Bench and System use
The CPX400SP and DP are a bus programmable version of the CPX400S and
CPX400D respectively, and include all of their manual control features plus
comprehensive digital interfaces and rear terminals.
Rear output terminals
Power and sense terminals are duplicated on the rear panel for rack mount
applications or other situations where rear connection is more appropriate.
Digital remote control
To meet the varying needs of today’s engineers, a comprehensive array of
interfaces is provided. RS-232, USB, GPIB and LAN (Ethernet) with LXI support
are provided as standard. Each of the digital bus interfaces provides full control
of voltage, current, and output on/off, plus read-back of voltage, current and
status. The interfaces are at ground potential and are opto-isolated from the
output terminals.
GPIB
The GPIB interface is compliant with IEEE-488.1 and IEEE-488.2. Currently
GPIB remains the most widely used interface for system applications.
RS-232
An RS-232/RS-423 interface is provided for use with legacy systems. This type
of serial interface remains in common useage and is perfectly satisfactory for
the control of power supplies because data speed is not an issue.
USB
USB provides a simple and convenient means of connection to a PC and is
particularly appropriate for small system use. A USB driver is provided which
supports Windows 2000, XP , Vista and Windows 7.
LAN (Ethernet)
The LAN interface uses a standard 10/100 base-T Ethernet hardware connection
with ICMP and TCP/IP Protocol for connection to a Local Area Network or direct
connection to a single PC. This interface supports LXI and is highly appropriate
for system use because of its scalable nature and low cost interconnection.
LXI Compliance
The LAN interface is compliant with LXI-C. LXI (LAN eXtensions for
Instrumentation) is the next-generation, LAN-based modular architecture
standard for automated test systems managed by the LXI Consortium, and
is expected to become the successor to GPIB in many systems. For more
information on LXI and how it replaces GPIB, or operates along side it, go to:
www.tti-test.com/go/lxi
IVI Driver
An IVI driver for Windows is included. This provides support for common high-
level applications such as LabView*, LabWindows*, and HP/Agilent VEE*.
CPX400DP Rear Panel
OUTPUT SPECIFICATIONS (each output)
Voltage/Current/Power Levels
Voltage Range: 0V to 60V.
Current Range: 0A to 20A.
Note: Actual maxima for voltage and current are typically 1% greater than the figures given above.
Power Range: Up to 420W subject to power envelope.
POWER ENVELOPE
The maximum current atany voltage settings is limited by the power envelope which is set to
give 7A at 60V rising to 20A at 20V under all ac supply conditions (both outputs loaded). At
lower output voltages the power is restricted by the 20 amps current maximum.
See PowerFlex power envelope graph on previous page.
Output Setting & Control
Voltage Setting: By coarse and fine controls.
Current Setting: By single logarithmic control.
Output Mode: Constant voltage or constant current with automatic cross-over.
CC indicator lit in constant current mode.
Output Switch: Electronic, non isolating. Preset voltage and current limit displayed
when Output is off. Output rise time no load <10ms.
View Settings: With the output On, the meters show actual voltage and current.
The preset levels can be viewed and adjusted at any time by pressing
the View Settings button.
Status Indication: LED indication of Output On, V/I Limits, CV, CI, Power Limit, Remote,
LAN status. Message on meter display for trip condition.
S-Lock
(Settings Lock) Voltage and current settings can be locked by a single button press.
Lock accuracy is equal to the meter accuracy (see Meter Specifications).
Output Performance
Ripple & Noise: Typically <3mV rms, <15mV pk-pk, (5mV rms max.) in CV mode.
Load Regulation: Voltage - <0·01% of maximum output for any load change within
the PowerFlex envelope (remote sense connected).
Current - <0·05% of maximum output for any load change within
the PowerFlex envelope.
Line Regulation: Voltage - <0·01% of maximum output for a 10% line change.
Current - <0·01% of maximum output for a 10% line change.
Transient Response: <250µs to within 50mV of setting for a 5% to 95% load change.
Temp. Coefficient: Typically <100ppm/°C
Output Protection
Output Protection: Forward protection by Over-voltage Protection (OVP) trip.
Reverse protection by diode clamp for currents to 3A.
OVP Setting/Range: Via screwdriver adjustable preset on front panel. Range 1V to 66V
Over-temperature: Output trips off for over-temperature.
Safety Interlocks: Operations that could cause an unexpected change in voltage or
current settings are interlocked with the output switch.
Output Connections
Output Terminals: Universal 4mm safety binding posts on 19mm (0·75”) spacing at front.
Screw terminals at rear (CPX400SA, SP & DP only).
Terminals can accept fixed shroud 4mm plugs, standard 4mm plugs, fork terminals and bare wires.
Remote Sense
Sense Selection: Voltage sensing is selected as Local or Remote by front panel switch.
Sense Terminals: Sprung loaded screw-less terminals at front.
Screw terminals at rear (CPX400SA, SP & DP only).
METER SPECIFICATIONS (each output)
Display Type: Dual 4-digit meters, 10mm (0·39”) LED.
Voltage Meter
Resolution/ Accuracy: 10mV / ± 0·1% of reading ± 2 digits
Current Meter
Resolution/ Accuracy: 10mA / ± 0·3% of reading ± 20mA
VOLTAGE TRACKING (CPX400D & DP)
Independent Mode
In the normal mode of operation, each output is fully independent and isolated. Operation
is equivalent to two single output power supplies.
Voltage Tracking Mode
The two outputs remain isolated, but the Slave voltage controls are disabled and the Slave
voltage is set equal to the Master voltage. This can be used to generate tracking bipolar
voltages, or tracking unipolar voltages relative to different grounds.
When voltages greater than 60V are required, the outputs can be wired in series to generate
0 to 120V with the voltage controlled from the Master.
When currents greater than 20A are required, the outputs can be wired in parallel to create
the equivalent of a 40A power supply with the voltage controlled from the Master.
Track Accuracy: Slave voltage = ± (0·1% of Master voltage setting + 10mV)
BOTH ON / BOTH OFF (CPX400D & DP)
Each output has an independent DC On/Off control, however, an additional control button is
provided which turn both outputs on or off simultaneously.
Specifications
CPX400SP and CPX400DP
Remote control via GPIB, RS-232, USB and LAN interfaces