GMS-10 RVS Series Oxygen Sensors Accurate oxygen measurement DESCRIPTION The GMS-10 is a high-accuracy oxygen sensor suitable for produces the 700 C required for the ZrO2 to operate as an O2 many applications. Originally developed for boiler pump. The time taken for the pump to achieve specific applications, the product range has been extended to offer minimum and maximum pressures within this chamber is a versions suitable for oxygen monitoring on board commercial measure of the partial oxygen pressure of the environment. and military aircraft, exhaust gas testing for automotive and medical test equipment. The sensor employs two ZrO2 discs The GMS-10 has to be operated by an electronic measuring with a small hermetically sealed chamber in between. One of circuit that controls the sensor operation and signal the discs functions as a reversible oxygen pump, which is processing. This circuit can either be incorporated into the used to successively fill and empty the chamber. The second customer's own electronics, or be purchased as a separate disc measures the ratio of the partial pressure difference and box of electronics from Honeywell. generates a corresponding sense voltage. A heat element FEATURES TYPICAL APPLICATIONS * High-accuracy measurement * Heating boiler control * Low power consumption * Industrial process control * No reference gas required * Control of aircraft on board oxygen generation systems * Linear output signal * Automotive exhaust gas diagnostics * No need for temperature stabilization * Service instruments * Function testing and calibration in ambient air * Monitoring environmental oxygen * Long life WARNING MISUSE OF DOCUMENTATION * The information presented in this product sheet is for reference only. Do not use this document as a product installation guide. * Complete installation, operation, and maintenance WARNING PERSONAL INJURY DO NOT USE these products as safety or emergency stop information is provided in the instructions supplied with devices or in any other application where failure of the each product. product could result in personal injury. Failur e to comply with these instr uctions could rresult esult in Failure instructions Failur e to comply with these instr uctions could rresult esult in Failure instructions death or serious injur y. injury death or serious injur y. injury MOUNTING DIMENSIONS IN MM/INCHES (FOR REFERENCE ONLY) WARRANTY/REMEDY Oxygen pressure range 2 mbar - 3 bar Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell's standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The for egoing is buyer emedy and foregoing buyer''s sole rremedy is in lieu of all war ranties, expr essed warranties, expressed or implied, including those of mer chantability and fitness for a merchantability par ticular purpose. In no event shall particular Honeywell be liable for consequential, special, or indir ect damages. indirect While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application. Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no Operational temperature 700 C responsibility for its use. Stand by temperature 500 C (typ.) Sensitivity 1.05 ms/mbar Accuracy < 5 mbar Response time < 15 s Warm up time < 100 s 83,0 3.27 150,0 5.91 5,0 0.20 O 12,0 0.47 WIRING DIAGRAM Heater (3&5) Yellow Pump (1) Red Common (2) Black Sense (4) Blue TECHNICAL DATA Electrical Characteristics Sensor voltage levels (recommended ), V 45-64-85 mV Pump current (recommended), A 40 uA Heater supply, V/A 4.0 V (1.7A) Heater supply (stand by) 2.0 V Pump resistance at 700 C, Ohm dc 1 kOhm (typ.) ac 1 kHz 120 Ohm (typ.) Operational specifications Warm up time (from stand by) < 20 s Permissable gas temperature -100 /+250 C Gas flow rate 0-10 m/s SALES AND SERVICE Honeywell serves its customers through a worldwide network of sales offices, representatives and distributors. For application assistance, current specifications, pricing or name of the nearest Authorized Distributor, contact your local sales office or: E-mail: info.sc@honeywell.com Mechanical characteristics Inter net: www.honeywell.com/sensing Internet: Repetitive permissable acceleration 5g Incidental permissable acceleration 30 g Phone and Fax: Europe UK +44 (0) 1698 481481 +44 (0) 1698 481676 Fax USA/Canada +1-800-537-6945 +1-815-235-6545 Fax International Automation and Contr ol Solutions Control Sensing and Control Honeywell Newhouse Industrial Estate Motherwell, ML1 5SB UK www.honeywell.com 100458-EN Issue 2 UK07 GLO September 2004 Copyright 2004 Honeywell International Inc. +1-815-235-6847 Honeywell Data Sheet GMS10-18 Oxygen Sensors - Screw Fit Probe FEATURES * Oxygen pressure range 2 mbar - 3bar * Zirconium dioxide (ZrO2) sensing elements * Non-consumption technology * Integral heating element * No need for temperature stabilisation * No reference gas required * High accuracy * Linear output signal * Operates with external interface boards * M18x1.5 screw mount GMS10-18A ELECTRICAL CONNECTION SPECIFICATIONS Heater Voltage* Porous Cap (15mm) Stand by Molex Connector 4.35VDC 0.1VDC (1.85A) 2VDC (0.85A) Pump resistance @ 700C** < 6k Permissible gas temperature -100 to 250C Gas flow rate 1: Pump (Red) 2: Common (Black) 3: Heater (Yellow) 4: Sense (Blue) 5: Heater (Yellow) 0 to 10 m/s Repetitive permissible acceleration 5g Incidental permissible acceleration 30 g * It is important to measure the heater voltage as close to the sensor as possible due to voltage drops in the supply cable. Heater can also be operated with an equivalent a.c. or PWM signals. ** The constant current source used in the pump circuit should be designed to drive a load of up to 6k (c) 2009 Honeywell International Inc. 1 www.honeywell.com Honeywell Data Sheet GMS10-18 Oxygen Sensors - Screw Fit Probe OUTLINE DRAWING Full Porous Cap (Length 15mm) Molex Connector Mass: 46g Molex mating connector 39-01-2061 (plastic plug) 39-00-0041 (pins) (c) 2009 Honeywell International Inc. X = 55mm for GMS10-18A X = 45mm for GMS10-18B X = 28mm for GMS10-18C 2 www.honeywell.com Honeywell Data Sheet GMS10-18 Oxygen Sensors - Screw Fit Probe PERFORMANCE CHARACTERISTIC Characteristics Min. Oxygen pressure range Typ. 2 Max. 3000 Accuracy 5 Internal operational temperature (full porous cap) mbar C 700 (4.35VDC) Response time (10-90% step, full porous cap) Unit 4 Warm up time (prior to sensor operation) 100 Warm up time (from stand by) 20 Output stabilisation time s ~180 ORDERING INFORMATION Part number Housing Termination GMS-18A Probe with M18 x 1.5 nut X = 55mm 0.15m Cable with Molex Connector GMS-18B Probe with M18 x 1.5 nut X = 45mm 0.15m Cable with Molex Connector GMS-18C Probe with M18 x 1.5 nut X = 28mm 0.15m Cable with Molex Connector WARNING Personal Injury DO NOT USE these products as safety or Emergengy Stop devices or in any other application Where failure of the product could result in Personal injury. Failure to comply with these instructions could Result in death or serious injury. (c) 2009 Honeywell International Inc. 3 www.honeywell.com Honeywell Data Sheet GMS10, KGZ12 and KGZ50 Oxygen Sensors - Probe Style FEATURES * Oxygen pressure range 2 mbar - 3bar * Zirconium dioxide (ZrO2) sensing elements * Non-consumption technology * Integral heating element * No need for temperature stabilisation * No reference gas required * High accuracy * Linear output signal * Operates with external interface boards * 80mm, 220mm and 400mm probes available GMS10RVS ELECTRICAL CONNECTION SPECIFICATIONS Heater Supply* Porous Lid Cap (17.8mm) Stand by Full Porous Cap (15.0mm) Stand by Molex Connector 4.00VDC 0.1VDC (1.7A) 1.65VDC (0.7A) 4.35VDC 0.1VDC (1.85A) 2VDC (0.85A) Pump resistance @ 700C** 1: Pump (Red) 2: Common (Black) 3: Heater (Yellow) 4: Sense (Blue) 5: Heater (Yellow) < 6k Permissible gas temperature Standard Temp. -100 to 250C Gas flow rate 0 to10 m/s Repetitive permissible acceleration 5g Incidental permissible acceleration 30 g Leadwires Pump (Red) Common (Black) Heater (Yellow) Sense (Blue) Heater (Yellow) * It is important to measure the heater voltage as close to the sensor as possible due to voltage drops in the supply cable. Heater can also be operated with an equivalent a.c. or PWM signals. ** The constant current source used in the pump circuit should be designed to drive a load of up to 6k (c) 2009 Honeywell International Inc. 1 www.honeywell.com Honeywell Data Sheet GMS10, KGZ12 and KGZ50 Oxygen Sensors - Probe Style OUTLINE DRAWINGS KGZ50 Porous Lid Cap Mass: 64g GMS10RVS Porous Lid Cap Mass: 20g GMS10WH Full Porous Cap Mass: 24g Molex mating connector 39-01-2061 (plastic plug) 39-00-0041 (pins) KGZ12SHORT Porous Lid Cap Mass: 105g KGZ12 Porous Lid Cap Mass: 195g Both KGZ12Short & KGZ12 Length = 200mm min 20 AWG All dimensions in mm. Sensor lengths are approximate. (c) 2009 Honeywell International Inc. 2 www.honeywell.com Honeywell Data Sheet GMS10, KGZ12 and KGZ50 Oxygen Sensors - Probe Style PERFORMANCE CHARACTERISTIC Characteristics Min. Oxygen pressure range Typ. Max. 2 3000 Accuracy 5 Internal operational temperature (Porous Lid Cap) 700 (4.00VDC) Internal operational temperature (Full Porous Cap) 700 (4.35VDC) mbar C Response time (10-90% step, Porous Lid Cap) 15 Response time (10-90% step, Full Porous Cap) 4 Warm up time (prior to sensor operation) 100 Warm up time (from stand by) 20 Output stabilisation time Unit s ~180 ORDERING INFORMATION Part number Probe Length Permissible Gas Temperature Sensor Response Termination GMS10RVS 80mm -100 to 250C < 15s 150mm Cable, Molex Connector GMS10WH 80mm -100 to 250C < 4s 150mm Cable, Molex Connector KGZ50 80mm -100 to 250C < 15s 1.2m Cable, Molex Connector KGZ12SHORT 220mm -100 to 250C < 15s Leadwires, 20 AWG, 200mm KGZ12 420mm -100 to 250C < 15s Leadwires, 20 AWG, 200mm WARNING Personal Injury DO NOT USE these products as safety or Emergengy Stop devices or in any other application Where failure of the product could result in Personal injury. Failure to comply with these instructions could Result in death or serious injury. (c) 2009 Honeywell International Inc. 3 www.honeywell.com MF010 Series Low Temperature and MF020 Series High Temperature Oxygen Sensors DESCRIPTION The MF010 Series Low Temperature Oxygen Sensors is designed to determine the oxygen concentration in air or inert gas mixtures with temperatures of -100 C to 250 C [-148 F to 482 F] max. The MF020 Series High Temperature Oxygen Sensors is designed to determine the oxygen concentration in air or inert gas mixtures with temperatures of -100 C to 400 C [-148 F to 752 F] max. These products are particularly suitable for measuring oxygen in areas that are not easily accessible, or in closed systems, such as ventilation pipes and containers. The MF010 and MF020 Series may both be supplied with a measuring range of 0.1 vol%O2 to 25 vol%O2 and 0.1 vol%O2 to 100 vol%O2. The entire measurement range is linear for both series. The sensor detects the gas mixture's oxygen partial pressure (absolute oxygen content) by using a dynamic process that occurs on two zirconium dioxide discs which form a hermetically sealed chamber. Because no chemical substances are used, the sensor exhibits significantly longer service life than an electrochemical sensor. The actual oxygen sensor is mounted in the tip of the bar probe and is protected by a stainless-steel sintered disk which serves as a flame back-flash stop. The water-proof plastic housing accommodates the electronics and is mechanically connected to the bar probe. * * * * * FEATURES No moving parts Low power consumption No need for temperature stabilization Functional testing and calibration in ambient air Long life Sensor electronics include: 1. Signal amplifier 2. Control for the ionic pump with an analog component 3. Test component 4. Internal monitoring logic component 5. Power supply for the sensor heating element and the analog and digital components 6. Reset 7. Voltage monitor 8. Analog output 9. Bi-directional digital output 10. Internal potentiometer The sensor outputs the measured values via an analog channel (4 mA to 20 mA or 0 V to 10 V) and a digital channel (also provides any error signals) which are evaluated and further processed in a user-specified external mounting device such as a fuel controller, display, measuring instrument, programmable logic controller or ventilation system. Both series may be calibrated manually or self-calibrated using atmospheric air instead of a reference gas. The sensor can also conduct external operational self-tests. Both series may also operate in a fault-proof mode. The sensor self-monitors during operation and provides error signal output. A second oxygen sensor is not necessary. * * * * * POTENTIAL APPLICATIONS Furnace/heating boiler control Industrial process control Combustion systems Compost systems Environmental control MF010 Series and MF020 Series TABLE 1. MF010 SERIES SPECIFICATIONS AND ORDER INFORMATION Characteristic Catalog Listing MF010-0-LC1 MF010-0-LC2 MF010-0-LC3 MF010-0-LC4 MF010-1-LC1 MF010-1-LC2 MF010-1-LC3 MF010-1-LC4 MF010-2-LC1 MF010-2-LC2 MF010-2-LC3 MF010-2-LC4 Temperature range at probe tip -100 C to 250 C [-148 F to 482 F] Output 0 Vdc to 10 Vdc 4 mA to 20 mA 1 mbar to 250 mbar O2 range (O2 partial pressure) 1 mbar to 1000 mbar 1 mbar to 1000 mbar 1 mbar to 250 mbar with equivalent O2 output 0.1 vol%O2 to 25 vol%O2 0.1 vol%O2 to 100 vol%O2 0.1 vol%O2 to 25 vol%O2 0.1 vol%O2 to 100 vol%O2 Probe length: 220 mm [8.7 in] -0-LXX 400 mm [15.7 in] -1-LXX 600 mm [23.6 in] -2-LXX TABLE 2. MF020 SERIES SPECIFICATIONS AND ORDER INFORMATION Catalog Listing Characteristic MF020-0-LC1 MF020-0-LC2 MF020-0-LC3 MF020-0-LC4 MF020-1-LC1 MF020-1-LC2 MF020-1-LC3 MF020-1-LC4 MF020-2-LC1 MF020-2-LC2 MF020-2-LC3 MF020-2-LC4 Temperature range at probe tip -100 C to 400 C [-148 F to 752 F] Output 0 Vdc to 10 Vdc 4 mA to 20 mA O2 range (O2 partial pressure) 1 mbar to 1000 mbar 1 mbar to 250 mbar 1 mbar to 1000 mbar 1 mbar to 250 mbar with equivalent O2 output 0.1 vol%O2 to 25 vol%O2 0.1 vol%O2 to 100 vol%O2 0.1 vol%O2 to 25 vol%O2 0.1 vol%O2 to 100 vol%O2 Probe length: 220 mm [8.7 in] -0-LXX 400 mm [15.7 in] -1-LXX 600 mm [23.6 in] -2-LXX TABLE 3. MF010 AND MF020 SERIES GENERAL SPECIFICATIONS Characteristic Parameter Supply voltage 24 Vdc 5% Supply current 500 mA Ambient temperature (includes solar radiation -10 C to 50 C [14 F to 122 F] Resolution 12 bit Heat-up time approx. 10 min Accuracy 2% FS (Full Scale) Reproducibility 1% FS (Full Scale) Gas flow rate 0 m/s to 10 m/s Reaction time approx. 3 s Incidental permissible acceleration 30 g Repetitive permissible acceleration 5g Bar probe material stainless steel 1.4301 Housing material Makrolon 8030 (30% GV), UL94 V-1 Housing sealing IP65 Connector sealing IP67 Weight (without bar probe) 150 g [5.3 oz] Termination (female connector) 99-4226-14-07 (Binder 693 series) Certification complies with EMC directives EN 50082-2 and EN 50081-1 and directives 89/336/EWG and 92/31/EWG 2 www.honeywell.com/sensing Oxygen Sensors FIGURE 1. DIMENSIONAL DRAWING AND PINOUT (For reference only: mm/in.) FIGURE 2. SYSTEM BLOCK DIAGRAM ORDER GUIDE Catalog Listing FIGURE 3. SENSOR MOUNTING Description MF010 Series Low temperature oxygen sensor (See Table 1 for specifics.) MF020 Series High temperature oxygen sensor (See Table 2 for specifics.) Honeywell Sensing and Control 3 WARNING WARNING PERSONAL INJURY MISUSE OF DOCUMENTATION DO NOT USE these products as safety or emergency stop * devices or in any other application where failure of the The information presented in this product sheet is for reference only. Do not use this document as a product product could result in personal injury. installation guide. Failure to comply with these instructions could result in death or serious injury. * Complete installation, operation, and maintenance information is provided in the instructions supplied with each product. Failure to comply with these instructions could result in WARRANTY/REMEDY Honeywell warrants goods of its manufacture as being free of death or serious injury. defective materials and faulty workmanship. Honeywell's standard product warranty applies unless agreed to otherwise SALES AND SERVICE by Honeywell in writing; please refer to your order Honeywell serves its customers through a worldwide network acknowledgement or consult your local sales office for specific of sales offices, representatives and distributors. For warranty details. If warranted goods are returned to Honeywell application assistance, current specifications, pricing or name during the period of coverage, Honeywell will repair or replace, of the nearest Authorized Distributor, contact your local sales at its option, without charge those items it finds defective. The office or: foregoing is buyer's sole remedy and is in lieu of all other warranties, expressed or implied, including those of E-mail: info.sc@honeywell.com merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages. Internet: www.honeywell.com/sensing Phone and Fax: While we provide application assistance personally, through Asia Pacific +65 6445-3033 Fax our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the +65 6355-2828 Europe +44 (0) 1698 481481 +44 (0) 1698 481676 Fax application. Latin America +1-305-805-8188 Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use. +1-305-883-8257 Fax USA/Canada +1-800-537-6945 +1-815-235-6847 +1-815-235-6545 Fax Sensing and Control Honeywell 1985 Douglas Drive North Minneapolis, MN 55422 www.honeywell.com/sensing 000670-2-EN IL50 GLO Printed in USA August 2008 Copyright (c) 2008 Honeywell International Inc. All rights reserved. Oxymac50, DE800 and Elecdit Oxygen Sensor Interface Boards Accurate oxygen measurement DESCRIPTION The Oxymac 50, Elecdit and DE800 sensor interface boards Oxymac50 is a recently upgraded interface that offers similar are control and measurement devices for Honeywell's range performance to the Elecdit board but operates from a 24 Vdc of dynamic highly accurate oxygen sensors. The ability of power supply. these sensors to operate without any reference gas makes them ideal for many applications including combustion and environmental control. Honeywell oxygen sensors using zirconia based sensing elements require external connection to an electronic measuring circuit to control sensor operation and signal The DE800 interface offers the features of the Oxymac50 but also contains the heating element power source. The Elecdit interface was developed for Industrial applications and offers a level of noise protection. It operates from a 15 V supply. processing. A heating element on the sensor produces the Oxymac50 and DE800 are the recommended interfaces for 700 C required for the ZrO2 to achieve its operational new applications. temperature. These interface boards supply the sensor with The interfaces are supplied ready for mounting into the necessary circuits to provide this control and processing, customers' enclosures. Oxymac50 a 15-pin D type connector, while also providing sensor calibration. The DE800 interface DE800 has screw connector block and Elecdit has 2,5 mm x also includes the heater supply, Oxymac50 and Elecdit 2,5 mm (0.1 in x 0.1 in) pin connector. require an external heater power supply . FEATURES BENEFITS * Provides the necessary circuits to control Honeywell oxygen * Removes the need to implement sensor control circuits sensors * Functional testing and calibration in ambient air within customers' equipment * Removes effects of barometric or application pressure changes and sensor drift for best accuracy * PCB board format * Ease of mounting and access for calibration * Linear output of measured oxygen content in voltage or * Flexible interconnection current format * Low power consumption TYPICAL APPLICATIONS * Heating boiler control * Industrial process control * Combustion systems * Compost systems * Environmental control * Medical * Aerospace * Lower running costs Oxymac50, DE800 and Elecdit TECHNICAL DATA Oxymac50 Measuring range DE800 Elecdit 0 % to 25 % oxygen (standard) 0 % to 100 % oxygen (option) Power source +24 Vdc 20 % +24 Vdc 20 % 15 Vdc 10 % Other inputs Ext Calibration, Sensor Pump, Ext Calibration, Sensor Pump, Ext Calibration, Sensor Pump, Sense and Common, Test Sense, Common, Heater, Test Sense and Common Power consumption (interface) <250 mW <500 mW <250 mW Outputs 0 V to 10 V, 4 mA to 20 mA(1) 4 mA to 20 mA, 0 V to 10 V(1), Pulse 0 V to 10 V, LED drive Accuracy 2 % full scale Resolution 0.04 V Operating temperature -10 C to 60 C Storage temperature -10 C to 50 C Dimensions 63 mm x 108 mm, 4 mm O mounting holes 6,5 mm O mounting holes Connector 15 pin male D connector Screw connector block Mates with type 80 mm x 100 mm, 60 mm x 42 mm x 17 mm 2 x 5 pin 2,5 mm x 2,5mm GMS-10, KGZ-10, KGZ-12 Notes: For warm up time and other sensor characteristics refer to datasheet 100470-EN (1) Factory set to voltage or current output OPERATING AND CALIBRATION NOTES Oxymac50 1. Reset of the Oxymac50 is possible by putting the power off/on. 2. The output signal can be 0 V to 10 V or 4 mA to 20 mA. The output signal is factory set by the solder bridge JP1. 3. A proper operation of a connected sensor and electronic circuit can be tested with the test switch or with an external signal. After switching the test function the output signal should indicate a 20 % lower value. 4. LED indication - proper sensor operation is indicated by blinking of LED D3. 5. Calibration - sensor and Oxymac circuit can be calibrated in normal air. After a loss of power the Oxymac50 circuit stores the calibration value. DE800 1. Heater adjustment - the heater voltage of the sensor can be adjusted with pot meter PC419. 2. Test function - after pressing the test button, the output has to go to a 20% lower output signal. Testing with an external (24 V) signal is possible via terminal TE. 3. Calibration - the sensor can be calibrated in normal air with the calibration button. Calibration is lost after power loss. 4. LED -if the sensor is functioning properly, the LED D4 is blinking. 5. Signal out - if the sensor is functioning properly a pulse signal can be measured at terminal K2. Elecdit Adjustment/calibration procedure. The interface is factory set, for sensor replacement follow these instructions. 1. Adjust with Pot1 12 V (internal 6 V) at Pin 2. 2. Adjust with Pot2 reference at Pin 1 to 10 V 0.2 V. 3. After approx 60 s an impulse (blinking) signal should be present at Pin 1. Apply at the sensor 95 mbar air, adjust with Pot5 at Pin 9 to 0.8 V. Then apply at the sensor 1013 mbar air and adjust with Pot3 at Pin 9 Uout to 8.3 V. 4. Adjust with Pot4 0 mV at Pin 10. 5. Adjust with Pot5 Uout to 0 V at Pin 9. 2 www.honeywell.com/sensing Oxygen Sensor Interface Boards MOUNTING DRAWINGS, FOR REFERENCE ONLY DE800 D3 TEST TA2 CAL CAL 1 2 9 10 9 4 11 10 5 12 9 6 13 7 7 8 14 8 15 PC419 6 Power supply voltage 24 V Measuring result TA1 TEST 0V 4 mA to 20 mA 0V 24 V System GND Red Black Blue H+ H- 5 3 +U GND K1 GND TE K2 PE JP1 Test (24 V) P PE D4 0 V and pulse C OXYMAC 50 System GND S OXYMAC50 Yellow Yellow + SENSOR (GMS-10) 0 S (Blue) 11 Calibration P (Red) 15 C (Black) 1 Test 5 SENSOR Heater voltage (Yellow) ORDER GUIDE ELECDIT Oxymac 50.V.1 IC2 measuring range 0 % to 25 % P2 P5 P3 P1 1 6 +12 V test out Output signal (+0 V to 10 V) +15 V power supply Oxymac 50.A.1 Oxymac50 circuit, output 4 mA to 20mA, measuring range 0 % to 100 % Test (amplified signal) 0 V power supply Oxymac50 circuit, output 0 V to 10V, 10 LED drive out -15 V power supply Oxymac 50.V.2 P4 5 Oxymac50 circuit, output 0 V to 10V, P (Red) measuring range 0 % to 25 % Oxymac 50.A.2 Oxymac50 circuit, output 4 mA to 20mA, DE800.V.1 DE800 circuit, output 0 V to 10V, measuring range 0 % to 100 % S (Blue) C (Black) measuring range 0 % to 25 % DE800.V.2 Power supply Heater 4.0 V (Yellow X 2) DE800 circuit, output 0 V to 10V, measuring range 0 % to 100 % Power supply Heater 0 V DE800.A.1 DE800 circuit, output 4 mA to 20mA, measuring range 0 % to 25 % SENSOR DE800.A.2 DE800 circuit, output 4 mA to 20mA, measuring range 0 % to 100 % Elecdit.V.1 Elecdit circuit, ouput 0 V to 10V, measuring range 0 % to 25 % Elecdit.V.2 Elecdit circuit, ouput 0 V to 10V, measuring range 0 % to 100 % Honeywell Sensing and Control 3 WARNING MISUSE OF DOCUMENTATION * The information presented in this product sheet is for reference only. Do not use this document as a product installation guide. * Complete installation, operation, and maintenance information is provided in the instructions supplied with each product. Failure to comply with these instructions could result in death or serious injury. WARRANTY/REMEDY SALES AND SERVICE Honeywell warrants goods of its Honeywell serves its customers through manufacture as being free of defective a worldwide network of sales offices, materials and faulty workmanship. representatives and distributors. For Honeywell's standard product warranty application assistance, current applies unless agreed to otherwise by specifications, pricing or name of the Honeywell in writing; please refer to nearest Authorized Distributor, contact your order acknowledgement or consult your local sales office or: your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge is in lieu of all war ranties, expr essed warranties, expressed PERSONAL INJURY or implied, including those of DO NOT USE these products as safety mer chantability and fitness for a merchantability or emergency stop devices or in any par ticular purpose. In no event shall particular other application where failure of the Honeywell be liable for product could result in personal injury. consequential, special, or indir ect indirect Failur e to comply with these Failure damages. instr uctions could rresult esult in death or instructions serious injur y. injury While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application. Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use. Automation and Control Solutions Sensing and Control Honeywell Newhouse Industrial Estate Motherwell, ML1 5SB, UK www.honeywell.com Inter net: www.honeywell.com/sensing Internet: Phone and Fax: Asia Pacific those items it finds defective. The for egoing is buyer emedy and foregoing buyer''s sole rremedy WARNING E-mail: info.sc@honeywell.com 100480-EN UK07 GLO May 2005 Copyright 2005 Honeywell International Inc. +65 6355-2828 +65 6445-3033 Fax Europe +44 (0) 1698 481481 +44 (0) 1698 481676 Fax Latin America +1-305-805-8188 +1-305-883-8257 Fax USA/Canada +1-800-537-6945 +1-815-235-6847 +1-815-235-6545 Fax