10MHz pulse generator with delay
0.1Hz to 10MHz frequency range
50ns minimum pulse width; fully variable pulse delay
Squarewave, double pulse and delayed pulse modes
Free-run, gated and triggered modes
50
output, variable 0.1V to 10V; TTL and sync outputs
THURLBY THANDAR INSTRUMENTS
TGP110
Designed and built in the U.K. by:
82100-0066
TGP110 10MHz pulse generator with delay
An essential instrument
The generation of pulses for the stimulus and control of elec-
tronic systems is beyond the capability of all but the most so-
phisticated of function generators.
The architecture of a dedicated pulse generator enables it to
generate pulses of a set width regardless of the repetition rate
offeringdutycycleswhichcanextenddownto1in100million.
In addition to continuous operation, single or multiple pulses
can be generated in response to trigger or gating signals with
precisely defined timing relationships.
A dedicated pulse-waveform output amplifier provides flat top
pulses with fast rise and fall times at variable amplitude.
Variable delay
The TGP110 offers selectable delay between the sync output
andpulseoutput. Intriggeredmodethisalsosetsthedelaybe-
tween a trigger signal and the start of the pulse.
Wide pulse range
TheTGP110cangeneratepulsewidthsinthe range50ns to5s.
There are eight overlapping decade ranges with vernier control
within each range.
The period range is 100ns to 10s, equivalent to a repetition rate
range of 10MHz to 0.1Hz. Delay is independently adjustable
over the same range as pulse width.
A complement switch inverts the mark-space polarity.
Squarewave and double pulse
Insquaremode, squarewavesaregeneratedat afrequencyset
bytheperiodcontrols alone. Thisprovidesaconvenientmeans
of generating variable period edges where the pulse width is
unimportant, for example.
Indoublepulse mode, asecondpulseis generated withinevery
periodataset delayafterthestartofthefirstpulse. Thedelayis
independently adjustable.
PERIOD, PULSE WIDTH, DELAY
Each parameter is variable within 8 overlapping decade ranges with a
vernier providing continuously variable control within each range.
PERIOD
Range: 100nsec to 10sec (10MHz to 0·1Hz).
Jitter: <0.1%.
PULSE WIDTH
Range: 50nsec to 5sec
Jitter: <0.1%.
DELAY
Range: 50nsec to 5sec
TRIGGER, GATE
RUN
Normal operational mode in which pulses are generated continuously at
0.1Hz to 10MHz.
TRIGGERED
DC to 10MHz pulse train in synchronism with external trigger pulses;
pulse width determined by pulse width controls. Trigger can be gener-
ated manually from front panel button.
GATED
0·1Hz to 10MHz pulse train, parameters set by period and pulse width
controls, starts synchronously with leading edge of gate input. Last pulse
is completed at the end of gating period. Gating signal can be generated
manually from front panel button.
PULSE MODES
NORMAL PULSE
One pulse is generated each period. The delay setting is ignored.
SQUAREWAVE
0·1Hz to 10MHz squarewave, frequency set by the period controls.
Pulse width and delay settings ignored.
Mark : Space ratio: 1:1 ±10%.
DOUBLE PULSE
A second pulse is generated after a delay set by the delay controls; the
delay is related to the leading edge of the first pulse.
DELAYED PULSE
A pulse is generated after a delay set by the delay controls; the delay is
related to rising edge of the trigger signal.
INPUTS
GATE/TRIG INPUT
Frequency range: DC - 10MHz
Signal range: TTL threshold; max. input ±10V.
Min. pulse width: >30nsec.
Input Impedance: Typically 10k.
OUTPUTS
50OUTPUT
Amplitude: Two switch selectable ranges of 0·1V - 1·0V and
1V - 10V from 50. (50mV to 500mV and 500mV to
5V into 50). Adjustable within ranges by a single
turn vernier.
Rise/Fall Times: Typically 10nsec into 50load. Maximum 15ns.
Aberrations: Typically <5% for output set at >20% of range maxi-
mum, into 50.
AUX OUTPUT
Duplicates 50output but at a fixed CMOS/TTL level.
SYNC OUTPUT
Amplitude: A positive going pulse at CMOS/TTL level.
Timing: Leading edge starts >20nsec before the TTL/50
output transition.
Duration: Typically 30nsec.
COMPLEMENT SWITCH
Inverts the AUX and 50outputs.
GENERAL
Power: 230V or 115V AC nominal 50/60Hz, adjustable inter-
nally; operating range ±14% of nominal; 20VA max.
Size: 140 x 220 x 230mm (HxWxD)
Weight: 1.6kg (3.5lb)
Operating Range: + 5°C to 40°C, 20-80% RH.
Storage Range: - 40°C to 70°C
Safety: Complies with EN61010-1.
EMC: Complies with EN55081-1 and EN50082-1.
ThurlbyThandarInstrumentsLtd. operatesa policyof continuousdevelopment andreserves
the right to alter specifications without prior notice.
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon. Cambs. PE18 7DX England
Tel: 01480 412451 Fax: 01480 450409 Email: sales@ttinst.co.uk
CYAN MAGENTA YELLOW BLACK CYAN MAGENTA YELLOW BLACK
10MHz pulse generator with delay
0.1Hz to 10MHz frequency range
50ns minimum pulse width; fully variable pulse delay
Squarewave, double pulse and delayed pulse modes
Free-run, gated and triggered modes
50
output, variable 0.1V to 10V; TTL and sync outputs
THURLBY THANDAR INSTRUMENTS
TGP110
Designed and built in the U.K. by:
82100-0066
TGP110 10MHz pulse generator with delay
An essential instrument
The generation of pulses for the stimulus and control of elec-
tronic systems is beyond the capability of all but the most so-
phisticated of function generators.
The architecture of a dedicated pulse generator enables it to
generate pulses of a set width regardless of the repetition rate
offeringdutycycleswhichcanextenddownto1in100million.
In addition to continuous operation, single or multiple pulses
can be generated in response to trigger or gating signals with
precisely defined timing relationships.
A dedicated pulse-waveform output amplifier provides flat top
pulses with fast rise and fall times at variable amplitude.
Variable delay
The TGP110 offers selectable delay between the sync output
andpulseoutput. Intriggeredmodethisalsosetsthedelaybe-
tween a trigger signal and the start of the pulse.
Wide pulse range
TheTGP110cangeneratepulsewidthsinthe range50ns to5s.
There are eight overlapping decade ranges with vernier control
within each range.
The period range is 100ns to 10s, equivalent to a repetition rate
range of 10MHz to 0.1Hz. Delay is independently adjustable
over the same range as pulse width.
A complement switch inverts the mark-space polarity.
Squarewave and double pulse
Insquaremode, squarewavesaregeneratedat afrequencyset
bytheperiodcontrols alone. Thisprovidesaconvenientmeans
of generating variable period edges where the pulse width is
unimportant, for example.
Indoublepulse mode, asecondpulseis generated withinevery
periodataset delayafterthestartofthefirstpulse. Thedelayis
independently adjustable.
PERIOD, PULSE WIDTH, DELAY
Each parameter is variable within 8 overlapping decade ranges with a
vernier providing continuously variable control within each range.
PERIOD
Range: 100nsec to 10sec (10MHz to 0·1Hz).
Jitter: <0.1%.
PULSE WIDTH
Range: 50nsec to 5sec
Jitter: <0.1%.
DELAY
Range: 50nsec to 5sec
TRIGGER, GATE
RUN
Normal operational mode in which pulses are generated continuously at
0.1Hz to 10MHz.
TRIGGERED
DC to 10MHz pulse train in synchronism with external trigger pulses;
pulse width determined by pulse width controls. Trigger can be gener-
ated manually from front panel button.
GATED
0·1Hz to 10MHz pulse train, parameters set by period and pulse width
controls, starts synchronously with leading edge of gate input. Last pulse
is completed at the end of gating period. Gating signal can be generated
manually from front panel button.
PULSE MODES
NORMAL PULSE
One pulse is generated each period. The delay setting is ignored.
SQUAREWAVE
0·1Hz to 10MHz squarewave, frequency set by the period controls.
Pulse width and delay settings ignored.
Mark : Space ratio: 1:1 ±10%.
DOUBLE PULSE
A second pulse is generated after a delay set by the delay controls; the
delay is related to the leading edge of the first pulse.
DELAYED PULSE
A pulse is generated after a delay set by the delay controls; the delay is
related to rising edge of the trigger signal.
INPUTS
GATE/TRIG INPUT
Frequency range: DC - 10MHz
Signal range: TTL threshold; max. input ±10V.
Min. pulse width: >30nsec.
Input Impedance: Typically 10k.
OUTPUTS
50OUTPUT
Amplitude: Two switch selectable ranges of 0·1V - 1·0V and
1V - 10V from 50. (50mV to 500mV and 500mV to
5V into 50). Adjustable within ranges by a single
turn vernier.
Rise/Fall Times: Typically 10nsec into 50load. Maximum 15ns.
Aberrations: Typically <5% for output set at >20% of range maxi-
mum, into 50.
AUX OUTPUT
Duplicates 50output but at a fixed CMOS/TTL level.
SYNC OUTPUT
Amplitude: A positive going pulse at CMOS/TTL level.
Timing: Leading edge starts >20nsec before the TTL/50
output transition.
Duration: Typically 30nsec.
COMPLEMENT SWITCH
Inverts the AUX and 50outputs.
GENERAL
Power: 230V or 115V AC nominal 50/60Hz, adjustable inter-
nally; operating range ±14% of nominal; 20VA max.
Size: 140 x 220 x 230mm (HxWxD)
Weight: 1.6kg (3.5lb)
Operating Range: + 5°C to 40°C, 20-80% RH.
Storage Range: - 40°C to 70°C
Safety: Complies with EN61010-1.
EMC: Complies with EN55081-1 and EN50082-1.
ThurlbyThandar InstrumentsLtd. operatesa policyof continuousdevelopment andreserves
the right to alter specifications without prior notice.
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon. Cambs. PE18 7DX England
Tel: 01480 412451 Fax: 01480 450409 Email: sales@ttinst.co.uk
CYAN MAGENTA YELLOW BLACK CYAN MAGENTA YELLOW BLACK