SPECIFICATION
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE Note.2
RATED CURRENT
CURRENT RANGE
RATED POWER (max.)
RIPPLE & NOISE (max.) Note.3
OUTPUT
VOLTAGE RANGE Note.7
FREQUENCY RANGE
POWER FACTOR (Typ.)
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INPUT
ENVIRONMENT
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT(max.)
SAFETY STANDARDS
SAFETY &
EMC
(Note. 9)
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSION
OTHERS
NOTE
PACKING
PROTECTION
GSM120B15-R7BGSM120B12-R7B GSM120B20-R7B GSM120B24-R7B GSM120B48-R7B
15V12V
GSM120B15GSM120B12
20V
GSM120B20
24V
GSM120B24
48V
GSM120B48
120mVp-p100mVp-p 150mVp-p 180mVp-p 200mVp-p
VOLTAGE TOLERANCE Note.4
LINE REGULATION Note.5
LOAD REGULATION
±5.0% ±5.0% ±3.0% ±2.5%
±1.0%±1.0% ±1.0% ±1.0% ±1.0%
±5.0%±5.0% ±4.0% ±3.0% ±2.5%
80 ~ 264VAC 113 ~ 370VDC
47 ~ 63Hz
PF>0.93 / 230VAC PF>0.97 / 115VAC at full load
89%88% 89.5% 90% 91.5%
1.4A / 115VAC 0.7A / 230VAC
Cold start 35A / 115VAC 70A / 230VAC
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
I/P-O/P: 4KVAC
I/P-O/P:100M Ohms / 500VDC / 25 / 70% RH℃
372K hrs min. MIL-HDBK-217F(25 )℃
-30 ~ +70 (Refer to "Derating Curve")℃
20% ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH non-condensing℃
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
167*67*35mm (L*W*H)
0.6Kg; 20pcs/13,0Kg/0.89CUFT
CONNECTOR
PLUG
CABLE See page 4~5; Other type available by customer requested
See page 4~5; Other type available by customer requested
OVERLOAD
OVER VOLTAGE
105 ~ 160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
105 ~ 135% rated output voltage
7A8.5A 6A 5A 2.5A
0 ~ 7A0 ~ 8.5A 0 ~ 6A 0 ~ 5A
0 ~ 2.5A
105W102W 120W 120W 120W
Protection type : Shut down o/p voltage, re-power on to recover
SETUP, RISE TIME Note.6
HOLD UP TIME (Typ.)
1500ms, 30ms / 230VAC 2000ms, 30ms / 115VAC at full load
40ms / 230VAC 24ms / 115VAC at full load
OVER TEMPERATURE Shut down o/p voltage, re-power on to recover
Touch current < 100 A/264VAC
±5.0%
±0.03% / ℃ (0 ~ 40℃)
ISOLATION LEVEL Primary-Secondary: 2xMOPP
℃
℃ ℃
For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
※:https://www.meanwell.com/serviceDisclaimer.aspx
GSM120B seri e s
120W AC-DC Reliable Green Medical Adaptor
OPERATING ALTITUDE Note.8
Parameter
Parameter
ESD
Standard
Standard
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Test Level / Note
Test Level / Note
Class B
Class B
Class A
-----
EMC IMMUNITY
EMC EMISSION
EN60601-1-2, EN61204-3
Level 4, 15KV air ; Level 4, 8KV contact
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
Level 3, 1KV/Line-Line
Level 4, 30A/m
Level 3, 10V
Level 3, 2KV
IEC60601-1/IEC60601-1-11, EN60601-1 EN , ANSI/AAMI ES60601-1 (3.1 version), / 60601-1-11 / ES60601-1-11
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved, EAC TP TC 004
μ
3000 meters
μ
RF field susceptibility
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Voltage dip, interruption
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
File Name:GSM120B-SPEC 2020-10-20
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EN55011 (CISPR11), FCC PART 15 / CISPR22,
CAN ICES-3(B)/NMB-3(B)
EN55011 (CISPR11), FCC PART 15 / CISPR22,
CAN ICES-3(B)/NMB-3(B)