Total Power at Input Mixer >0 dBm
Note: Mixer power level (dBm) = input power (dBm) – input attenuation (dB).
Displayed Average Noise Level (DANL)
0 dB RF Attenuation, RBW=100 Hz,
VBW=1Hz, RMS Average Detector, Trace
Average ≥ 50, Input Impedance=50 Ω,
Tracking Generator Off.
DANL
(Preamplier Off)
100 kHz to 10 MHz:
<-85 dBm-3 x (f/1 MHz) dB, typical -125 dBm
10 MHz to 2.5 GHz:
<-127 dBm+3 x (f/1GHz) dB,
typical -130 dBm
2.5 GHz to 3 GHz:<-115 dBm
100 kHz to 10 MHz:
<-75 dBm-3 x (f/1 MHz) dB, typical -115 dBm
10 MHz to 2.5 GHz:
<-117 dBm+3 x (f/1 GHz) dB,
typical -120 dBm
2.5 GHz to 3 GHz:<-105 dBm
DANL
(Preamplier On)
100 kHz to 1 MHz:<-103 dBm
1 MHz to 10 MHz:<-103 dBm-3 x (f/1 MHz) dB,
typical -143 dBm
10 MHz to 2.5 GHz:<-145 dBm+3 x (f/1 GHz)
dB, typical -148 dBm
2.5 GHz to 3 GHz:<-133 dBm
100 kHz to 1 MHz:<-93 dBm
1 MHz to 10 MHz:<-93 dBm-3 x (f/1 MHz)
dB, typical -133 dBm
10 MHz to 2.5 GHz:<-135 dBm+3 x (f/1
GHz) dB, typical -138 dBm
2.5 GHz to 3 GHz:<-123 dBm
Level Display Range
Log Scale 1 dB to 200 dB
Linear Scale 0 to Reference Level
Number of Display Points Normal: 601; Full Screen: 751
Number of Traces 3 + Math Trace
Trace Detectors Normal, Positive-peak, Negative-peak, Sample, RMS, Voltage Average, Quasi-Peak
Trace Functions Clear Write, Max Hold, Min Hold, Average, Freeze, Blank
Scale Units dBm, dBmV, dBμV, V, W
Frequency Response
10 dB RF Attenuation, Relative to 50 MHz, 20℃ to 30℃
Frequency Response Preamplier Off , 100 kHz to 3 GHz: <0.7 dB
Preamplier On , 1 MHz to 3 GHz: <1.0 dB
Preamplier Off , 100 kHz to 3 GHz: <1.0 dB
Preamplier On , 1 MHz to 3 GHz: <1.4 dB
Input Attenuation Switching Uncertainty
Setting Range 0 to 50 dB, in 1 dB step
Switching Uncertainty (fc=50 MHz,
relative to 10 dB, 20 ºC to 30 ºC) < (0.3 + 0.01 x attenuator setting) dB <0.8 dB
Absolute Amplitude Uncertainty
Uncertainty
(fc=50 MHz, peak detector,
preamplier off, 10 dB RF
attenuation, input signal=-10 dBm,
20 ºC to 30 ºC)
±0.4 dB
RBW Switching Uncertainty
10 Hz to 1 MHz, relative
to 1 kHz RBW 100 Hz to 1 MHz, relative to 1 kHz RBW
Uncertainty <0.1 dB
Reference Level
Range -100 dBm to +30 dBm, in 1 dB step
Resolution Log Scale: 0.01 dB; Linear Scale: 5 digits
Level Measurement Uncertainty
95% condence level, S/N>20 dB, RBW=VBW=1 kHz, preamplier off, 10 dB RF attenuation, -50 dBm<reference level<0,10
MHz<fc<3 GHz, 20 ºC to 30 ºC
Level Measurement Uncertainty <1.0 dB, nominal <1.5 dB, nominal
RF Input VSWR
10 dB RF Attenuation
VSWR
100 kHz to 10 MHz: <1.8, nominal
10 MHz to 2.5 GHz: <1.5, nominal
2.5 GHz to 3 GHz: <1.8, nominal
0 dB RF Attenuation, RBW=10
Hz, VBW=1Hz, RMS Average
Detector, Trace Average ≥ 50,
Input Impedance=50 Ω, Tracking
Generator Off.
5RIGOL