INFRARED CO2 CONTROLLER DATA SHEET ZFP9 This is a non-dispersion type infrared CO2 gas analyzer. It accurately measures the concentration of CO2 gas in the air, indicates the result and simultaneously outputs an analog signal. Since upper and lower limit alarm functions are standard it can also be used as a controller. This instrument is suitable for use as a CO2 meter in green houses, building ventilation systems, CA (Controlled Atmosphere) storage facilities, and so on. FEATURES 1. 2. 3. 4. Use of high-performance infrared-ray system Excellent stability and easy maintenance Control with upper/lower limit alarm functions possible Compact and lightweight (Approx. 3kg) SPECIFICATIONS Functional specifications Measuring system: NDIR (non-dispersion infrared ray system) Measured gas: CO2 in the air Measuring range: 0 to 0.2%, 0 to 0.3%, 0 to 1%, 0 to 5%, 0 to 10%, 0 to 20% CO2 (as specified) Output signal: 4 to 20mA DC nonlinear (max. load resistance 350) or 0 to 100mV DC nonlinear (output resistance 100) Note) Linear ouput is available only for 0 to 0.2% range, 4 to 20mA DC or 0 to 100 mV DC. Alarm function: Alarm setting range; 0 to 100% FS Setting method; set value display on indicator with internal upper/lower limit setting VR's Output; relay contact (2 x 1c), 250V AC, 3A (resistive load) Alarm hysteresis width; 4% of scale length (at 50% FS) Indication; upper limit indicator lamp (red) -- lights up when CO2 concentration exceeds upper limit. lower limit indicator lamp (red) -- lights up when CO2 concentration is below the lower limit. Indicator: CO2 concentration actual scale, moving coil type, JIS 2.5 class Power supply: 100, 115, 200, 220V AC 10%, 50/60Hz (as specified) Power consumption: Approx. 18VA Ambient temperature: 0 to 40C Ambient humidity: 90% RH or less Storage temperature: -20 to 50C Sample gas temperature: 0 to 50C Gas sampling: Aspirating pump (with power ON-OFF switch), membrane filter built in, sample gas flow rate approx. 0.6r/min Performance specifications Repeatability: Zero point drift: Response time: Warmup time: Linearity: 1% FS Within 10%/6 months Within 10 seconds (for 90% response) Approx. 30 minutes 3% FS (only for 0 to 0.2% range) Structure and material Enclosure: Case material: Indoor dust-proof type Case cover (ABS resin), base plate (steel plate) Surface color: Case (Munsell 3.1Y7.0/1.1) Lid (Munsell 9.2YR4/1) Material of parts contacting gas: SUS304, corrosion-resistant aluminum Mounting: Wall mounting Outer dimensions (H x W x D): 257 x 220 x 85mm Mass {Weight}: Approx. 3kg Piping: 4.8mm dia. hose end (sample gas inlet) Wiring: M3 screw terminal EDS3-104b Date Apr. 30, 1997 ZFP9 SCOPE OF DELIVERY CODE SYMBOLS Controller and standard accessories (filter paper - 5 sheets, fuse - 2) 1 2 3 4 5 6 7 8 Z F P 9 Standard gas for zero/span calibration (type ZBM) Pressure regulator for standard gas (type ZBD610) Flowmeter for zero/span calibration (type ZBD472) OUTLINE DIAGRAM Measuring range 0 to 0.3% 0 to 1% 0 to 5% 0 to 10% 0 to 20% 0 to 0.2% A B C D E G ITEMS TO BE ORDERED SEPARATELY Output signal DC 0 to 100mV nonlinear DC 4 to 20mA nonlinear DC 0 to 100mV linear* DC 4 to 20mA linear* A B C D 1 2 4 5 (Unit:mm) Description 1 Power supply 100V AC 50/60Hz 115V AC 50/60Hz 220V AC 50/60Hz 200V AC 50/60Hz * Linear output is available only for 0 to 0.2% range. (5th code : G) 220 257 85 Rear panel diagram 7 137 23 8 16 R3.5 12mm dia. 35 Mounting hole 36 44 Sample gas inlet (for 4.8mm dia. hose end) 21.5 22.5 137 Cable inlets (22mm dia.) (with rubber bushing) 35 28 120 7mm dia. mounting hole Cable inlets (22mm dia.) (with rubber bushing) Sample gas outlet (for 4.8mm dia. hose end) (attached for more than 0 to 1% range) CONNECTION DIAGRAM 1 2 3 100V, 115V Ground 200V, 220V terminal AC 50/60Hz power supply (as specified) 4 5 6 7 8 9 NO COM NC NO COM NC 10 OUTPUT Lower limit Upper limit Output signal 0 to 100mV DC or 4 to 20mA DC Alarm output Lower limit alarm:4 - 5 become conductive when CO2 sample gas drops below set value, while 5 - 6 tive when concentration exceeds set value. Upper limit alarm:8 - 9 become conductive when CO2 sample gas drops below set value, while 7 - 8 tive when concentration exceeds set value. CAUTIONS ON WIRING Use vinyl insulated cable with a cross-sectional area of 1.25mm2 or more for the power supply and alarm contact output. Use two-core shielded cable for the output signal cable in order to avoid the influence of inductive noise. Be sure to connect shielded wires to the ground terminal. Keep the output signal cable away from the power line and alarm contact output line. The product conforms to the requirements of the Electromagnetic compatibility Directive 89/336/EEC as detailed within the technical construction file number TZ734574. The applicable standards used to demonstrate compliance are : EN 50081-1 : 1991 CLASS A Conducted and Radiated emissions EN 50082-1 : 1992 11 Radiated immunity, ESD and FBT ZFP concentration in become conducconcentration in become conduc- ZFP9 Caution on Safety *Before using this product, be sure to read its instruction manual in advance. Head office 6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan http://www.fesys.co.jp Sales Div. International Sales Dept. No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187 http: //www.fic-net.co.jp Information in this catalog is subject to change without notice. Printed in Japan