Agilent L4411A
‘Low profile’ 6½ Digit Multimeter
Leading the Industry in High-Performance
System Test
Product Overview
Agilent L4411A
6½-Digit High-
Performance DMM
• 50,000 readings/sec @ 4½ digits
direct to PC
• 10,000 readings/sec @ 5½ digits
direct to PC
• 1,000 readings/sec @ 6½ digits
direct to PC
• Analog level triggering
• Programmable pre/post triggering
• LAN extensions for instruments
(LXI), USB & GPIB standard
• 30 PPM 1 year basic DC accuracy
• DCV, ACV, DCI, ACI, 2-wire and
4-wire resistance, frequency,
period, and diode test
• Capacitance & temperature
measurements
• Expanded measurement ranges
Dramatic system performance
Whether it’s raw reading speed or
fast system throughput, the L4411A
sets a new benchmark in performance.
Using a new A/D technology, the
L4411A achieves an impressive
50,000 readings a second at 4½ digits,
and can stream readings to your
computer at this same speed!
Transactional I/O (single reading—
measurement and PC transfer time) is
3x faster than other popular modular
DMMs, significantly enhancing your
test throughput. Triggering is fast and
precise, with both trigger latency and
trigger jitter less than 1 µs, while bus
query response is less than 500 µs.
ACV measurements are faster as
well thanks to a digital measurement
technique that additionally improves
accuracy at high and low frequencies.
A new standard for modular
system DMMs
The L4411A 6½ digit high performance
DMM expands Agilent’s industry
leading offering of LXI system
products. For the test system integra-
tor looking for the next generation
modular DMM, this new meter offers
the industry’s best measurement
speed and throughput, a reduced size
(1 rack unit high), superior measure-
ment performance, and a choice of
computer interfaces, including LXI,
providing high performance, easy
to use, economical I/O. A simple
display, including ‘latest reading’
and LAN address, allows the system
integrator to quickly integrate and
debug the test system. And finally,
the DMM comes with a compatibility
mode, requiring little-to-no code
change to upgrade your test system
with next generation capabilities.
2
DMM dimensions (mm)
LXI—Class C
LAN Extensions for Instruments (LXI)
provides the next generation I/O for
system applications requiring high
throughput. Transfer rates of over
250,000 readings/sec are attainable
ensuring even the most data intensive
measurements are fast, without the
overhead cost of an instrument
mainframe. LXI provides a built-in
Graphical Web Interface that allows
you to interactively control the DMM
without the hassle of programming,
making it great for debugging your
system. The L4411A DMM is LXI —
Class C compliant.
Enhanced measurement
capabilities
The L4411A offers temperature and
capacitance capabilities in addition to
those measurements you have come
to expect such as DCV, ACV, DCI,
ACI, 2-wire and 4-wire resistance,
frequency, period and diode test. You
also get offset compensated Ohms,
allowing you to accurately measure
resistance in the presence of volt-
ages. Measurement ranges have been
expanded as well; for example, DC
and AC current ranges now go down
to 100 µA, resulting in 100 pA resolu-
tion. Real-time math and statistics are
included, and a peak-detect capability
allows you to capture peaks as short
as 20 µs.
System integration
When deciding on your next system
DMM you can’t go wrong with the
L4411A. Choose from LAN (LXI),
USB or GPIB interfaces, all standard
on the L4411A, to connect to your
computer. The 1U size is perfect
for space constrained applications
like aerospace/defense depot test.
Concerned about the viability of your
existing software programs? This
new DMM responds to standard
commands for programmable instru-
mentation (SCPI). Additionally there is
a 34401A/E1412A emulation mode to
ensure the easiest upgrade possible,
virtually eliminating costly software
and documentation changes. The
autoranging power supply allows
you to connect to any input power
without selecting input voltages or
changing fuses. Agilent’s I/O Library
Suite ships with the L4411A to help
you quickly establish an error-free
connection between your PC and
instrument. It provides robust instru-
ment control and works with the
software development environment
you choose.
Companion LXI switch modules
Need a switch to route your signal
to the L4411A? Consider Agilent’s
LXI switch modules. Choose from a
40-channel armature relay multiplexer
(L4421A), a dual/quad 4x8 reed relay
matrix (L4433A) or a 32-channel Form
A/C general purpose relay switch
module (L4437A). Additionally, Agilent
has LXI DAC, digital I/O and multi-
function modules to help complete
your test system requirements. All
from the leader in LXI instrumenta-
tion, Agilent Technologies.
Built to last
Our new DMM was designed to high
standards of ruggedness and reliabil-
ity. From the robust, compact package
to careful selection of components
and conservative circuit design, this
meter is built to last. Calculated mean
time between failure (MTBF) is in
excess of 100,000 hours. Backed by
a 1-year warranty and a worldwide
network of service centers, you can
buy with confidence.
Go to the web
For the latest information on these
or other Agilent DMMs, go to
www.agilent.com/find/dmm
DMM dimensions (mm)
212.3
363.2
40.9
3
Accuracy Specifications ± (% of reading + % of range) 1
1. Specifications are for 90 minute warm-up and 100 PLC.
2. Relative to calibration standards.
3. 20% overrange on all ranges, except DCV 1000 V, ACV 750 V, DCI and
ACI 3 A ranges.
4. For each additional volt over ± 500 V add 0.02 mV of error.
5. Specifications are for sinewave input > 0.3% of range and > 1 mVrms.
Add 30 µV error for frequencies below 1 kHz. 750 VAC range limited to 8 x 107
Volts-Hz. For each additional volt over 300 Vrms add 0.7 mVrms of error.
6. Specifications are for 4-wire resistance measurements, or 2-wire using
Math Null. Without Math Null, add 0.2 Ω additional error in 2-wire resistance
measurements.
7. Specifications are for sinewave input > 1% of range and > 10 µArms.
Frequencies > 5 kHz are typical for 1 A and 3 A ranges.
8. Specifications are for 1-hour warm-up using Math Null. Additional errors may
occur for non-film capacitors.
9. For total measurement accuracy, add temperature probe error.
10. Accuracy specifications are for the voltage measured at the input terminals only.
1 mA test current is typical. Variation in the current source will create some
variation in the voltage drop across a diode junction.
Function Range 3 Frequency,
test current or
burden voltage
24 hour 2
Tcal ±1°C
90 day
Tcal ±5°C
1 year
Tcal ±5°C
Temperature coefficient/°C
0 °C to (Tcal -5°C)
(Tcal +5°C) to 55°C
DC voltage 100.0000 mV 0.0030 + 0.0030 0.0040 + 0.0035 0.0050 + 0.0035 0.0005 + 0.0005
1.000000 V 0.0020 + 0.0006 0.0030 + 0.0007 0.0035 + 0.0007 0.0005 + 0.0001
10.00000 V 0.0015 + 0.0004 0.0020 + 0.0005 0.0030 + 0.0005 0.0005 + 0.0001
100.0000 V 0.0020 + 0.0006 0.0035 + 0.0006 0.0040 + 0.0060 0.0005 + 0.0001
1000.000 V 4 0.0020 + 0.0006 0.0035 + 0.0006 0.0040 + 0.0060 0.0005 + 0.0001
True RMS
AC voltage 5
100.0000 mV
to 750.000 V
3 Hz – 5 Hz 0.50 + 0.02 0.50 + 0.03 0.50 + 0.03 0.010 + 0.003
5 Hz – 10 Hz 0.10 + 0.02 0.10 + 0.03 0.10 + 0.03 0.008 + 0.003
10 Hz – 20 kHz 0.02 + 0.02 0.04 + 0.02 0.045 + 0.02 0.005 + 0.002
20 kHz – 50 kHz 0.04 + 0.025 0.08 + 0.025 0.090 + 0.025 0.010 + 0.0025
50 kHz – 100 kHz 0.10 + 0.040 0.20 + 0.040 0.200 + 0.040 0.020 + 0.0040
100 kHz – 300 kHz 1.00 + 0.250 1.20 + 0.250 1.200 + 0.250 0.120 + 0.0200
Resistance 6 100.0000 Ω 1 mA 0.0030 + 0.0030 0.008 + 0.004 0.010 + 0.004 0.0006 + 0.0005
1.000000 kΩ 1 mA 0.0020 + 0.0005 0.007 + 0.001 0.010 + 0.001 0.0006 + 0.0001
10.00000 kΩ 100 µA 0.0020 + 0.0005 0.007 + 0.001 0.010 + 0.001 0.0006 + 0.0001
100.0000 kΩ 10 µA 0.0020 + 0.0005 0.010 + 0.001 0.012 + 0.001 0.0006 + 0.0001
1.000000 MΩ 5 µA 0.0020 + 0.0010 0.030 + 0.001 0.040 + 0.001 0.0010 + 0.0002
10.00000 MΩ 500 nA 0.0100 + 0.0010 0.600 + 0.001 0.800 + 0.001 0.0030 + 0.0004
100.0000 MΩ 500 nA || 10 MΩ 0.200 + 0.001 6.000 + 0.001 8.000 + 0.001 0.1000 + 0.0001
1.000000 GΩ 500 nA || 10 MΩ 2.000 + 0.001 1.0000 + 0.0001
DC current 100.0000 µA < 0.03 V 0.010 + 0.020 0.040 + 0.025 0.050 + 0.025 0.0020 + 0.0030
1.000000 mA < 0.30 V 0.007 + 0.006 0.030 + 0.006 0.050 + 0.006 0.0020 + 0.0005
10.00000 mA < 0.03 V 0.007 + 0.020 0.030 + 0.020 0.050 + 0.020 0.0020 + 0.0020
100.0000 mA < 0.30 V 0.010 + 0.004 0.030 + 0.005 0.050 + 0.005 0.0020 + 0.0005
1.000000 A < 0.80 V 0.050 + 0.006 0.080 + 0.010 0.100 + 0.010 0.0050 + 0.0010
3.000000 A < 2.0 V 0.100 + 0.020 0.120 + 0.020 0.150 + 0.020 0.0050 + 0.002
True RMS
AC current 7
100.0000 µA
to 3.00000 A
3 Hz – 5 kHz 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006
5 kHz – 10 kHz 0.20 + 0.04 0.20 + 0.04 0.20 + 0.04 0.030 + 0.006
Frequency
or period
100 mV
to 750 V
3 Hz – 5 Hz 0.070 + 0.000 0.070 + 0.000 0.070 + 0.000 0.005 + 0.000
5 Hz – 10 Hz 0.040 + 0.000 0.040 + 0.000 0.040 + 0.000 0.005 + 0.000
10 Hz – 40 Hz 0.020 + 0.000 0.020 + 0.000 0.020 + 0.000 0.001 + 0.000
40 Hz – 300 kHz 0.005 + 0.000 0.006 + 0.000 0.007 + 0.000 0.001 + 0.000
Capacitance 8 1.0000 nF 500 nA 0.50 + 0.50 0.50 + 0.50 0.50 + 0.50 0.05 + 0.05
10.000 nF 1 µA 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01
100.00 nF 10 µA 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01
1.0000 µF 10 µA 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01
10.000 µF 100 µA 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01
Temperature 9
RTD
-200 °C to 600 °C
0.06 °C
0.06 °C
0.06 °C
0.003 °C
Thermistor -80 °C to 150 °C 0.08 °C 0.08 °C 0.08 °C 0.002 °C
Diode test 10 1.0000 V 1 mA 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020
4
A-to-D converter noise performance
Integration time
(NPLC)
Resolution
(ppm of range) 1
Normal mode
rejection (dB) 2
Readings/second 4
0.001 30 0 50,000
0.002 15 0 25,000
0.006 6 0 10,000
0.02 3 0 3,000
0.06 1.5 0 1,000
0.2 0.7 0 300
1 0.3 55 60 (50)
2 0.2 110 3 30 (25)
10 0.1 110 3 6 (5)
100 0.03 110 3 0.6 (0.5)
1. Resolution is defined as the typical DCV 10 V range RMS noise.
Auto-zero on for NPLC ≥ 1. See manual for additional noise characteristics.
2. Normal mode rejection for power line frequency ± 0.1%.
3. For power-line frequency ± 1% 75 dB and for ± 3% 55 dB.
4. Maximum rate with auto-zero off for 60 Hz and (50 Hz) operation.
System reading and throughput rates
DMM memory to PC (maximum reading rate out of memory) 1
Drawing – Path B
Reading format GPIB
readings/sec
USB 2.0
readings/sec
LAN (VXI-11)
readings/sec
LAN (sockets)
readings/sec
ASCII 4,000 8,500 7,000 8,500
32-bit binary 89,000 265,000 110,000 270,000
64-bit binary 47,000 154,000 60,000 160,000
Direct I/O measurements (single reading – measure and I/O time) 1
Drawing – Path C
Maximum reading
rate into memory
or to direct I/O
(readings/sec)
Drawing–Path A or C
Function Resolution
(NPLC)
GPIB
msec
USB 2.0
msec
LAN
(VXI-11)
msec
LAN
(sockets)
msec
DCV/2-wire
resistance
0.001 2.6 2.9 4.6 3.2 50,000
ACV/
frequency
Fast filter
1 ms gate
10.0 10.0 10.0 10.0 500
1. ½ scale input signal, immediate trigger, trigger delay 0, auto-zero off, auto-range off, no math, null off,
60 Hz line frequency. See manual for performance on other functions.
System performance
Function
change
(msec) 1
Range change
LAN/GPIB
(msec) 2
Auto-range
(msec) 3
Maximum
external
trigger rate
Maximum
internal
trigger rate
DCV/2-wire
resistance
22 3.9/2.6 7.5 5,000/s 50,000/s
ACV/frequency 37 6.5/6.4 19 500/s 500/s
1. Time to change from 2-wire Resistance to this specified function, or DCV to 2-wire resistance using the
SCPI ”FUNC” command
2. Time to change from one range to the next higher range, ≤ 10 V, ≤ 10 MΩ
3. Time to automatically change one range and be ready for the new measurement, ≤ 10 V, ≤ 10 MΩ
DMM
input DMM memory
50 k readings
non-volatile
Path ‘A’
Path ‘C’ Path ‘B’
PC
USB
LAN
GPIB
ADC
L4411A/
34410A/
34411A
DMM
System reading architecture
100
10
1
0.1
0.01
0.001 0.01 0.1 1 10 100
Integration time (NPLC)
RMS noise (ppm of range)
System reading architecture
5
Measurement Characteristics
DC voltage
Measurement
method:
Continuously integrating multi-slope IV A/D
converter
Linearity: 0.0002% of reading
(10 V range) + 0.0001% of range
Input
resistance: 0.1 V, 1 V, 10 V ranges 10 MΩ or > 10 GΩ
(selectable)
100 V, 1000 V ranges 10 MΩ ±1% (fixed)
Input bias current: < 50 pA at 25 °C
Input protection: 1000 V
DC CMRR: 140 dB 1
True RMS AC voltage
Measurement
method:
AC-coupled True RMS measurement.
Digital sampling with anti-alias filter.
Crest factor: No additional error for crest factors < 10.
Limited by peak input and 300 kHz bandwidth.
Peak input: 300% of range or 1100 V
Overload
ranging:
Will select higher range if peak input
overload is detected during auto range.
Overload is reported in manual ranging.
AC CMR: 70 dB 2
Maximum input: 400 Vdc, 1100 Vpk
Input impedance: 1 MΩ ± 2% in parallel with < 150 pF
Input protection: 750 Vrms all ranges
Resistance
Measurement
method:
Selectable 2-wire or 4-wire.
Current source referenced to LO input.
Offset
compensation:
Selectable on the 100 Ω, 1 kΩ, and 10 kΩ
ranges
Max. lead
resistance
(4-wire):
10% of range per lead for 100 Ω, 1 kΩ. 1 kΩ
per lead on all other ranges
Input protection: 1000 V on all ranges
DC current
Current shunt: 200 Ω for 100 µA, 1 mA
2 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
Input protection: 3 A, 250 V fuse
True RMS AC current
Measurement
method:
AC-coupled true RMS measurement.
Directly coupled to the fuse and shunt.
Digital sampling with anti-alias filter.
Current shunt: 200 Ω for 100 µA, 1 mA
2 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
Maximum input: The peak value of the DC + AC current must
be < 300% of range. The RMS current must
be < 3 A including the DC current content.
Input protection: 3 A, 250 V fuse
Frequency and period
Measurement
method:
Reciprocal-counting technique. AC-coupled
input using the AC voltage measurement
function.
Input impedance: 1 MΩ ± 2% in parallel with < 150 pF
Input protection: 750 Vrms all ranges
Capacitance
Measurement
method:
Current input with measurement of
resulting ramp.
Connection type: 2-wire
Temperature
Thermistor: 2.2 kΩ, 5 kΩ, and 10 kΩ
RTD: α = 0.00385
Ro from 49 Ω to 2.1 kΩ
Diode test
Response time: 300 samples/sec
Continuity
threshold:
Fixed at 10 Ω
Operating characteristics
Maximum readings/second
Digits
Function 3 4.5 5.5 6.5
DCV 50 k 10 k 1 k
2-wire Ω 50 k 10 k 1 k
DCI 50 k 10 k 1 k
Frequency 500 90 10
Period 500 90 10
Filter setting fast med slow
ACV 500 150 50
ACI 500 150 50
Additional specifications
Resolution: See table on page 4
Overall bandwidth,
DCV & DCI:
15 kHz typical @ 20 µs aperture (-3 dB)
Triggering: Pre/Post, Analog Level, Int/Ext, Pos/Neg
Timebase
resolution:
19.9524 µs 0.01% accuracy
Trigger jitter: 2 µs (p-p), 20 µs (p-p) when pre-triggered
1. For 1 kΩ unbalanced in LO lead, ± 500 V peak maximum
2. For 1 kΩ unbalanced in LO lead and < 60 Hz, ± 500 V peak maximum
3. Maximum rate for DCV, DCI, and resistance functions
(using zero settling delay, autozero off, manual range)
6
Ordering Information
Spurious-free dynamic range & signal to noise
distortion ratio
Function Range Spur-free SNDR
DCV 1 V -75 dB 60 dB
10 V 1-70 dB 60 dB
100 V -75 dB 60 dB
1. 10 V range: 2 V (p-p) <signal< 16 V (p-p)
Triggering and memory
Reading hold
sensitivity:
1% of reading
Samples per
trigger:
1 to 1,000,000
Trigger delay: 0 to 3600 sec (20 µs step size)
External trigger: Programmable edge, low-power TTL
compatible
Delay: < 1 µs
Jitter: < 1 µs
Max rate: 5,000/sec
Min pulse width: 1 µs
Voltmeter
complete:
3 V logic output, 2 µs pulse with
programmable edge
Nonvolatile
memory:
50,000 readings
Volatile memory: 1,000,000 readings
Sample timer:
Range: 0 to 3600 s (20 µs step sizes)
Jitter: < 100 ns
General specifications
Power supply: 90 V – 264 V @ 45-66 Hz
90 V – 134 V @ 360 – 440 Hz
Power line
frequency:
Automatically sensed at power-on
Power
consumption:
50 VA peak (18 W average)
Operating
environment:
Full accuracy for 0 °C to 55 °C,
80% R.H. at 40 °C non-condensing
Storage
temperature:
-40 °C to 70 °C
Weight: 1.9 kg (4.25 Lbs)
Dimensions: (W x H x D) 40.9 mm x 212.3 mm x 363.2 mm
Safety: IEC 61010-1, EN 61010-1, UL 61010-1,
CAN/CSA-C22.2 No. 61010-1, refer to
Declarations of Conformity for current
revisions. Measurement CAT II 300V, CAT I
1000V. Pollution Degree 2
EMC: IEC 61326, EN 61326, CISPR 11, ICES-001,
AS/NZS 2064.1, refer to Declaration of
Conformity for current revisions.
Vibration & shock: MIL-T-28800E, Type III, Class 5 (Sine only)
LXI compliance: LXI Class C, ver. 1.0
Warranty: 1 year
Agilent L4411A Multimeter
Accessories included
Test report, power cord, LAN cross-over interface cable.
Product reference CD-ROM (34410-13601):
• Software
IntuiLink software
IVI-COM driver
LabView driver
Example programs
• Online documentation
Programmer’s reference
Getting started guide
User’s guide
Service guide
Localized manuals
Agilent I/O Libraries CD-ROM (E2094-60003)
Options
Opt. 001 Front measurement terminals ONLY
Opt. A6J ANSI Z540 compliant calibration
Agilent accessories
Y1133A Measurement & trigger cable kit
Y1160A Rack mount kit for L4400A series instruments
racks 1 or 2 instruments side-by-side on
sliding tray
11059A Kelvin probe set
11060A Surface mount device (SMD) test probes
11062A Kelvin clip set
34134A DC coupled current probe
34136A High voltage probe
34138A Test lead set
34171B Input terminal connector (sold in pairs)
34172B Input calibration short (sold in pairs)
34330A 30 A current shunt
E2308A 5 kΩ thermistor probe
www.agilent.com
www.agilent.com/find/L4411A
www.agilent.com/find/dmm
For more information on Agilent Technol-
ogies’ products, applications or services,
please contact your local Agilent office.
The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada (877) 894 4414
Brazil (11) 4197 3500
Mexico 01800 5064 800
United States (800) 829 4444
Asia Pacific
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 375 8100
Europe & Middle East
Belgium 32 (0) 2 404 93 40
Denmark 45 70 13 15 15
Finland 358 (0) 10 855 2100
France 0825 010 700*
*0.125 €/minute
Germany
49 (0) 7031 464 6333
Ireland 1890 924 204
Israel 972-3-9288-504/544
Italy 39 02 92 60 8484
Netherlands 31 (0) 20 547 2111
Spain 34 (91) 631 3300
Sweden 0200-88 22 55
United Kingdom 44 (0) 131 452 0200
For other unlisted Countries:
www.agilent.com/find/contactus
Revised: June 8, 2011
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2011
Published in USA, November 16, 2011
5989-6303EN
Agilent Advantage Services is committed
to your success throughout your equip-
ment’s lifetime. To keep you competitve,
we continually invest in tools and
processes that speed up calibration and
repair and reduce your cost of ownership.
You can also use Infoline Web Services
to manage equipment and services more
effectively. By sharing our measurement
and service expertise, we help you create
the products that change our world.
www.agilent.com/quality
www.agilent.com/find/advantageservices
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
www.lxistandard.org
LAN eXtensions for Instruments puts
the power of Ethernet and the Web
inside your test systems. Agilent is a
founding member of the LXI consortium.
Agilent Channel Partners
www.agilent.com/find/channelpartners
Get the best of both worlds: Agilent’s
measurement expertise and product
breadth, combined with channel
partner convenience.