RIGOL
Data Sheet
DM3058/DM3058E Digital Multimeter
Note[1]: The difference between DM3058 and DM3058E is only that DM3058E doesn’t support LAN and GPIB
interfaces. Nov. 2015
RIGOL TECHNOLOGIES, INC.
Product Overview
DM3058/DM3058E[1] is a digital
multimeter designed with 5½ di gi ts
readings resolution and dual-display
especially fitting to the needs of
high-precision, multifunction and
automatic measurement.
Applications
Research&Development Laboratory
Scientific Research and Educ at i o n
Detection and Maintenance
Quality Test
Automatic Prod uction T est
Powerful Measurement Functions
Basic Measurement Function
DC Vol tage: 200 mV ~ 1000 V
DC Current: 200 μA ~ 10 A
AC Voltage: True-RMS, 200 mV ~ 750 V
AC Current: True-RMS, 20 mA ~ 10 A
2-Wire, 4-Wire Resistance: 200 Ω ~ 100
MΩ
Capacitance Meas ure ment: 2 nF ~ 10000
μF
Continuity Test: Range is fixed at 2 kΩ
Diode Test: Range is fixed at 2.0 V
Frequency Measurement: 20 Hz ~ 1 MHz
Period Measurement: 1 μs ~ 0.05 s
Any Sensor Me a s urement: Support for 6
types of sensor (DCV, DCI , F req, 2WR,
4 WR an d TC)
Math Function
Max, Min, Ave ra ge, Sta n dard De v i a t i on,
Pass/Fail, dBm, dB, Relative Measurement
and Histogram
Main Features
Real digits readings res olution
Up to 123 rdgs/s measurement speed
True-RMS AC Voltage and AC Current
measuring
Quickly Save or Re call the 10 groups of
Preset Configuration
Preset 10 groups of Standard Sensor
Configuration, built-in cold terminal
compensation f or thermocouple
Clone the instrument configurations into
other DM3058/DM3058E via USB storage
device
The first digit digital multimeter
passing LXI Criterion in industry, which
can achieve system integration easily
With easy, convenient and flexible any
sensor measurement control software:
UltraSensor
Standard configuration interface: USB
Device, USB Host, LAN (only fo r DM3058),
RS-232, GPIB (onl y fo r DM 3 0 58)
Sup port remo te cont rol via commands and
compatible wit h comm ands of main
stream multimeters
Easy to Use Design
256×64 LCD
Sup port double display, Chinese and
English menu
Built-in help system makes information
acquisition more easier
File management (su pp ort for USB-disk
and local storage)
1
Double Display Mirror Image Configuration
Preset Mode Pass/Fail
Any Sensor Measurement LXI & Web Control
Easy, Convenient and Flexible Control Software
Dual-display function makes your
measurements achieve maximum results with
little effort. It can dis play two performances of
a signal simultaneously instead of two
multimeters or measurements in the past.
Any sensor measurement is a new
conception being forward to meet user s
requirements. By this particular fun ction , you
can easily connect pressure sens or or flux
sensor or temperature sensor.
The multimeter supports 6 kinds of sensor
(DCV, DCI , Freq, 2WR, 4WR and TC).
Meanwhile, there are 10 groups of standard
sensor configuration preset within the
instrument.
Thermocouple Col d Terminal compensation is
built in the multimet er.
configuration) can be cloned into other
DM3058/DM3058E on prod u ct line via USB
storage device to impr ove work efficiency.
Based on Preset Mode, the worker operation
on product line could be greatly simplified.
Besides, 10 groups of Preset Configuration
are available to be stored and recall e d.
Pass/Fail test function can prompt for signals
beyond the range based o n specified higher
and lower limit, and make test result more
obvious.
DM3058 is the first digit d igital multim ete r
passing LXI Criterion in industry that makes
system integration easier.
Remote Control for DM3058 can be achieved
by virtual panel from Web page in the same
operation way with front panel of instrument.
UltraSensor is used to control any sensor. The main
functions include:
Create a project for any sensor measurement
which can be download to the multi meter;
Connect with the multimeter to achieve any
sensor measurement;
Monitor the sen sor data in real time and sho w
them in the form of figure;
Enable to save data in the format of CSV and TXT
and reference curve in the format of BMP.
UltraSensor Software Interface
2
Specifications
DC Characteristics Accuracy ± (% of reading + % of range)[1]
Function
Range[2]
Test current or
Load voltage
1 Year
23
± 5
Temperature
Coefficient
0
- 18
28
- 50
DC V oltag e
200.000 mV
0.015 + 0.004
0.0015 + 0.0005
2.00000 V
0.015 + 0.003
0.0010 + 0.0005
20.0000 V
0.015 + 0.004
0.0020 + 0.0005
200.000 V
0.015 + 0.003
0.0015 + 0.0005
1000.00 V[4]
0.015 + 0.003
0.0015 + 0.0005
DC Current
200.000 μA
<8 mV
0.055 + 0.005
0.003 + 0.001
2.00000 mA
<80 mV
0.055 + 0.005
0.002 + 0.001
20.0000 mA
<0.05 V
0.095 + 0.020
0.008 + 0.001
200.000 mA
<0.5 V
0.070 + 0.008
0.005 + 0.001
2.00000 A
<0.1 V
0.170 + 0.020
0.013 + 0.001
10.0000 A[5]
<0.3 V
0.250 + 0.010
0.008 + 0.001
Resistance[3]
200.000 Ω
1 mA
0.030 + 0.005
0.0030 + 0.0006
2.00000
1 mA
0.020 + 0.003
0.0030 + 0.0005
20.0000
100 μA
0.020 + 0.003
0.0030 + 0.0005
200.000
10 μA
0.020 + 0.003
0.0030 + 0.0005
2.00000
1 μA
0.040 + 0.004
0.0040 + 0.0005
10.0000
200 nA
0.250 + 0.003
0.0100 + 0.0005
100.000
200 nA || 10
1.75 + 0.004
0.2000 + 0.0005
Diode Test
2.0000 V[6]
1 mA
0.05 + 0.01
0.0050 + 0.0005
Continuity
Test
2000 Ω
1 mA
0.05 + 0.01
0.0050 + 0.0005
Remarks
[1] Specifications are for 0.5 hour warm-up, Slow” measurement rate and calibration temperature 18 ~ 28.
[2] 20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
[3] Specif ic ations are for 4-wire measure or 2-wire measu re under REFoperation. ± 0.2 Ω of ext ra errors wil l be
generated if perform 2-wire measure without REFoperation.
[4] Plus 0.02 mV of error per 1 V after the first ±500 VDC.
[5] 30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or AC RMS
7 A.
[6] Accuracy specifications are only for voltage measuring at input terminal. The typical value of current under
measure is 1 mA. Voltage drop at the diode junction may vary with c u r ren t su pply.
DC Voltage
Input Resistance
200 mV and 2 V 10 or >10 selectable
(Input si gnals that exceed ± 2.5 V i n the se ranges will pass the 1 00 (typical) clamp
resistance.)
20 V, 200 V and 1000 V 10 MΩ ± 2%
Input Bias Current
<90 pA, 25
Input Protection
1000 V on all ranges
CMRR (common mode
rejection ratio)
120 dB (For the 1 kΩ unbalanced resistance in LO lead, maximum ± 500 VDC
NMRR (normal mode
rejection ratio)
60 dB at slowme asu r ement rate
20 dB are added if open the AC filter(Settling time will add 0.35 s while source
impedance nears zero)
Resistance
Testing Method
4-wire resistance or 2-wire resistance optional
For current source refer to LO input
Open-circuit V oltage
P/F in <8 V
Maximum Lead
Resistance
(4-wire resistance)
10% of ranges , on the range of 200 Ω, 1 , for each lead
1 kΩ, on all other ra ng es , for each lead
Input Protection
1000 V, on all ranges
3
DC Current
Shunt Resistor
200 μA sampling voltage < 8 mV
2 mA sampling volta ge < 80 mV
1Ω for 20 mA, 200 mA
0.008 Ω for 2 A, 10 A
Input Protection
Rear panel: accessible 10 A, 250 V fast-melt fuse
Internal: 12 A, 250 V slow-melt fuse
Continuity / Diode Test
Measurement Method
1 mA ± 5% constant-current source, < 8 V open-circuit voltage
Response Time
123 samples/sec, with beeper
Continuit y Thre shold
Adjustable between 1 Ω and 2000 Ω
Input Protection
1000 V
Setup time Attentions
The s etu p time about vol tage measurement is in f luenced by sour c e resi st an ce and m edi a c harac ter istic s o f c abl e as
well as input signal.
AC Characteristics Accuracy ± (% of reading + % of range) [1]
Function
Range[2]
Frequency Range
1 Year
23
±5
Temperature
Coefficient
0
- 18
28
- 50
True RMS
AC Voltage[3]
200.000 mV
20 Hz 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz 100 kHz
3.0 + 0.05
0.05 + 0.010
2.00000 V
20 Hz 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz 100 kHz
3.0 + 0.05
0.05 + 0.010
20.0000 V
20 Hz 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz 100 kHz
3.0 + 0.05
0.05 + 0.010
200.000 V
20 Hz 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz 100 kHz
3.0 + 0.05
0.05 + 0.010
750.000 V
20 Hz 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz 100 kHz
3.0 + 0.05
0.05 + 0.010
True RMS
AC Current[5]
20.0000 mA
20Hz45 Hz
1.5 + 0.10
0.015 + 0.015
45 Hz - 2 kHz
0.50 + 0.10
0.015 + 0.006
2 kHz -10 kHz
2.50 + 0.20
0.015 + 0.006
200.000 mA
20 Hz - 45 Hz
1.50 + 0.10
0.015 + 0.005
45 Hz2 kHz
0.30 + 0.10
0.015 + 0.005
2 kHz - 10 kHz
2.50 + 0.20
0.015 + 0.005
2.00000 A
20 Hz45 Hz
1.50 + 0.20
0.015 + 0.005
45 Hz - 2 kHz
0.50 + 0.20
0.015 + 0.005
2 kHz10 kHz
2.50 + 0.20
0.015 + 0.005
10.0000 A[5]
20 Hz45 Hz
1.50 + 0.15
0.015 + 0.005
45 Hz - 2 kHz
0.50 + 0.15
0.015 + 0.005
2 kHz – 5 kHz
2.50 + 0.20
0.015 + 0.005
Additional wave crest factor error (not Sine)[6]
Wave crest coefficient
Error (% range)
1 - 2
0.05
2 - 3
0.2
Remarks:
[1] Specifications are for 0.5 hour warm-up, Slowmeasure and calibration temperature 18 - 28.
[2] 20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
4
[3] Specifications are for amplitude of sine wave input >5% of range. 750 V range limited to 8x107 Volt-Hz. For inputs
from 1% to 5% of range and <50 kHz, add 0.1% of range extra error. For 50 kHz to 100 kHz, add 0.13%.
[4] Specifications are for sine wave input >5% of range. 0.1% errors will be added when the range of input sine wave
is 1% ~ 5%.
[5] 30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or AC RMS
7 A.
[6] F or frequency<100 Hz.
True RMS AC Voltage
Measurement Method
AC coupled true RMS measure - up to 1000 V DC bias are permitted on every range
Wave Crest Factor
≤ 3 at full scale
Input Impedance
1 ± 2% in parallel with <100 pF on all ranges
AC Filter Ban dwidt h
20 Hz ~ 100 kHz
CMRR (common mode
rejection ratio)
60 dB (Fo r th e 1 imbalance resi stance among Lo l ead and <60 Hz, maximum ±500
VDC)
True RMS AC Current
Measurement Method
DC coupled to the fuse and shunt; AC coupled the True RMS measurement (measures the
AC components only)
Wave Crest Factor
≤ 3 at full scale
Maximum Input
The DC + AC current peak va lue <300% of range . The RMS current includ ing DC current
is <10 A
Shunt Resistor
1 Ω for 20 mA, 200 mA
0.008 Ω for 2 A, 10 A
Input Protection
Rear panel: accessible 10 A, 250 V fast-melt fuse
Internal: 12 A, 250 V slow-melt fuse
Setup Time Attentions
Mak e sure that the RC return at input terminal has been in a stable state completely (higher than 1 s) before accurate
measurement;
Input >300 Vrms (or >5 Arms) will cause the self heating of the signal conditioning component to generate error, this
error is included in the characteristics of the instrument. Internal temperature variation results from the self heating
will cause additional an error on ac range that is lower than 0.03% of readings, and it will disappear after a few
minutes.
Frequency a n d Period Characteris tics Accuracy ± (% of reading + % of range) [1]
Function
Range
Frequency Range
1 Year
23
±5
Temperature
Coefficient
0
- 18
28
- 50
Frequency/Period
200 mV
750 V[2]
20 Hz - 2 kHz
0.01 + 0.003
0.002 + 0.001
2 kHz - 20 kHz
0.01 + 0.003
0.002 + 0.001
20 kHz - 200 kHz
0.01 + 0.003
0.002 + 0.001
200 kHz - 1 MHz
0.01 + 0.006
0.002 + 0.002
20 mA
10 A[3]
20 Hz - 2 kHz
0.01 + 0.003
0.002 + 0.001
2 kHz - 10 kHz
0.01 + 0.003
0.002 + 0.001
Remarks:
[1] Specifications are for 0.5 hour warm-up.
[2] Except for special marks, the AC input volta ge is 15% to 120% of range when <100 kHz a nd 40% to 120% of
range when >100 kHz. 750 V ra ng e is limite d to 750 VRMS. 200 mV ranges is for ful l sc a l e or h igher. For inputs
from 30 mV to 200 mV, multiply total % of reading error by 10.
[3] For AC input current from 15% to 120% of range except where noted. 20 mA range specifications are for full scale.
For inputs from 5 mA to 20 mA, mu lt i ply tot al % of reading er ror by 10. 10 A rang e is for AC input current from
25% to 100% of range.
Frequency and Period
Measurement Method: Reciprocal-counting technique, AC-coupled input, AC voltage or AC current measurement
function
Measure Attentions
Generally, errors are leaded into all frequency counters when measuring low voltage or low frequency signal.
Shielding input can extremely help to reduce measuring errors caused by exterior noise.
Setup Time Attentions
5
If the variational DC components appeared in signals under measure, errors may be caused while measuring period
or frequency. Please ensure that the RC loop at inp ut terminal must be stable during exact measuring (higher than 1
sec).
Capacitan ce Characteristics Accuracy ± (% of reading + % of range) [1]
Function
Range[2]
Maximum Testing
Current
1 Year
23
±5
Temperature Coefficient
0
- 18
28
- 50
Capacitance
2.000 nF
200 nA
3 + 1.0
0.08 + 0.002
20.00 nF
200 nA
1 + 0.5
0.02 + 0.001
200.0 nF
2 μA
1 + 0.5
0.02 + 0.001
2.000 μF
10 μA
1 + 0.5
0.02 + 0.001
200 μF
100 μA
1 + 0.5
0.02 + 0.001
10000 μF
1 mA
2 + 0.5
0.02 + 0.001
Remarks:
[1] Specifications are for 0.5 hour warm-up and REFoperation. Using of non-film capacitor may generate additional
errors.
[2] Specifications are for from 1% to 120% on 2 nF range and ranges from 10% to 120% on other ranges.
Capacitance Measuring
Measurement Method
Measure the rate of change of voltage generated during the current flowing the
capacitance
Connection Type
2-wire
Input Protection
1000 V on all ranges
Measure Attentions:
Small capacitance is easily influenced by external noise and thus causes errors while measuring, shielding input can
extremely help to reduce this kind of errors.
Other Me asu rin g C hara cte ri s tics
Triggering and Memory
Samples/Trigger
1 ~ 2000
Trigger Delay
8 ms ~ 2000 ms optional
External Trigger Input
Input Level
TTL compatible (High level when left input
terminal is hanging in the air)
Trigger Condition
Risin g edg e/ fallin g ed ge selectable
Input Impendence
>20 kΩ, in parallel with 400 pF, DC -coupled
Min Pulse
500 μs
VMC Output
Electric Level
TTL compatible ( input >=1 kohm loads
Output Polarity
straight polarity and negative polarity optional
Input Impedance
200 ohm, typical
Arbitrary Sensor
Suppor t for m ulti ply typ es of sen sor such a s Th ermoc ouple, DC Voltage, DC C urrent, Resis tance (2-wire or 4-wire)
and Frequency outp ut; With thermocouple compensation at cold junction.
Output P olarity: straight pol arity and negative polarity optional
Preset ITS-90 trans for m of B , E, J , K, N, R, S an d T t herm ocoup le and tr a nsf orm of pla tinum Pt100, Pt385 resistance
temperatu r e sen so r
Math Functions
Pass/Fail, REL (RELative), Min/Max/Average, dBm, dB, Hold, Historgram, Standard deviation
History Records
Volatile Memory
2000 readings of history records
Nonvolatile Memory
10 gourps of history records (2000 readings/group); 10 groups of sensor records: (1000
readings/group); 10 groups of setting records of instrument; 10 groups of setting records
of arbitrary sensor; support USB-disk external storage
6
General Specifications
Power Supply
AC 100 V ~ 120 V
45 Hz ~ 440 Hz
AC 200 V ~ 240 V
45 Hz ~ 66 Hz
Consumption
20 VA p eak value
Mechanism
Dimension
107.0 mm×231.6 mm×290.5 mm
Weight 2.5 kg
Other Characteristics
Display Screen
LCD display with 256×64 lattices, support for Double display, Menu display, Operating
help and En gl ish/ Ch i n ese bi li ngual.
Operating Environment
Full accuracy from 0
to 50
; 80% R.H. and 40
, non condensing
Storage Temperature: -20
~ 70
Shock and Vibration: conformi ng to MIL-T-28800E, III, 5 level (only for sine)
height abo ve sea le vel: up to 3000 meters
Safety
Conform i ng to IEC61010-1: 2001. Measure CAT I 1000 V/CAT II 600 V
Class of pollution: 2
Remote Interface
GPIB (only for DM3058), 10/100Mbit LAN (only for DM3058), USB2.0 Full Speed Device &
(USB-disk available), RS-232C
Programmer Lan gu age RIGOL 3058 SCPI, FLUKE45, Agilent34401A
LXI Compatibility LXI Class C, Version1.1 (only for DM3058)
Warm Up Time 30 minutes
7
Ordering Information
Name of Product
RIGOL DM3058/DM3058E Digital Multimeter
Standard Accessories
A Power Cord that fits the stand ard of
destination country
Two Test Le a ds (black an d red )
Two Alligator Clips (black and red)
An USB Data Cable
A Backup Fuse
A Quick Guide
User’s Guide and Application software
(CD-ROM)
Optional Accessories
Kelvin Test Clips
RS232 Cable
Warranty
Thank you very much for choosing RIGOL
products!
RIGOL warrants that the product mainframe
and produ ct a ccesso ries will be free from defects
in materials and workmanship within the
warran ty period.
If a prod uct pr ove s defective within the
respective period, RIGOL guarantees free
repla cement or repai r of an y defecti ve products
within a reasonable period of time. To get repair
service, please contact with your neare s t
RIGOL sales or service office.
There is no othe r wa rr ant y, expressed o r im plied,
except such as is expressly set forth herein or
other applicable warranty card. There is no
implied wa rranty of merchant ability or fitness for
a particular purpose. Under no circumstances
shall RIGOL be liable for any conse quential,
indirect, ensuing or special damages for any
breach of warranty in any case.
Contact Us
If you have any problem or requirement when using our products or this manual, please contact
RIGOL.
E-mail: service@rigol.com
Websites: www.rigol.com