525B Temperature/ Pressure Calibrator Superior accuracy and functionality in an economical benchtop package Technical Data enable you to calibrate other process calibrators and a wide variety of other instruments with accuracy that rivals any calibrator in its price range. * Key features: * * * * * The Fluke Calibration 525B Temperature/ Pressure Calibrator gives you a workhorse combination of high accuracy and broad functionality for temperature and pressure instrument calibration. Compact and economical, the 525B has an interface for automated calibration, providing wide workload coverage in instrument shops and calibration labs, as well as in ATE applications. The 525B is the most accurate Fluke Calibration temperature calibrator, sourcing and measuring a complete range of RTDs, thermocouples, and the YSI 400 thermistor. It also measures pressure covering common ranges from 1 inch (6900 Pa) of water up to 10,000 PSI (69 MPa) using the Fluke 700 Series and 525A-P Series Pressure Modules. Plus, the dc voltage and current specifications of the 525B * * * * * Simulates and measures all ANSI thermocouples, as well as L and U types, and provides cold junction compensation to enable calibration of a wide variety of thermocouple instrumentation. Direct input for storage of ITS-90 RTD constants. RTD source uncertainties to 0.03 C. Direct measurement of all Fluke 700 Series and 525A-P Series Pressure Modules covering the most common pressure ranges from 1 H20 to 10,000 PSI. Converts easily to any pressure unit directly from the front panel or through remote communications. Sources dc voltage and current for multifunction workload coverage, enabling calibration of data loggers, strip chart recorders, multimeters, and other industrial instruments. Supports automation using Fluke Calibration's MET/CAL(R) Plus Calibration Management Software or custom automated programs with standard RS-232-C and IEEE-488 interface. Eight user programmable setpoints allow quick recall of values for Zero, Span and linearity checks during calibration. Programmable setpoint dwell times for automated calibration and troubleshooting. * Selectable internal or external CJC (Cold junction compensation) provides you with the ability to make exacting thermocouple measurements using remote junctions. Easy five-way binding posts for low-loss convenient hookup. Banana plug, screw terminal, spade lug, wraparound terminal, wire pass-through terminal. NIST-traceable calibration with data is included. At home in the cal lab The 525B is a compact bench instrument well suited for traditional calibration laboratories with a large temperature workload. Its 0 to 100 V output and 0 to 100 mA current capabilities--plus its ability to serve as a highly-accurate pressure standard--make it a versatile performer. Filling the gap in the instrument shop In process plant environments, instrument shop technicians responsible for maintaining the equipment pool need to calibrate and repair a wide range of instruments, transmitters, field calibrators, P/l converters, transducers, and temperature probes. With its ability to simulate almost any temperature sensor, precise pressure monitoring capability, and dc voltage and current accuracy rivaling any calibrator in its price range, the 525B fills the gap between lower accuracy handheld field calibrators and more expensive high-end multifunction calibrators. A good fit for ATE test applications With precise calibration-grade instrument control enabled via an IEEE bus, the 525B is a hardworking addition to automated test systems in quality and manufacturing test applications requiring capabilities such as voltage and current sourcing, sensor simulation, and temperature monitoring and validation. Powerful functionality that is easy to use From start to finish, the 525B is designed for ease of operation. The intuitive front panel design features large keypads and display that help reduce training time and make the 525B comfortable to use even for long periods of time. Plus, you can store frequently used constants for a variety of probes in memory for faster setup on the job. requirements, including those for ISO 9000 and similar quality standards. They report on forward and reverse traceability, allowing you to easily document traceability from any asset to the equipment calibrated. It even includes measurement uncertainty reporting to meet the requirements of ISO/IEC 17025. Pressure modules An optional set of external pressure modules provides pressure measurement capability. The 525B can accept either the Fluke 700 Series or the 525A-P pressure modules. Both module types plug directly into the calibrator's front panel Lemo connector; the 525B firmware auto-detects the type and value of the attached module. World-class temperature and pressure calibration solutions MET/CAL(R) Plus automates 525B calibration and documentation management In addition to the 525B Temperature/Pressure Calibrator, Fluke Calibration offers a broad range of solutions in temperature and You can use the 525B with humidity calibration as well as MET/CAL Plus calibration software pressure and flow calibration. (V.6.11 or later) to meet the strinFluke Calibration makes everygent documentation and reporting thing you need for calibrating requirements imposed by quality temperature sensors including standards such as ISO 9000. SPRTs, PRTs, thermistors, and MET/CAL Plus is a powerful thermocouples. The constant software environment for creating, temperature baths and dry-block editing, testing, and documenting calibrators offer unmatched stabilcalibration procedures, and for ity, and the primary temperature performing automated calibrations. standards are used in national At the core of MET/CAL Plus metrology institutes around the (R) is MET/TRACK software, a world. For precision thermometry, database management system Fluke Calibration thermometer for test and measurement assets readouts and probes are excepcreated specifically to manage tionally accurate and easy to use, equipment. It enables you to and the 1620A "DewK" temperatrack the information you need to ture and humidity data logger maintain quality calibrations. And handles your environmental it supports the traceability and monitoring needs. record-keeping requirements of Fluke Calibration also offers a modern quality and accreditation range of high-end products focusstandards. ing on pressure and flow calibraMET/TRACK provides you with a tors and standards. More than 30 variety of standard report formats national measurement institutes prepared with Crystal ReportsTM rely on Fluke Calibration pressure Professional, which is included standards to establish their preswith MET/CAL Plus. These report sure references. Fluke Calibration formats cover a wide range of pressure standards, transfer staninformation to help you meet dards, pressure controller/calibramany of your documentation tors and monitors are recognized 2 Fluke Calibration 525B Temperature/Pressure Calibrator for their advanced features and the realistic specifications that are the hallmark of a true metrology company. Calibration service, repair, and support Fluke Calibration provides extensive calibration support and service to maximize the value of your calibration investment. Our worldwide network of Calibration Centers has traceability to national standards. Service and support are just a telephone call or fax away. We offer fast, quality repair and calibration services including a module exchange program, comprehensive training, and full support in setting up your lab. General specifications Warm-up time Twice the time since last warmed up, to a maximum of 30 minutes Settling time Less than 5 seconds for all functions and ranges except as noted Standard interface Standard, RS-232; IEEE-488 (GPIB) Temperature performance Operating: 0 C to 50 C Calibration (tcal): 15 C to 35 C Storage: -20 C to 70 C Electromagnetic compatibility Meets CE requirements, including susceptibility Temperature coefficient Temperature coefficient for temperatures outside tcal +5 C is 10 % of the 90 day specification (or 1 year, as applicable) per C Relative humidity Operating: < 80 % to 30 C, < 70 % to 40 C, < 40 % to 50 C Storage: < 95 %, non-condensing Altitude Operating: 3,050 m (10,000 ft) maximum Non-operating: 12,200 m (40,000 ft) maximum Safety EN 61010 Second, ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92, NRTL Analog low isolation 20 V EMC Designed to comply with IEC 61326-1/1997 (EMC) Line power Line voltage (selectable): 100 V/120 V or 220 V/240 V Line frequency: 47 Hz to 63 Hz Line voltage variation: 10 % about line voltage setting Power consumption 15 VA maximum Dimensions Height: 13.3 cm (5.25 in) plus 1.5 cm (0.6 in) four feet on bottom Width: 0.75 in standard rack width Depth: 47.3 cm (18.6 in) overall Weight (without options) 4 kg (9 lb) Line Frequency ..........................................................47 to 63 Hz Line Voltage Variation .............................................. 10 % about line voltage setting Power consumption.....................................................15 VA maximum Dimensions Height ..........................................................................13.3 cm (5.25 in) plus 1.5 cm (0.6 in) four feet on bottom Width...........................................................................3/4 standard rack width Depth ...........................................................................47.3 cm (18.6 in) overall Weight (without options)..............................................4 kg (9 lb) Electrical Specifications DC Voltage Specifications, Output Absolute Uncertainty, tcal 5 C (ppm of output + V) [1] Ranges 90 days Stability 24 hours, 1 C (ppm of output + V) 3 5+2 10 4 + 10 100 4 + 100 1 mV 5 + 1 mV TC Output 3 5+2 Resolution Maximum [2] Burden 1 V 10 V 100 V 1 mV 10 mA 10 mA 10 mA 1 mA 1 V 10 1 year 0 to 100.000 mV 0 to 1.00000 V 0 to 10.0000 V 0 to 100.000 V 25 25 25 25 3 10 100 1 mV 30 30 30 30 -10 to 75.000 mV 25 3 30 Notes: [1] All outputs are positive only. [2] Remote sensing is not provided. Output resistance is < 1. Noise Ranges Bandwidth 10 Hz to 10 kHz (V rms) 6 V 60 V 600 V 20 mV Bandwidth 0.1 to 10 Hz (ppm of output + V p-p) 1 V 10 V 100 V 10 ppm+1 mV 0 to 100.000 mV 0 to 1.00000 V 0 to 10.0000 V 0 to 100.000 V DC Current Specifications, Output [1] Ranges 0 to 100.000 mA Absolute Uncertainty tcal 5 C (ppm of output + A) 90 days 1 year 40 1 50 1 Resolution Maximum Compliance Voltage Maximum Inductive Load 1 A 12 V 100 mH Note: [1] All outputs are positive only. Noise Ranges 0 to 100.000 mA Bandwidth 0.1 to 10 Hz p-p 2000 nA Bandwidth 10 Hz to 10 kHz rms 20 A Resistance Specifications, Output [1] Ranges 5 to 400.00 5 to 4.0000 k Absolute Uncertainty tcal 5C, 90 days 1 year 0.012 0.015 0.25 0.3 [2] Resolution Allowable Current 0.001 0.01 1 to 3 mA 100 A to 1 mA Notes: [1] Continuously variable from 0 to 4 k. [2] For currents lower than shown, the floor adder increases by Floor(new) = Floor(old) x Imin/Iactual. For example, a 500 A stimulus measuring 100 has a floor uncertainty of 0.015 x 1 mA/500 A = 0.03 . Resistance Specifications, Input [1] Ranges 5 to 400.00 5 to 4.00000 k Absolute Uncertainty tcal 5C, (ppm of output + ) 90 days 1 year 20 0.0035 20 0.004 20 0.0035 20 0.004 Resolution Stimulus Current 0.001 0.01 1 mA 0.1 mA Note: [1] 4-wire mode. 525B Temperature/Pressure Calibrator Fluke Calibration 3 Thermocouple Specification, Output and Input Range (C) TC Type B C E J K L N R S T U mV Minimum 600 C 800 C 1000 C 1550 C 0 C 150 C 650 C 1000 C 1800 C -270 C -100 C -25 C 350 C 650 C -270 C -100 C -25 C 120 C 1000 C -270 C -100 C -25 C 120 C 1000 C -200 C -100 C 800 C -270 C -100 C -25 C 120 C 410 C -50 C 250 C 400 C 1000 C 0 C 250 C 1000 C 1400 C -270 C -150 C 0 C 120 C -200 C 0 C Maximum 800 C 1000 C 1550 C 1820 C 150 C 650 C 1000 C 1800 C 2316 C -100 C -25 C 350 C 650 C 1820 C -100 C -30 C 150 C 760 C 1200 C -100 C -25 C 120 C 1000 C 1372 C -100 C 800 C 900 C -100 C -25 C 120 C 410 C 1300 C 250 C 400 C 1000 C 1760 C 250 C 1000 C 1400 C 1750 C -150 C 0 C 120 C 400 C 0 C 600 C -10 to 75.000 mV Note: [1] Does not include thermocouple wire error. 4 Fluke Calibration 525B Temperature/Pressure Calibrator Absolute Uncertainty [1] tcal 5 C, (C) Output/Input 90 days 1 year 0.42 C 0.46 C 0.39 C 0.39 C 0.40 C 0.40 C 0.44 C 0.45 C 0.25 C 0.30 C 0.21 C 0.26 C 0.23 C 0.31 C 0.38 C 0.50 C 0.63 C 0.84 C 0.38 C 0.50 C 0.16 C 0.18 C 0.14 C 0.15 C 0.14 C 0.16 C 0.16 C 0.21 C 0.20 C 0.27 C 0.18 C 0.20 C 0.14 C 0.16 C 0.14 C 0.17 C 0.18 C 0.23 C 0.25 C 0.33 C 0.19 C 0.22 C 0.14 C 0.16 C 0.19 C 0.26 C 0.30 C 0.40 C 0.37 C 0.37 C 0.26 C 0.26 C 0.17 C 0.17 C 0.33 C 0.40 C 0.20 C 0.24 C 0.16 C 0.19 C 0.14 C 0.18 C 0.21 C 0.27 C 0.58 C 0.58 C 0.34 C 0.35 C 0.31 C 0.33 C 0.30 C 0.40 C 0.56 C 0.56 C 0.36 C 0.36 C 0.30 C 0.37 C 0.35 C 0.46 C 0.51 C 0.63 C 0.18 C 0.24 C 0.13 C 0.16 C 0.12 C 0.14 C 0.56 C 0.56 C 0.27 C 0.27 C RTD and Thermistor Specification, Output Range (C) RTD Type Pt 385, 100 Pt 3926, 100 Pt 3916, 100 Pt 385, 200 Pt 385, 500 Pt 385, 1000 Ni 120, 120 Cu 427, 10 YSI 400 [2] Minimum -200 C -80 C 0 C 100 C 300 C 400 C 630 C -200 C -80 C 0 C 100 C 300 C 400 C -200 C -190 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C -200 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C -200 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C -200 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C -80 C 0 C 100 C -100 C 15 C Maximum -80 C 0 C 100 C 300 C 400 C 630 C 800 C -80 C 0 C 100 C 300 C 400 C 630 C -190 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C 630 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C 630 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C 630 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C 630 C 0 C 100 C 260 C 260 C 50 C Absolute Uncertainty [1] tcal 5 C, (C) 90 days 1 year 0.03 C 0.04 C 0.04 C 0.05 C 0.04 C 0.05 C 0.03 C 0.04 C 0.04 C 0.04 C 0.04 C 0.05 C 0.04 C 0.05 C 0.03 C 0.04 C 0.03 C 0.04 C 0.03 C 0.04 C 0.03 C 0.04 C 0.03 C 0.04 C 0.04 C 0.05 C 0.03 C 0.03 C 0.03 C 0.04 C 0.03 C 0.04 C 0.03 C 0.04 C 0.03 C 0.04 C 0.03 C 0.04 C 0.03 C 0.04 C 0.04 C 0.05 C 0.04 C 0.05 C 0.31 C 0.38 C 0.32 C 0.38 C 0.33 C 0.39 C 0.33 C 0.39 C 0.36 C 0.43 C 0.36 C 0.43 C 0.42 C 0.50 C 0.42 C 0.50 C 0.13 C 0.15 C 0.13 C 0.15 C 0.13 C 0.16 C 0.14 C 0.17 C 0.14 C 0.17 C 0.15 C 0.18 C 0.16 C 0.19 C 0.16 C 0.19 C 0.06 C 0.07 C 0.06 C 0.08 C 0.07 C 0.08 C 0.07 C 0.08 C 0.07 C 0.09 C 0.07 C 0.09 C 0.08 C 0.09 C 0.08 C 0.09 C 0.02 C 0.02 C 0.02 C 0.02 C 0.01 C 0.02 C 0.30 C 0.38 C 0.005 C 0.007 C Notes: [1] 2-wire output. [2] Based on MINCO Application Aid No. 18. 525B Temperature/Pressure Calibrator Fluke Calibration 5 RTD and Thermistor Specification, Input Range (C) RTD Type Pt 385, 100 Pt 3926, 100 Pt 3916, 100 Pt 385, 200 Pt 385, 500 Pt 385, 1000 PtNi 385, 120 (Ni120) Cu 427, 10 YSI 400 SPRT [2] Minimum -200 C -80 C 0 C 100 C 300 C 400 C 630 C -200 C -80 C 0 C 100 C 300 C 400 C -200 C -190 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C -200 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C -200 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C -200 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C -80 C 0 C 100 C -100 C 15 C -200 C Maximum -80 C 0 C 100 C 300 C 400 C 630 C 800 C -80 C 0 C 100 C 300 C 400 C 630 C -190 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C 630 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C 630 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C 630 C -80 C 0 C 100 C 260 C 300 C 400 C 600 C 630 C 0 C 100 C 260 C 260 C 50 C 660 C Notes: [1] 4-wire mode. Uncertainties listed do not include probe uncertainties. [2] Based on MINCO Application Aid No. 18. 6 Fluke Calibration 525B Temperature/Pressure Calibrator Absolute Uncertainty [1] tcal 5 C, (C) 90 days 1 year 0.011 C 0.012 C 0.018 C 0.020 C 0.018 C 0.020 C 0.027 C 0.030 C 0.031 C 0.035 C 0.042 C 0.047 C 0.050 C 0.057 C 0.011 C 0.011 C 0.014 C 0.015 C 0.018 C 0.019 C 0.026 C 0.029 C 0.031 C 0.034 C 0.041 C 0.046 C 0.006 C 0.006 C 0.011 C 0.012 C 0.014 C 0.015 C 0.018 C 0.019 C 0.025 C 0.028 C 0.026 C 0.029 C 0.031 C 0.034 C 0.040 C 0.045 C 0.042 C 0.047 C 0.008 C 0.009 C 0.012 C 0.013 C 0.015 C 0.017 C 0.020 C 0.022 C 0.050 C 0.053 C 0.053 C 0.057 C 0.070 C 0.075 C 0.071 C 0.076 C 0.007 C 0.008 C 0.019 C 0.020 C 0.023 C 0.025 C 0.030 C 0.033 C 0.032 C 0.035 C 0.037 C 0.041 C 0.047 C 0.052 C 0.048 C 0.053 C 0.011 C 0.012 C 0.014 C 0.015 C 0.019 C 0.020 C 0.025 C 0.028 C 0.027 C 0.030 C 0.030 C 0.034 C 0.041 C 0.045 C 0.042 C 0.047 C 0.009 C 0.010 C 0.010 C 0.011 C 0.011 C 0.012 C 0.067 C 0.069 C 0.005 C 0.007 C 0.05 C 0.06 C 700 Series Pressure Modules Specifications Differential Gage Absolute Vacuum Dual High Model Range/ Resolution Range (approx)/ Resolution Reference1 uncertainty (23 3 C) High2 side media Low2 side media Fitting material Fluke 700P00 1 in. H2O/0.001 0.25 kPa/0.0002 0.300 % Dry Dry 316 SS Fluke 700P01 10 in. H2O/0.01 2.5 kPa/0.002 0.200 % Dry Dry 316 SS Fluke 700P02 1 psi/0.0001 6900 Pa/0.7 0.150 % Dry Dry 316 SS Fluke 700P22 1 psi/0.0001 6900 Pa/0.7 0.100 % 316 SS Dry 316 SS Fluke 700P03 5 psi/0.0001 34 kPa/0.001 0.050 % Dry Dry 316 SS Fluke 700P23 5 psi/0.0001 34 kPa/0.001 0.025 % 316 SS Dry 316 SS Fluke 700P04 15 psi/0.001 103 kPa/0.01 0.025 % Dry Dry 316 SS Fluke 700P24 15 psi/0.001 103 kPa/0.01 0.025 % 316 SS Dry 316 SS Fluke 700P05 30 psi/0.001 207 kPa/0.01 0.025 % 316 SS N/A 316 SS Fluke 700P06 100 psi/0.01 690 kPa/0.07 0.025 % 316 SS N/A 316 SS Fluke 700P27 300 psi/0.01 2070 kPa/0.1 0.025 % 316 SS N/A 316 SS Fluke 700P07 500 psi/0.01 3400 kPa/0.1 0.025 % 316 SS N/A 316 SS Fluke 700P08 1000 psi/0.1 6900 kPa/0.7 0.025 % 316 SS N/A 316 SS Fluke 700P09 1500 psi/0.1 10 M Pa/0.001 0.025 % 316 SS N/A 316 SS Fluke 700PA3 5 psi/0.0001 34 kPa/0.001 0.050 % 316 SS N/A 316 SS Fluke 700PA4 150 psi/0.001 103 kPa/0.001 0.050 % 316 SS N/A 316 SS Fluke 700PA5 30 psi/0.001 207 kPa/0.01 0.050 % 316 SS N/A 316 SS Fluke 700PA6 100 psi/0.01 690 kPa/0.001 0.050 % 316 SS N/A 316 SS Fluke 700PV3 -5 psi/0.0001 -34 kPa/0.001 0.040 % 316 SS Dry 316 SS Fluke 700PV4 -15 psi/0.001 -103 kPa/0.01 0.040 % 316 SS Dry 316 SS Fluke 700PD2 1 psi/0.0001 6900 Pa/0.7 0.150 % 316 SS Dry 316 SS Fluke 700PD3 5 psi/0.0001 34 kPa/0.001 0.040 % 316 SS Dry 316 SS Fluke 700PD4 15 psi/0.001 103 kPa/0.01 0.025 % 316 SS Dry 316 SS Fluke 700PD5 -15/30 psi/0.001 -100/207 kPa/0.01 0.025 % 316 SS N/A 316 SS Fluke 700PD6 -15/100 psi/0.01 -100/690 kPa/0.07 0.025 % 316 SS N/A 316 SS Fluke 700PD7 -15/200 psi/0.01 -100/1380 kPa/0.1 0.040 % 316 SS N/A 316 SS Fluke 700P29 3000 psi/0.1 20.7 M Pa/0.001 0.050 % C276 N/A C276 Fluke 700P30 5000 psi/0.1 34 M Pa/0.001 0.050 % C276 N/A C276 Fluke 700P31 10000 psi/1 69 M Pa/0.007 0.050 % C276 N/A C276 1 Total uncertainty, one year for temperature range O C to +5O C. Total uncertainty, 1.0% of full span for temperature range -10 C to O C. For POO module only, compensated temperature range is 15 C to 35 C. 2 "Dry" indicates dry air or non-corrosive gas as compatible media. "316SS" indicates media compatible with Type 316 Stainless Steel. "C276" indicates media compatible with Hastelloy C276. Use of pressure zero is required prior to measurement or source. Max overpressure specification includes common mode pressure. Modules are CE rated. Metric adapter(s}: 1/4" NPr female to male BSP/ISO 1/419, tapered thread, included with all modules except P29, P3O, and P31, all modules include a NIST traceable certificate and test data. 525A-P Series Precision Pressure Transducers Type Model Range/Resolution Range/Resolution Differential 525A-P02 1 psi/0.00001 6900 Pa/0.01 Reference Uncertainty (23 3 C) 0.008 % FS Gage 525A-P03 5 psi/0.00001 34 kPa/0.001 0.008 % FS Gage 525A-P04 15 psi/0.001 103 kPa/0.001 0.008 % FS Gage 525A-P05 30 psi/0.0001 207 kPa/0.001 0.008 % FS Gage 525A-P06 100 psi/0.001 690 kPa/0.001 0.008 % FS Gage 525A-P07 500 psi/0.001 3400 kPa/0.01 0.008 % FS Gage 525A-P08 1000 psi/0.01 6900 kPa/0.01 0.008 % FS Gage 525A-P29 3000 psi/0.01 20.7 M Pa/0.0001 0.008 % FS Absolute 525A-PA4 15 psi/0.0001 103 kPa/0.001 0.008 % FS Absolute 525A-PA5 30 psi/0.0001 207 kPa/0.001 0.008 % FS Absolute 525A-PA6 100 psi/0.001 690 kPa/0.001 0.008 % FS Absolute 525A-PA7 500 psi/0.001 3400 kPa/0.01 0.008 % FS Absolute 525A-PA8 1000 psi/0.01 6900 kPa/0.01 0.008 % FS Vacuum 525A-PV4 -15 TO 0 psi/0.0001 -34 kPa/0.001 0.008 % FS 525B Temperature/Pressure Calibrator Fluke Calibration 7 Pressure Measurement The Calibrator can accept either the Fluke 700 or 525A-P series pressure modules. Pressure modules plug directly into the front panel Lemo connector with the Calibrator firmware autodetecting the type and value of the module you are attaching. Range ............................................Determined by pressure module Accuracy/Resolution.................Determined by pressure module Units PSI ..............................................pounds per square inch inH2O4C ..................................inches of water at 4 degrees Celsius inH2O20C................................inches of water at 20 degrees Celsius cmH2O4C.................................centimeters of water at 4 degrees Celsius cmH2O20C ..............................centimeters of water at 20 degrees Celsius BAR ............................................bars mBAR .........................................millibars KPAL ..........................................kilopascals inHG 0C ...................................inches of mercury at 0 degrees Celsius mmHG 0C ................................millimeters of mercury at 0 degrees Celsius Kg/cm2 ......................................kilograms per square centimeter Model Description 525B Temperature/Pressure Calibrator Software MET/CAL Plus Automated Calibration Software (IEEE-488 and RS-232) Accessories 5520A-525A/ Leads Thermocouple and test lead set 1 meter TC wire with male mini-plugs, All ANSI, L and U types; TC shorting plug; 1, 80AK-8001 TC adapter; 1, bead thermocouple type K; 1 meter shielded test lead with double banana plugs (M); 2, 30 A test leads, banana type connectors Y525 19" Rack Mount Kit Fluke 700PD3 Dual Pressure Module 5 psi / 34 kPa Fluke 700P31 High Pressure Module 10000 psi/69 M Pa Fluke 700P22 Differential Pressure Module 1 psi / 6900 Pa Fluke 700P24 Differential Pressure Module 15 psi / 103 kPa Fluke 700PD2 Dual Pressure Module 1 psi / 6900 Pa Fluke 700P08 Gage Pressure Module 1000 psi / 6900 kPa Fluke 700PA4 Absolute Pressure Module 15 psi / 103 kPa Fluke 700P30 High Pressure Module 5000 psi / 34 M Pa Fluke 700PD7 Dual Pressure Module -15/200 psi / -100/1380 kPa Fluke 700P05 Gage Pressure Module 30 psi / 207 kPa Fluke 700P06 Gage Pressure Module 100 psi / 690 kPa Fluke 700P29 High Pressure Module 3000 psi / 20.7 M Pa Fluke 700PD5 Dual Pressure Module -15 / 30 psi / -100 / 207 kPa Fluke 700PV3 Vacuum Pressure Module -5 psi / -34 kPa Fluke 700P23 Differential Pressure Module 5 psi / 34 kPa Fluke 700PA5 Absolute Pressure Module 30 psi / 207 kPa Model Description Fluke 700P03 Differential Pressure Module 5 psi / 34 kPa Fluke 700PD6 Dual Pressure Module -15/100 psi / -100/690 kPa Fluke 700P04 Differential Pressure Module 15 psi / 103 kPa Fluke 700PA6 Absolute Pressure Module 100 psi / 6900 kPa Fluke 700P01 Differential Pressure Module 10.00 in. H2O / 2.5 kPa Fluke 700PD4 Dual Pressure Module 15 psi / 103 kPa 525A-P02 Differential 1 psi/6900 Pa 525A-P03 Gage 5 psi/34 kPa 525A-P04 Gage 15 psi/103 kPa 525A-P05 Gage 30 psi/207 kPa 525A-P06 Gage 100 psi/690 kPa 525A-P07 Gage 500 psi/3400 kPa 525A-P08 Gage 1000 psi/6900 kPa 525A-P29 Gage 3000 psi/20.7 M Pa 525A-PA4 Absolute 15 psi/103 kPa 525A-PA5 Absolute 30 psi/207 kPa 525A-PA6 Absolute 100 psi/690 kPa 525A-PA7 Absolute 500 psi/3400 kPa 525A-PA8 Absolute 1000 psi/6900 kPa 525A-PV4 Vacuum -15 to 0 psi/-34 kPa Support MET/SUPPORT Gold Software Support Program, one-year membership MET/SUPPORT GLDNW MET/SUPPORT Gold Program, up to four workstations, one-year membership Accessories continued Fluke Calibration. Precision, performance, confidence.TM Fluke Corporation PO Box 9090, Everett, WA 98206 U.S.A. Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands 8 Fluke Calibration 525B Temperature/Pressure Calibrator (c)2001-2012 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 6/2012 1611141D_EN Pub_ID: 10884-eng Rev 02 For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1(425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com