QUICK START GUIDE FOR DEMONSTRATION CIRCUIT 919
16-BIT 10 TO 105 MSPS ADC
5
non-harmonically related spurs and broad band
noise. Low phase noise Agilent 8644B generators
are used with TTE band pass filters for both the
Clock input and the Analog input. In the case of the
LTC2203/2 we use a divide by 4.
This demo board is, populated by default for the dc
input path. There is a transformer mounted at T1,
but C4, C6, R9 and R13 are not populated. If a sin-
gle ended AC input is required, the DC input paths
must be disconnected by removing R26,27,31 &
32, and the above components installed.
If the transformer input is required, C4 & C6
should be 0.1uF X5R 0402, R9 and R13 should be
10 ohm 0402 resistors.
Note that this transformer (ETC1-1T) is poor below
1 MHz.
Reduced amplitude signals can be applied to 300
KHz. Applications below 300 KHz must be driven
directly via DC inputs.
As there are a significant number of these boards
that are customized, please confirm the population
of your board. The schematic below shows the de-
fault population, the photograph shows the popula-
tion of a transformer coupled version. This board
may also be populated with LTC2204-2207 for DC
drive applications, in which case, R33 is a 0.1uF
capacitor.
Apply the analog input signal of interest to the SMA
connector on the DC919 demonstration circuit
board marked “J2 ANALOG INPUT”. These inputs
are capacitive coupled to a Flux coupled transform-
ers ETC1-1T. In some cases, where these devices
are to be used in under-sampling scenarios, this
transformer should be replaced with an ETC1-1-13
Balun.
The DC919 can be modified for direct DC drive from
a suitable differential signal source.
This may be done by yourself or at special request
when you order the demo board.
If the DC input paths are populated with low value
(5.1 ohm) resistors at both ends of these transmis-
sion lines, you must provide a reasonably well bal-
anced differential drive with 1.25V common mode.
The spacing of these SMA connectors (0.8”) allows
them to be mated directly with demo boards for
devices such as the LT1993, LT1994, LT5514 and
others.
It is not recommended to drive this ADC in a single
ended fashion into a single DC input.
An internally generated conversion clock output is
available on pin 3 of J1 and the data samples are
available on Pins 7-37 of J1 which can be collected
via a logic analyzer, cabled to a development system
through a SHORT 2 to 4 inch long 40-pin ribbon
cable or collected by the DC718 QuickEval-II Data
Acquisition Board using the
PScope System Soft-
ware
provided or down loaded from the Linear
Technology website at
http://www.linear.com/software/. If a DC718 was
provided, follow the DC718 Quick Start Guide and
the instructions below.
If data is to be collected by a logic analyzer, pin 40
must be strapped to OVDD or 2.5V.
(Please see schematic.)
To start the data collection software if
“
PScope.exe
”, is installed (by default) in
\Program Files\LTC\PScope\, double click the
PScope Icon or bring up the run window under the
start menu and browse to the PScope directory and
select PScope.
Configure PScope for the appropriate variant of the
DC919 demonstration circuit by selecting the cor-
rect A/D Converter as installed on the DC919. Un-
der the “Configure” menu, go to “Device.” Under
the “Device” pull down menu, select device, either
LTC2202, LTC2203 or any of the pin-compatible 14-
Bit and 16-Bit parts up to LTC2207. Select the part
in the Device List and PScope will automatically
blank the last two LSBs when using a DC919 sup-
plied with a 14-Bit part. If you are operating with a