Name of document: 30910036 Index C
Status: 03/2013
Platinum Resistance Temperature Detector SMD 0603 (V)
The PRTD SMD 0603 is designed for automatic mounting in large volume applications on printed circuit boards where long
time stability, interchangeability combined with low costs are important.
Specification DIN EN 60751
Tolerance Class B (R
0
: ±0.12%)
Class 2B (R
0
: ±0.24%)
Temperature range -50°C to +150°C
(Application temperatures of +150°C are only
possible with the use of expansion-matched
circuit board material; up to +130°C with circuit
board material not matched for expansion)
Temperature coefficient TCR
= 3850 ppm/K
Soldering connection End-termination galvanic tin plated with Ni- barrier layer
Long term stability max R
0
-Drift 0,06% after 250h with 130°C
Environmental conditions unhoused for dry environments only
Insulation resistance > 100 MΩ at 20°C;
> 2 MΩ at 150°C (glass cover)
Measuring current 1000Ω: 0.1 up to 0.3mA
(self heating has to be considered)
Self heating 0.8 K/mW at 0°C
Reaction time Flowing water (v= 0.4m/s): t
0.5
< 0.1s
t
0.9
< 0.25s
Air flow (v= 2m/s): t
0.5
< 2.5s
t
0.9
< 8 s
Processing instructions face up-mounting: reflow soldering or wave soldering,
e. g. double wave < 8s / 235°C
Storage life Min. 9 months (in dry environment)
Packaging „Face-up“ in blister reel, 4000 pcs
Note Other tolerances, values of resistance are available on request.
Please have a look for the Information about the tested soldering profile on the next page
Nominal
Resistance
R0
Tolerance
DIN EN 60751
1996-07
Tolerance
DIN EN 60751
2009-05
Order Number
1000 Ohm at 0°C Class 2B
Class B F 0.6
F 0.3 32 207 637
32 207 638
We reserve the right to make alterations and technical data printed. All technical data serves as a guideline and does not guarantee
particular properties to any products.
1901 Route 130
North Brunswick, NJ 08902
Phone 732-940-4400 Fax 732-940-4445
Email info.hst-us@heraeus.com
http://heraeus-sensor-technology-us.com
Heraeus Sensor Technology USA