3/8
PERFORMANCE SPECIFICATIONS
Ambient Temperature: 25°C (unless otherwise specified)
PARAMETERS MIN TYP MAX UNITS NOTES
Accuracy -0.5 0.5 %F.S. BFSL 001psiG @ 25°C
(combined non linearity, hysteresis, and repeatability) -0.25 0.25 %F.S. BFSL 005psi @ 25°C
-0.1 0.1 %F.S. BFSL >005psi, <01Kpsi @ 25°C
-0.25 0.25 %F.S. BFSL ≥01Kpsi, ≤05Kpsi @ 25°C
Isolation, Body to any Lead 1 MΩ @ 25Vdc
Pressure Cycles 1.00E+6
FS Cycles
Proof Pressure 3X Rated
Burst Pressure 4X Rated
Long Term Stability (1 year) -0.25 0.25 %F.S. 001psiG
-0.1 0.1 %F.S. ≥005psi, ≤05Kpsi
Total Error Band (over compensated range) -1.25 1.25 %F.S. 001psiG
-1 1 %F.S. 005psi
-0.75 0.75 %F.S. >005psi, ≤05Kpsi
Compensated Temperature 0 50 °C 001psiG
0 70 °C 005psi
-20 +85 °C ≥005psi, ≤05Kpsi
Operating Temperature -40 +125 °C 001psi
-40 +125 °C ≥005psi, ≤05Kpsi, Cable 105°C
Storage Temperature -40 +125 °C Cable 105°C
Load Resistance (R
L
) R
L
> 100k Ω Voltage Output
Bandwidth DC to 1KHz (Typical)
Shock (11ms) 50g, 11msec Half Sine Shock per MIL-STD-202G, Method 213B, Condition A
Vibration (20 to 200kHz) ±20g, MIL-STD-810C, Procedure 514.2, Fig 514.2-2, Curve L
For custom configurations, consult factory.
CE Compliance
EN55022 Emissions Class A & B
IEC61000-4-2 Electrostatic Discharge Immunity (6kV contact/8kV air)
IEC61000-4-3 EM Field Immunity (30V/m)
IEC61000-4-4 Electrical Fast Transient Immunity (1kV)
IEC61000-4-5 Surge (1kV)
IEC61000-4-6 Conducted Immunity (10V)
IEC61000-4-9 Pulsed Magnetic Field Immunity (100A/m)