04 | Keysight | N2780B Series AC/DC Current Probes A wide selection of current probes to meet your application’s needs - Data Sheet
Compatible Oscilloscopes
Any oscilloscope offering 1 MΩ BNC
input including Keysight Technologies,
Inc. 1000, 3000 InfiniiVision 2000X,
3000X, 5000, 6000 and 7000 Series,
and Infiniium 8000 and 9000 Series.
You must select the input impedance
of the oscilloscope to be 1 MΩ in order
to make accurate measurements. If the
oscilloscope you are using has a 50Ω
input impedance setting only, you can
purchase the Keysight E2697A 50Ω to 1
MΩ adapter for use with the Infiniium
80000 or 90000 Series or the N5449A
high impedance probe adapter for use
with the Infiniium 90000 X-Series.
Ordering Information
N2780B
2 MHz/500A AC/DC current probe
N2781B
10 MHz/150A AC/DC current probe
N2782B
50 MHz/30A AC/DC current probe
N2783B
100 MHz/30A AC/DC current probe
N2779A
3-channel power supply for
N2780B Series current probes
Note*: The amplitude accuracy specifications are guaranteed at 23°C ± 3°C (or 73°F ± 5°F)
**: Insulated conductor should be used.
Model number N2780B N2781B N2782B N2783B
Bandwidth (-3dB) DC to 2 MHz DC to 10 MHz DC to 50 MHz DC to 100 MHz
Risetime 175 ns or less 35 ns or less 7 ns or less 3.5 ns or less
Maximum current (continuous) RMS 500 A 150 A 30 A 30 A
Maximum peak current (non-continuous) 700 A peak 300 A peak; 50 A peak 50 A peak
Lowest measurable current (at ±3% accuracy
at DC, scope set to 1 mV/div and
high-resolutionmode on)
20 mA 20 mA 5 mA 5 mA
Output voltage rate 0.01 V/A (100:1) 0.01 V/A (100:1) 0.1 V/A (10:1) 0.1 V/A (10:1)
Max input voltage** 600 V CAT II,
300 V CAT III
600 V CAT II,
300 V CAT III
300 V CAT I 300 V CAT I
Amplitude accuracy*
(DC and 45 to 66 Hz, rated current)
±1.0% rdg.
± 500 mA
±1.0% rdg.
± 100 mA
±1.0% rdg.
± 10 mA
±1.0% rdg.
± 10 mA
Noise (measured with 20 MHz
bandwidth limit filter on the scope)
Equivalent to
25 mArms or less
Equivalent to
25 mArms or less
Equivalent to
2.5 mArms or less
Equivalent to
2.5 mArms or less
Temperature coefficient for sensitivity (within a
range of 0°C to 40 °C or 32 °F to 104 °F)
±2% or less ±2% or less ±2% or less ±2% or less
Effect of external magnetic fields
(in a DC to 60 Hz, 400 A/m magnetic field)
Equivalent to a
maximum of
800 mA
Equivalent to a
maximum of
150 mA
Equivalent to a
maximum of 20 mA
Equivalent to a
maximum of 5 mA
Maximum rated power 7.2 VA (with rated
current)
5.5 VA (with rated
current)
5.6 VA (with rated cur-
rent)
5.3 VA (with rated cur-
rent)
Rated supply voltage DC ±12 V ±0.5 V DC ±12 V ±1 V DC ±12 V ±0.5 V DC ±12 V ±0.5 V
Diameter of measurable conductors 20 mm dia. (0.79” dia.) 20 mm dia. (0.79”
dia.) )
5 mm dia. (0.2” dia.) 5 mm dia. (0.2” dia.)
Diameter of measurable conductors Sensor cable:
Approx. 2 m (78.7”)
Power supply cable:
Approx. 1 m (39.4”)
Sensor cable:
Approx. 2 m (78.7”)
Power supply cable:
Approx. 1 m (39.4”)
Sensor cable:
Approx. 1.5 m (78.7”)
Power supply cable:
Approx. 1 m (39.4”)
Sensor cable
Approx. 1.5 m (78.7”)
Power supply cable:
Approx. 1 m (39.4”)