4 Fluke Corporation Fluke 3000 FC Series Test Tools. The Fluke Wireless Team.
Frequency
Range Resolution Accuracy [1]
99.99Hz 0.01Hz
0.1% + 1
999.9Hz 0.1Hz
9.999kHz 0.001kHz
99.99kHz 0.01kHz
1 Frequency is specified up to 99.99kHz in volts and up to 10kHz in amps.
Input characteristics
Function
Overload
Protection
Input Impedance
(nominal) Common Mode Rejection Ratio (1 kΩ unbalance) Normal Mode Rejection
1100 V rms > 10 MΩ < 100 pF > 120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
1100 V rms > 10 MΩ < 100 pF > 60 dB, dc to 60 Hz
1100 V rms > 10 MΩ < 100 pF > 120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
Open circuit test voltage Full scale voltage Typical short circuit current
To 6 MΩ 50 MΩ
1100 V rms < 2.7 V dc < 0.7 V dc < 0.9 V dc < 350 mA
1100 V rms < 2.7 V dc 2.000 V dc < 1.1 mA
Function Overload Protection Overload
mA Fused, 44/100A, 1000V FAST Fuse 600mA overload for 2minutes maximum, 10minutes rest minimum
MIN/MAX recording
Function Accuracy
DC functions The specified accuracy of the measurement function ± 12 counts for changes > 350 ms in duration.
AC functions The specified accuracy of the measurement function ± 40 counts for changes > 900 ms in duration.
General specifications (Not compatible with Fluke CNX test tools)
Maximum voltage between any
terminal and earth ground
1000 V dc or ac rms
Ω fuse protection from A inputs 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only
Display (LCD) Update rate: 4/sec
Votls, amps, ohms: 6000 counts
Frequency: 10,000 counts
Capacitance: 1,000 counts
Battery type Three AA Alkaline batteries, NEDA 15A IEC LR6
Battery life 250 hours minimum
RF communications 2.4 GHZ ISM Band
RF communication range Open air, unobstructed: Up to 20m
Obstructed, sheetrock wall: Up to 6.5m
Obstructed, concrete wall, or steel electrical enclosure: Up to 3.5m
Temperature Operating: -10 °C to 50 °C
Storage: -40 °C to 60 °C
Temperature coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C)
Relative humidity 0 % to 90 % (0 °C to 35 °C), 0 % to 75 % (35°C to 40°C), 0% to 45% (40°C to 50°C)
Altitude Operating: 2,000 m
Storage: 12,000 m
Electromagnetic compatibilty
EMI, RFI, EMC, RF
EN 61326-1:2006, EN 61326-2-2:2006. ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489
V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249,
FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE
Safety compliance ANSI/ISA 61010-1 / (82.02.01): 3rd edition
CAN/CSA-C22.2 No 61010-1-12: 3rd edition
UL 61010-1: 3rd edition
IEC/EN 61010-1:2010
Certifications CSA, FCC, CE
Ingress Protection (IP) rating IP54
Pollution Degree 2
Size (H x W x L) 4.75 cm x 9.3 cm x 20.7 cm (1.87 in x 3.68 in x 8.14 in)
Weight 340 g (12 oz)