Technical Data
Fluke i2000 Flex
AC Current Clamp
Electrical specifications
Current ranges: 20 A, 200 A, 2000 A ACRMS
Output sensitivity (ac-coupled): 100 mV, 10 mV,
1 mV per A
Load impedance: 100 kW minimum
Accuracy (at +25 °C): ± 1 % of range (45 Hz to
65 Hz)
Linearity (10 % to 100 % of range): ± 0.2 %
of reading
Noise: 8 mV ac rms (30 A), 2 mV ac rms (300 A,
3000 A)
Frequency range: 10 Hz to 20 kHz (-3 dB)
Phase error: < ± 1 º (45 Hz to 65 Hz), : ± 10 º
(at 20 kHz)
Position sensitivity (with cable > 25 mm from
the coupling): ± 2 % of range
External field (with cable > 200 mm from the
head): ± 1 % of range
The i2000 is an ac current probe Utilizing
the Rogowski principle. It can be used to
measure currents up to 2000 A when used
in conjunction with multimeters, recorders
or data loggers. The flexible and lightweight
measuring head allows quick and easy
installation in hard to reach areas.
Power supply: 2 x AA/MN 1500/LR6 Alkaline,
200 hours, low battery indicator
Temperature coefficient: ± 0.08 % of reading
per °C
Working voltage (see Safety standards section):
600 V ac rms or dc
General specifications
Head cable length (double insulated): 610 mm
(24 in)
Head cable diameter: 14.3 mm (0.56 in)
Head bend radius: 38.1 mm (1.5 in)
Cable length (head to electronics): 2 m (6.56 ft)
Output connection: BNC with BNC to Banana
adapter included
Operating temperature range: 0 °C to +70 °C
(32 °F to 158 °F) (head and electronics)