The Model DMM7510 combines all the advantages of a precision digital multimeter, a graphical touchscreen display, and a high speed, high resolution digitizer to create an industry first: a graphical sampling multimeter. The digitizer gives the Model DMM7510 unprecedented signal analysis flexibility; the five-inch capacitive touchscreen display makes it easy to observe, interact with, and explore measurements with "pinch and zoom" simplicity. This combination of high performance and high ease of use offers unparalleled insight into your test results. * Precision multimeter with 31/2to 71/2-digit resolution * 14 PPM basic one-year DCV accuracy Capture Waveforms with the Built-in 1MS/sec Digitizer Capturing and displaying waveforms and transient events just got easier with the DMM7510's voltage or current digitizing function. The built-in 1MS/sec, 18-bit digitizer makes it possible to acquire waveforms without the need to use a separate instrument. The digitizing functions employ the same ranges that the DC voltage and current functions use to deliver exceptional dynamic measurement range. In addition, the voltage digitizing function uses the same DC voltage input impedance (10G or 10M) levels to reduce loading significantly on the DUT. * 100mV, 1, and 10A ranges offer the sensitivity needed for measuring low level signals Graphical sampling DMM 71/2-Digit Graphical Sampling Multimeter * Make accurate low resistance measurements with offset compensated ohms, four-wire, and dry circuit functions * Capture and display waveforms or transients with 1MS/sec digitizer * Large internal memory buffer; store over 11 million readings in standard mode or 27.5 million in compact mode The high speed digitizing function allows capturing and displaying voltage and current waveforms. Advanced triggering options make it possible to capture a signal at precisely the right point. * Auto-calibration feature improves accuracy and stability by minimizing temperature and time drift * Display more with five-inch, high resolution touchscreen interface * Readings and screen images can be saved quickly via the front panel USB memory port * Multiple connectivity options: GPIB, USB, and LXI-compliant LAN interfaces * Two-year specifications allow for longer calibration cycles The built-in graphing utility supports displaying and comparing measurements or waveforms from up to four reading buffers at once. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company DIGITAL MULTIMETERS & SYSTEMS Graphical sampling DMM DMM7510 DIGITAL MULTIMETERS & SYSTEMS DMM7510 71/2-Digit Graphical Sampling Multimeter DMM7510-NFP 71/2-Digit Graphical Sampling Multimeter, with No Front Panel DMM7510-RACK 71/2-Digit Graphical Sampling Multimeter, with No Handle DMM7510-NFP-RACK 71/2-Digit Graphical Sampling Multimeter, with No Front Panel and No Handle Accessories Supplied 1756 Test Leads USB-B-1 USB Cable, Type A to Type B, 1m (3.3 ft) CA-180-3A TSP-Link/Ethernet Cable Documentation CD DMM7510 QuickStart Guide Test Script Builder Software (available at www.keithley.com ) KickStart Startup Software (available at www.keithley.com) LabVIEW and IVI Drivers (available at www.keithley.com) Make Demanding Measurements with Confidence The Model DMM7510's design makes the most of Keithley's low level measurement expertise. Features like the low noise input stage and the 32-bit A-to-D converter allow this instrument to deliver DC accuracies typically only found in metrology-grade instrumentation--but at about half the price of those solutions. The Model DMM7510's 100mV, 10, and 10A ranges deliver the sensitivity needed to measure low signals with confidence when characterizing today's demanding electronic designs. In addition to one- and two-year accuracy specifications, an auto-calibration function ensures greater accuracy between calibration cycles. One Power Line Cycle Noise Performance The DMM7510 provides noise performance equivalent to or better than many 81/2-digit DMMs. Graphical sampling DMM Graphical sampling DMM Ordering Information 71/2-Digit Graphical Sampling Multimeter 2V 1V Volts DMM7510 0V -1V -2V 61/2 Digit 81/2 Digit DMM7510 Comparison of the Model DMM7510's 1V DC noise performance with that of typical 61/2- and 81/2-digit multimeters. All data was taken at 1 NPLC with a low thermal short applied to the input. 15 Measurement Functions The DMM7510 provides 15 basic measurement functions. In addition to the digitizing voltage and current functions, it includes capacitance, ACV and ACI, temperature (RTD, thermistor, and thermocouple), 2- and 4-wire resistance, dry circuit ohms, period, frequency, diode test, and DC voltage ratio. The instrument's flat menu structure allows for fast configuration and improves usability. Its intuitive design lets you learn how to operate the instrument and begin making device measurements faster and with greater confidence. RTD, Thermistor, Thermocouple 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company 71/2-Digit Graphical Sampling Multimeter DMM7510 1f DC Voltage AC Voltage DC Current AC Current Resistance (2-wire) Resistance (4-wire) Dry Circuit Resistance Capacitance Period Frequency 1p 1n 1 1m 1 1k 10nV 100nV 1pA 1G 10.1A 10.1A 1nA 1 100n 1 1.2G 1.2G 2.4k 0.1pF 1.2mF 2s Digitize Voltage Digitize Current 1M 1010V 707VRMS 330ms 3Hz 500kHz 1V 1010V 0.1nA 10.1A Linear Scale Thermocouple Thermistor 3-Wire RTD 4-Wire RTD 1820C -200C -80C -200C -200C -500 150C 850C 850C 0 500 1000 1500 2000 Designed for Higher Testing Productivity In addition to its advanced touchscreen, the Model DMM7510's front panel offers a variety of features that enhance its speed, user-friendliness, and learnability, including a USB 2.0 memory I/O port, a HELP key, a rotary navigation/control knob, and front/rear input selector button. All front-panel buttons are backlit to enhance visibility. Five-inch touchscreen displays more information simultaneously Rotary control knob offers an alternative to touchscreen navigation Online HELP is just a button press away Store test results and screen images quickly via the USB 2.0 Memory port Lower swipe screen speeds access to often-used features REPLACEMENT FUSES DMM7510-FUSE-10A 11A Current Fuse For DMM7510 DMM7510-FUSE-3A 3.5A Current Fuse For DMM7510 CABLES, CONNECTORS, ADAPTERS CA-18-1 Shielded Dual Banana Cable, 1.2m (4 ft.) COMMUNICATION INTERFACES & CABLES KPCI-488LPA IEEE-488 Interface for PCI Bus KUSB-488B IEEE-488 USB-to-GPIB Interface Adapter 7007-1 Shielded GPIB Cable, 1m (3.2ft) 7007-2 Shielded GPIB Cable, 2m (6.5ft) CA-180-3A CAT5 Crossover Cable for TSP-Link / Ethernet USB-B-1 USB Cable, Type A to Type B, 1m (3.3 ft) TRIGGERING 2450-TLINK 8501-1 8501-2 8503 AND CONTROL DB-9 to Trigger Link Connector Adapter Trigger Link Cable, DIN-to-DIN, 1m (3.2 ft.) Trigger Link Cable, DIN-to-DIN, 2m (6.5 ft.) DIN-to-BNC Trigger Cable RACK MOUNT KITS 4299-8 Single Fixed Rack Mount Kit 4299-9 Dual Fixed Rack Mount Kit 4299-10 Dual Fixed Rack Mount Kit. Mount One DMM7510 and One Series 26xxB Instrument 4299-11 Dual Fixed Rack Mount Kit. Mount One DMM7510 and One Instrument from Series 2400, Series 2000, etc. 4299-12 Dual Fixed Rack Mount Kit. Mount One DMM7510 and One Keysight Instrument. DMM7510-BenchKit Ears and Handle for DMM7510-NFP-RACK and DMM7510-RACK Models DIGITAL MULTIMETERS & SYSTEMS Graphical sampling DMM Logarithmic Scale ACCESSORIES AVAILABLE TEST LEADS AND PROBES 1752 Premium Safety Test Lead Kit 1754 2-Wire Universal 10-Piece Test Lead Kit 1756 General Purpose Test Lead Kit 5804 Kelvin (4-Wire) Universal 10-Piece Test Lead Kit 5805 Kelvin (4-Wire) Spring-Loaded Probes 5806 Kelvin Clip Lead Set 5808 Low Cost Single-pin Kelvin Probe Set 5809 Low Cost Kelvin Clip Lead Set 8606 High Performance Modular Probe Kit 8610 Low Thermal Shorting Plug Graphical sampling DMM DMM7510 Measurement Capabilities Front/rear input selector 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company 71/2-Digit Graphical Sampling Multimeter DIGITAL MULTIMETERS & SYSTEMS Graphical sampling DMM SERVICES AVAILABLE EXTENDED WARRANTIES DMM7510-EW 1 Year Factory Warranty Extended to 2 Years from Date of Shipment DMM7510-3Y-EW 1 Year Factory Warranty Extended to 3 Years from Date of Shipment DMM7510-5Y-EW 1 Year Factory Warranty Extended to 5 Years from Date of Shipment DMM7510-NFP-EW 1 Year Factory Warranty Extended to 2 Years from Date of Shipment DMM7510-NFP-3Y-EW 1 Year Factory Warranty Extended to 3 Years from Date of Shipment DMM7510-NFP-5Y-EW 1 Year Factory Warranty Extended to 5 Years from Date of Shipment CALIBRATION CONTRACTS C/DMM7510-3Y-17025 KeithleyCare(R) 3 Year ISO-17025 Calibration Plan C/DMM7510-3Y-DATA KeithleyCare 3 Year Calibration w/ Data Plan C/DMM7510-3Y-STD KeithleyCare 3 Year Std Calibration Plan C/DMM7510-5Y-17025 KeithleyCare 5 Year ISO-17025 Calibration Plan C/DMM7510-5Y-DATA KeithleyCare 5 Year Calibration w/ Data Plan C/DMM7510-5Y-STD KeithleyCare 5 Year Std Calibration Plan C/DMM7510-NFP-3Y-17025 KeithleyCare 3 Year ISO-17025 Calibration Plan C/DMM7510-NFP-3Y-DATA KeithleyCare 3 Year Calibration w/ Data Plan C/DMM7510-NFP-3Y-STD KeithleyCare 3 Year Std Calibration Plan C/DMM7510-NFP-5Y-17025 KeithleyCare 5 Year ISO-17025 Calibration Plan C/DMM7510-NFP-5Y-DATA KeithleyCare 5 Year Calibration w/ Data Plan C/DMM7510-NFP-5Y-STD KeithleyCare 5 Year Std Calibration Plan C/NEW DATA Calibration Data for New Units C/NEW DATA ISO ISO-17025 Calibration Data for New Units Easily accessible current fuses LAN/LXI interface USB 2.0 interface Measure Complete and External Trigger (Model 8503 BNC Accessory Available) Six user-defined digital I/Os GPIB interface The rear panel of the DMM7510 provides connections and controls that simplify configuring multi-instrument test solutions, including input connectors, remote control interfaces (GPIB, USB 2.0, and LXI/Ethernet), a D-sub 9-pin digital I/O port (for internal/external trigger signals and handler control), and TSP-Link(R) jacks for connecting to other TSP-enabled instruments. Flexible System Integration and Programming To offer users maximum programming flexibility and simplify configuring multi-instrument test systems, the DMM7510 includes Keithley's powerful Test Script Processor (TSP(R)) system and SCPI programming mode. The embedded scripting capability allows running powerful test scripts directly on the instrument, without the need for an external PC controller. These test scripts are complete test programs based on an easy-to-use yet highly efficient and compact scripting language, Lua (www.lua.org). Scripts are a collection of instrument control commands and/or program statements. Program statements control script execution and provide facilities such as variables, functions, branching, and loop control. This allows you to create powerful measurement applications with significantly reduced development times. Test scripts can contain any sequence of routines that are executable by conventional programming languages (including decision-making algorithms), so the instrument can manage every facet of the test without the need to communicate with a PC for decision-making. This eliminates delays due to GPIB, Ethernet or USB traffic congestion and greatly improves overall test times. TSP technology also offers "mainframe-less channel expansion." The TSP-Link channel expansion bus and a 100 Base T Ethernet cable allow connecting multiple DMM7510s with other TSP-enabled instruments in a master-slave configuration so they operate as a single integrated system. These instruments include the Model 2450 and Model 2460 Interactive SourceMeter(R) SMU instruments, Series 2600B SourceMeter SMU instruments, and the Series 3700A Switch/Multimeter systems. TSP-Link supports up to 32 units per GPIB or IP address, so it's easy to scale a system to fit the requirements of an application. A standard SCPI programming mode supports taking advantage of all of the DMM7510's new features when programming remotely. In addition, the instrument is code-compatible with the SCPI language, which many other DMMs use. This code compatibility avoids the need to rewrite code that is normally associated with upgrading to a new instrument with new c apabilities. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company Graphical sampling DMM DMM7510 DMM7510 71/2-Digit Graphical Sampling Multimeter Graphical sampling DMM Graphical sampling DMM Free Instrument Control Startup Software Keithley's KickStart instrument control startup software lets you begin taking measurements in minutes. * Connect using any remote interface (GPIB, USB, LAN) * Save instrument setups * Screenshot capture * Command line dialog box Ready-to-use Instrument Drivers Simplify Programming Need to create your own customized application software? Native National Instruments LabVIEW(R), IVI-C, and IVI-COM drivers are available for downloading at www.keithley.com to simplify the programming process. DIGITAL MULTIMETERS & SYSTEMS KickStart combines a wide range of functions to enhance testing productivity: * Instrument-specific UI panel * Manual instrument configuration * Basic reading display and tabular viewing of data * Datalogging * Native X-Y data graphing * Panning & zooming * Basic statistics (native to instrument, mX+b) * Saving/exporting data 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company 71/2-Digit Graphical Sampling Multimeter Specification Conditions DC Voltage This document contains specifications and supplemental information for the Model DMM7510 71/2-Digit Graphical Sampling Multimeter instrument. Specifications are the standards against which the Model DMM7510 is tested. Upon leaving the factory, the Model DMM7510 meets these specifications. Supplemental and typical values are nonwarranted, apply at 23C (73F), and are provided solely as useful information. Measurement accuracies are specified at the Model DMM7510 terminals under these conditions: * Temperature 23 5C, 5% to 80% relative humidity, noncondensing. * After a 90-minute warmup period. * 1 PLC or 5 PLC; for NPLC settings less than 1 PLC, add appropriate ppm of range for peak noise uncertainty from the RMS noise table. * Autozero enabled unless otherwise noted. * Remote sense operation or properly zeroed local operation. * Calibration period: One year or two years (calibration period may vary depending on customer requirements). * TACAL = Ambient temperature of last automatic calibration. * TCAL = Ambient temperature of last external calibration; factory calibration performed at 23 1C. ACCURACY (INPUT IMPEDANCE AUTO) Accuracy (ppm of reading + ppm of range) Input Resolution Impedance 2 >10 G or 10 nV 10 M 1 % >10 G or 100 nV 10 M 1 % >10 G or 1 V 10 M 1 % Range 1 100.00000 mV 4 1.0000000 V 4 10.000000 V 4 24 Hour TCAL 1C 2 90 Day TCAL 5C 1 Year TCAL 5C 2 Year TCAL 5C Temperature Coefficient 3 6+9 12 + 9 18 + 9 29 + 9 0.1 + 2.5 4+1 9+2 15+ 2 26 + 2 0.1 + 0.5 2 + 0.7 100.00000 V 4 10 V 10 M 1 % 8+3 1000.0000 V 4, 6 100 V 10 M 1 % 8+3 9 + 1.2 14 + 1.2 22 + 1.2 0.1 + 0.05 (18 + 5) 5 35 + 5 (19 + 5) 5 35 + 5 (22 + 5) 5 40 + 5 (23 + 5) 5 40 + 5 (30 + 5) 5 45 + 5 (31 + 5) 5 45 + 4 (0.15 + 0.05) 5 2.0 + 0.5 (0.15 + 0.05) 5 2.0 + 0.5 RMS NOISE (additional peak noise uncertainty) 7 * Applies to ppm of range. * Peak noise uncertainty is included in DC specifications for 1 PLC. * Add peak noise uncertainty to measurements for <1 PLC. * Input impedance set to Auto. Examples: * 10V at 0.006 PLC: 1.2 (from Accuracy table) + 11 (additional peak noise uncertainty) = 12.2 ppm of range. * 10V at 1 PLC: 1.2 + 0 = 1.2 ppm of range. NPLC 5 1 0.2 8 0.2 0.06 0.006 0.0005 Digits 71/2 71/2 61/2 61/2 51/2 41/2 31/2 100 mV 0.5 0.5 2 (10) 2 (12) 3 (17) 6 (42) 30 (220) 1V 0.08 0.09 0.2 (1.6) 0.2 (1.6) 0.4 (2.7) 3 (18) 20 (150) 10 V 0.06 0.07 0.1 (1.1) 0.1 (1) 0.3 (2.1) 1 (11) 20 (130) 100 V 0.3 0.4 1.1 (9.4) 1.1 (8.9) 3 (17) 20 (100) 120 (690) 1000 V 0.06 0.07 0.1 (1) 0.2 (1.1) 0.3 (2.4) 3 (18) 20 (150) DC VOLTAGE SENSE ACCURACY Accuracy (ppm of reading + ppm of range) Range 100.00000 mV 1.0000000 V 10.00000 V 24 Hour TCAL 1C 6 + 14 4 + 1.5 2 + 1.0 90 Day TCAL 5C 12 + 14 9+3 9 + 1.8 1 Year TCAL 5C 18 + 14 15 + 3 14 + 1.8 2 Year TCAL 5C 29 + 14 26 + 3 22 + 1.8 Temperature Coefficient 9 0.1 + 2.5 0.1 + 0.5 0.1 + 0.05 DC VOLTAGE RATIO DIGITAL MULTIMETERS & SYSTEMS For input signals 1% of the range, ratio accuracy = [[V INPUT ppm of reading + V INPUT ppm of range * (V INPUT range/V INPUT input)] + [VSENSE ppm of reading + VSENSE ppm of range * (VSENSE range/VSENSE input)]]. 1. 2. 3. 4. 5. 6. 7. 8. 9. 20% overrange on all ranges except 1% for 1000V range. Relative to calibration accuracy. Add per degree from TCAL 5C. When properly zeroed using the Rel function with external cables. Specified within 30 days of autocalibration, TOPER 5C from TACAL. For signal levels greater than 500V, add 0.02 ppm/V to the ppm of the readings specification for measurements exceeding 500V. Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. V RMS noise is typical. Additional peak noise is guaranteed. With line sync on. Add per degree from TCAL 5C. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company Model DMM7510 condensed specifications Model DMM7510 condensed specifications DMM7510 71/2-Digit Graphical Sampling Multimeter DC VOLTAGE CHARACTERISTICS NORMAL MODE REJECTION ADC LINEARITY: 1.0 ppm of reading + 1.0 ppm of range. INPUT IMPEDANCE: 100mV to 10V Ranges: Selectable >10G || <400pF (auto) or 10M 1% (10M). 100V to 1000V Ranges: 10M 1%. INPUT BIAS CURRENT: <50pA at 23C under the following conditions: Autozero off or input impedance 10M. COMMON MODE CURRENT: <2.1A peak-peak in 1MHz bandwidth. <100nA peak-peak in 1kHz bandwidth. COMMON MODE VOLTAGE: 500Vpeak LO terminal to chassis maximum. DC VOLTAGE AUTOZERO OFF ERROR: For 1C and 10 minutes, add (8ppm of reading + 15V). For DC voltage, line frequency 0.1%. Line Sync On Line Sync Off 5 PLC 110 dB 60 dB 1 PLC 90 dB 60 dB 0.2 PLC 45 dB -- 0.01 PLC -- -- For DC voltage and 1kW unbalanced in LO terminal; AC CMRR is 70dB. NPLC Line Sync CMRR 5 On 140 dB 1 On 140 dB 0.2 On 120 dB 0.2 Off 80 dB Resistance ENHANCED ACCURACY (within 30 days of autocalibration, TOPER 5C from TACAL) 10 Accuracy (ppm of reading + ppm of range) Range 11 1.0000000 10.000000 100.00000 1.0000000 k 10.000000 k 15 100.00000 k 15, 16 1.0000000 M 15, 17 10.000000 M 18 100.00000 M 18 1.0000000 G 18 Resolution 0.1 1 10 100 1 m 10 m 100 m 1 10 100 Test Current 12 (5%) 10 mA 10 mA 1 mA 1 mA 100 A 10 A 10 A 0.69 A || 10 M 0.69 A || 10 M 0.69 A || 10 M 24 Hour TCAL 1C 13 15 + 50 15 + 5 12 + 4 12 + 3 13 + 3 13 + 3 14 + 3 150 + 6 800 + 30 9000 + 100 Resolution 0.1 1 10 100 1 m 10 m 100 m 1 10 100 Test Current 21 (5%) 10 mA 1 mA 1 mA 1 mA 100 A 10 A 10 A 0.69 A || 10 M 0.69 A || 10 M 0.69 A || 10 M 24 Hour TCAL 1C 22 15 + 50 15 + 5 12 + 4 12 + 3 10 + 3 13 + 3 14 + 3 150 + 6 800 + 30 9000 + 200 90 Day TCAL 5C 30 + 50 30 + 5 27 + 4 24 + 3 30 + 3 30 + 3 30 + 4 200 + 10 2000 + 30 9000 + 100 1 Year TCAL 5C 30 + 50 30 + 5 27 + 4 24 + 3 30 + 3 30 + 3 30 + 4 200 + 10 2000 + 30 9000 + 100 2 Year TCAL 5C 30 + 50 30 + 5 27 + 4 24 + 3 30 + 3 30 + 3 30 + 4 200 + 10 2000 + 30 9000 + 100 Temperature Coefficient 14 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 70 + 1 385 + 1 3000 + 1 Model DMM7510 condensed specifications COMMON MODE REJECTION ACCURACY 19 Accuracy (ppm of reading + ppm of range) Range 20 1 10 100 1 k 10 k 24 100 k 24, 25 1 M 24, 26 10 M 27 100 M 27 1 G 27 90 Day TCAL 5C 40 + 50 40 + 5 35 + 4 30 + 3 30 + 3 38 + 3 38 + 5 200 + 10 2000 + 30 9000 + 200 10. Specifications are for 4-wire resistance, offset compensation on for 10k measurements, and offset compensation off for 10k measurements. 1 range is 4-wire only. For 2-wire, with Rel, add 50m to ppm of range uncertainty. Without Rel and with Model 1756 test leads, add 100m to ppm of range uncertainty. 11. 20% overrange on all ranges. 12. Test current with offset compensation off, 5%. 13. Relative to calibration accuracy. 14. Add per degree from TCAL 5C. 15. Specifications are for external cable and load capacitance <1nF. 16. For offset compensation on, add 10ppm uncertainty to ppm of reading. 17. For 4-wire 1M, open lead detector on, add 10 ppm uncertainty to ppm of reading. 18. Specified for <10% lead resistance mismatch in HI and LO. 1 Year TCAL 5C 50 + 50 50 + 5 47 + 4 41 + 3 42 + 3 50 + 3 50 + 5 400 + 10 2000 + 30 13000 + 200 2 Year TCAL 5C 70 + 50 70 + 5 65 + 4 65 + 3 65 + 3 65 + 3 65 + 5 600 + 12 2600 + 30 14000 + 200 Temperature Coefficient 23 2.5 + 5 2.5 + 0.5 5 + 0.25 5 + 0.25 2.5 + 0.25 5+1 5+1 70 + 1 385 + 1 3000 + 1 19. Specifications are for 4-wire resistance, offset compensation on for 10k measurements, and offset compensation off for 10k measurements. 1 range is 4-wire only. For 2-wire, with Rel, add 50m to ppm of range uncertainty. Without Rel and with Model 1756 test leads, add 100m to ppm of range uncertainty. 20. 20% overrange on all ranges. 21. Test current with offset compensation off. 22. Relative to calibration accuracy. 23. Add per degree from TCAL 5C. 24. Specifications are for external cable and load capacitance <1nF. 25. For offset compensation on, add 10ppm of uncertainty to ppm of reading. 26. For 4-wire, 1M, open lead detection on, add 10ppm uncertainty to ppm of reading. 27. Specified for <10% lead resistance mismatch in HI and LO. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company DIGITAL MULTIMETERS & SYSTEMS Model DMM7510 condensed specifications DMM7510 71/2-Digit Graphical Sampling Multimeter DMM7510 RESISTANCE OPEN CIRCUIT DC VOLTAGE 28 2-WIRE OHMS Range 20 1 10 100 , 1 k 10 k 100 k, 1 M 10 M to 1 G 2-wire - 9.2 V 14.0 V 9.5 V 12.7 V 6.9 V Offset compensation on 4-wire 9.5 V 9.5 V 14.3 V 0.0 V 0.0 V (100 k range only) - * Applies to ppm of range. * Peak noise uncertainty is included in DC specifications for 1 PLC. * Add peak noise uncertainty to measurements for <1 PLC. EXAMPLES * 10k at 0.006 PLC: 3 (from Accuracy table) + 5 (50m with Rel ) + 43 (additional peak noise uncertainty) = 51 ppm of range. * 10k at 1 PLC: 3 + 5 + 0 = 8 ppm of range. NPLC 5 1 0.2 30 0.2 0.06 0.006 0.0005 4-WIRE OHMS (10k) Offset Compensation On RMS NOISE (additional peak noise uncertainty) 29 * Applies to ppm of range. * Peak noise uncertainty is included in DC specifications for 1 PLC. * Add peak noise uncertainty to measurements for <1 PLC. EXAMPLES * 1k at 0.006 PLC: 3 (from Accuracy table) + 26 (additional peak noise uncertainty) = 29 ppm of range. * 1k at 1 PLC: 3 + 0 = 3 ppm of range. NPLC 5 1 0.2 30 0.2 0.06 0.006 0.0005 Digits 71/2 71/2 61/2 61/2 51/2 41/2 31/2 1 2.8 4.2 30 (160) 50 (250) 110 (490) 110 (710) 520 (3420) 10 0.3 0.4 3 (13) 5 (22) 11 (47) 10 (70) 50 (340) 100 0.3 0.4 3 (13) 5 (22) 11 (46) 10 (70) 50 (340) 1 k 0.07 0.12 0.4 (2.6) 0.6 (3.2) 1.1 (6.6) 4 (26) 40 (220) 10 k 0.3 0.5 1.2 (8.2) 1.2 (8.3) 2 (16) 10 (60) 50 (300) Dry Circuit Resistance Digits 71/2 71/2 61/2 61/2 51/2 41/2 31/2 10 1.1 0.6 2 (17) 2 (17) 3 (22) 6 (50) 30 (300) 100 0.8 0.6 2 (10) 2 (14) 3 (19) 6 (50) 30 (230) 1 k 0.1 0.09 0.2 (1.5) 0.3 (1.6) 0.4 (3.7) 3 (21) 20 (150) 10 k 0.2 0.4 0.8 (6.3) 0.8 (6.4) 2 (12) 6 (43) 30 (210) RESISTANCE CHARACTERISTICS MAXIMUM 4-WIRE OHMS LEAD RESISTANCE: 5 per lead for 1 range, 10% of range per lead for 10 to 1k ranges; 1k per lead for all other ranges. OFFSET COMPENSATION: Selectable on 4-wire, 1 to 100k ranges. OPEN LEAD DETECTOR: Default is off. AUTOZERO OFF ERROR: For 2-wire ohms, 1C and 10 minutes, add (8ppm of reading) + 1.5m for 10, 15m for 100 and 1k ranges, 150m for 10k range, 1.5 for 100 k range, and 15 for all other ranges. For 4-wire ohms, 1C and 10 minutes, add (8ppm of reading). INPUT CURRENT LIMIT: For signals with a magnitude of +12V to +40V or -12V to -40V: 13mA source or sink, typical. For signals with a magnitude of greater than +40V or -40V: 130A source or sink, typical. ENHANCED ACCURACY (within 30 days of autocalibration, TOPER 5C from TACAL) Accuracy (ppm of reading + ppm of range) Range 31 1.000000 10.00000 100.0000 1.000000 k 10.00000 k Resolution 1 10 100 1 m 10 m Test Current 35 (5%) 10 mA 1 mA 100 A 10 A 5 A Open Circuit DUT Voltage 32 25 mV 25 mV 25 mV 25 mV 25 mV 24 Hour TCAL 1C 33 25 + 80 25 + 80 25 + 80 25 + 80 25 + 80 90 Day TCAL 5C 50 + 80 50 + 80 90 + 80 180 + 80 320 + 80 Resolution 1 10 100 1 m 10 m Test Current 35 (5%) 10 mA 1 mA 100 A 10 A 5 A Open Circuit DUT Voltage 32 25 mV 25 mV 25 mV 25 mV 25 mV 24 Hour TCAL 1C 33 25 + 80 25 + 80 25 + 80 25 + 80 25 + 80 90 Day TCAL C 50 + 80 50 + 80 90 + 80 180 + 80 320 + 80 1 Year TCAL 5C 50 + 80 50 + 80 90 + 80 180 + 80 320 + 80 2 years TCAL 5C 50 + 80 50 + 80 90 + 80 180 + 80 320 + 80 Temperature Coefficient 34 1.5 + 0.1 1.5 + 0.1 1.5 + 0.1 1.5 + 0.1 1.5 + 0.1 DIGITAL MULTIMETERS & SYSTEMS ACCURACY Accuracy (ppm of reading + ppm of range) Range 31 1.000000 10.00000 100.0000 1.000000 k 10.00000 k 28. Open circuit voltage is typical, measured from input HI to LO, SHI and SLO open. For 1 to 1M ranges using an external digital multimeter (DMM) set to 10M input impedance; for 10M to 1G ranges, set external DMM to >10G input impedance. 29. Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is guaranteed. 30. With line sync on. 31. 32. 33. 34. 35. 1 Year TCAL 5C 70 + 80 70 + 80 140 + 80 400 + 80 800 + 80 2 Year TCAL 5C 90 + 80 90 + 80 200 + 80 600 + 80 1300 + 80 Temperature Coefficient 34 2.5 + 1 5+1 2.5 + 1 5+1 8+1 20% overrange on all ranges, except 2.4k for the 10k range. Maximum clamp voltages are DC, typical accuracy is 20%. Add 20% for offset compensation on. Relative to calibration accuracy. Add per degree from TCAL 5C. Test current with offset compensation off. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company Model DMM7510 condensed specifications Model DMM7510 condensed specifications RMS NOISE (additional peak noise uncertainty) 29 Offset compensation off 4-wire 9.2 V 9.2 V 14.2 V 9.5 V 14.3 V 6.9 V 71/2-Digit Graphical Sampling Multimeter NPLC 5 1 0.2 37 0.2 0.06 0.006 0.0005 Digits 71/2 71/2 61/2 61/2 51/2 41/2 31/2 1 10 9 30 (130) 60 (220) 70 (350) 130 (750) 520 (3550) 10 11 9 30 (120) 60 (230) 70 (350) 120 (830) 530 (3520) 100 6 7 30 (120) 50 (190) 50 (290) 110 (700) 530 (3380) 1 k 5 7 30 (120) 50 (190) 50 (280) 100 (690) 500 (3370) DRY CIRCUIT RESISTANCE CHARACTERISTICS MAXIMUM 4-WIRE OHMS LEAD RESISTANCE: 0.5 per lead for 1 range. 10% of range per lead for 10 to 100 ranges. 50 per lead for 1k to 10k ranges. INPUT CURRENT LIMIT: For signals greater than 20mV, current limited, 13mA typical. OFFSET COMPENSATION: Selectable on 1 to 10k ranges. AUTOZERO OFF ERROR: For 1C and 10 minutes, add 8 ppm of reading. 10 k 0.9 0.8 3 (16) 9 (35) 20 (90) 20 (110) 100 (670) DC Current ENHANCED ACCURACY (within 30 days of autocalibration, TOPER 5C from TACAL) Accuracy (ppm of reading + ppm of range) Range 38 10.000000A 100.00000 A 1.0000000 mA 10.000000 mA 100.00000 mA 1.0000000 A 3.000000 A 10.000000 A 41 Resolution 1 pA 10 pA 100 pA 1 nA 10 nA 100 nA 1 A 1 A Maximum Burden Voltage 15 mV 15 mV 15 mV 20 mV 200 mV 400 mV 1300 mV 650 mV 24 Hour TCAL 1C 39 30 + 30 20 + 5 30 + 5 40 + 5 50 + 18 150 + 50 200 + 40 700 + 275 90 Day TCAL 5C 75 + 30 60 + 9 60 + 9 60 + 9 150 + 30 400 + 50 400 + 40 800 + 275 1 Year TCAL 5C 75 + 30 60 + 9 60 + 9 60 + 9 150 + 30 400 + 50 400 + 40 1500 + 275 2 Year TCAL 5C 75 + 30 60 + 9 60 + 9 60 + 9 150 + 30 400 + 50 400 + 40 2000 + 275 Temperature Coefficient 40 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 0.15 + 0.1 50 + 10 Model DMM7510 condensed specifications RMS NOISE (additional peak noise uncertainty) 36 * Applies to ppm of range. * Peak noise uncertainty is included in DC specifications for 1 PLC. * Add peak noise uncertainty to measurements when < 1 PLC. EXAMPLES: * 10 at 0.2 PLC: 80 (from Accuracy table) + 230 (additional peak noise uncertainty) = 310 ppm of range. * 10 at 1 PLC: 80 + 0 = 80 ppm of range. ACCURACY Range 38 10.000000A 100.00000 A 1.0000000 mA 10.000000 mA 100.00000 mA 1.0000000 A 3.000000 A 10.000000 A 41 36. 37. 38. 39. 40. 41. Resolution 1 pA 10 pA 100 pA 1 nA 10 nA 100 nA 1 A 1 A Maximum Burden Voltage 15 mV 15 mV 15 mV 20 mV 200 mV 400 mV 1300 mV 650 mV Accuracy (ppm of reading + ppm of range) 24 Hour TCAL 1C 39 30 + 30 20 + 5 30 + 5 40 + 5 50 + 18 150 + 50 200 + 40 700 + 275 90 Day TCAL 5C 100 + 30 75 + 12 75 + 12 75 + 12 300 + 30 400 + 50 400 + 40 800 + 275 1 Year TCAL 5C 125 + 40 100 + 15 100 + 15 100 + 15 400 + 30 450 + 50 450 + 40 1500 + 275 2 Year TCAL 5C 175 + 50 150 + 20 150 + 20 150 + 20 500 + 30 500 + 50 500 + 40 2000 + 275 Temperature Coefficient 40 10 + 8 10 + 3 10 + 3 10 + 3 50 + 5 10 + 10 10 + 10 50 + 10 DIGITAL MULTIMETERS & SYSTEMS Model DMM7510 condensed specifications DMM7510 Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is guaranteed. With line sync on. 20% overrange supported for all ranges except for 3A and 10A, which are 1% supported. Relative to calibration accuracy. Add per degree from TCAL 5C. Rear input terminals only. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company 71/2-Digit Graphical Sampling Multimeter DMM7510 NPLC Digits 5 71/2 1 71/2 0.2 61/2 44 61/2 0.2 0.06 51/2 0.006 41/2 0.0005 31/2 10 A 0.15 0.4 0 (220) 120 (260) 130 (280) 130 (350) 260 (2110) 100 A 0.14 0.13 0 (23) 12 (26) 12 (29) 14 (42) 30 (300) 1 mA 0.09 0.1 0.2 (3.4) 1.2 (3.8) 1.3 (5.6) 3 (20) 20 (150) 10 mA 0.1 0.1 0.2 (1.6) 0.3 (1.8) 0.4 (3.9) 2 (20) 20 (160) 100 mA 0.3 0.5 2 (10) 1.9 (9.8) 2 (14) 4 (30) 30 (190) 1A 0.3 0.5 2 (11) 2 (10) 2 (14) 4 (31) 30 (190) 3A 0.2 0.3 0.7 (4.6) 0.8 (5) 1.2 (7.7) 7 (51) 70 (510) 10A 43 0.8 1.2 4 (32) 8 (37) 10 (59) 20 (110) 60 (420) DC CURRENT CHARACTERISTICS Range Effective Internal Shunt Value 45 Autozero Off Error: For 1C and 10 minutes add (8 ppm of reading + range error) Overload Recovery: For each additional sustained amp beyond 1.5A, add the following initial ppm of range error until thermally settled after overload recovery 10 A 1 k 100 A 100 1 mA 10 10 mA 1 100 mA 0.1 1A 0.1 3A 0.1 10 A 43 0.005 150 pA 1.5 nA 15 nA 150 nA 15 A 150 A 150 A 3 mA 15500 1800 150 150 6500 200 -- -- Temperature 4-WIRE RTD OR 3-WIRE RTD THERMOCOUPLE TYPES: 100 platinum PT100, D100, F100, PT385, PT3916; or user-configurable 0 to 10k. TYPES: B, E, J, K, N, R, S, T Accuracy C Type 4-Wire RTD 3-Wire RTD 47 Range -200 to 850 C -200 to 850 C Resolution 0.01 C 0.01 C 2 Year TCAL 5C 0.06 C 0.75 C Resolution 0.01 C 2 Year TCAL 5C 0.08 C Temperature Coefficient 46 0.003 C/C 0.003 C/C THERMISTOR TYPES: 2.252k, 5k, and 10k. Accuracy C Type Thermistor DIGITAL MULTIMETERS & SYSTEMS Model DMM7510 condensed specifications Model DMM7510 condensed specifications RMS NOISE (additional peak noise uncertainty) 42 * Applies to ppm of range. * Peak noise uncertainty is included in DC specifications for 1 PLC. * Add peak noise uncertainty to measurements when <1 PLC. EXAMPLES: * 1mA at 0.006 PLC: 9 (from Accuracy table) + 20 (additional peak noise uncertainty) = 29 ppm of range. * 1mA at 1 PLC: 9 +0 = 9 ppm of range. Range -80 to +150 C Type B E J K N R S T Range 350 to +1820 C -200 to +1000 C -200 to +760 C -200 to +1372 C -200 to +1300 C 0 to +1768 C 0 to +1768 C -100 to +400 C Resolution 0.1 C 0.001 C 0.001 C 0.001 C 0.001 C 0.1 C 0.1 C 0.001 C Accuracy C 2 Year, TCAL 5C 48 Simulated Reference Temperature Junction Coefficient 46 0.6 C 0.03 C/C 0.2 C 0.03 C/C 0.2 C 0.03 C/C 0.2 C 0.03 C/C 0.2 C 0.03 C/C 0.6 C 0.03 C/C 0.6 C 0.03 C/C 0.2 C 0.03 C/C Temperature Coefficient 46 0.002 C/C 42. Noise values are based on 1000 readings with autozero on and AMPS terminal open. RMS noise is typical. Additional peak noise is guaranteed. 43. Rear input terminals only. 44. With line sync on. 45. Values are typical and guaranteed by design. 46. Add per degree from TCAL 5C; specifications without autocalibration. 47. For 3-wire RTD, accuracy is for <0.1 lead resistance mismatch for input HI and LO. Add 0.25C/0.1 of HI-LO lead resistance mismatch. 48. Exclusive of cold-junction errors. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company 71/2-Digit Graphical Sampling Multimeter DMM7510 Resolution 100 m Test Current 1 mA Open Circuit Voltage 14.0 V Accuracy (ppm of reading + ppm of range) 2 Year TCAL 5C 100 + 100 Temperature Coefficient 50 2.5 + 1 Accuracies specified for additional cable and stray capacitance properly zeroed with the Rel function. ACCURACY Range 51 1.0000 nF 10.000 nF 100.00 nF 1.0000 F 10.000 F 100.00 F 1000.0 F CONTINUITY CHARACTERISTICS CONTINUITY HIGH LIMIT: User-selectable; default 10. Charge Resolution Current 52, 53 0.1 pF 1.1 A 1 pF 1.1 A 10 pF 10 A 0.1 nF 100 A 1 nF 100 A 10 nF 1 mA 0.1 F 10 mA Accuracy Maximum (% of reading + % of range) 2 years Temperature Circuit TCAL 5C Coefficient 50 Voltage 2.8 V 1 + 0.2 0.15 + 0.05 2.8 V 1 + 0.1 0.15 + 0.01 3V 0.4 + 0.1 0.01 + 0.01 3V 0.4 + 0.1 0.01 + 0.01 3V 0.4 + 0.1 0.01 + 0.01 3V 0.4 + 0.1 0.01 + 0.01 3V 0.5 + 0.1 0.01 + 0.01 Diode Voltage Measure Range 51 10.000000 V Accuracy (ppm of reading + ppm of range) Resolution 1 V Bias Level (Selectable) 10 A / 100 A / 1 mA 90 Day TCAL 5C 20 + 5 1 Year TCAL 5C 30 + 5 2 Year TCAL 5C 45 + 5 Temperature Coefficient 50 2.5 + 1 Digitize Voltage ACCURACY (Input Impedance AUTO) Accuracy (ppm of reading + ppm of range) Range 54, 55 100.000 mV 1.00000 V 10.0000 V 100.000 V 59 1000.00 V 60 Resolution 56 1 V 10 V 0.1 mV 1 mV 10 mV Input Impedance 57 >10 G or 10 M 1% >10 G or 10 M 1% >10 G or 10 M 1% 10 M 1% 10 M 1% 49. 50. 51. 52. 53. 54. 90 Day TCAL 5C 210 + 100 110 + 75 110 + 75 110 + 75 110 + 75 1 Year TCAL 5C 220 + 100 120 + 75 120 + 75 120 + 75 120 + 75 Specifications exclude lead resistance. Add per degree from TCAL 5C; specifications without autocalibration. 20% overrange on all ranges. Charging current values are typical, guaranteed by design. Discharge current limited to <13mA. For DC coupling, 20% overrange for 100mV to 100V. For AC coupling, 500% overrange 100mV to 100V. 1% for 1000V range DC and AC coupling. 55. Accuracy with sample rate 1k per second, aperture auto, and 100 reading buffer average. 2 Year TCAL 5C 230 + 100 130 + 75 130 + 75 130 + 75 130 + 75 Temperature Coefficient 58 15 + 20 15 + 20 10 + 20 15 + 20 10 + 20 56. 57. 58. 59. Power up default is 41/2 digits. User-selectable. Add per degree from TCAL 5%. For 100V range, input impedance auto and without ACAL, add 100ppm of range additional uncertainty and 15ppm/C additional uncertainty for "of range" temperature coefficient for operation outside of TCAL 5C. 60. For signal levels greater than 500V, add 0.02 ppm/V to the ppm of the readings specification for measurements exceeding 500V. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company DIGITAL MULTIMETERS & SYSTEMS Model DMM7510 condensed specifications Range 49 1.0000 k Capacitance Model DMM7510 condensed specifications Continuity 71/2-Digit Graphical Sampling Multimeter DMM7510 SIGNAL CHARACTERISTICS 61, 62, 63 Range 100.000 mV 1.00000 V 10.0000 V Analog Bandwidth (-3dB) 600 kHz 600 kHz 600 kHz Maximum Flatness Error 3 Hz to 20 kHz 64 0.015 dB 0.01 dB 0.01 dB THD 20 kHz Signal (-1dB FS) 65 0.04 % 0.03 % 0.01 % DC-coupled Settling Time (0.5%) 5 s 6 s 4 s AC-coupled Filter FAST Settling Time (0.5%) 80 ms 80 ms 80 ms AC-coupled Filter SLOW Settling Time (0.5%) 2.3 s 2.5 s 2.5 s AC Coupling Low Frequency (-3dB) point 66 1 Hz 1 Hz 1 Hz TYPICAL DC COUPLED Analog Bandwidth (-3dB) 20 kHz 67 20 kHz Range 100.000 V 1000.00 V Maximum Flatness Error 3 Hz to 1 kHz 64 0.1 dB 0.1 dB Total Harmonic Distortion (THD) 1 kHz Signal (-1dB FS) 65 1.3 % 1.8 % Settling Time (0.5%) 160 s 80 s TYPICAL AC COUPLED Analog Bandwidth (-3dB) 600 kHz 600 kHz Range 100.000 V 1000.00 V Maximum Flatness Error 3 Hz to 20 kHz 64 0.1 dB 0.1 dB Filter FAST Settling Time (0.5%) 80 ms 80 ms Filter SLOW Settling Time (0.5%) 2.3 s 2.3 s DC-COUPLED ADDITIONAL NOISE UNCERTAINTY, TYPICAL 68 Low Frequency Coupling Point 66 (-3dB) 1 Hz 1 Hz DC-COUPLED TOTAL HARMONIC DISTORTION (THD), TYPICAL 69 Total Harmonic Distortion, Digitize Voltage RMS Noise, Digitize Voltage 1000 -50 -60 -70 100mV 10 1V 10V dB of Range ppm of Range 100 100mV -80 1V -90 10V 100V 100V 1000V 1 -100 1000V -110 0.1 1k 2k 5k 10k 20k 50k 100k 200k 500k 1M DIGITAL MULTIMETERS & SYSTEMS Samples per Second 61. 62. 63. 64. Accuracy with sample rate 1M per second and aperture 1s. Verified with sine wave input and DC content 3% of range. For AC coupling, maximum crest factor of 5. For DC coupled, 0dB reference frequency is 3Hz. For AC coupled, 0dB reference frequency is 1kHz. For AC coupled operation below 1kHz, add 0.1dB. 65. Exclusive of source input noise. 66. With AC coupling frequency = 3Hz and AC coupling filter = Slow. -120 1k 2k 5k 10k 20k 50k 100k 200k Samples per Second 67. For input impedance auto, bandwidth is 6kHz. 68. Specified with aperture auto and 4-wire short on input terminals. For 100V range, input impedance 10M, multiply by 2.5. For all ranges and sample rate >1k, add an additional 3x RMS noise uncertainty to ppm of range. 69. Specified with aperture Auto, 100 Hz sine wave for sample rate 5 k, and 1 kHz sine wave for sample rate 10 k. Distortion is calculated using first five harmonics. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company Model DMM7510 condensed specifications Model DMM7510 condensed specifications TYPICAL AC AND DC COUPLED 71/2-Digit Graphical Sampling Multimeter DMM7510 DC-COUPLED EFFECTIVE NUMBER OF BITS (ENOB), TYPICAL 70 Effective Number of Bits (ENOB), Digitize Voltage 18 Model DMM7510 condensed specifications 14 Bits 100mV 1V 12 10V 10 8 1k 2k 5k 10k 20k 50k 100k 200k Samples per Second Digitize Current DC ACCURACY 71 Accuracy (ppm of reading + ppm of range) Range 72 10.0000 A 100.000 A 1.00000 mA 10.0000 mA 100.000 mA 1.00000 A 3.00000 A 10.0000 A 75 Resolution 73 0.1 nA 1 nA 10 nA 100 nA 1 A 10 A 100 A 100 A Burden Voltage 15 mV 15 mV 15 mV 20 mV 200 mV 400 mV 1300 mV 650 mV 90 Day TCAL 5C 150 + 75 150 + 75 150 + 75 150 + 75 340 + 100 400 + 110 650 + 150 950 + 350 1 Year TCAL 5C 160 + 75 160 + 75 160 + 75 160 + 75 450 + 100 500 + 110 900 + 150 1500 + 350 2 Year TCAL 5C 170 + 75 170 + 75 170 + 75 170 + 75 560 + 100 600 + 110 900 + 150 2000 + 350 Temperature Coefficient 74 30 + 15 30 + 15 30 + 15 30 + 15 50 + 20 50 + 25 50 + 25 50 + 25 SIGNAL CHARACTERISTICS, TYPICAL 76 Range 72 10.0000 A 100.000 A 1.00000 mA 10.0000 mA 100.000 mA 1.00000 A 77 3.0000 A 77 10.0000 A 75, 77, 78 Maximum Flatness Error 3 Hz to 20 kHz 0.15 dB 0.15 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB Analog Bandwidth (-3dB) 100 kHz 100 kHz 100 kHz 100 kHz 100 kHz 100 kHz 100 kHz 100 kHz Total Harmonic Distortion (THD) 20 kHz Signal (-1dB FS) 0.02 % 0.01 % 0.01 % 0.01 % 0.02 % 0.02 % 0.02 % 0.02 % 70. Specified with aperture Auto, 100Hz sine wave for sample rate 5k, and 1kHz sine wave for sample rate 10k. For the 100V and 1000V ranges, use the 1V and 10V range ENOB, respectively; guaranteed by design. 71. Accuracy with sample rate 1k per second, aperture auto, and 100 reading buffer average. 72. 20% overrange on all ranges except 3.3% for 3A and 10A ranges. 73. Power up default is 41/2 digits. DC-coupled Settling Time (0.5%) 8 s 7 s 3 s 8 s 5 s 6 s 6 s 6 s 74. 75. 76. 77. 78. Add per degree from TCAL 5C. Rear input terminals only. Verified with sine wave input and DC content 3 % of range. 0 dB reference frequency is 3 Hz. 10A range is available only on the rear input terminals. 10A flatness verified to 10kHz; 100kHz guaranteed by design. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company DIGITAL MULTIMETERS & SYSTEMS Model DMM7510 condensed specifications 16 71/2-Digit Graphical Sampling Multimeter DMM7510 ADDITIONAL NOISE UNCERTAINTY, TYPICAL 79 TOTAL HARMONIC DISTORTION (THD), TYPICAL 80 RMS Noise, Digitize Current Total Harmonic Distortion, Digitize Current 1000 -70 10A 100A 1mA 10mA 10A 100A 1mA -100 100mA, 1A 10 3A 10mA 100mA -110 10A 1 -90 -120 1k 2k 5k 1k 10k 20k 50k 100k 200k 500k 1M 2k 5k 10k 20k 50k 100k 200k Samples per Second Samples per Second EFFECTIVE NUMBER OF BITS (ENOB), TYPICAL 81 Effective Number of Bits (ENOB), Digitize Current 18 16 14 Bits 10uA 100uA 1mA 12 10mA 100mA 10 8 1k 2k 5k 10k 20k 50k 100k 200k DIGITAL MULTIMETERS & SYSTEMS Samples per Second Digitizer Characteristics MAXIMUM RESOLUTION: 18 bits. MEASUREMENT INPUT COUPLING: DC or AC (voltage only). SAMPLING RATE 82: Programmable 1k through 1 million. VOLATILE SAMPLE MEMORY WITH TIMESTAMP: 27.5 million . MINIMUM RECORD TIME: 1s. TIMESTAMP RESOLUTION: 1ns with standard or full buffer style. 1s with compact buffer style. TIMESTAMP ACCURACY: With standard or full buffer style, 20ns between adjacent readings, with total buffer time <2s. With compact buffer style, 2s adjacent readings, with total buffer buffer time <2s. MAXIMUM RECORD LENGTH: 8 million. 79. Specified with aperture Auto and open input terminals. For all ranges and for 1k sample rate, add an additional 3x RMS noise uncertainty to ppm of range. 80. Specified with aperture Auto, 100 Hz sine wave for sample rate 5 k, and 1 kHz sine wave for sample rate 10 k. Distortion is calculated using first five harmonics. For the 1 A, 3 A, and 10 A ranges, use the 100 mA range accuracy; guaranteed by design. 81. Specified with aperture Auto, 100Hz sine wave for sample rate 5k, and 1kHz sine wave for sample rate 10k. For the 1A, 3A, and 10A ranges, use the 100mA ENOB; guaranteed by design. 82. Sample rate is not continuously adjustable. For valid discrete settings, see the Model DMM7510 Reference Manual. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company Model DMM7510 condensed specifications 100 dB of Range ppm of Range Model DMM7510 condensed specifications -80 71/2-Digit Graphical Sampling Multimeter DMM7510 True RMS AC Voltage and AC Current Temperature Coefficient/C (all ranges) Resolution 0.1 V 1 V 10 V 100 V 1 mV 3 Hz to 5 Hz 1.0 + 0.03 1.0 + 0.03 1.0 + 0.03 1.0 + 0.03 1.0 + 0.03 5 Hz to 10 Hz 0.30 + 0.03 0.30 + 0.03 0.30 + 0.03 0.30 + 0.03 0.30 + 0.03 10 Hz to 20 kHz 0.06 + 0.03 0.06 + 0.03 0.06 + 0.03 0.06 + 0.03 0.06 + 0.03 20 kHz to 50 kHz 0.14 + 0.05 0.14 + 0.05 0.14 + 0.05 0.14 + 0.05 0.14 + 0.05 50 kHz to 100 kHz 0.6 + 0.08 0.6 + 0.08 0.6 + 0.08 0.6 + 0.08 0.6 + 0.08 100 kHz to 300 kHz 4.0 + 0.5 4.0 + 0.5 4.0 + 0.5 4.0 + 0.5 4.0 + 0.5 - - 0.01 + 0.003 0.03 + 0.003 0.005 + 0.003 0.006 + 0.005 0.01 + 0.006 0.03 + 0.01 Range 83 1.000000 mA 10.00000 mA 100.0000 mA 1.000000 A 3.000000 A 10.00000 A 85 Resolution 1 nA 10 nA 100 nA 1 A 1 A 10 A 3 Hz to 5 Hz 1.0 + 0.04 1.0 + 0.04 1.0 + 0.04 1.0 + 0.04 1.0 + 0.05 1.0 + 0.05 5 Hz to 10 Hz 0.30 + 0.04 0.30 + 0.04 0.30 + 0.04 0.30 + 0.04 0.30 + 0.05 0.40 + 0.05 10 Hz to 2 kHz 0.08 + 0.03 0.08 + 0.03 0.08 + 0.03 0.20 + 0.04 0.20 + 0.05 0.40 + 0.05 2 kHz to 5 kHz 0.09 + 0.03 0.09 + 0.03 0.09 + 0.03 0.88 + 0.04 0. 88 + 0.05 0. 88 + 0.05 5 kHz to 10 kHz 0.09 + 0.03 0.09 + 0.03 0.09 + 0.03 2.0 + 0.04 2.0 + 0.05 2.0 + 0.05 - - 0.10 + 0.004 0.030 + 0.004 0.005 + 0.003 0.006 + 0.005 0.006 + 0.005 1-Year Accuracy: (% of reading + % of range) TCAL 5C Function Current 84 Temperature Coefficient/C (all ranges) ADDITIONAL AC UNCERTAINTIES - LOW FREQUENCY UNCERTAINTY Additional Uncertainty (% of reading), Lower Frequency Uncertainty 20 Hz to 30 Hz 30 Hz to 50 Hz 50 Hz to 100 Hz 100 Hz to 200 Hz 200 Hz to 300 Hz 300 Hz to 500 Hz > 500 Hz 3 BW (3 Hz to 300 kHz) 0 0 0 0 0 0 0 Detector Bandwidth (BW) 30 BW 300 BW (30 Hz to 300 kHz) (300 Hz to 300 kHz) 0.3 -- 0 -- 0 4.0 0 0.72 0 0.18 0 0.07 0 0 ADDITIONAL AC VOLTAGE CREST FACTOR UNCERTAINTIES 86 ADDITIONAL UNCERTAINTY: (% of reading). Input Signal Frequency 3 Hz to 5 Hz 5 Hz to 10 Hz 10 Hz to 30 Hz 5 Hz to 100 Hz 100 Hz to 300 Hz 100 Hz to 300 Hz 500 Hz to 10 kHz Detector Bandwidth 3 Hz 3 Hz 3 Hz 30 Hz 30 Hz 300 Hz 300 Hz Maximum Crest Factor: 5 at Range Full Scale 1 to 2 2 to 3 3 to 4 4 to 5 1.00 4.00 4.80 5.00 0.50 1.20 1.30 1.40 0.20 0.30 0.60 0.90 0.20 0.30 0.60 0.90 0.05 0.15 0.30 0.40 0.50 1.20 1.30 1.50 0.05 0.15 0.30 1.20 83. 20% overrange on AC functions except 1% on 700V, 3.33% on 3A, and 1% on 10A. Default resolution is 61/2 digits. 84. Specifications are for detector bandwidth of 3Hz and sine wave inputs >5% of range. Detector bandwidth of 3Hz and 30Hz are multisample A/D conversions. Detector bandwidth of 300Hz is a single A/D conversion, programmable from 0.0005 PLC to 15 PLC (60Hz), 12 PLC (50Hz). Default condition set to 1 PLC. 85. Rear terminals only. 86. Applies for non-sine wave inputs, DC content 3% of range, maximum crest factor 5.0. For bandwidth 30Hz, autozero off, 61/2 digits at 1 PLC, 31/2 digits at 0.0005 PLC. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company DIGITAL MULTIMETERS & SYSTEMS Model DMM7510 condensed specifications Voltage 84 Range 83 100.0000 mV 1.000000 V 10.00000 V 100.0000 V 700.000 V Model DMM7510 condensed specifications 1-Year Accuracy: (% of reading + % of range) TCAL 5C Function Model DMM7510 condensed specifications AC VOLTAGE CHARACTERISTICS AC CURRENT CHARACTERISTICS MEASUREMENT METHOD: AC-coupled, true RMS. INPUT IMPEDANCE: 1M 2%|| <150pF. VOLT*HERTZ PRODUCT: <2.1 x 107 V*Hz verified; input frequency verified for <300kHz. MEASUREMENT METHOD: AC-coupled, true RMS. 1 mA <16 mV 10 mA <20 mV 100 mA <0.2 V 1A <0.4 V 3A <1.3 V 10 A 87 <0.65 V 0.006 0.006 0.12 0.05 -- -- Frequency and Period MEASUREMENT ACCURACY 88 Aperture 250 ms 100 ms 10 ms Measurement Resolution 0.1 ppm 0.1 ppm 0.1 ppm THRESHOLD LEVEL ACCURACY 89 Accuracy (ppm of reading + ppm of aperture time) Frequency: 3 Hz to 500 kHz Period: 333 ms to 2 s 1 Year, TCAL 5C 80 + 0.333 80 + 3.33 80 + 33.3 2 Year, TCAL 5C 160 + 0.333 160 + 3.33 160 + 33.3 Threshold Range 100 mV to 700 V Threshold Resolution 0.05% Accuracy (% of reading) 2 Year, TCAL 5C 1.0% FREQUENCY AND PERIOD CHARACTERISTICS MEASUREMENT METHOD: Reciprocal counting technique. APERTURE: 10ms to 273ms; default is 10ms. TYPICAL READING RATES, 60Hz (50Hz) OPERATION 90, 91, 92, 93 NPLC 1 0.2 0.06 0.006 0.0005 Digits 71/2 61/2 51/2 41/2 31/2 Functions: DC Voltage (10 V), 2-wire Ohms (10k), DC Current (1 mA) Measurements Measurements Into Buffer Into Computer 59.8 (49.8) 58 (48) 295 (240) 250 (210) 965 (810) 950 (800) 7500 (6700) 7300 (6500) 26000 (26000) 24000 (24000) Detector Bandwidth (Hz) 3 30 300 94 300 94 DIGITAL MULTIMETERS & SYSTEMS Range Burden Voltage (RMS) Overload Recovery: For each additional sustained ampere beyond 1.5A, add the following initial % of range error until thermally settled after overload recovery Digits 61/2 61/2 61/2 31/2 Functions: 4-wire ohms (1 k), 4-wire/3-wire RTD Measurements Measurements Into Buffer Into Computer 29 (24) 28 (24) 128 (109) 119 (100) 310 (280) 315 (280) 750 (730) 740 (720) 860 (860) 860 (860) Functions: ACV, ACI Measurements Measurements Into Buffer Into Computer 0.5 (0.5) 0.5 (0.5) 3.3 (3.3) 3.3 (3.3) 59.8 (49.8) 55 (46) 26200 (26200) 24500 (24500) 87. Rear input terminals only. 88. Specified for square wave inputs. Input signal must be >10% of ACV range. If input is<20mV on the 100mV range, then the frequency must be >10Hz. For sine wave inputs, frequency must be >100Hz. For frequencies 100Hz, threshold level 50% of input signal and 7Hz, threshold level 3% of range. 89. Threshold range is voltage RMS and threshold level voltage peak. Specified with 1kHz square wave. 100V and 700V threshold ranges guaranteed by design. Functions: Thermistor Measurements Measurements Into Buffer Into Computer 57 (48) 57 (48) 230 (200) 230 (200) 900 (750) 900 (750) 6800 (6000) 6800 (6000) 18000 (18000) 18000 (18000) Functions: Dry Circuit (1 k) Measurements Measurements Into Buffer Into Computer 27 (23) 26 (22) 100 (89) 96 (85) 190 (180) 190 (180) 295 (290) 295 (290) 310 (310) 310 (310) DIGITIZE, TYPICAL Sampling Rate 10 kS/s 20 kS/s 50 kS/s 100 kS/s 1 MS/s Digits 51/2 41/2 41/2 41/2 31/2 Resolution 18 16 16 15 12 Measurements Into Computer 93 9700 19000 44400 80000 108000 90. Reading speeds for autozero off, fixed range, autodelay off. Offset compensation off and open lead detector off where applicable. 91. Buffer measurements: For <0.2 PLC, multisample, single buffer transfer binary reading only. 92. PC measurements: For 1 and 0.2 PLC single reading and single transfer to computer (USB). 93. Reading rates using factory default operating conditions and autorange off, autodelay off. Speeds include measurement and data transfer out of the USB. 1000 readings with binary transfer over USB. 94. For bandwidth 300Hz, autozero off, 61/2 digits at 1 PLC, 31/2 digits at 0.0005 PLC. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company Model DMM7510 condensed specifications 71/2-Digit Graphical Sampling Multimeter DMM7510 Triggering MODE: 31/2-digit, autozero off, 0.0005 PLC, excludes measurement time. Time includes function change from DC voltage or 2-wire ohms to listed function. TIME BASE ACCURACY: 25ppm. TRIGGER SOURCE: Analog DCV, DCI, or any system trigger. TRIGGER COUPLING: DC or AC (DCV function only). INPUT TRIGGER LATENCY 96, 97, 98: <225ns. INPUT TRIGGER JITTER 96, 97: <50ns. SAMPLE PERIOD JITTER 96, 97: <1ns. Function DC Voltage or 2-wire ohms (<10 k) 4-wire ohms (<10 k) DC Current Frequency or Period 95 AC Voltage or AC Current Digitize Voltage or Current Function Change (ms) 6 7 7 7 7 7 Range Change (ms) 1.3 1.3 1.3 1.3 1.3 1.3 RANGES FOR FUNCTION CHANGE TIMES Function change times apply to the ranges listed in the table below. Function DC Voltage 2-wire or 4-wire Ohms DC Current Dry-circuit Ohms Thermocouple Thermistor AC Current AC Voltage Range 10 V 1 k 1 mA 10 Use DC Voltage rates Use 2-wire Ohms rates 1 mA 1V Buffer Transfer Speed (Binary) Average for 1000 readings Average for 1000 readings with timestamp DMM REAR-PANEL TRIGGERS EXT TRIG IN AND OUT: 0V to 5V logic signal input and output, TTL compatible. EXT TRIGGER LATENCY (IN and OUT): <400ns. EXT TRIGGER LATENCY (IN or OUT): <200ns (guaranteed by design). ANALOG TRIGGERING 99 ANALOG LEVEL, EDGE, OR WINDOW TRIGGER TYPES 100 Trigger Characteristics Input Resolution Basic Accuracy (TACAL 5C) 101, 102 Measurements into Computer (per second) USB LAN GPIB 280000 270000 190000 170000 140000 100000 Voltage Input 100 mV to 1000 V 0.05% 1% Current Input 10 A to 10 A 0.05% 1% ANALOG TRIGGER LATENCIES Positive Logic Negative Logic Digital I/O 800 ns + 40 ns jitter 800 ns + 40 ns jitter External 930 ns + 40 ns jitter 840 ns + 40 ns jitter WINDOW FILTER AND MEMORY (BUFFER) WINDOW FILTER SIZE: 0 to 10% of reading, where 0 averages all readings. MEMORY: Up to 27.5 million timestamped readings with the compact buffer style, with additional memory available using an external USB flash drive. MAXIMUM INTERNAL MEMORY (Buffer): 27.5 million readings with the compact buffer style (61/2-digit without formatting), 11 million readings with the standard or full buffer style. 95. For DC voltage or 2-wire ohms to frequency or period, 10ms aperture. For AC current or AC voltage, detector bandwidth is 300Hz. 96. Guaranteed by design; for digital I/O only. 97. Stimulus command required to meet specifications. 98. If using trigger model, add 200ns uncertainty. 99. For DC or AC coupled, the trigger level can be set up to 100% of measure range. 100. Rising or falling edge triggering supported. Window trigger requires setting two independent levels. 101. Trigger event occurs after the threshold crossing at a time determined by total trigger latencies. 102. Accuracy specifications require user ACAL and are verified with level trigger amplitude set to 50% of range with a 100Hz sine wave at 100% full scale of range. High frequency rejection is off. NPLC 0.0005 (DC voltage/DC current) or aperture 1s for digitize voltage or digitize current. Specified for fixed range, autozero off. For digitized DC voltage AC coupled, add 0.5 %. For DC current and digitized DC current 3A or 10A ranges, add an additional 2%. 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company DIGITAL MULTIMETERS & SYSTEMS Model DMM7510 condensed specifications System Performance, Typical Model DMM7510 condensed specifications 71/2-Digit Graphical Sampling Multimeter DMM7510 DMM7510 71/2-Digit Graphical Sampling Multimeter DIGITAL MULTIMETERS & SYSTEMS SPECIFICATION CONDITIONS: This document contains specifications and supplemental information for the Model DMM7510 Precision Sampling Digital Multimeter instrument. Specifications are the standards against which the Model DMM7510 is tested. Upon leaving the factory, the Model DMM7510 meets these specifications. Supplemental, typical, and characteristic values are non-warranted, apply at 23C, and are provided solely as useful information. All specifications apply to front or rear terminal inputs, except 10 A specifications (rear terminals only). INPUT PROTECTION: 1010 V DC (715 V RMS V AC) all ranges and functions on HI and LO terminals; 350V all ranges and functions on sense HI, sense LO terminals; 250V rated current input terminal; fused 3A and 10A ranges; current input terminals protected to 1kV. 3A INPUT FUSE PROTECTION: 3.5A, 1kV fast blow type; Keithley part number DMM7510-FUSE-3A. 10A INPUT FUSE PROTECTION: 11A, 1kV fast blow type; Keithley part number DMM7510-FUSE-10A. AC VOLTAGE INPUT: Maximum DCV: 1000V on any AC voltage range. COMMON MODE ISOLATION: 500VDC or ACVpeak LO to chassis. All terminals >10G, <350pF any terminal to chassis. POWER LINE: Universal input, 100V to 240V. LINE FREQUENCY: 50Hz or 60Hz, automatically sensed at power-up. POWER CONSUMPTION: 60VA. OPERATING ENVIRONMENT: Specified for 0 to 50C, 80% relative humidity at 35C, altitude up to 2000 meters. STORAGE ENVIRONMENT: -30 to 70C. REAL TIME CLOCK: Lithium battery backup (3+ years battery life). EMC: Conforms to European Union EMC Directive. SAFETY: NRTL listed to UL61010-1, and CSA C22.2 No 61010-1; conforms with European Union Low Voltage Directive. VIBRATION: MIL-PRF-28800F Class 3, Random. WARM-UP: 90 minutes to rated accuracy. INPUT SIGNAL CONNECTIONS: Front and rear safety banana jacks. COOLING: Forced air, fixed speed. DIMENSIONS: Without handle and bumpers: 88mm high x 213mm wide x 410mm deep (3.46 in. x 8.39 in. x 16.13 in.). With handle and bumpers (bench configuration): 106mm high x 255mm wide x 425mm deep (4.18 in. x 10.05 in. x 16.75 in.). SHIPPING WEIGHT (with bumpers and handle): 4.08kg (9.0 lb.). SHIPPING WEIGHT (without bumpers and handle): 3.63kg (8.0 lb.). DIGITAL I/O: Connector: 9-pin female D. 5V Power Supply Pin: Limited to 500 mA at > 4 V (solidstate fuse protected). Lines: Six input/output, user-defined, for digital I/O or triggering. Input Signal Levels: 0.7V (maximum logic low) 3.7V (minimum logic high). Input Voltage Limits: -0.25V (absolute minimum) +5.25V (absolute maximum). Maximum Source Current: +2.0mA at >2.7V (per pin). Maximum Sink Current: -50mA at 0.7V (per pin, solid-state fuse protected). Handler: User-defined start of test, end of test, four category bits MATH FUNCTIONS: Rel, dB, Limit Test, Percentage, 1/x, and mX + b. REMOTE INTERFACE: LAN: RJ-45 connector, 10/100BT; Virtual Front Panel. IP Configuration: Static or DHCP. GPIB: IEEE-488.1 compliant. Supports IEEE-488.2 common commands and status model topology. USB Device (rear panel, type B): 2.0 full speed, USBTMC compliant. USB Host (front panel, type A): USB 2.0, support for flash drives, FAT 32. LXI COMPLIANCE: LXI version 1.4 Core 2011. LANGUAGE: Embedded Test Script Processor (TSP) accessible from any host interface; responds to high-speed test scripts comprised of remote commands and statements (for example, branching, looping, math); able to execute high-speed test scripts stored in memory without host intervention; also SCPI (default command set). ACCESSORIES SUPPLIED: Product Information CD-ROM, Model DMM7510 Quick Start Guide, Kickstart Software Quick Start Guide, power cord, 1 m USB cable (type A to type B), 3 m LAN cable, and Model 1756 Standard Test Lead Kit. ACCESSORIES AVAILABLE: (Calibration / Data / ISO 17025), software IVI/VISA drivers for Microsoft(R) Visual Basic(R), Visual C/C++(R), National Instruments (NITM) LabVIEWTM, Keithley Test Script Builder, Keithley KickStart, and NI LabWindowsTM/CVI. DISPLAY: Five-inch capacitive touch, color thin-film-transistor (TFT) WVGA (800x480) with LED backlight. PASSWORD PROTECTION: 30 characters. EXPANSION INTERFACE: The TSP-Link(R) expansion interface allows TSP-enabled instruments to trigger and communicate with each other. IP CONFIGURATION: Static or DHCP (manual or automatic). 1.888.KEITHLEY (U.S. only) www.keithley.com A Greater Measure of Confidence A Tektronix Company Model DMM7510 condensed specifications Model DMM7510 condensed specifications GENERAL INSTRUMENT SPECIFICATIONS