Keysight Technologies InfiniiVision 2000 X-Series Oscilloscopes Data Sheet 2 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Breakthrough technology for budget conscious customers Overview of the Keysight InfiniiVision X-Series oscilloscopes Analog channels InfiniiVision 2000 X-Series InfiniiVision 3000 X-Series InfiniiVision 4000 X-Series 2 and 4 2 and 4 2 and 4 Bandwidth (upgradable) 70, 100, 200 MHz 100, 200, 350, 500 MHz, 1 GHz 200, 350, 500 MHz, 1 GHz, 1.5 GHz Digital channels 8 (MSO models or upgrade)* 16 (MSO models or upgrade) 16 (MSO models or upgrade) Maximum sample rate 2 GSa/s 5 GSa/s (1-GHz models) 4 GSa/s (100-500 MHz models) 5 GSa/s Maximum memory depth 100 kpts/channel (standard) 1 Mpt/channel (optional) 2 Mpts (standard), 4 Mpts (option) 4 Mpts (standard) Waveform update rate > 50,000 waveforms per second > 1,000,000 waveforms per second > 1,000,000 waveforms per second Display 8.5-inch display 8.5-inch display 12.1-inch capacitive touch display InfiniiScan Zone touch trigger No No Standard WaveGen 20-MHz function/ arbitrary waveform generator Single-channel function only (option) Single-channel AWG (option) Dual-channel AWG (option) Integrated digital voltmeter Yes (option) Yes (option) Yes (option) Search and navigate Yes (serial) Yes Yes Serial protocol analysis Yes (optional: CAN, LIN, IC, SPI, RS232/UART)* Yes (optional: ARINC 429, CAN FlexRay, I2C, I2S, LIN, MILSTD-1553, SPI, UART/RS232) Yes (optional: ARINC 429, CAN, FlexRay, I2C, I2S, LIN, MIL-STD-1553, SPI, UART/RS232, USB 2.0) Segmented memory Yes (option) Yes (option) Standard Mask/limit testing Yes (option) Yes (option) Yes (option) Power analysis No Yes (option) Yes (option) USB 2.0 signal quality test No No Yes (option) HDTV analysis No Yes (option) Yes (option) Advanced waveform math No Yes (option) Standard Connectivity Standard USB 2.0 (LAN/Video option) (GPIB option) Standard USB2.0 (LAN/Video option) (GPIB option) Standard USB2.0, LAN, video out (GPIB option) * The digital channels and serial protocol analysis cannot be used simultaneously on 2000 X-Series. Need bigger display and state-of-the-art usability? See the InfiniiVision 4000 X-Series, which redefines the oscilloscope experience. -- Industry's first 12.1-inch capacitive touch display -- InfiniiScan Zone touch trigger capability -- 200 MHz - 1.5 GHz DSO and MSO models -- 1,000,000 wfms/sec -- Standard segmented memory -- Fully upgradable 5 instrument in 1 -- Digital channels (MSO) -- Protocol analysis including USB 2.0 -- 20 MHz Dual-channel WaveGen with arbitrary waveform and modulation support -- Digital voltmeter (DVM) -- Application analysis including power analysis, Xilinx FPGA, and USB 2.0 signal quality -- N8900A Infiniium Offline software support -- Standard LAN, USB, and video out connectivity See www.keysight.com/find/4000X-Series for more details. 3 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet More scope The InfiniiVision 2000 X-Series offers entry-level price points to fit your budget with superior performance and optional capabilities that are not available in any other oscilloscope in its class. This Keysight Technologies, Inc. breakthrough technology delivers more scope for the same budget. With more scope, you can: -- See more of your signal more of the time with the largest screen in its class, the deepest memory and the fastest waveform update rates -- Do more with the power of 5 instruments in 1: oscilloscope, logic timing analyzer, WaveGen built-in 20 MHz function generator (optional), serial protocol triggering and decode (optional), and digital voltmeter (optional) -- Get more investment protection with the classes only fully upgradable scope, including memory and bandwidth, and a standard 5 year warranty.* *5-year warranty applies to all orders on or after 1/1/2013. 4 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet See more of your signal, more of the time Largest display Engineering for the best signal visibility starts with the largest display. Our 8.5inch WVGA display offers 50% more viewing area with 3.5 times the resolution (WVGA 800 x 480 versus 7-inch WQVGA 480 x 234). Notice that the Keysight 2000 X-Series allows you to see more of your signals, and captures the infrequent glitch that you are unable to see on other oscilloscopes in this class. Fastest update rate With Keysight-designed MegaZoom IV custom ASIC technology, the InfiniiVision 2000 X-Series family delivers up to 50,000 waveforms per second. With this speed you can see signal detail and infrequent anomalies more of the time. How does Keysight do that? Keysight-designed MegaZoom IV custom ASIC technology combines the capabilities of an oscilloscope, logic analyzer, and WaveGen built-in function generator in a compact form factor at an affordable price. 4th generation MegaZoom technology enables the industry's fastest waveform update rate with responsive deep memory acquisitions. 5 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Do more with the power of 5 instruments in 1 Best-in-class oscilloscope The InfiniiVision 2000 X-Series features Keysight's patented MegaZoom IV smart memory technology that is always enabled and always responsive providing the industry's fastest update rate at up to 50,000 waveforms per second, with no compromise if you turn on measurements or add digital channels. In addition, the 2000 X-Series offers 23 automated measurements such as voltage, time, and frequency as well as five waveform math functions including add, subtract, multiply, divide, and FFT. Industry's first economy-class mixed signal oscilloscope (MSO) The 2000 X-Series is the first instrument in its class to offer an integrated logic timing analyzer. Digital content is everywhere in today's designs and with an additional 8 integrated digital timing channels, you now have up to 12 channels of time-correlated triggering, acquisition and viewing on the same instrument. Buy a 2 or 4 channel DSO and at any time, upgrade it yourself to a MSO with a license to turn on those integrated 8 digital timing channels. Industry's first WaveGen built-in 20 MHz function generator with a modulation capability An industry first, the 2000 X-Series offers an integrated 20 MHz function generator, now available with the signal modulation capability. Ideal for educational or design labs where bench space and budget are at a premium, the integrated function generator provides stimulus output of sine, square, ramp, pulse, DC and noise waveforms to your device under test. No need to buy a separate function generator when you can get one integrated in your new oscilloscope. Turn on WaveGen at any time by ordering the DSOX2WAVEGEN option and install the license yourself. Hardware-based serial protocol decode and triggering -- Embedded serial triggering and analysis (IC, SPI) -- Computer serial triggering and analysis (RS232/422/485/UART) -- Automotive and industrial serial triggering and analysis (CAN,LIN) Keysight's InfiniiVision Series oscilloscopes are the industry's first scopes to use hardware-based serial protocol decoding. Other vendors' oscilloscopes use software post-processing techniques that slow down both waveform and decode update rate. That's especially true when using deep memory, which is often required to capture multiple packetized serial bus signals. Faster decoding with hardware-based technology enhances scope usability and, more importantly, the probability of capturing infrequent serial communication errors. After capturing a serial bus communication, you can easily perform a searchand-navigation operation based on specific criteria of your interest. Sometimes it may be necessary to correlate data from one serial bus to another. You can turn on the serial trigger and decode or digital channels simultaneously on the 2000 X-Series. Integrated digital voltmeter An industry first, the 2000 X-Series offers an integrated 3-digit voltmeter (DVM) and 5-digit frequency counter inside the oscilloscopes. The voltmeter operates through the same probes as the oscilloscope channels, however, the measurements are de-coupled from the oscilloscope triggering system so that both the DVM and triggered oscilloscope measurements can be made with the same connection. The voltmeter results are always displayed, keeping these quick characterization measurements at your fingertips. Turn on DVM at any time by ordering the DSOXDVM option. 6 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Get more investment protection with the industry's only fully upgradable oscilloscope Upgradability: Project needs change, but traditional oscilloscopes are fixed - you get what you pay for at the time of purchase. With the 2000 X-Series, your investment is protected. If you need more bandwidth (up to 200 MHz), digital channels, memory, WaveGen, integrated digital voltmeter, or measurement applications in the future, you can easily add them all after the fact. See page 21 for more information on upgradable products. Mask testing Whether performing pass/fail tests to specified standards in manufacturing or testing for infrequent signal anomalies in R&D debug, the mask test option can be a valuable productivity tool. The 2000 X-Series features hardwarebased mask testing and can perform up to 50,000 tests per second. Segmented memory When capturing low-duty cycle pulses or data bursts, you can use segmented memory acquisition to optimize acquisition memory. Segmented memory acquisition lets you selectively capture and store important segments of signals without capturing unimportant signal idle/deadtime. Segmented memory acquisition is ideal for applications including packetized serial pulses, pulsed laser, radar bursts and high-energy physics experiments. Up to 250 segments can be captured on the 2000 X-Series models with a minimum re-arm time under 19 s. 30-day trial license The 2000 X-Series comes with a one-time 30-day alloptional-features trial license. You can choose to start the 30-day trial at any time. In addition you can redeem individual optional feature 30-day trial licenses at any time by visiting www.keysight.com/find/30daytrial. This enables you to receive in effect 60 days of trial license of each optional feature. Add at the time of your purchase or upgrade later: -- Bandwidth -- Digital channels (MSO) -- Memory -- WaveGen built-in 20 MHz function generator -- Integrated digital voltmeter (DVM) -- Serial protocol analysis -- Measurement applications -- Mask testing -- Segmented memory -- Educators' lab kit 7 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Other productivity tools Reference waveforms Store up to two waveforms in the scope's non-volatile reference waveform memory locations. Compare these reference waveforms with live waveforms, and perform post analysis and measurements of stored data. You can also store waveform data on a removable USB memory device that can be recalled back into one of the available two reference memories of the scope for full waveform measurement and analysis. Save and/or transfer waveforms as XY data pairs in a comma-separated values format (*.csv) for PC analysis. Save screen images to a PC for documentation purposes in a variety of formats including: 8-bit bitmaps (*.bmp), 24-bit bitmaps (*.bmp), and PNG 24-bit images (*.png). Localized GUI and help Operate the scope in the language most familiar to you. The graphical user interface, built-in help system, front panel overlays, and user's manual are available in 13 languages. Choose from: English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish, Russian, Portuguese, Thai, Polish and Italian. During operation, access the built-in help system just by pressing and holding any button. Probe solutions Get the most out of your 2000 X-Series scope, by using the right probes and accessories for your application. Keysight offers a complete family of innovative probes and accessories for the InfiniiVision 2000 X-Series scopes. For the most up-to-date and complete information about Keysight's probes and accessories, please visit our Web site at www.keysight.com/find/scope_probes. Autoscale Quickly display any active signals and automatically set the vertical, horizontal and trigger controls for optimal viewing with the press of the autoscale button. (This feature can be disabled or enabled for the education environment via a USB thumb drive file with a SCPI remote command). 8 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Other productivity tools (continued) Connectivity and LXI compatibility Built-in USB host (one front, one back) and USB device ports make PC connectivity easy. Operate the scope from your PC and save and recall stored waveforms as well as set-up files via LAN. An optional LAN/VGA module gives you network connectivity and complete LXI class C support as well as the ability to connect to an external monitor. An optional GPIB module is also available. Only one module may be used at a time. 34840B BenchVue lets you visualize the 3000 X-Series and multiple measurements simultaneously. Save time with the ability to export measurement data to Excel, Word and MATLAB in three clicks. Monitor and control your 3000 XSeries with a mobile device from anywhere. Learn more at www.keysight.com/find/BenchVue. View Scope enables simple and free time-correlated me asurements between a 2000 X-Series oscilloscope and a Keysight 16900, 16800, 1690 or 1680 Series logic analyzer. Virtual front panel In addition to the traditional VNC virtual front panel remote operation through your favorite PC Web browser, the InfiniiVision X-Series supports remote oscilloscope control from your tablet devices. The tablet virtual front panel looks and acts as the real front panel on the oscilloscope. Control the setting, save/recall data, get image, and more. Secure erase The secure erase feature comes standard with all InfiniiVision X-Series models. At the press of a button, internal nonvolatile memory is clear of all setup, reference waveforms, and user preferences, ensuring the highest level of security in compliance with National Industrial Security Program Operation Manual (NISPOM) Chapter 8 requirements. 9 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Other productivity tools (continued) Infiniium Offline oscilloscope analysis software (N8900A) Keysight's Infiniium Offline PC-based analysis oscilloscope software allows you to do additional signal viewing, analysis and documentation tasks away from your scope. Capture waveforms on your scope, save to a file, and recall the waveforms into Infiniium Offline. The application supports a variety of popular waveform formats from multiple oscilloscope vendors and includes the following features: View and analyze away from your scope and target system Navigate -- Pan and zoom to anywhere in the data record. Navigate in time, or between bookmarks. View -- Up to 8 waveforms simultaneously, 1, 2, or 4 grids (stacked, side by side, custom layout, zoom) Measurements -- Over 50 automated measurements -- View up to 20 simultaneously -- User-customizable result window (size, position, information) -- X & Y markers with dynamic delta values Analyze -- 20 math operators including FFT and filters -- Up to four independent/cascaded math functions -- Measurement histogram Use familiar scope controls to quickly navigate and zoom in to any event of interest. View windows -- Analog, math, spectral, measurement results (simultaneous, tabbed, or undocked) Documentation -- Right-click to copy -- Up to 100 bookmarks -- Annotated axis values -- Markers with dynamic delta value updates when moved -- One step save/load setup and all waveforms Analysis upgrades (optional) -- Protocol decode for I2C/SPI, RS232/UART, CAN/ LIN/ FlexRay, SATA, 8B/10B, digRF v4, JTAG , MIPI D-Phy, SVID, Ethernet 10G KR, PCIe 1, 2, 3, USB 2, 3, HSIC -- Jitter analysis -- Serial data analysis Add bookmarks and call outs to produce friendly and useful documentation. 10 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Other productivity tools (continued) Keysight Spectrum Visualizer (ASV) software This PC-based software package connects to the scope via USB or ethernet connection and uses the Keysight I/O libraries to communicate. It provides advanced FFT frequency domain analysis at a cost-effective price as well as spectrum and spectrogram analysis with an intuitive user interface that RF engineers are familiar with. Tools include: Spectrum measurements -- Power (dBm) vs. frequency -- Horizontal (x-axis): Specify center frequency and frequency span, or start and stop frequencies -- Vertical (y-axis): Specify reference level (dBm) and scale (dB/div) -- Settable resolution bandwidth -- Flat top, Gaussian, or Hanning windows applied to the time domain data for the FFT analysis -- Marker to peak amplitude, and marker to center frequency. -- Marker peak search can be enabled for time-varying signals -- Multiple marker, with delta X and delta Y readouts Waterfall View for ASV Spectrogram Measurement Acquisition and display modes -- Free Run (continuous ), Triggered, Stop, Single, Preset -- Triggered mode: specify trigger power level (dBm), single or continuous sweep -- Enable/disable y-axis label -- Enable/disable main trace display -- Max hold display mode -- Gated Measurements -- Multiple viewing options - Spectrogram - waterfall - 3D -- Changeable scaling settings on main window -- Local language support -- Multiple oscilloscopes can be configured to allow user to rapidly switch between multiple instruments Arbitrary waveform generator source control -- 20 MHz sine wave -- 10 MHz square wave -- Pulsed waveform -- WaveGen source settings can be altered while ASV is running for interactive signal source and analysis capability Close-Up Detail on Frequency Shift Keying (FSK) Characteristics with the ASV Spectrogram Measurement 11 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Designed with education in mind Quickly and easily set up or upgrade a teaching lab Teach your students what an oscilloscope is and how to perform basic measurements with the Educator's Oscilloscope Training Kit (DSOXEDK). It includes training tools created specifically for electrical engineering and physics undergraduate students and professors. It contains an array of built-in training signals, a comprehensive oscilloscope lab guide and tutorial written specifically for the undergraduate student, and an oscilloscope fundamentals PowerPoint slide set for professors and lab assistants. For more information, refer to www.keysight.com/find/EDK. Also available are DreamCatcher's full semester application-specific courseware written around Keysight test and measurement equipment: www.dreamcatcher.asia/cw. With features such as the ability to disable autoscale and the 50- Ohm input data path, the InfiniiVision X-Series is a perfect choice for education. Get your students to quickly put the scope to work Intuitive localized front panel design with pushable knobs for quick access to commonly used oscilloscope functions helps students spend more time learning the concepts and less time learning how to use the oscilloscope. Enable your students to answer their own questions with the localized built-in help system that provides quick access by simply pressing and holding any button. Keysight Technologies DSOXEDK Educator's Oscilloscope Training Kit Stretch your budget over the long term Save money with an industry-exclusive built-in 20 MHz WaveGen, instead of a separate function generator. Buy what you need today and protect your investment in the future with the only oscilloscopes in this class with upgradable bandwidth, 8 digital channels (MSO), WaveGen, integrated digital voltmeter and measurement applications. Get long scope life and keep repair costs to a minimum with a standard 5-year warranty*, and an instrument reliability you've come to expect from the leader in test and measurement equipment. *Applies to all orders on or after 1/1/2013. Optimize lab bench space With 5 instruments in 1, you will save on precious lab bench space by getting an oscilloscope, logic timing analyzer, serial protocol analyzer, WaveGen function generator and integrated digital voltmeter all in one innovative instrument with a footprint that is only 5.57 inches deep. With the large 8.5inch WVGA display, you can easily view all signals on one screen with enough viewing area for more than one student to view. Lab Guide and Tutorial for Undergraduate Electrical Engineering and Physics Students 12 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Oscilloscope shown actual size Oscilloscope shown actual size www.keysight.com/find/BenchVue 8.5-inch high 8.5inch high resolution resolution wide wide screen display screensubtle display reveals details reveals that most subtle scopesdetails that show mostyou scopes don't don't show you WaveGen - - WaveGen Industry first first Industry built-in built-in function function generator generator Get eightintegrated integrated Get up up to eight digital digital channels channels Built-in Built-inUSB USBport port makes it makes iteasy easytoto save save your yourwork workand and update your system update your system software softwarequickly quickly 13 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Navigation front panel controls make it easy to play, stop, rewind and fast forward through waveforms Dedicated keys for quick access to digital channels, serial analysis, math functions and reference waveforms Quick summary display of sample rate, channel settings and measurements All front panel knobs are pushable Demo and training signals Quickly pan and zoom for analysis with MegaZoom IV's instant response and optimum resolution Autoscale lets you quickly display any analog or digital active signals, automatically setting the vertical, horizontal and trigger controls for the best display, while optimizing memory Integrated digital voltmeter 14 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Configuring your InfiniiVision X-Series oscilloscope Step 1. Choose your bandwidth and channel count. InfiniiVision 2000 X-Series scopes 2002A Bandwidth* (-3 dB) Calculated rise time (10-90%) Input channels DSOX MSOX 2004A 2012A 70 MHz 5 ns 2 2+8 2014A 2022A 100 MHz 3.5 ns 4 4+8 2 2+8 2024A 200 MHz 1.75 ns 4 4+8 2 2+8 4 4+8 * For example, if you chose 100 MHz, 2+8 channels, the model number will be MSOX2014A. Step 2. Tailor your scope with measurement applications to save time and money.1 Application 2000 X-Series 1 Megapoint memory upgrade DSOX2MEMUP (-010) Embedded serial triggering and analysis (IC, SPI) DSOX2EMBD (-LSS)* Computer serial triggering and analysis (RS232/422/485/UART) DSOX2COMP (-232)* Automotive serial triggering and analysis (CAN, LIN) DSOX2AUTO (-AMS)* WaveGen (built-in function generator) DSOX2WAVEGEN (-001) Integrated digital voltmeter DSOXDVM (-DVM) Educator's kit DSOXEDK (-EDK) Mask testing DSOX2MASK (-LMT) Segmented memory DSOX2SGM (-SGM) InfiniView oscilloscope analysis software N8900A Keysight spectrum visualizer (ASV) 64997A 1. See pages 20-21 for more detailed information on upgradability, and installation process. * Serial trigger and decode application will not run simultaneously with digital channels. Step 3. Choose your probes2 Probes 2000 X-Series N2862B 150 MHz 10:1 passive probe Standard one per channel for 70 and 100 MHz models N2863B 300 MHz, 10:1 passive probe Standard one per channel for 200 MHz models N6459-60001 8-channel logic probe and accessory kit Standard on MSO models or with DSOX2MSO upgrade N2889A 350 MHz 10:1/1:1 passive probe Optional 10070D 20 MHz 1:1 passive probe with probe ID Optional 10076A 250 MHz 100:1, 4 kV high-voltage passive probe with probe ID Optional N2791A 25 MHz, 700 V high-voltage differential probe Optional N2792A 200 MHz 10:1 differential probe Optional 1146A 100 kHz, 100 A, AC/DC current probe Optional 2. See page 20 for probe compatibilty table. For more information on probes and accessories, see the Keysight literature 5968-8153EN Step 4. Add the final touches. Recommended accessories LAN/VGA connection module 2000 X-Series DSOXLAN GPIB connection module DSOXGPIB Rack mount kit N6456A Soft carrying case and front panel cover N6457A Hard copy manual N6458A Front panel cover (only) N2747A 15 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Performance characteristics Specification overview 2002A 2004A 2012A 2014A 2022A 2024A Bandwidth * (-3 dB) 70 MHz 100 MHz 200 MHz "Calculated rise time (10-90%)" 5 ns 3.5 ns 1.75 ns Input channels DSOX 2 4 2 4 2 4 MSOX 2+8 4+8 2+8 4+8 2+8 4+8 Maximum sample rate 2 GSa/s half-channel interleaved, 1 GSa/s per channel Maximum memory depth 100 kpts per channel (standard), 1 Mpt per channel (optional with DSOX2MEMUP) Display size and type 8.5-inch WVGA with 64 levels of intensity grading Waveform update rate > 50,000 waveforms per second Vertical system analog channels Input coupling AC, DC Input sensitivity range 1 mV/div to 5 V/div** Input impedance 1 M 2% (11 pF) Vertical resolution 8 bits (measurement resolution is 12 bits with averaging) Dynamic range 8 divisions from center screen Maximum input voltage CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk CAT II 300 Vrms, 400 Vpk; with N2862B or N2863B 10:1 probe: 300 Vrms DC vertical accuracy [DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] ** DC vertical gain accuracy* 3% full scale ( 10 mV/div); 4% full scale (< 10 mV/div) ** DC vertical offset accuracy Channel-to-channel isolation Position/offset range 0.1 div 2mV 1% of offset setting 200 MHz ~ 1 GHz 40 dB from DC to maximum specified bandwidth of each model 1.5 GHz 40 dB from DC to 1 GHz, 35 dB from 1 GHz to 1.5 GHz 1 M 1 mV to 200 mV/div: 2 V, > 200 mV to 5 V/div: 50V Hardware bandwidth limits Approximately 20 MHz (selectable) Horizontal system analog channels 2002A Time base range 2012A 2014A 5 ns/div to 50 s/div Horizontal resolution 2022A 2024A 2 ns/div to 50 s/div 2.5 ps Time base accuracy* Time base delay time range 2004A 25 ppm 5 ppm per year (aging) Pre-trigger Post-trigger Channel-to-channel deskew range Time accuracy (using cursors) Greater of 1 screen width or 200 s (400 s in interleaving mode) 1 s to 500 s 100 ns (time base accuracy* reading) (0.0016 * screen width) 100 ps * Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and from 10 C firmware calibration temperature. ** 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity setting. 16 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Performance characteristics (continued) Acquisition modes Normal Peak detect Capture glitch as narrow as 500 ps at all timebase settings. Averaging Select from 2,4,8,16, 64... to 65,536 High resolution mode 12 bits of resolution when 20 s/div Segmented Re-arm time= 19 s (minimum time between trigger events) Trigger system Trigger modes -- Normal (triggered): requires trigger event for scope to trigger -- Auto: triggers automatically in absence of trigger event -- Single: triggers only once on a trigger event, press [Single] again for scope to find another trigger event, or press [Run] to trigger continuously in either Auto or Normal mode -- Force: front panel button that forces a trigger Trigger coupling Coupling selections: AC, DC, noise reject, LF reject and HF reject. Trigger source Each analog channel, each digital channel (MSO models or DSOX2MSO upgrade, Ext, WaveGen, line) Trigger sensitivity (internal)* < 10 mV/div: greater of 1 div or 5 mV; 10 mV/div: 0.6 div Trigger sensitivity (external)* 200 mV (DC to 100 MHz); 350 mV (100 MHz - 200 MHz) External trigger input Included on all models Trigger type selections All 2000 X-Series models Edge Trigger on a rising, falling, alternating or either edge of any source Pulse width Trigger on a pulse on a selected channel, whose time duration is less than a value, greater than a value, or inside a time range -- Minimum duration setting: 2 ns- 10 ns (depends on bandwidth) -- Maximum duration setting: 10 s Pattern Trigger when a specified pattern of high, low, and don't care levels on any combination of analog, digital, or trigger channels is [entered | exited]. Pattern must have stabilized for a minimum of 2 ns to qualify as a valid trigger condition. Video Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC, PAL, SECAM, PAM-M) IC (optional) Trigger on IC (Inter-IC bus) serial protocol at a start/stop condition or user defined frame with address and/or data values. Also trigger on missing acknowledge, address with no accq, restart, EEPROM read, and 10-bit write. SPI (optional) Trigger on SPI (Serial Peripherial Interface) data pattern during a specific framing period. Supports positive and negative Chip Select framing as well as clock Idle framing and userspecified number of bits per frame. CAN (optional) Trigger on CAN (controller area network) version 2.0A and 2.0B signals. Trigger on the start of frame (SOF) bit (standard). Remote frame ID (RTR), data frame ID (~RTR), remote or data frame ID, data frame ID and data, error frame, all errors, acknowledge error and overload frame. LIN (optional) Trigger on LIN (Local Interconnect Network) sync break, sync frame ID, or frame ID and data. RS232/422/485/UART (optional) Trigger on Rx or Tx start bit, stop bit or data content * Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and from 10 C firmware calibration temperature. 17 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Performance characteristics (continued) Cursors Types Amplitude, time , frequency (FFT), manual, tracking, binary, HEX Measurements T, 1/T, V/X, 1/X, Y, Phase and Ratio Cursors** -- Single cursor accuracy: [DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] -- Dual cursor accuracy: [DC vertical gain accuracy + 0.5% full scale]* Automatic waveforms measurements Voltage Snapshot all, maximum, minimum, peak-to-peak, top, base, amplitude, overshoot, preshoot, average- N cycles, average-full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N cycles Time Period, frequency, rise time, fall time, + width, - width, duty cycle, delay AB (rising edge), delay AB (falling edge), phase AB (rising edge,) and phase AB (falling edge) Waveform math Operators Add, subtract, multiply, FFT FFT Windows: Hanning, flat top, rectangular; Blackman-Harris - up to 64 kpts resolution Sources Math functions available between any two channels Display characteristics Display 8.5-inch WVGA Resolution 800 (H) x 480 (V) pixel format (screen area) Interpolation Sin(x)/x interpolation (using FIR filter; used when there is less than one sample per column of the display) Persistence Off, infinite, variable persistence (100ms-60s) Intensity gradation 64 intensity levels Modes Normal XY - XY mode changes the display from voltage versus time scale to a volts versus volts scale Roll - Displays the waveform moving across the screen from right to left much like a strip chart recorder MSO (digital channels) Upgradable from DSO Yes MSO channels 8 channels (D0 to D7) Maximum sample rate 1 GSa/s Maximum record length 500 kpts per channel (digital channels only) 125 kpts per channel (analog and digital channels) Threshold selections TTL (+1.4 V), CMOS (+2.5 V), ECL (-1.3 V), User-definable (8.0 V in 10 mV stops) Threshold accuracy* (100 mV + 3% of threshold settings) Maximum input dynamic range 10 V about threshold Minimum voltage swing 500 mVpp Input impedance 100 k 2% at probe tip, ~8 pF Minimum detectable pulse width 5 ns Channel-to-channel skew 2 ns (typical), 3 ns (maximum) * Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and from 10 C firmware calibration temperature. ** 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting. 18 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Performance characteristics (continued) WaveGen - built-in function generator (specifications are typical) Waveforms Sine, square, pulse, triangle, ramp, noise, DC Sine -- Frequency range: 0.1 Hz to 20 MHz -- Amplitude flatness: 0.5 dB (relative to 1 kHz) -- Harmonic distortion: -40 dBc -- Spurious (non harmonics): -40 dBc -- Total harmonic distortion: 1% -- SNR (50 ohm load, 500 MHz BW) : 40 dB (Vpp 0.1 V); 30 dB (Vpp <0.1 V) Square wave/pulse -- Frequency range: 0.1 Hz to 10 MHz -- Duty cycle: 20 to 80% -- Duty cycle resolution: Larger of 1% or 10 ns -- Pulse width: 20 ns minimum -- Pulse width resolution: 10 ns or 5 digits, whichever is larger -- Rise/fall time: 18 ns (10 to 90%) -- Overshoot: < 2% -- Asymmetry (at 50% DC): 1% 5 ns -- Jitter (TIE RMS): 500 ps Ramp/triangle wave -- Frequency range: 0.1 Hz to 100 kHz -- Linearity: 1% -- Variable symmetry: 0 to 100% -- Symmetry resolution: 1% Noise Bandwidth: 20 MHz typical Frequency -- Sine wave and ramp accuracy: -- 130 ppm (frequency < 10 kHz) -- 50 ppm (frequency > 10 kHz) -- Square wave and pulse accuracy: -- [50+frequency/200] ppm (frequency < 25 kHz) -- 50 ppm (frequency 25 kHz) -- Resolution: 0.1 Hz or 4 digits, whichever is larger Amplitude -- Range: -- 20 mVpp to 5 Vpp into Hi-Z -- 10 mVpp to 2.5 Vpp into 50 ohms -- Resolution: 100 V or 3 digits, whichever is larger -- Accuracy: 2% (frequency = 1 kHz) DC offset -- Range: -- 2.5 V into Hi-Z -- 1.25 V into 50 ohms -- Resolution: 100 V or 3 digits, whichever is larger -- Accuracy: 1.5% of offset setting 1.5% of amplitude 1 mV Trigger output Trigger output available on Trig out BNC 19 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Performance characteristics (continued) WaveGen - built-in function generator (specifications are typical) (Continued) Modulation Modulation types: AM, FM, FSK Carrier waveforms: sine, ramp Modulation source: Internal (no external modulation capability) AM: Modulation waveform: sine, square, ramp Modulation frequency (1Hz to 20 kHz) Depth: 0% to 100% FM: Modulation: sine, square, ramp (1 Hz to 20 kHz) Modulation frequency (1 Hz to 20 kHz) Minimum carrier frequency: 10 kHz Minimum deviation: 1 Hz Maximum deviation: 100 kHz or (carrier frequency - 9 kHz), whichever is smaller. FSK: Modulation: 50% duty cycle square wave FSK rate: 1 Hz to 20 kHz Minimum carrier frequency: 10 kHz Minimum hop frequency: 2 * FSK rate Integrated Digital Voltmeter Functions ACrms, DC, DCrms, Frequency Resolution ACV/DCV: 3 digits Frequency: 5.5 digits Measuring rate 100 times/ second Autoranging Automatic adjustment of vertical amplification to maximize the dynamic range of measurements. Range meter Graphical display of most recent measurement, plus extrema over the previous 3 seconds. Measurement range (specifications are typical) Frequency Range Vertical Range Vertical Accuracy ACRms 20 Hz - 100 KHz [DC vertical gain accuracy + 0.5% full scale DCRms 20 Hz - 100 KHz DC NA 100 MHz to 500 MHz: 1 mV/div to 5 V/ div** (1 M and 50 Ohm) 1 GHz model: 1 mV/div to 5 V/div** (1 M), 1mV/div to 1V/div (50 Ohm) Frequency counter 1Hz - BW of Scope <10 mV/div: greater of 1 div or 5 mV; 10 mV/div: 0.6 div 25 ppm 5 ppm per year (aging) [DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] [DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] 20 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet InfiniiVision X-Series physical characteristics Connectivity Standard ports One USB 2.0 hi-speed device port on rear panel. Supports USBTMC protocol Two USB 2.0 hi-speed host ports, front and rear panel Supports memory devices, printers and keyboards Optional ports GPIB, LAN, WVGA video out General and environmental characteristics Power line consumption 100 watts Power voltage range 100-120V, 50/60/400 Hz; 100-240V, 50/60 Hz 10% auto ranging Temperature Operating: 0 to +55 C Non-operating: -30 to +71 C Humidity Operating: Up to 80% RH at or below +40 C; up to 45% RH up to +50 C Non-operating: Up to 95% RH up to 40 C; up to 45% RH up to 50 C Altitude Operating: up to 4,000 m, Non-operating 15,300 m Electromagnetic compatibility Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN 61326-1:2006 Group 1 Class A requirement CISPR 11/EN 55011 IEC 61000-4-2/EN 61000-4-2 IEC 61000-4-3/EN 61000-4-3 IEC 61000-4-4/EN 61000-4-4 IEC 61000-4-5/EN 61000-4-5 IEC 61000-4-6/EN 61000-4-6 IEC 61000-4-11/EN 61000-4-11 Canada: ICES-001:2004 Australia/New Zealand: AS/NZS Safety UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04 Dimensions 381 mm (15 in) W x 204 mm (8 in) H x 142 mm (5.6 in) D Weight Net: 3.9 kg (8.5 lbs), shipping: 4.1 kg (9.0 lbs) Nonvolatile storage Reference waveform display 2 internal waveforms or USB thumb drive Waveform storage Set up, .bmp, .png, .csv, ASCII, XY, reference waveforms, .alb, .bin, lister, mask , HDFS Max USB flash drive size Supports industry standard flash drives Set ups without USB flash drive 10 internal setups Set ups with USB flash drive Limited by size of USB drive Included standard with oscilloscope Standard 5-year warranty* (90 days for unserialized accessories) Standard secure erase Standard Probe N2862B 150 MHz 10:1 passive probe Standard one per channel for 70 and 100 MHz models N2863B 300 MHz, 10:1 passive probe Standard one per channel for 200 MHz models N6459-60001 8-channel logic probe and accessory kit Standard on MSO models or with DSOX2MSO upgrade Built-in help language support for English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish, Russian, Portuguese and Italian, Certificate of Calibration, Documentation CD Interface language support GUI menus: English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish, Russian, Portuguese, Thai, Polish and Italian Localized power cord *Applies to all orders on or after 1/1/2013. For MET/CAL procedures, click on the Cal Labs solutions link below Cal Labs Solutions http://www.callabsolutions.com/products/Keysight/ These procedures are FREE to customers. 21 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet License-only bandwidth upgrades and measurement applications Bandwidth upgrade models Process description Process description 2000 X-Series 1 DSOX2BW12 70 MHz to 100 MHz, 2 ch, License only DSOX2BW14 70 MHz to 100 MHz, 4 ch, License only DSOX2BW22 100 MHz to 200 MHz, 2 ch, License only DSOX2BW24 100 MHz to 200 MHz, 4 ch, License only Measurement applications DSOX2MEMUP Upgrade to 1 Mpts per channel DSOX2COMP Computer serial triggering and analysis (RS232/422/485/UART) DSOX2AUTO Automotive serial triggering and analysis (CAN, LIN) DSOX2EMBD Embedded serial triggering and analysis (I2C, SPI) DSOX2WAVEGEN WaveGen (built-in function generator) DSOXDVM Integrated digital voltmeter DSOXEDK Educator's kit DSOX2MASK Mask testing DSOX2SGM Segmented memory DSOX2MSO Upgrade to 8 digital timing channels 2 3 4 5 6 Place order for a license only bandwidth upgrade or measurement appliction product to a Keysight sales partner. If multiple bandwidth upgrade steps are needed, order all the corresponding upgrade products required to get from current bandwidth to desired bandwidth. In the case where the new bandwidth requires higher bandwidth passive probes, they are included with the upgrade. For the DSOX2BW22 and DSOX2BW24, the N2863B 10:1 300 MHz passive probes (1 per channel) will be sent with the upgrade. Receive a paper or electronic .pdf Entitlement Certificate document for any of the orderable measurement applications For bandwidth upgrades only, you receive a stick-on label document indicating upgraded bandwidth specification. Use Entitlement Certificate or electronic .pdf document containing instructions and certificate number needed to generate a license file for a particular 2000 or 3000 X-Series oscilloscope model number and serial number unit. Receive the licensed file and installation instructions via email. Copy license file (.lic extension) from email to a USB drive and follow instructions in email to install the purchased bandwidth upgrade or measurement application on the oscilloscope. For bandwidth upgrades only, attach bandwidth upgraded stick-on labels to front and rear panels of the oscilloscope. Model number and serial number of the oscilloscope do not change. 22 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet Keysight Technologies Oscilloscopes Multiple form factors from 20 MHz to > 90 GHz | Industry leading specs | Powerful applications 23 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.axiestandard.org AdvancedTCA(R) Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Keysight is a founding member of the AXIe consortium. ATCA(R), AdvancedTCA(R), and the ATCA logo are registered US trademarks of the PCI Industrial Computer Manufacturers Group. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Keysight is a founding member of the LXI consortium. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight's commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight's measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/2000x-series For more information on Keysight Technologies' products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada Brazil Mexico United States (877) 894 4414 55 11 3351 7010 001 800 254 2440 (800) 829 4444 Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Other AP Countries 1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 (65) 6375 8100 Europe & Middle East Austria Belgium Finland France Germany Ireland Israel Italy Luxembourg Netherlands Russia Spain Sweden Switzerland United Kingdom 0800 001122 0800 58580 0800 523252 0805 980333 0800 6270999 1800 832700 1 809 343051 800 599100 +32 800 58580 0800 0233200 8800 5009286 800 000154 0200 882255 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-09-24-14) This information is subject to change without notice. (c) Keysight Technologies, 2012 - 2014 Published in USA, August 2, 2014 5990-6618EN www.keysight.com