. (25% Load Step)
- IT / Multimedia Equipment
EN 62368-1
IEC 62368-1
UL 62368-1
EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
- Certification Documents
www.tracopower.com/overview/thm3wi
EN 60601-1-2 edition 4 (Medical Devices)
EN 55011 class A (internal filter)
EN 55011 class B (with external filter)
EN 55032 class A (internal filter)
EN 55032 class B (with external filter)
FCC Part 18 class A (internal filter)
FCC Part 18 class B (with external filter)
EN 55011 class A (internal filter)
EN 55011 class B (with external filter)
EN 55032 class A (internal filter)
EN 55032 class B (with external filter)
FCC Part 18 class A (internal filter)
FCC Part 18 class B (with external filter)
External filter proposal:
www.tracopower.com/overview/thm3wi
EN 60601-1-2 edition 4 (Medical Devices)
- Electrostatic Discharge
EN 61000-4-2, ±15 kV, perf. criteria A
EN 61000-4-2, ±8 kV, perf. criteria A
- RF Electromagnetic Field
EN 61000-4-3, 10 V/m, perf. criteria A
EN 61000-4-4, ±2 kV, perf. criteria A
EN 61000-4-5, ±2 kV, perf. criteria A
5 Vin models: KY 1000 µF // Vishay V10P45
24 Vin models: KY 470 µF
48 Vin models: KY 330 µF
- Conducted RF Disturbances
EN 61000-4-6, 10 Vrms, perf. criteria A
EN 61000-4-8, 100 A/m, perf. criteria A
EN 61000-4-8, 1000 A/m, perf. criteria A
95% max. (non condensing)
All specifications valid at nominal voltage, full load and +25°C after warm-up time unless otherwise stated.