Agilent E2307A Thermocouple Bead Temperature Probe
•Exposed welded thermocouple junction with PTFE
insulation wire, 36 inches (0.9 m) long, terminated in a
0.75 inch dual-banana plug.
•Thermocouple type: #24 gage ANSI-K type calibration.
•Accuracy: –260°C to –110°C (–436°F to –200°F) 2% of reading
–110°C to 260°C (–200°F to 500°F) ±2.2°C (±4°F)
•Maximum temperature: 260°C (500°F)
•For use with: Agilent 973A or any other device that accepts type-K
thermocouples and a dual-banana plug.
Agilent E2308A Thermistor Temperature Probe
•Stainless steel encapsulated thermistor (5000Ω @ 25°C),
36 inch (0.9 m) leads, terminated in a 0.75 inch dual-banana plug.
•Accuracy: –80°C to –40°C (–112°F to –40°F) ±1°C (2°F)
–40
°C to 0°C (–40°F to 32°F) ±0.4°C (0.8°F)
0°C to 70°C (32°F to 158°F) ±0.2°C (0.4°F)
70°C to 100°C (158°F to 212°F) ±0.3°C (0.6°F)
100
°C to 150°C (212°F to 302°F) ±1°C (2°F)
•Maximum temperature: 150°C (302°F)
•Time constant: 3 seconds typical in well-stirred water
•Isolation: 42 volts peak max
•Stability (for 10 months): 25°C <0.01°C
100°C <0.2°C
150°C <1.5°C
•Stainless Steel Probe: 1.5 in. L x 0.125 in. dia. (38 mm L x 3.2 mm dia.)
•For use with: Agilent 971A/972A/973A/974A and Agilent 3458A.
Can also be used with the Agilent 34401A with application software.
Agilent E2307A Thermocouple Bead Probe
Agilent E2308A Thermistor Probe
E2307-90002
Copyright © Agilent Technologies 2003-2012
Printed in Malaysia April 2012 E0412