BCD.00286_AA_December-3-2013 1 www.power-one.com
DATA SHEET
Dialysis
Monitoring
Pumps
Surgical Devices
Home Health
Portable Device
MBC40 SERIES MEDICAL 40W AC/DC
40 W convection cooled
-20 to 50 deg C full load operation
2 x MOPP
90-264 VAC input
2” x 4” x 1.2” (50.8mm x 101.6 mm x 30.5 mm)
No minimum load required - single output models
Conducted EMI EN 55022-B, FCC Part 15 Level B
Medical Safety Agency Approvals
No load power < 0.3 W
FEATURES
APPLICATIONS
MBC40 Series 2 www.power-one.com
DATA SHEET
TECHNICAL DATA:
Input
DESCRIPTION/CONDITION
Universal Input
90 - 264 Vac
47 to 400 Hz (3)
230 Vac (cold start)
60 A max.
264 Vac 50/60 Hz
250 μA max
120 Vac @ 40 W
230 Vac @ 40 W
0.85 A rms
0.45 A rms
Output
PARAMETER
DESCRIPTION/CONDITION
Voltage Adjustment
V1
± 10%
Transient Response
Main output 50 to 100% load change, 50/60
Hz, 50% duty cycle, 0.1A / uSec
< 10%, recovery time < 5 mS
Over Voltage Protection
130% Typical (V1 only)
Short Circuit Protection
130% Typical
Remote Sense
V1
0.5 Vdc compensation
Efficiency
>85%
Set Point Tolerance
V1: ± 0.5%, V2 & V3: ± 5%
Rise Time
< 100 mSec
Ordering Information
PRODUCT
FAMILY
VOLTS
(VDC)
OUTPUT CURRENT
MAX (A)
MINIMUM
LOAD (A) (4)
RIPPLE &
NOISE
TOTAL
REGULATION
OVP THRESHOLD
MBC40-1005G
5.1
8.0
0
1%
± 0.8%
130% Typical
MBC40-1012G
12
3.5
0
1%
± 0.8%
130% Typical
MBC40-1015G
15
2.7
0
1%
± 0.8%
130% Typical
MBC40-1024G
24
1.7
0
1%
± 0.8%
130% Typical
MBC40-1048G
48
0.83
0
1%
± 0.8%
130% Typical
MBC40-3000G
5.2
6.0
0.5
1%
± 0.8%
130% Typical
12.5
2.0
0.1
1%
± 5.3%
-12.8
0.5
0.0
1%
± 5.3%
MBC40-3001G
5.2
6.0
0.5
1%
± 0.8%
130% Typical
24
1.0
0.1
1%
± 5.3%
-12.8
0.5
0.0
1%
± 5.3%
MBC40-3002G
5.2
6.0
0.5
1%
± 0.8%
130% Typical
14.6
1.5
0.1
1%
± 5.3%
-14.8
0.5
0.0
1%
± 5.3%
MBC40-3003G
3.3
6.0
1.0
1.5%
± 0.8%
130% Typical
5.2
3.0
0.1
1%
± 5.3%
-12.8
0.5
0.0
1%
± 5.3%
BCD.00286_AA_December-3-2013 3 www.power-one.com
DATA SHEET
Notes:
1. Maximum outputs for each output. Max power rating should not be exceeded.
2. Output noise measurement is made with a 20 MHz bandwidth using a 6” twisted pair, terminated with a 10 uF tantalum
capacitor in parallel with a 0.1 uF ceramic capacitor.
3. Safety Approved: 47 to 63 Hz.
4. Minimum load specified to meet cross regulation.
5. Specifications are for nominal input voltage, 25° C and max load unless otherwise stated.
6. Class 1 models have Earthing tab J4 Class 2 models (-2 suffix) have no Earthing tab.
7. Specifications subject to change without notice.
8. Warranty 2 years.
General Specifications
PARAMETER
DESCRIPTION/CONDITION
Hold Up Time
6 mSec
120 Vac input
MTBF
>100 khrs
MIL-HDBK-217F
Switching Frequency
67 kHz Typical
Isolation Voltage
Input to Output: Min 4242 Vdc
Input to Ground: Min 2120 Vdc
Weight
150 g (0.33 lbs)
Environmental
PARAMETER
DESCRIPTION/CONDITION
Operating Temperature
Operating
-20 to +7 C. See derating chart below.
Storage
-40 to +85° C
Altitude
Operating 10,000 ft.
Non Operating 40,000 ft
Relative Humidity
95%
Non Condensing
Conducted Emissions
EN55022, FCC part 15 Level B
Harmonic Current Correction
EN61000-3-2 Compliant Level 3
Electromagnetic Susceptibility
EN61000-4-2 Level 3
2, 3, 4, 5 Level 3
Conducted Immunity
EN61000-4-6
Safety
PARAMETER
DESCRIPTION/CONDITION
EN / UL / CSA
60601-1 3rd Edition
MBC40 Series 4 www.power-one.com
DATA SHEET
Figure 1 Output Power Vs. Temperature
Figure 2 Dimension Drawing (Top and Side View)
Mechanical
INPUT = J1
EARTHING TAB = J4
DC OUTPUT = J2
REMOTE SENSE= J3
Pin 1: AC Neutral
Pin 2: Removed
Pin 3: AC Line
Molex: 19705-4301
Pin 1 = V1
Pin 2 = V1
Pin 3 = RTN
Pin 4 = RTN
Pin 5 = V3
Pin 6 = V2
Pin 1 = +V1 Sense
Pin 2 = -V1 Sense
Mating Connector:
Molex: 09-50-3031
Pins: 08-50-0106
Mating Connector:
Molex: 190030001
Mating Connector:
Tyco: 647402-6; Pins: 3-647409
Mating Connector:
Molex: 22-01-2021
Copyright © 2010 Power-One Inc. All rights reserved. Words and logos that are identified as trademarks and/or service marks are, unless noted otherwise, the
trademarks and service marks of Power-One Inc. in the U.S. and other countries. All other product or service names are the property of their respective holders.
Power-One products are protected under numerous U.S. and foreign patents and pending applications, maskwork rights, and copyrights. Power-One reserves the
right to make changes to any products and services at any time without notice. Power-One assumes no responsibility or liability arising out of the application or use
of any information, product, or service described herein except as expressly agreed to in writing by Power-One Inc.
NUCLEAR AND MEDICAL APPLICATIONS - Power-One products are not designed, intended for use in, or authorized for use as critical components in life
support systems, equipment used in hazardous environments, or nuclear control systems without the express written consent of the respective divisional president
of Power-One, Inc.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date
manufactured. Specifications are subject to change without notice.