914X Series Field Metrology Well User's Guide Revision 7A1901-EN Limited Warranty & Limitation of Liability Each product from Fluke Corporation, Hart Scientific Division ("Hart") is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year for Field Metrology Wells. The warranty period begins on the date of the shipment. Parts, product repairs, and services are warranted for 90 days. The warranty extends only to the original buyer or end-user customer of a Hart authorized reseller, and does not apply to fuses, disposable batteries or to any other product, which in Hart's opinion, has been misused, altered, neglected, or damaged by accident or abnormal conditions of operation or handling. Hart warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Hart does not warrant that software will be error free or operate without interruption. Hart does not warrant calibrations on the Field Metrology Well. Hart authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Hart. Warranty support is available if product is purchased through a Hart authorized sales outlet or Buyer has paid the applicable international price. Hart reserves the right to invoice Buyer for importation costs of repairs/replacement parts when product purchased in one country is submitted for repair in another country. Hart's warranty obligation is limited, at Hart's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Hart authorized service center within the warranty period. To obtain warranty service, contact your nearest Hart authorized service center or send the product, with a description of the difficulty, postage, and insurance prepaid (FOB Destination), to the nearest Hart authorized service center. Hart assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Hart determines that the failure was caused by misuse, alteration, accident or abnormal condition or operation or handling, Hart will provide an estimate or repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point). THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL. OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision. Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive * American Fork, UT 84003-9775 * USA Phone: +1.801.763.1600 * Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Specifications subject to change without notice. * Copyright (c) 2007 * Printed in USA ii Table of Contents 1 Before You Start........................................................................1 1.1 1.2 1.3 1.4 Introduction................................................................................................ 1 Unpacking................................................................................................. 2 Symbols Used............................................................................................ 3 Safety Information...................................................................................... 4 1.4.1 1.4.2 1.5 CE Comments............................................................................................ 8 1.5.1 1.5.2 1.5.3 1.5.4 1.6 2 Authorized Service Centers....................................................................... 9 Specifications.......................................................................................... 13 Environmental Conditions........................................................................ 15 Quick Start...............................................................................17 3.1 3.2 Setup........................................................................................................ 17 Parts and Controls................................................................................... 18 3.2.1 3.2.2 3.2.3 3.2.4 3.3 Display Panel.................................................................................................19 Display...........................................................................................................20 Power Panel...................................................................................................22 -P Option Panel (-P models only)...................................................................24 Languages............................................................................................... 26 3.3.1 3.3.2 4 EMC Directive..................................................................................................8 Immunity Testing..............................................................................................8 Emission Testing...............................................................................................9 Low Voltage Directive (Safety).........................................................................9 Specifications and Environmental Conditions.....................13 2.1 2.2 3 Warnings..........................................................................................................5 Cautions...........................................................................................................7 Language Selection.......................................................................................26 Reset to English Language............................................................................27 Menu Structure........................................................................29 4.1 4.2 Temp Setup Menu.................................................................................... 29 Prog Menu............................................................................................... 30 4.2.1 4.2.2 4.3 4.4 Switch Test Parameters..................................................................................31 Switch Test Description..................................................................................31 System Menu........................................................................................... 33 Input Setup (-P only)................................................................................ 34 iii 5 Maintenance............................................................................35 5.1 iv Field Metrology Well Performance Analysis............................................ 35 Tables Table 1 Symbols used......................................................................................... 3 Table 2 Base Unit Specifications...................................................................... 13 Table 3 -P Option Specifications ..................................................................... 14 v Figures Figure 1 Clamp-on ferrite installation.................................................................. 9 Figure 2 914X Field Metrology Well.................................................................. 18 Figure 3 Display panel and keys ..................................................................... 20 Figure 4 914X display....................................................................................... 21 Figure 5 9142 power panel............................................................................... 23 Figure 6 9143 and 9144 power panel............................................................... 23 Figure 7 -P option panel .................................................................................. 24 Figure 8 Probe connector wiring ..................................................................... 25 Figure 9 Jumper locations for 3-wire and 2-wire connections.......................... 26 Figure 10 Steps to language selection............................................................. 27 Figure 11 Main Menu - Temp SetUp................................................................. 29 Figure 12 Main Menu - Prog Menu................................................................... 30 Figure 13 Auto and manual switch test operation example............................. 32 Figure 14 Main Menu - System Menu............................................................... 33 Figure 15 Main Menu - Input Setup ................................................................. 34 vi Before You Start Introduction 1 Before You Start 1.1 Introduction Field Metrology Wells (9142, 9143, and 9144) are designed to be reliable, stable heat sources that can be used in the field or laboratory. They offer accuracy, portability, and speed for nearly every field calibration application. The instruments have been designed with the field user in mind and are easy to use while maintaining stability, uniformity, and accuracy comparable to some laboratory instruments. Special built-in features make Field Metrology Wells extremely adaptable. The exclusive Voltage Compensation allows the technician to plug into mains power with voltage from 90 V ac to 250 V ac without degradation to the instrument. The Ambient Temperature Compensation (Patent Pending) provides the largest operating range in the industry (0C to 50C) with the largest guaranteed temperature range (13C to 33C). The Gradient Temperature Compensation (Patent Pending) keeps the axial gradient within specification over the entire temperature range of the instrument and over the specified guaranteed operating temperature range. These combined features along with the rugged design, light weight, and small size make this line of instruments ideal for field applications. Unique Patent Pending safety features make these the safest field heat sources available. The unique Air Flow Design (Patent Pending) keeps the probe handle cool protecting delicate instruments and the user. The Block Temperature Indicator (Patent Pending) shows the user when the well temperature is above 50C letting the user know when it is safe to remove the insert or move the instrument. The indicator light illuminates when the instrument is energized and the well is above 50C. If the instrument is removed from mains power, the indicator light flashes until the well has cooled to less than 50C. The optional "Process" version ("914X-P") combines the heat source with a built-in readout eliminating the need for the technician to take two instruments to the field. The readout is perfect for transmitter loop, comparison calibration, or a simple check of a thermocouple sensor. There is no need to carry additional tools into the field with the "Process" option of a built-in readout for resistance, voltage, and mA measurement, 24V loop power, and on-board documentation. The convenient smart reference connector automatically transfers and stores the probe coefficients. The Field Metrology Wells' controller uses a PRT sensor and thermoelectric modules or heaters to achieve stable, uniform temperatures throughout the block. The LCD display continuously shows many useful operating parameters including the block temperature, the current set-point, block stability, and heating and cooling status. For the Process version, the reference temperature and secondary input type (UUT) readings are displayed. The display can be set to show the information in one of eight different languages; English, Japanese, Chinese, German, Spanish, French, Russian, and Italian. The instrument's rugged design and special features make them ideal for the field or the laboratory. With proper use, the instrument provides continued accurate calibration of temperature 1 914X Field Metrology Wells Unpacking sensors and devices. Before use, the user should be familiar with the warnings, cautions, and operating procedures of the calibrator as described in the User's Guide. 1.2 Unpacking Unpack the instrument carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately. Verify that the following components are present: 9142 9142 Field Metrology Well 9142-INSX Insert (X=A, B, C, D, E, or F) Power Cord RS-232 Cable User Guide Technical Manual CD Report of Calibration and calibration label 6-pin DIN Connector (-P model only) Test Lead Kit (-P model only) Well Insulator Clamp-on ferrites (3) [-P model only] Tongs (insert removal tool) 9930 Interface-it Software and User's Guide 9143 2 9143 Field Metrology Well 9143-INSX Insert (X=A, B, C, D, E, or F) Power Cord RS-232 Cable User Guide Technical Manual CD Report of Calibration and calibration label 6-pin DIN Connector (-P model only) Test Lead Kit (-P model only) Clamp-on ferrites (3) [-P model only] Tongs (insert removal tool) 9930 Interface-it Software and User's Guide Before You Start Symbols Used 9144 9144 Field Metrology Well 9144-INSX Insert (X=A, B, C, D, E, or F) Power Cord RS-232 Cable User Guide Technical Manual CD Report of Calibration and calibration label 6-pin DIN Connector (-P model only) Test Lead Kit (-P model only) Clamp-on ferrites (3) [-P model only] Tongs (insert removal tool) 9930 Interface-it Software and User's Guide If all items are not present, contact an Authorized Service Center (see Section 1.6 Authorized Service Centers on page 9). 1.3 Symbols Used Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this guide. Table 1 Symbols used Symbol Description AC (Alternating Current) AC-DC Battery Complies with European Union directives DC Double Insulated 3 914X Field Metrology Wells Safety Information Symbol Description Electric Shock Fuse PE Ground Hot Surface (Burn Hazard) Read the User's Guide (Important Information) Off On Canadian Standards Association C-TICK Australian EMC mark The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/ EC) mark. 1.4 Safety Information Field Metrology Wells are designed in accordance with IEC 61010-1, IEC 61010-2-010 and CAN/CSA 22.2 No 61010.1-04. Use this instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired. Refer to the safety information in the Warnings and Cautions sections below. The following definitions apply to the terms "Warning" and "Caution". 4 "Warning" identifies conditions and actions that may pose hazards to the user. "Caution" identifies conditions and actions that may damage the instrument being used. Before You Start Safety Information 1.4.1 Warnings To avoid personal injury, follow these guidelines. GENERAL DO NOT use this instrument in environments other than those listed in the User's Guide. Inspect the instrument for damage before each use. Inspect the case. Look for cracks or missing plastic. DO NOT use the instrument if it appears damaged or operates abnormally. Follow all safety guidelines listed in the User's Guide. Calibration equipment should only be used by trained personnel. If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Before initial use, or after transport, or after storage in humid or semi-humid environments, or anytime the instrument has not been energized for more than 10 days, the instrument needs to be energized for a "dry-out" period of 2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-2. If the product is wet or has been in a wet environment, take necessary measures to remove moisture prior to applying power such as storage in a low humidity temperature chamber operating at 50C for 4 hours or more. DO NOT use this instrument for any application other than calibration work. The instrument was designed for temperature calibration. Any other use of the instrument may cause unknown hazards to the user. DO NOT place the instrument under a cabinet or other structure. Overhead clearance is required. Always leave enough clearance to allow for safe and easy insertion and removal of probes. Use of this instrument at HIGH TEMPERATURES for extended periods of time requires caution. Completely unattended high temperature operation is not recommended due to safety hazards that can arise. This instrument is intended for indoor use only. Follow all safety procedures for the test and calibration equipment you use. If used, inspect the test leads for damaged insulation or exposed metal. Check for test lead continuity. Replace damaged test leads as necessary. Do not use the instrument if it operates abnormally. Protection may be impaired. When in doubt, have the instrument serviced. Do not apply more than the rated voltage, as marked on the instrument, between terminals or between any terminal and earth ground. 5 914X Field Metrology Wells Safety Information Never touch the probes to a voltage source when the test leads are plugged into the current terminals. Select the proper function and range for each measurement. Disconnect the test leads before changing to another measure or source function. DO NOT operate the Field Metrology Well around explosive gas, vapor, or dust. DO NOT operate instrument at orientations other than upright. Tilting the instrument or laying it down on its side during use could create a fire hazard. BURN HAZARD The instrument is equipped with a Block Temperature Indicator (front panel LED HOT indicator - Patent Pending) even when the instrument is unplugged. When the indicator is flashing, the instrument is disconnected from mains power and the temperature of the block is above 50C. When the indicator is illuminated, always on, the instrument is powered and the block temperature is above 50C. DO NOT turn the instrument upside down with the inserts in place; the inserts will fall out. DO NOT operate near flammable materials. Use of this instrument at HIGH TEMPERATURES for extended periods of time requires caution. DO NOT touch the well access surface of the instrument. The block vent may be very hot due to the fan blowing across the heater block of the instrument. The temperature of the well access is the same as the actual display temperature, e.g. if the instrument is set to 600C and the display reads 600C, the well is at 600C. Probes and inserts may be hot and should only be inserted and removed from the instrument when the instrument indicates temperatures less than 50C. DO NOT turn off the instrument at temperatures higher than 100C. This could create a hazardous situation. Select a set-point less than 100C and allow the instrument to cool before turning it off. The high temperatures present in Field Metrology Wells designed for operation at 300C and higher may result in fires and severe burns if safety precautions are not observed. ELECTRICAL HAZARD These guidelines must be followed to ensure that the safety mechanisms in this instrument operate properly. This instrument must be plugged into an AC only electric outlet according to Table 2, Specifications . The power cord of the instrument is equipped with a three-pronged grounding plug for your protection against electrical shock hazards. It must be plugged directly into a properly grounded three-prong receptacle. The receptacle must be installed in accordance with local codes and ordinances. Consult a qualified electrician. DO NOT use an extension cord or adapter plug. 6 Before You Start Safety Information If supplied with user accessible fuses, always replace the fuse with one of the same rating, voltage, and type. Always replace the power cord with an approved cord of the correct rating and type. HIGH VOLTAGE is used in the operation of this equipment. SEVERE INJURY or DEATH may result if personnel fail to observe safety precautions. Before working inside the equipment, turn power off and disconnect power cord. -P Model Only When using test leads, keep fingers behind the finger guards on the test leads. DO NOT apply more than the rated voltage, as marked on the instrument, between the terminals, or between any terminal and earth ground (30 V 24 mA max all terminals). Never touch the probe to a voltage source when the test leads are plugged into current terminals. Select the proper function and range for your measurement. Inspect the test leads for damaged insulation or exposed metal. Check test leads continuity. Replace damaged test leads before you use the calibrator. 1.4.2 Cautions To avoid possible damage to the instrument, follow these guidelines: DO NOT leave the inserts in the instrument for prolonged periods. Due to the high operating temperatures of the instrument, the inserts should be removed after each use and buffed with a Scotch-Brite(R) pad or emery cloth (see Section 5 Maintenance on page 35). Always operate this instrument at room temperature between 41F and 122F (5C to 50C). Allow sufficient air circulation by leaving at least 6 inches (15 cm) of clearance around the instrument. Overhead clearance of 1 meter (3 ft) is required. DO NOT place instrument under any structure. Component lifetime can be shortened by continuous high temperature operation. DO NOT apply any type of voltage to the display hold terminals. Applying a voltage to the terminals may cause damage to the controller. DO NOT use fluids to clean out the well. Fluids could leak into electronics and damage the instrument. Never introduce any foreign material into the probe hole of the insert. Fluids, etc. can leak into the instrument causing damage. Unless recalibrating the instrument DO NOT change the values of the calibration constants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the calibrator. 7 914X Field Metrology Wells CE Comments DO NOT allow the probe sheath or inserts to drop into the well. This type of action can cause a shock to the sensor and affect the calibration. The instrument and any thermometer probes used with it are sensitive instruments that can be easily damaged. Always handle these devices with care. DO NOT allow them to be dropped, struck, stressed, or overheated. DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty environment. Always keep the well and inserts clean and clear of foreign material. The Field Metrology Well is a precision instrument. Although it has been designed for optimum durability and trouble free operation, it must be handled with care. Always carry the instrument in an upright position to prevent the inserts from dropping out. The convenient handle allows for hand carrying the instrument. If a mains supply power fluctuation occurs, immediately turn off the instrument. Power bumps from brown-outs could damage the instrument. Wait until the power has stabilized before reenergizing the instrument. The probe and the block may expand at different rates. Allow for probe expansion inside the well as the block heats. Otherwise, the probe may become stuck in the well. Most probes have handle temperature limits. If the probe handle limits are exceeded, the probe may be permanently damaged. Due to a unique Air Flow Design (Patent Pending), Field Metrology Wells protect the probe handle temperature and provide a safer temperature handle for the user. 1.5 CE Comments 1.5.1 EMC Directive Hart Scientific's equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration of Conformity for your instrument lists the specific standards to which the instrument was tested. The instrument was designed specifically as a test and measuring device. Compliance to the EMC directive is through IEC 61326-1 Electrical equipment for measurement, control and laboratory use. As noted in the IEC 61326-1, the instrument can have varying configurations. The instrument was tested in a typical configuration with shielded RS-232 cables. 1.5.2 Immunity Testing Using Clamp-On Ferrites For the -P model only, clamp-on ferrites are provided for use in improving its electromagnetic (EM) immunity in environments of excessive EM interference. During EMC testing we found that ferrites clamped around probe cables for the Reference PRT, the PRT/RTD input, and the thermocouple (TC) input reduced the risk the EM interference affects measurements. Therefore, 8 Before You Start Authorized Service Centers we recommend that the clamp-on ferrites provided be used on the cables of probes attached to the readout, especially if the product is used near sources of EM interference such as heavy industrial equipment. To attach a ferrite to a probe cable, make a loop in the cable near the connector and clamp the ferrite around half of the loop as shown in the diagram. The ferrite can be easily snapped open and moved to a new probe when needed. connector clamp-on ferrite probe Figure 1 Clamp-on ferrite installation 1.5.3 Emission Testing The instrument fulfills the limit requirements for Class A equipment. The instrument was not designed to be used in domestic establishments. 1.5.4 Low Voltage Directive (Safety) In order to comply with the European Low Voltage Directive (2006/95/EC), Hart Scientific equipment has been designed to meet the EN 61010-1 and the EN 61010-2-010 standards. 1.6 Authorized Service Centers Please contact one of the following Authorized Service Centers to coordinate service on your Hart product: Fluke Corporation Hart Scientific Division 799 E. Utah Valley Drive American Fork, UT 84003-9775 USA Phone: +1.801.763.1600 Telefax: +1.801.763.1010 E-mail: support@hartscientific.com 9 914X Field Metrology Wells Authorized Service Centers Fluke Nederland B.V. Customer Support Services Science Park Eindhoven 5108 5692 EC Son NETHERLANDS Phone: +31-402-675300 Telefax: +31-402-675321 E-mail: ServiceDesk@fluke.nl Fluke Int'l Corporation Service Center - Instrimpex Room 2301 Sciteck Tower 22 Jianguomenwai Dajie Chao Yang District Beijing 100004, PRC CHINA Phone: +86-10-6-512-3436 Telefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Office Service Center 60 Alexandra Terrace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65-6799-5588 Telefax: +65-6799-5589 E-mail: anthony.ng@fluke.com When contacting a Service Centers for support, please have the following information available: 10 Model Number Serial Number Voltage Complete description of the problem 11 Specifications and Environmental Conditions Specifications 2 Specifications and Environmental Conditions 2.1 Specifications Table 2 Base Unit Specifications Base Unit Specifications 9142 9143 9144 -25 C to 150 C (-13 F to 302 F) 33 C to 350 C (91 F to 662 F) 50 C to 660 C (122 F to 1220 F) Display Accuracy 0.2 C Full Range 0.2 C Full Range 0.35 C at 50 C 0.35 C at 420 C 0.5 C at 660 C Stability 0.01 C Full Range 0.02 C at 33 C 0.02 C at 200 C 0.03 C at 350 C 0.03 C at 50 C 0.05 C at 420 C 0.05 C at 660 C Axial Uniformity at 40 mm (1.6 in) 0.05 C Full Range 0.04 C at 33 C 0.1 C at 200 C 0.2 C at 350 C 0.05 C at 50 C 0.35 C at 420 C 0.5 C at 660 C Axial Uniformity at 60 mm (2.4 in) 0.07 C Full Range 0.04 C at 33 C 0.2 C at 200 C 0.25 C at 350 C 0.1 C at 50 C 0.6 C at 420 C 0.8 C at 660 C Radial Uniformity 0.01 C Full Range 0.01 C at 33 C 0.015 C at 200 C 0.02 C at 350 C 0.02 C at 50 C 0.05 C at 420 C 0.1 C at 660 C Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes) 0.006 C Full Range 0.015 C Full Range 0.015 C at 50 C 0.025 C at 420 C 0.035 C at 660 C Loading Effect (versus display with 6.35 mm probes) 0.08 C Full Range 0.2 C Full Range 0.1 C at 50 C 0.2 C at 420 C 0.2 C at 660 C 0.025 C 0.03 C 0.1 C Temperature Range at 23 C Hysteresis Operating Conditions 0 C to 50 C, 0 % to 90 % RH (non-condensing) Environmental conditions for all specifications except temperature range 13 C to 33 C Immersion (Well) Depth Insert OD 150 mm (5.9 in) 30 mm (1.18 in) 25.3 mm (1.00 in) 24.4 mm (0.96 in) Heating Time 16 min: 23 C to 140 C 23 min: 23 C to 150 C 25 min: -25 C to 150 C 5 min: 33 C to 350 C 15 min: 50 C to 660 C Cooling Time 15 min: 23 C to -25 C 25 min: 150 C to -23 C 32 min: 350 C to 33 C 14 min: 350 C to 100 C 35 min: 660 C to 50 C 25 min: 660 C to 100 C Resolution Display Key Pad Size (H x W x D) 0.01 LCD, C or F user-selectable Arrows, Menu, Enter, Exit, 4 soft keys 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in) 13 914X Field Metrology Wells Specifications Base Unit Specifications Weight 9142 9143 9144 8.16 kg (18 lbs) 7.3 kg (16 lbs) 7.7 kg (17 lbs) Power Requirements 100 V to 115 V ( 10 %) 50/60 Hz, 635 W 230 V ( 10 %) 50/60 Hz, 575 W 100 V to 115 V ( 10 %), 50/60 Hz, 1400 W 230 V ( 10%), 50/60 Hz, 1800 W System Fuse Ratings 115 V: 6.3 A T 250 V 230 V: 3.15 A T 250 V 115 V: 15 A F 250 V 230 V: 10 A F 250 V 4-20 mA Fuse (-P model only) Computer Interface Safety 50 mA F 250V RS-232 and 9930 Interface-it control software included EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 Table 3 -P Option Specifications -P Specifications 14 Built-in Reference Thermometer Readout Accuracy (4-Wire Reference Probe) 0.013 C at -25 C 0.015 C at 0 C 0.020 C at 50 C 0.025 C at 150 C 0.030 C at 200 C 0.040 C at 350 C 0.050 C at 420 C 0.070 C at 660 C Reference Resistance Range 0 ohms to 400 ohms Reference Resistance Accuracy 0 ohms to 42 ohms: 0.0025 ohms 42 ohms to 400 ohms: 60 ppm of reading Reference Characterizations ITS-90, CVD, IEC-751, Resistance Reference Measurement Capability 4-wire Reference Probe Connection 6 Pin Din with Infocon Technology Built-in RTD Thermometer Readout Accuracy NI-120: 0.015 C at 0 C PT-100 (385): 0.02 C at 0 C PT-100 (3926): 0.02 C at 0 C PT-100 (JIS): 0.02 C at 0 C RTD Resistance Range 0 ohms to 400 ohms Resistance Accuracy 0 ohms to 25 ohms: 0.002 ohms 25 ohms to 400 ohms: 80 ppm of reading RTD Characterizations PT-100 (385),(JIS),(3926), NI-120, Resistance RTD Measurement Capability 2-,3-,4-wire RTD w\ Jumpers only RTD Connection 4 terminal input Specifications and Environmental Conditions Environmental Conditions -P Specifications Built-in TC Thermometer Readout Accuracy Type J: 0.7 C at 660 C Type K: 0.8 C at 660 C Type T: 0.8 C at 400 C Type E: 0.7 C at 660 C Type R: 1.4 C at 660 C Type S: 1.5 C at 660 C Type M: 0.6 C at 660 C Type L: 0.7 C at 660 C Type U: 0.75 C at 600 C Type N: 0.9 C at 660 C Type C: 1.1 C at 660 C TC Millivolt Range -10 mV to 75 mV Voltage Accuracy 0.025 % of reading +0.01mV Internal Cold Junction Compensation Accuracy 0.35 C (ambient of 13 C to 33 C) TC Connection Small connectors Built-in mA Readout Accuracy 0.02% of reading + 0.002 mA mA Range Cal 4-22 mA, Spec 4-24 mA mA Connection 2 terminal input Loop Power Function 24 VDC loop power Built-in Electronics Temperature Coefficient (0 C to 13 C, 33 C to 50 C) 0.005 % of range per C The temperature range may be limited by the reference probe connected to the readout. The BuiltIn Reference Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires. 2.2 Environmental Conditions Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance section. The instrument operates safely under the following environmental conditions: ambient temperature range: 0-50C (32-122F) ambient relative humidity: 0 % to 90 % (non-condensing) mains voltage: within 10% of nominal vibrations in the calibration environment should be minimized altitude: less than 2,000 meters indoor use only 15 Quick Start Setup 3 Quick Start 3.1 Setup Note: The instrument will not heat, cool, or control until the "SET PT." parameter is "Enabled". Place the calibrator on a flat surface with at least 6 inches of free space around the instrument. Overhead clearance is required. DO NOT place under a cabinet or structure. Plug the instrument power cord into a mains outlet of the proper voltage, frequency, and current capability (see Section 2.1 Specifications on page 13 for power details). Observe that the nominal voltage corresponds to that indicated on the front of the calibrator. Carefully place the insert into the well. Inserts should be of the smallest hole diameter possible still allowing the probe to slide in and out easily. Various insert sizes are available. Contact an Authorized Service Center for assistance (see Section 1.6 Authorized Service Centers on page 9). The well must be clear of any foreign objects, dirt and grit before an insert is installed. The insert is installed with the two small tong holes positioned upward. Turn on the power to the calibrator by toggling the switch on the power entry module. After a brief self-test, the controller should begin normal operation. The main screen appears within 30 seconds. If the instrument fails to operate, please check the power connection. The display shows the well temperature, and waits for user input before further operation. Press "SET PT." and use the arrow keys to set the desired set-point temperature. Press "ENTER" to save the desired set-point and enable the instrument. After five (5) seconds, the instrument should start to operate normally and heat or cool to the designated set-point. 17 914X Field Metrology Wells Parts and Controls Figure 2 914X Field Metrology Well 3.2 Parts and Controls This section describes the exterior features of the Field Metrology Well. All interface and power connections are found on the front of the instrument (see Figure 2). 18 Quick Start Parts and Controls 3.2.1 Display Panel Figure 3 on next page shows the layout of the display panel. Display (1) The display is a 240 x 160 pixel monochrome graphics LCD device with a bright LED backlight. The display is used to show current control temperature, measurements, status information, operating parameters, and soft key functions. udlr Arrow Keys (2) The Arrow Keys allow you to move the cursor on the display, change the display layout, and adjust the contrast of the display. The contrast can only be adjusted using the u and d arrow keys while viewing the main display window. Enter Key (3) The Enter Key allows you to select menus and accept new values. SET PT. (4) The Set Pt. Key allows you to enable the instrument to heat or cool to a desired set-point. Until this key is enabled, the instrument will not heat or cool. It is in a "sleep" state for safety of the operator and instrument. C/F Key (5) The C/F Key allows you to change the displayed temperature units from C to F and vice versa. Menu Key (6) The Menu Key allows the user to access all parameter and settings menus. From the main menu, the user can use the soft keys to access submenus and functions. Exit Key (7) The Exit Key allows you to exit menus and cancel newly entered values. Soft Keys (8) The Soft Keys are the four buttons immediately below the display (labeled F1 to F4). The functions of the soft keys are indicated on the display above the buttons. The function of the keys may change depending on the menu or function that is selected. Switch Connector (9) The switch hold connector posts are located on the left side of the display panel. 19 914X Field Metrology Wells Parts and Controls Block Temperature Indicator (10) [Patent Pending] The Block Temperature Indicator lamp allows users to know when the block temperature is safe (50C to 60C) to remove inserts or move the Field Metrology Well. The indicator light is lit continuously once the block has exceeded approximately 50C (varies 50C to 60C). The indicator light stays lit until the block cools to less than approximately 50C. If the instrument is disconnected from mains power, the indicator light flashes until the block temperature is less than approximately 50C. Figure 3 Display panel and keys 3.2.2 Display The front panel display is shown in detail in Figure 4 on opposite page. Heat Source Temperature (1) The most recent block temperature measurement is shown in large digits in the box at the top of the screen. Set-point Temperature (2) The current set-point temperature is displayed just below the Process Temperature. Reference Thermometer Temperature (3) [-P models only] When installed, the most recent reference thermometer measurement is shown on the screen. 20 Quick Start Parts and Controls Stability Status (4) On the right hand side of the screen, you will find a graph displaying the current status of the stability of the Field Metrology Well. Heating/Cooling Status (5) Just below the stability graph there is a bar graph that will indicate HEATING, COOLING, or CUTOUT. This status graph indicates the current level of heating or cooling if the instrument is not in cutout mode. UUT Output (6) [-P models only] When installed, the most recent UUT output measurement is shown. The value displayed depends on the output type selected: mA, RTD, or TC. Soft Key Functions (7) The four texts at the bottom of the display (not shown) indicate the functions of the soft keys (F1-F4). These functions change with each menu. Editing Windows While setting up and operating the instrument, you are often required to enter or select parameters. Editing windows appear on the screen when necessary to show the values of parameters and allow edits. Figure 4 914X display 21 914X Field Metrology Wells Parts and Controls 3.2.3 Power Panel The following are found on the lower front panel of the instrument (see Figures 5 and Figure 6 on opposite page). Power Cord Plug (1) The power supply cord attaches to the lower front power panel. Plug the cord into an AC mains supply appropriate for the voltage range as specified in the specifications tables. Power Switch (2) For the 9142, the power switch is located on the power entry module of the unit at the lower center of the power panel. For the 9143 and 9144, the power switch is located between the RS-232 and the fuses. Serial Connector (3) On the 9142, the serial connector is a 9-pin subminiature D type located on the power panel above the power entry module. On the 9143 and 9144, the serial connector is a 9-pin subminiature D type located on the power panel to the left of the power switch. The serial (RS-232) interface can be used to transmit measurements and control the operation of the instrument. Fuses (4) For the 9142, the fuses are located inside the power entry module of the unit (Figure 5 on opposite page). For the 9143 and 9144, the fuses are separate from the power connector (Figure 6 on opposite page). If necessary, fuses must be replaced according to Specifications (see Section 2.1 Specifications on page 13. 22 Quick Start Parts and Controls Figure 5 9142 power panel Figure 6 9143 and 9144 power panel 23 914X Field Metrology Wells Parts and Controls 3.2.4 -P Option Panel (-P models only) The -P (process version) panel is the readout portion of the instrument and is only available with -P models. MEASURE 2 mA LOOP 3 COM REF PRT TC 1 4 5 S FU E 50 mA F 250 V S FU E 30 V, 24 mA MAX ALL TERMINALS Figure 7 -P option panel Reference Thermometer Connection (1) The 6-pin DIN smart connector on the front panel allows a reference probe to be attached to the instrument for use with the reference thermometer function of the instrument. The smart connector stores probe calibration coefficients. The 6-pin DIN accepts traditional connectors and the probe coefficients can be entered into the readout or an appropriate characterization curve can be selected through the user interface (see Section 1.5.2 Immunity Testing on page 8 for information on using clamp-on ferrites). A PRT is the only type of probe that is supported by the reference thermometer input. The PRT (RTD or SPRT) probe connects to the reference thermometer input using a 6-pin DIN connector. Figure 8 shows how a four-wire probe is wired to the 6-pin DIN connector. One pair of wires attaches to pins 1 and 2 and the other pair attaches to pins 4 and 5 (pins 1 and 5 source current and pins 2 and 4 sense the potential). If a shield wire is present, it should be connected to pin 3, which is also used for the memory circuit. Pin 6 is only used for the memory circuit. 24 Quick Start Parts and Controls Probe Connector 1 2 6 5 3 4 Shield RTD Sensor Figure 8 Probe connector wiring A two-wire probe can also be used with the reference thermometer. It is connected by attaching one wire to both pins 1 and 2 of the plug and the other wire to both pins 4 and 5. If a shield wire is present, it should be connected to pin 3. Accuracy may be significantly degraded using a two-wire connection because of lead resistance. 4-20mA Connectors (2) The 4-20mA connectors allow current and/or voltage probes to be connected for measurement of associated devices. PRT/RTD Connector (3) The 4-wire PRT/RTD connectors allow the user to connect 3-wire and 2-wire (with jumpers, see Figure 9 on next page) PRT/RTDs to the readout. The correct wiring for the 4-wire PRT/ RTD is shown on the instrument. Figure 9 shows the correct wiring for a 2 or 3-wire PRT/RTD (see Section 1.5.2 Immunity Testing on page 8 for information on using clamp-on ferrites). 25 914X Field Metrology Wells Languages Figure 9 Jumper locations for 3-wire and 2-wire connections Thermocouple (TC) Connector (4) The TC connector allows for the use of subminiature TC connectors (see CE Comments on page 8 for information on using clamp-on ferrites). Fuse (5) Fuse for the 4-20 mA circuit. Always replace with a fuse of the appropriate rating (see Section 2.1 Specifications on page 13). 3.3 Languages The display on Field Metrology Wells can be set to different languages depending on the configuration. 3.3.1 26 European: English, French, Spanish, Italian, German Russian: Russian, English Asian: English, Chinese, Japanese Language Selection Select the language to be displayed by following the steps shown in Figure 10 on opposite page. Quick Start Languages Figure 10 Steps to language selection 3.3.2 Reset to English Language If you are in a language and need a short cut exit, press F1 and F4 simultaneously to reset the display to English. To reset to your originally selected language after resetting to English, follow the steps in Figure 10 on this page. Note: The F1 and F4 English shortcut override is temporary. If you toggle the power off, the instrument jwill return to the language selected in the DISPLAY SETUP menu rather than coming up in English. 27 Menu Structure Temp Setup Menu 4 Menu Structure 4.1 Temp Setup Menu 9142/9143/9144 MAIN MENU F1 F2 TEMP SETUP PROG MENU F1 SETUP F2 CUTOUT F3 SYSTEM MENU F4 INPUT SETUP (-P only) SCAN RATE: Degrees per minute at which the Field Metrology Well increases or decreases temperature STABLE LIMIT: The defined minimum stability over a 2 minute period before alarm is activated STABLE ALARM: F1 - EDIT PRESET <1-8> F4 - SAVE/DISABLE Up/Down Arrow Keys Up Key: Darker Down Key: Lighter F1 & F4 Keys (same time) F1 & F3 Keys (same time) 1 Beep - Valid key entry 2 Beep - Invalid key entry Code Update Mode Keys ENTER & EXIT Keys (hold during power up) Allows instrument software to be updated Figure 11 Main Menu - Temp SetUp 29 914X Field Metrology Wells Prog Menu 4.2 Prog Menu 9142/9143/9144 MAIN MENU F1 TEMP SETUP F2 PROG MENU F1 RUN PROGram F2 RAMP/SOAK F4 INPUT SETUP (-P only) TEST STATUS: Runs selected test RECORD DATA: SOAK TIME: Length of time in minutes at which the Field Metrology Well remains at the set point temperature Number O.f CYCLES: <1-999> PASS TOLERANCE: (-P only) Pass/Fail criteria for the ramp/soak program F1 F3 F3 SYSTEM MENU SET POINTS SETPOINT 1: SETPOINT 2: SETPOINT 3: SETPOINT 4: SETPOINT 5: SETPOINT 6: SETPOINT 7: SETPOINT 8: SWITCH TEST TEST METHOD: Displays test results DATE: TIME: PRINT TESTS: < NO, YES> Erases all test data Figure 12 Main Menu - Prog Menu 30 Menu Structure Prog Menu 4.2.1 Switch Test Parameters SWITCH TEMP The SWITCH TEMP parameter is the nominal change temperature of the switch. UPPER TEMP The UPPER TEMP parameter is the temperature during a cycle at which the Field Metrology Well begins to heat or cool at the rate specified in "Scan Rate" found in MAIN MENU|TEMP SETUP|SETUP|SCAN RATE. LOWER TEMP The LOWER TEMP parameter is the temperature at which the Field Metrology Well heats or cools, in order to begin testing if the test is just starting or the temperature at which the instrument begins to heat to start a cycle. APPROACH The APPROACH parameter controls the use of the Scan Rate during the approach to the setpoint. During the test, the controller uses the system Scan Rate until the temperature is within the approach temperature of either the high temp or low temp parameters. NO. CYCLES The NO. CYCLES parameter determines how many times the instrument heats and cools allowing a thermal switch or batch of switches to be tested. 4.2.2 Switch Test Description CAUTION: The switch, switch wires, switch components and switch accessories can be damaged if the Field Metrology Well exceeds their temperature limits. The SWITCH TEST is used to select, set up, execute and view the results of switch tests. The Switch Test function allows thermal switches to be tested for open and/or close temperatures. The Switch Test allows for Auto or Manual operation. Figure 13 on next page shows a graphical representation of how a switch test works. For Auto operation, enter the Prog Menu. Under Switch Test, select Auto Test. Enter the SWITCH TEMP. Set the Test Method to AUTO. Exit to the Run Prog menu. Ensure that Run Test is set to SWITCH TEST. Set Test Status to RUN. Press Enter and the instrument will engage and start the 3-cycle test within a few seconds. Exit to the main screen to view the test progress, refer to the Menu Structure. For Manual operation, in the Temp Setup menu, select Setup and enter the SCAN RATE. Exit to the Prog Menu. Under Switch Test, select Manual Test. Enter the UPPER TEMP, LOWER TEMP, APPROACH LIMIT, and NO. CYCLES parameters. Set the Test Method to MANUAL. Exit to the Run Prog menu. Ensure that the Run Test is set to SWITCH TEST. Set Test Status to 31 914X Field Metrology Wells Prog Menu RUN. Press Enter and the instrument will engage and start the test within a few seconds. Exit to the main screen to view the test progress, refer to the Menu Structure. When the switch resets, the test completes and the values of the switch OPEN, switch CLOSE, and switch BAND are displayed for the user to record. The values may also be recorded internally in the instrument by selecting the option to record the data (-P model only). Figure 13 Auto and manual switch test operation example 32 Menu Structure System Menu 4.3 System Menu 9142/9143/9144 MAIN MENU F1 F2 TEMP SETUP F1 PROG MENU F4 INPUT SETUP (-P only) SYSTEM SETUP F1 DISPLaY SETUP F2 COMMunication SETUP F3 DATE TIME (-P only) LANGUAGE: KEY AUDIO: LINEFEED: TIME STAMP: F1 CALPOINTS F2 CONTRoLler F3 CALibration REFerence (-P only) F4 CALibration UUT (-P only) TEMPerature 1: TEMPerature 2: TEMPerature 3: GRADient 1: GRADient 2: GRADient 3: GRADient 4: (9144 only) GRADient 5: (9144 only) CALibration DATE: TEMPerature Proportional Band: TEMPerature INTegration: TEMPerature DERivative: REF C0: REF C100: INPUT CALibration DATE: < yyyy.mm.dd > F1 CALibration TC F2 CALibration mA TC C0: TC C100: TC CRJ: INPUT CALibration DATE: < yyyy.mm.dd > mA C4: mA C22: INPUT CALibration DATE: < yyyy.mm.dd > SYSTEM INFO MODEL: SERIAL number: FirmWare VERsion: CALibration DATE: INPUT CALibration DATE: : (-P only) Figure 14 Main Menu - System Menu 33 914X Field Metrology Wells Input Setup (-P only) 4.4 Input Setup (-P only) 9142/9143/9144 MAIN MENU F1 F2 TEMP SETUP PROG MENU F1 SELECT INPUT F2 SETUP INPUT F3 F3 SYSTEM MENU F4 INPUT SETUP (-P only) SENSOR TYPE: RTD TYPE: LOOP POWER: INPUT: TEMPerature: REFerence INPUT F1 PROGram PROBE SERIAL number: CALibration DATE: PROBE TYPE: