Digital Storage Oscilloscopes
TBS1000 Series Datasheet
The TBS1000 Digital Storage Oscilloscope Series provides you with
affordable performance in a compact design. Packed with standard
features - including USB connectivity, 16 automated measurements, limit
testing, data logging, and context-sensitive help - the TBS1000 Series
oscilloscopes help you get more done, in less time.
Key performance specifications
150 MHz, 100 MHz, 60 MHz, 40 MHz and 25 MHz bandwidth models
4 and 2-channel models
Up to 1 GS/s sample rate on all channels
2.5k point record length on all channels
Advanced triggers including pulse width trigger and line-selectable
video trigger
Key features
16 automated measurements, and FFT analysis for simplified
waveform analysis
Built-in waveform limit testing
Automated, extended data logging feature
Autoset and signal auto-ranging
Built-in context-sensitive help
Probe check wizard
Multiple-language user interface
5.7 in. (144 mm) Active TFT Color Display
Small footprint and lightweight - Only 4.9 in. (124 mm) deep and 4.4 lb.
(2 kg)
Connectivity
USB 2.0 host port on the front panel for quick and easy data storage
USB 2.0 device port on rear panel for easy connection to a PC or direct
printing to a PictBridge®-compatible printer
Includes Tektronix OpenChoice® software for connecting your bench
Digital precision for accurate measurements
With up to 150 MHz bandwidth and 1 GS/s maximum sample rate, no other
digital storage oscilloscope offers as much bandwidth and sample rate for
the price. Tektronix proprietary sampling technology provides real-time
sampling with the stated sampling rate on all channels, all the time to
accurately capture your signals. Sampling performance is not reduced
when using multiple channels.
See all the details other oscilloscopes might miss with Tektronix proprietary digital real-
time sampling
The TBS1000 Digital Oscilloscope Series is especially well suited to meet
the needs of today's schools and universities. Packed with features and
built-in tools, the TBS1000 is easy to learn and simple to operate - ideal for
first-time oscilloscope users and students. Featuring the same user
interface as the Tektronix TDS Oscilloscope Family, your students will learn
to operate the world's most popular oscilloscope platform, with over
500,000 oscilloscopes in operation worldwide.
www.tektronix.com 1
Quickly and easily capture waveforms
To simplify integration with your existing curriculum, the TBS1000 also
includes an Education Resource CD filled with tools to help your students
master the use of an oscilloscope. The TBS1000 offers the tools and
performance you need at a price you can afford.
Critical tools for troubleshooting your device
Advanced triggers - rising/falling edge, pulse width, and video - help you
quickly isolate your signals of interest. Once you've captured a signal,
advanced math capabilities and automated measurements can speed your
analysis. Quickly perform an FFT or add, subtract, or multiply waveforms.
Sixteen automated measurements quickly and reliably calculate important
signal characteristics such as frequency or rise time, while the built-in Limit
Test function enables you to easily identify problems in your signal.
Quickly perform an FFT with the advanced math functions
Designed to make your work easy
The TBS1000 series oscilloscopes are designed with the ease of use and
familiar operation you have come to expect from Tektronix.
Intuitive operation
The intuitive user interface with dedicated per-channel vertical controls,
auto-setup, and auto-ranging makes these instruments easy to use,
reducing learning time and increasing efficiency.
Help when you need it, where you need it
The context-sensitive help system provides important information specific to the task you
are working on
The built-in Help menu provides you with important information on your
oscilloscope's features and functions. Help is provided in the same
languages as the user interface.
Probe check wizard
Check out your probe compensation before making measurements with just
one button that starts a fast, easy procedure.
Datasheet
2 www.tektronix.com
Limit test
Limit test provides a quick Pass/Fail comparison of any triggered input signal to a user-
defined template
The oscilloscope can automatically monitor source signals and output Pass
or Fail results by judging whether the input waveform is within predefined
boundaries. Specific actions can be triggered on violation including
stopping waveform acquisition, stopping Limit Test functions, saving the
failed waveform data or screen image to a USB memory device, or any
combination of the above. This is an ideal solution for manufacturing or
service applications where you need to make decisions quickly.
Flexible data transfer
Data logging enables automatic saving of triggered waveforms
The USB host port on the front panel enables you to save your instrument
settings, screenshots, and waveform data in a flash. The built-in Data
Logging feature means you can set up your oscilloscope to save user-
specified triggered waveforms to a USB memory device for up to 24 hours.
You can also select the "Infinite" option for continuous waveform
monitoring. With this mode you can save your triggered waveforms to an
external USB memory device without a duration limitation until the memory
device is full. The oscilloscope will then guide you to insert another USB
memory device to continue saving waveforms.
Easy PC connectivity
Easily capture, save, and analyze measurements results by connecting to
your PC with the rear-panel USB device port and the included copy of
OpenChoice PC Communications Software. Simply pull screen images and
waveform data into the stand-alone desktop application or directly into
Microsoft Word and Excel. Alternatively, if you prefer not to use your PC,
you can simply print your image directly to any PictBridge-compatible
printer.
Easily capture, save and analyze measurement results with OpenChoice PC
communications software
Performance you can count on
In addition to industry-leading service and support, every TBS1000 series
oscilloscope comes backed with a 5-year warranty as standard.
TBS1000 Digital Storage Oscilloscopes
www.tektronix.com 3
Educational resources
Every TBS1000 model includes an education resource CD filled with tools
to help your students master the use of an oscilloscope. The education
resource CD includes two student labs and instructor's guides, and two
primers. The Introduction to Oscilloscopes student lab and instructor's
guide explains the basics of oscilloscope operation complete with hands-on
exercises for your students. The Introduction to Oscilloscope Probes
student lab and instructor's guide explains the fundamentals of probing and
how probes can affect measurement quality. The two primers included are
the most popular and widely-used from Tektronix - the XYZs of
Oscilloscopes and ABCs of Probes.
The included education resource CD is filled with tools to help students master the use of
an oscilloscope
Datasheet
4 www.tektronix.com
Specifications
All specifications apply to all models unless noted otherwise.
Model overview
TBS1022 TBS1042 TBS1062 TBS1064 TBS1102 TBS1104 TBS1152 TBS1154
Bandwidth 125 MHz 40 MHz 60 MHz 60 MHz 100 MHz 100 MHz 150 MHz 150 MHz
Channels 2 2 2 4 2 4 2 4
Sample rate on each channel 500 MS/s 500 MS/s 1.0 GS/s 1.0 GS/s 1.0 GS/s 1.0 GS/s 1.0 GS/s 1.0 GS/s
Record length 2.5k points at all time bases
Vertical system — Analog channels
Vertical resolution 8 bits
Input sensitivity range 2 mV to 5 V/div on all models with calibrated fine adjustment
DC gain accuracy ±3%, from 10 mV/div to 5 V/div
Maximum input voltage 300 VRMS CAT II; derated at 20 dB/decade above 100 kHz to 13 Vp-p AC at 3 MHz
Offset range 2 mV to 200 mV/div: ±1.8 V
>200 mV to 5 V/div: ±45 V
Bandwidth limit 20 MHz
Input coupling AC, DC, GND
Input impedance 1 MΩ in parallel with 20 pF
Vertical zoom Vertically expand or compress a live or stopped waveform
Horizontal system — Analog channels
Time base range 5 ns to 50 s/div
Time base accuracy 50 ppm
Horizontal zoom Horizontally expand or compress a live or stopped waveform
1Bandwidth is 20 MHz at 2 mV/div
TBS1000 Digital Storage Oscilloscopes
www.tektronix.com 5
Input/Output ports
USB interface USB host port on front panel supports USB flash drives USB device port on back of instrument supports connection to PC and all
PictBridge®-compatible printers
GPIB interface Optional
Data storage
Nonvolatile storage
Reference waveform display 2.5K point reference waveforms
Waveform storage without
USB flash drive
2.5K point
Maximum USB flash drive size 64 GB
Waveform storage with USB
flash drive
96 or more reference waveforms per 8 MB
Setups without USB flash
drive
10 front-panel setups
Setups with USB flash drive 4000 or more front-panel setups per 8 MB
Screen images with USB flash
drive
128 or more screen images per 8 MB (the number of images depends on file format selected)
Save All with USB flash drive 12 or more Save All operations per 8 MB
A single Save All operation creates 3 to 9 files (setup, image, plus one file for each displayed waveform)
Acquisition system
Acquisition modes
Peak Detect High-frequency and random glitch capture. Captures glitches as narrow as 12 ns (typical) at all time base settings from 5 μs/div to
50 s/div
Sample Sample data only
Average Waveform averaged, selectable: 4, 16, 64, 128
Single Sequence Use the Single Sequence button to capture a single triggered acquisition sequence
Roll At acquisition time base settings of >100 ms/div
Trigger system
External trigger input Included on all models
Trigger modes Auto, Normal, Single Sequence
Trigger types
Edge (Rising/Falling) Conventional level-driven trigger. Positive or negative slope on any channel. Coupling selections: AC, DC, Noise Reject, HF
Reject, LF Reject
Video Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC, PAL, SECAM)
Pulse Width (or Glitch) Trigger on a pulse width less than, greater than, equal to, or not equal to, a selectable time limit ranging from 33 ns to 10 s
Trigger source Two channel models: CH1, CH2, Ext, Ext/5, AC Line
Four channel models: CH1, CH2, CH3, CH4, Ext, Ext/5, AC Line
Trigger view Displays trigger signal while Trigger View button is depressed.
Trigger signal frequency readout Provides a frequency readout of the trigger source.
Datasheet
6 www.tektronix.com
Waveform measurements
Cursors
Types Amplitude, Time
Measurements ΔT, 1/ΔT, ΔV
Automatic measurements Period, Frequency, +Width, –Width, Rise Time, Fall Time, Max, Min, Peak-to-Peak, Mean, RMS, Cycle RMS, Cursor RMS, Duty
Cycle, Phase, and Delay
Waveform math
Arithmetic Add, Subtract, Multiply
Math functions FFT
FFT Windows: Hanning, Flat Top, Rectangular 2048 sample points
Sources Two channel models: CH1 – CH2, CH2 – CH1, CH1 + CH2, CH1 × CH2
Four channel models: CH1 – CH2, CH2 – CH1, CH1 + CH2, CH1 × CH2, CH3 – CH4, CH4 – CH3, CH3 + CH4, CH3 × CH4
Autoset
Autoset menu Single-button, automatic setup of all channels for vertical, horizontal, and trigger systems, with undo Autoset
Square wave Single Cycle, Multicycle, Rising or Falling Edge
Sine wave Single Cycle, Multicycle, FFT Spectrum
Video (NTSC, PAL, SECAM) Field: All, Odd, or Even Line: All or Selectable Line Number
Autorange
Automatically adjust vertical and/or horizontal oscilloscope settings when probe is moved from point to point, or when the signal
exhibits large changes.
Display system
Interpolation Sin (x)/x
Waveform styles Dots, vectors
Persistence Off, 1 s, 2 s, 5 s, infinite
Format YT and XY
TBS1000 Digital Storage Oscilloscopes
www.tektronix.com 7
Physical characteristics
Dimensions mm in.
Height 158.0 6.22
Width 326.3 12.85
Depth 124.2 4.89
Shipping dimensions mm in.
Height 266.7 10.5
Width 476.2 18.75
Depth 228.6 9.0
Weight kg lb.
Instrument only 2.0 4.4
...with accessories 2.2 4.9
RM2000B rackmount mm in
Width 482.6 19.0
Height 177.8 7.0
Depth 108.0 4.25
Environmental
Temperature
Operating 0 to +50 ºC
Nonoperating –40 to +71 ºC
Humidity
Operating and nonoperating Up to 85% RH at or below +40 ºC
Up to 45% RH up to +50 ºC
Altitude
Operating and nonoperating Up to 3,000 m (9,843 ft.)
Regulatory
Electromagnetic compatibility Meets Directive 2004/108/EC, EN 61326-2-1 Class A; Australian EMC Framework
Safety UL61010-1:2004, CSA22.2 No. 61010-1:2004, EN61010-1:2001, IEC61010-1:2001
Datasheet
8 www.tektronix.com
Ordering information
Models
TBS1022 25 MHz, 2 Ch, 500 MS/s, TFT DSO
TBS1042 40 MHz, 2 Ch, 500 MS/s, TFT DSO
TBS1062 60 MHz, 2 Ch, 1 GS/s, TFT DSO
TBS1064 60 MHz, 4 Ch, 1 GS/s, TFT DSO
TBS1102 100 MHz, 2 Ch, 1 GS/s, TFT DSO
TBS1104 100 MHz, 4 Ch, 1 GS/s, TFT DSO
TBS1152 150 MHz, 2 ch, 1 GS/s, TFT DSO
TBS1154 150 MHz, 4 ch, 1 GS/s, TFT DSO
Language options
Translated front-panel overlays included with their respective user manuals. 2
Language Description
L0 English (front-panel overlay on instrument)
L1 French (front-panel overlay)
L2 Italian (front-panel overlay)
L3 German (front-panel overlay)
L4 Spanish (front-panel overlay)
L5 Japanese (front-panel overlay)
L6 Portuguese (front-panel overlay)
L7 Simple Chinese (front-panel overlay)
L8 Standard Chinese (front-panel overlay)
L9 Korean (front-panel overlay)
L10 Russian (front-panel overlay)
2User manuals (PDF) in 11 languages are available on the CD and for download from www.tektronix.com . There are no printed user manuals.
TBS1000 Digital Storage Oscilloscopes
www.tektronix.com 9
Power plug options
Opt. A0 North America power plug (115 V, 60 Hz)
Opt. A1 Universal Euro power plug (220 V, 50 Hz)
Opt. A2 United Kingdom power plug (240 V, 50 Hz)
Opt. A3 Australia power plug (240 V, 50 Hz)
Opt. A5 Switzerland power plug (220 V, 50 Hz)
Opt. A6 Japan power plug (100 V, 110/120 V, 60 Hz)
Opt. A10 China power plug (50 Hz)
Opt. A11 India power plug (50 Hz)
Opt. A12 Brazil power plug (60 Hz)
Opt. A99 No power cord
Service options
Opt. D1 Calibration Data Report
Probes and accessories are not covered by the oscilloscope warranty and Service Offerings. Refer to the datasheet of each probe and accessory model for its unique warranty
and calibration terms.
Standard accessories
Accessory Description
Passive probes, one per channel TPP0101: 100 MHz passive probe for: TBS1022, TBS1042, TBS1062, TBS1064, TBS1102, and TBS1104
TPP0201: 200 MHz passive probe for: TBS1152 and TBS1154
Power cord (Please specify plug option)
NIM/NIST Traceable certificate of calibration
Printed documentation Installation and safety manual
(English, Japanese, and Simplified Chinese)
CD with customer documentation and
OpenChoice PC communications
software
Customer documentation including detailed user manuals (English, French, German, Italian, Japanese, Korean, Portuguese,
Russian, Simplified Chinese, Spanish, Traditional Chinese)
Software for fast and easy communication between an MS Windows PC and the TBS1000 series using USB to transfer and save
settings, waveforms, measurements, and screen images
Educators Classroom and Lab
Resource CD
Contains lab experiments and primers for both oscilloscopes and probes
5-year warranty Covers labor and parts for defects in materials and workmanship for 5 years, excluding probes and accessories (probes and
accessories are not covered by the oscilloscope warranty and service offerings. refer to the data sheet of each probe and
accessory model for its unique warranty and calibration terms)
Datasheet
10 www.tektronix.com
Recommended accessories
Accessory Description
TEK-USB-488 GPIB-to-USB converter
AC2100 Soft carrying case for instrument
HCTEK4321 Hard plastic carrying case for instrument (requires AC2100)
RM2000B Rackmount kit
077-0444-xx Programmer manual – English only
077-0772-xx Service manual – English only
174-4401-xx USB host to device cable, 3 ft. long
Recommended probes
Probe Description
TPP0101 10X passive probe, 100 MHz bandwidth
TPP0201 10X passive probe, 200 MHz bandwidth
P2220 1X/10X passive probe, 200 MHz bandwidth
P6101B 1X passive probe (15 MHz, 300 VRMS CAT II rating)
P6015A 1000X high-voltage passive probe (75 MHz)
P5100A 100X high-voltage passive probe (500 MHz)
P5200A 50 MHz, 50X/500X high-voltage differential probe
P6021A 15 A, 60 MHz AC current probe
P6022 6 A, 120 MHz AC current probe
A621 2000 A, 5 to 50 kHz AC current probe
A622 100 A, 100 kHz AC/DC current probe/BNC
TCP303/TCPA300 150 A, 15 MHz AC/DC current probe/amplifier
TCP305A/TCPA300 50 A, 50 MHz AC/DC current probe/amplifier
TCP312A/TCPA300 30 A, 100 MHz AC/DC current probe/amplifier
TCP404XL/TCPA400 500 A, 2 MHz AC/DC current probe/amplifier
Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar.
Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats.
TBS1000 Digital Storage Oscilloscopes
www.tektronix.com 11
Datasheet
ASEAN / Australasia (65) 6356 3900 Austria 00800 2255 4835* Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Belgium 00800 2255 4835* Brazil +55 (11) 3759 7627 Canada 1 800 833 9200
Central East Europe and the Baltics +41 52 675 3777 Central Europe & Greece +41 52 675 3777 Denmark +45 80 88 1401
Finland +41 52 675 3777 France 00800 2255 4835* Germany 00800 2255 4835*
Hong Kong 400 820 5835 India 000 800 650 1835 Italy 00800 2255 4835*
Japan 81 (3) 6714 3010 Luxembourg +41 52 675 3777 Mexico, Central/South America & Caribbean 52 (55) 56 04 50 90
Middle East, Asia, and North Africa +41 52 675 3777 The Netherlands 00800 2255 4835* Norway 800 16098
People's Republic of China 400 820 5835 Poland +41 52 675 3777 Portugal 80 08 12370
Republic of Korea 001 800 8255 2835 Russia & CIS +7 (495) 6647564 South Africa +41 52 675 3777
Spain 00800 2255 4835* Sweden 00800 2255 4835* Switzerland 00800 2255 4835*
Taiwan 886 (2) 2722 9622 United Kingdom & Ireland 00800 2255 4835* USA 1 800 833 9200
* European toll-free number. If not accessible, call: +41 52 675 3777 Updated 10 April 2013
For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com.
Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price
change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies.
02 Jul 2013 3GW-28557-3
www.tektronix.com
Mouser Electronics
Authorized Distributor
Click to View Pricing, Inventory, Delivery & Lifecycle Information:
Tektronix:
TBS1064 TBS1104 TBS1154