Agilent 34410A and 34411A Multimeters Data Sheet See Agilent's Truevolt Series of DMMs * Display DMM results in ways you never have before * Measure with unquestioned Truevolt confidence * Move to the next generation 34401A DMM with 100% assurance www.agilent.com/find/dmm Agilent 34410A 61/2-Digit High-Performance DMM * 10,000 readings/s @ 5 1/2 -digits direct to PC * 1,000 readings/s @ 6 1/2 -digits direct to PC * 30 PPM 1-year Basic DC accuracy * LAN, USB & GPIB standard * DCV, ACV, DCI, ACI, 2-wire and 4-wire Resistance, Frequency, Period, Continuity, and Diode Test * Capacitance & Temperature measurements * Expanded measurement ranges * Data logger with 50 k reading non-volatile memory Agilent 34411A 61/2-Digit Enhanced-Performance DMM All the features of the 34410A, plus: * 50,000 readings/s @ 41/2 -digits direct to PC * 1 million volatile reading memory * Analog level triggering * Programmable Pre/Post triggering The Best Just Got Better The Agilent 34410A and 34411A 61/2-Digit DMMs represent the latest generation of multimeters from Agilent Technologies. Building on the phenomenal success of the industrystandard Agilent 34401A, these new meters offer improved accuracy, expanded measurement capability, dramatically improved measurement speed and throughput, and modern computer interfaces including LAN and USB. The dual display offers both dual measurement capabilities and ease of use when setting up and configuring the DMM. Improvements have been made in every facet of the 34401A to make the best even better, whether you use it on the bench or in a system. Dramatic Speed Improvements Whether it's raw reading speed or fast system throughput, the 34410A sets a new benchmark in performance. Using a new A/D technology, the 34410A achieves an impressive 10,000 readings a second at 51/2-digits, and can stream readings to your computer at this same speed! Triggering is fast and precise, with both trigger latency and trigger jit- ter less than 1 s, while bus query response is less than 500 s. ACV measurements are faster as well thanks to a digital measurement technique that additionally improves accuracy at high and low frequencies. For even greater reading speeds, select the 34411A, which achieves 50,000 readings a second at 41/2-digits. Enhanced Measurement Performance The 34410A and 34411A offer Temperature and Capacitance capabilities, in addition to those measurements you have come to expect, such as DCV, ACV, DCI, ACI, 2-wire and 4-wire Resistance, Frequency, Period, Continuity and Diode Test. You also get Offset Compensated Ohms, allowing you to accurately measure resistance in the presence of voltages. Measurement ranges have been expanded as well; for example, DC and AC Current Ranges now go down to 100 A, resulting in 100 pA resolution. Real-time math and statistics are included, and a peakdetect capability allows you to capture peaks as short as 20 s. Even Greater Performance with the 34411A The 34411A has all the features of the 34410A, plus additional performance that makes it even more powerful. With the ability to make 50,000 readings per second at 41/2 -digits, Analog Level Triggering, programmable Pre- and Post-Trigger and 1 million readings of volatile memory in addition to 50,000 reading of non-volatile memory, you now have the ability to capture low-frequency waveforms, characterize device performance and transfer results for analysis on your computer. Data Logger Function A front panel data logger function allows you to set the meter up to make unattended, paced measurements over a fixed time or number of events, then pull up the results later for review or transfer to a computer for analysis. Set the meter up to take measurements every 10 seconds for an hour, go have lunch, and check the results upon your return. The contextual front panel sequences make setup and read back a breeze. Improved Ease of Use From the inclusion of a second display to configuring setups for each of the measurement functions, these new DMMs offer significant improvements in usability. Simple things are simple; more complicated setups are easier than ever. There is even a new probe set designed to more easily probe today's finepitch components. Finally, there is a built-in Graphical Web Interface that allows you to interactively control the DMM without the hassle of programming at all! Modern I/O for Improved Connectivity When connecting to a computer, select LAN, USB or GPIB interfaces; all three are standard on the 34410A and 34411A. Concerned about the viability of your existing software programs? These new DMMs respond to Standard Commands for Programmable Instrumentation (SCPI), and there is even a 34401A Emulation Mode to ensure the easiest upgrade possible. Agilent's I/O Library Suite ships with the 34410A and 34411A to help you quickly establish an error-free connection between your PC and instrument. It provides robust instrument control and works with the software development environment you choose. Product Reference CD-ROM with soft documentation and software: LXI - Class C LAN Extensions for Instruments (LXI) provides the next generation I/O for system applications requiring the highest throughput. Transfer rates of over 250,000 readings/s are attainable ensuring even the most data intensive measurements are fast, without the overhead cost of an instrument cardcage. Both the 34410A and the 34411A are LXI - Class C compliant. Bench Dimensions: Built To Last Our new DMMs were designed to high standards of ruggedness and reliability. From the robust package with its shock absorbing bumpers to careful selection of components and conservative circuit design, these meters are built to last. Calculated Mean Time Between Failure (MTBF) is in excess of 100,000 hours. Backed by a 1-year warranty and a worldwide network of service centers, you can buy with confidence. * * * * * * * Optional Printed Documentation: * Quick Start Tutorial * User's Guide * Service Guide 261.2 mm 103.8 mm 303.2 mm System Dimensions: Go to the Web For the latest information on these or other Agilent DMMs, go to www.agilent.com/find/dmm Accessories Included: * Test Lead Kit with probes and SMT attachments. * Test report, power cord, and USB interface cable. Programmer's Reference Help Quick Start Tutorial User's Guide Service Guide Programming Examples IntuiLink for Multimeters LabVIEW and IVI-COM drivers 212.8 mm 88.3 mm 272.3 mm 2 Accuracy Specifications (% of reading + % of range)1 Function Range 3 Frequency, 24 Hour 2 90 Day 1 Year Temperature Test Current or Burden Voltage Tcal 1 C Tcal 5 C Tcal 5 C 0 C to (Tcal -5 C) (Tcal +5 C) to 55 C 0.0005 + 0.0005 0.0005 + 0.0001 0.0005 + 0.0001 0.0005 + 0.0001 0.0005 + 0.0001 0.010 + 0.003 0.008 + 0.003 0.005 + 0.003 0.010 + 0.005 0.020 + 0.008 0.120 + 0.020 0.0006 + 0.0005 0.0006 + 0.0001 0.0006 + 0.0001 0.0006 + 0.0001 0.0010 + 0.0002 0.0030 + 0.0004 0.1000 + 0.0001 1.0000 + 0.0001 0.0020 + 0.0030 0.0020 + 0.0005 0.0020 + 0.0020 0.0020 + 0.0005 0.0050 + 0.0010 0.0050 + 0.0020 0.015 + 0.006 0.030 + 0.006 0.005 + 0.000 0.005 + 0.000 0.001 + 0.000 0.001 + 0.000 0.05 + 0.05 0.05 + 0.01 0.01 + 0.01 0.01 + 0.01 0.01 + 0.01 Coefficient/C DC Voltage True RMS AC Voltage5 Resistance6 DC Current True RMS AC Current 7 Frequency or Period Capacitance8 100.0000 mV 1.000000 V 10.00000 V 100.0000 V 1000.000 V 4 100.0000 mV to 750.000 V 100.0000 1.000000 k 10.00000 k 100.0000 k 1.000000 M 10.00000 M 100.0000 M 1.000000 G 100.0000 A 1.000000 mA 10.00000 mA 100.0000 mA 1.000000 A 3.000000 A 100.0000 A to 3.00000 A 100 mV to 750 V 1.0000 nF 10.000 nF 100.00 nF 1.0000 F 10.000 F Temperature9 RTD Thermistor Continuity -200 C to 600 C -80 C to 150 C 1000.0 Diode Test10 1.0000 V 3 Hz - 5 Hz 5 Hz - 10 Hz 10 Hz - 20 kHz 20 kHz - 50 kHz 50 kHz - 100 kHz 100 kHz - 300 kHz 1 mA 1 mA 100 A 10 A 5 A 500 nA 500 nA || 10 M 500 nA || 10 M < 0.03 V < 0.3 V < 0.03 V < 0.3 V < 0.8 V < 2.0 V 3 Hz - 5 kHz 5 kHz - 10 kHz 3 Hz - 5 Hz 5 Hz - 10 Hz 10 Hz - 40 Hz 40 Hz - 300 kHz 500 nA 1 A 10 A 10 A 100 A 1 mA 1 mA 0.0030 + 0.0030 0.0020 + 0.0006 0.0015 + 0.0004 0.0020 + 0.0006 0.0020 + 0.0006 0.50 + 0.02 0.10 + 0.02 0.02 + 0.02 0.05 + 0.04 0.20 + 0.08 1.00 + 0.50 0.0030 + 0.0030 0.0020 + 0.0005 0.0020 + 0.0005 0.0020 + 0.0005 0.0020 + 0.0010 0.0100 + 0.0010 0.200 + 0.001 2.000 + 0.001 0.010 + 0.020 0.007 + 0.006 0.007 + 0.020 0.010 + 0.004 0.050 + 0.006 0.100 + 0.020 0.10 + 0.04 0.20 + 0.04 0.070 + 0.000 0.040 + 0.000 0.020 + 0.000 0.005 + 0.000 0.50 + 0.50 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.0040 + 0.0035 0.0030 + 0.0007 0.0020 + 0.0005 0.0035 + 0.0006 0.0035 + 0.0006 0.50 + 0.03 0.10 + 0.03 0.05 + 0.03 0.09 + 0.05 0.30 + 0.08 1.20 + 0.50 0.008 + 0.004 0.007 + 0.001 0.007 + 0.001 0.007 + 0.001 0.010 + 0.001 0.030 + 0.001 0.600 + 0.001 6.000 + 0.001 0.040 + 0.025 0.030 + 0.006 0.030 + 0.020 0.030 + 0.005 0.080 + 0.010 0.120 + 0.020 0.10 + 0.04 0.20 + 0.04 0.070 + 0.000 0.040 + 0.000 0.020 + 0.000 0.006 + 0.000 0.50 + 0.50 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.0050 + 0.0035 0.0035 + 0.0007 0.0030 + 0.0005 0.0040 + 0.0006 0.0040 + 0.0006 0.50 + 0.03 0.10 + 0.03 0.06 + 0.03 0.10 + 0.05 0.40 + 0.08 1.20 + 0.50 0.010 + 0.004 0.010 + 0.001 0.010 + 0.001 0.010 + 0.001 0.012 + 0.001 0.040 + 0.001 0.800 + 0.001 8.000 + 0.001 0.050 + 0.025 0.050 + 0.006 0.050 + 0.020 0.050 + 0.005 0.100 + 0.010 0.150 + 0.020 0.10 + 0.04 0.20 + 0.04 0.070 + 0.000 0.040 + 0.000 0.020 + 0.000 0.007 + 0.000 0.50 + 0.50 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.06 C 0.08 C 0.06 C 0.08 C 0.06 C 0.08 C 0.003 C 0.002 C 0.002 + 0.010 0.002 + 0.010 0.008 + 0.020 0.008 + 0.020 0.010 + 0.020 0.010 + 0.020 0.0010 + 0.0020 0.0010 + 0.0020 1 2 3 4 5 Specifications are for 90 minute warm-up and 100 PLC. Relative to calibration standards. 20% overrange on all ranges, except DCV 1000 V, ACV 750 V, DCI and ACI 3 A ranges. For each additional volt over 500 V add 0.02 mV of error. Specifications are for sinewave input > 0.3% of range and > 1 mVrms. Add 30 V error for frequencies below 1 kHz. 750 VAC range limited to 8 x 10 7 Volts-Hz. For each additional volt over 300 Vrms add 0.7 mVrms of error. 6 Specifications are for 4-wire resistance measurements, or 2-wire using Math Null. Without Math Null, add 0.2 additional error in 2-wire resistance measurements. 7 Specifications are for sinewave input > 1% of range and > 10 Arms. Frequencies > 5 kHz are typical for all ranges. For the 3 A range (all frequencies) add 0.05% of reading + 0.02% of range to listed specifications. 8 Specifications are for 1-hour warm-up using Math Null. Additional errors may occur for non-film capacitors. 9 For total measurement accuracy, add temperature probe error. 10 Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current is typical. Variation in the current source will create some variation in the voltage drop across a diode junction. 3 A-to-D Converter Noise Performance 1 2 3 4 5 Resolution (ppm of range)1 Normal Mode Rejection (dB)2 Readings/Second4 0.0015 0.002 5 0.006 0.02 0.06 0.2 1 2 10 100 30 15 6 3 1.5 0.7 0.3 0.2 0.1 0.03 0 0 0 0 0 0 55 110 3 110 3 110 3 50,000 25,000 10,000 3,000 1,000 300 60 (50) 30 (25) 6 (5) 0.6 (0.5) 100 RMS Noise (ppm of range) Integration Time (NPLC) 10 1 0.1 0.01 0.001 0.01 0.1 1 10 Integration Time (NPLC) Resolution is defined as the typical DCV 10 V range RMS noise. Auto-zero on for NPLC 1. See manual for additional noise characteristics. Normal mode rejection for power line frequency 0.1%. For power-line frequency 1% 75 dB and for 3% 55 dB. Maximum rate with auto-zero off for 60 Hz and (50 Hz) operation. Only available for the 34411A. System Reading and Throughput Rates System Reading Architecture DMM memory to PC (Maximum reading rate out of memory)1 Drawing - Path B Reading Format ASCII 32-bit Binary 64-bit Binary GPIB Readings/s USB 2.0 Readings/s 4,000 89,000 47,000 8,500 265,000 154,000 LAN (VXI-11) Readings/s LAN (Sockets) Readings/s 7,000 110,000 60,000 8,500 270,000 160,000 Direct I/O Measurements (Single reading - measure and I/O time)1 Drawing - Path C Function DCV/2-wire Resistance ACV/ Frequency 1 Maximum Reading Rate into Memory or to Direct I/O Resolution (NPLC) GPIB ms USB 2.0 ms LAN (VXI-11) ms LAN (Sockets) ms 0.006 (0.001) 2.6 2.9 4.6 3.2 10,000 (50,000) Fast Filter 1 ms Gate 10.0 10.0 10.0 10.0 500 (Readings/s) Drawing - Path A or C 1/2 scale input signal, immediate trigger, trigger delay 0, auto-zero off, auto-range off, no math, null off, 60 Hz line frequency. Specifications are for 34410A or (34411A). See manual for performance on other functions. System Performance Function Change (ms)1 Range Change (ms)2 LAN/GPIB Autorange (ms)3 Maximum External Trigger Rate Maximum Internal Trigger Rate4 DCV/2-wire Resistance 22 3.9/2.6 7.5 5,000/s 10,000/s (50,000/s) ACV/ Frequency 37 6.5/6.4 19 500/s 500/s 1 Time to change from 2-wire Resistance to this specified function, or DCV to 2-wire Resistance using the SCPI "FUNC" command. 2 Time to change from one range to the next higher range, 10 V, 10 M. 3 Time to automatically change one range and be ready for the new measurement, 10 V, 10 M. 4 Specifications are for 34410A or (34411A). 4 100 Measurement Characteristics DC Voltage True RMS AC Current Additional 34411A Specifications Measurement Method: Continuously integrating multi-slope IV A/D converter Measurement Method: AC-coupled True RMS measurement. Directly coupled to the fuse and shunt. Digital sampling with anti-alias filter. Resolution: See table on page 4 Linearity: (10 V range) 0.0002% of reading + 0.0001% of range Input Resistance: 0.1 V, 1 V, 10 V Ranges 100 V, 1000 V Ranges 10 M or > 10 G (Selectable) 10 M 1% (Fixed) Current Shunt: 200 for 100 A, 1 mA 2 for 10 mA, 100 mA 0.1 for 1 A, 3 A Trigger Jitter: 2 s (p-p), 20 s (p-p) when pre-triggered Input Bias Current: < 50 pA at 25 C Input Protection: 1000 V Input Protection: 3 A, 250 V fuse 140 dB1 Frequency and Period True RMS AC Voltage Measurement Method: AC-coupled True RMS measurement. Digital sampling with anti-alias filter. Crest Factor: No additional error for crest factors < 10. Limited by peak input and 300 kHz bandwidth. Peak Input: 300% of range or 1100 V Overload Ranging: Will select higher range if peak input overload is detected during auto range. Overload is reported in manual ranging. AC CMR: Triggering: Pre/Post, Int/Ext, Pos/Neg Timebase Resolution: 19.9524 s 0.01% accuracy Maximum Input: The peak value of the DC + AC current must be < 300% of range. The RMS current must be < 3 A including the DC current content. DC CMRR: Overall Bandwidth, DCV & DCI: 15 kHz typical @ 20 s aperture (-3 dB) Measurement Method: Reciprocal-counting technique. AC-coupled input using the AC voltage measurement function. Input Impedance: 1 M 2% in parallel with < 150 pF Input Protection: 750 Vrms all ranges Spurious-Free Dynamic Range & Signal to Noise Distortion Ratio Function DCV Range Spur-Free SNDR 1V -75 dB 60 dB 1 10 V -70 dB 60 dB 100 V -75 dB 60 dB 1 10 V range: 2 V (p-p) < signal < 16 V (p-p) Triggering and Memory Reading Hold Sensitivity: 1% of reading Samples per Trigger: 1 to 50,000 (34410A) 1 to 1,000,000 (34411A) Trigger Delay: 0 to 3600 s (20 s step size) External Trigger: Programmable edge, Low-power TTL compatible Delay: < 1 s Max rate: 5,000/s Jitter: < 1 s Min Pulsewidth: 1 s Capacitance Measurement Method: Current input with measurement of resulting ramp. Voltmeter Complete: 3 V Logic output, 2 s pulse with programmable edge Connection Type: 2-wire Nonvolatile Memory: 50,000 readings 70 dB 2 Maximum Input: 400 Vdc, 1100 Vpk Temperature Input Impedance: 1 M 2% in parallel with < 150 pF Thermistor: 2.2 k, 5 k, and 10 k Input Protection: 750 Vrms all ranges RTD: = 0.00385 Ro from 49 to 2.1 k Volatile Memory: 50,000 readings (34410A) 1,000,000 readings (34411A) Sample Timer: Range: 0 to 3600 s (20 s step sizes) Jitter: < 100 ns Resistance Measurement Method: Selectable 2-wire or 4-wire. Current source referenced to LO input. Offset Compensation: Selectable on the 100 , 1 k, and 10 k ranges Input Protection: 1000 V on all ranges Function3 DCV 2-wire DCI Frequency Period Filter setting ACV ACI Input Protection: 1 2 Digits 4.5 50 k4 50 k4 50 k4 500 500 fast 500 500 5.5 10 k 10 k 10 k 90 90 med 150 150 3 A, 250 V fuse For 1 k unbalanced in LO lead and < 60 Hz, 500 V peak maximum Maximum rate for DCV, DCI, and resistance functions (using zero settling delay, autozero off, manual range) 4 34411A only Power Consumption: 25 VA peak (16 W average) Operating Environment: Full accuracy for 0 C to 55 C, 80% R.H. at 40 C non-condensing 6.5 1k 1k 1k 10 10 slow 50 50 Storage Temperature: -40 C to 70 C Weight: 3.72 kg (8.2 lbs) Safety: IEC 61010-1, EN 61010-1, UL 61010-1, CAN/CSA-C22.2 No. 61010-1, Refer to Declarations of Conformity for current revisions. Measurement CAT II 300 V, CAT I 1000 V. Pollution Degree 2 EMC: IEC 61326, EN 61326, CISPR 11, ICES-001, AS/NZS 2064.1, Refer to Declaration of Conformity for current revisions. Vibration & Shock: MIL-T-28800E, Type III, Class 5 (Sine Only) For 1 k unbalanced in LO lead, 500 V peak maximum 3 Power Line Frequency: 45 Hz to 66 Hz and 360 Hz to 440 Hz, Automatically sensed at power-on Continuity Threshold: Fixed at 10 Operating Characteristics Maximum readings/second Current Shunt: 200 for 100 A, 1 mA 2 for 10 mA, 100 mA 0.1 for 1 A, 3 A Power Supply: 100 V/120 V/220 V/240 V 10% Response Time: 300 samples/sec with audible tone Max. Lead Resistance (4-wire): 10% of range per lead for 100 , 1 k. 1 k per lead on all other ranges DC Current General Specifications Continuity/Diode Test LXI Compliance: LXI Class C, ver. 1.0 5 Warranty: 1 year Ordering Information Agilent 34410A and 34411A Multimeters Accessories included Test lead kit with probes and SMT attachments, calibration certificate, power cord, and USB interface cable. Product Reference CD-ROM with soft documentation and software: * Programmer's Reference Help * Quick Start Tutorial * User's Guide * Service Guide * Programming Examples * IntuiLink for Multimeters * LabVIEW and IVI-COM drivers Options Opt. A6J ANSI Z540 compliant calibration Probes/Leads/Clip Accessories 11059A Kelvin probe set 11060A Surface mount device (SMD) test probes 11062A Kelvin clip set 34133A Precision electronic test leads 34134A DC coupled current probe 34136A High voltage probe 34138A Test lead set 34171B Input terminal connector (sold in pairs) 34172B Input calibration short (sold in pairs) 34308A Thermistor kit 34330A 30 A current shunt E2308A 5 k thermistor probe Y1133A Low-thermal external digital multimeter scanning kit Rack Mount Kits 34190A Rackmount kit: designed for use with only one instrument, mounted on either the left or the right side of the rack. 34191A 2U dual flange kit: secures the instrument to the front of the rack. This kit can be used with the 34194A dual lock link kit to mount two half-width, 2U height instruments side-by side. 34194A Dual lock link kit: recommended for side-by-side combinations and includes links for instruments of different depths. This kit can be used with the 34191A 2U dual flange kit to mount two half-width, 2U height instruments side-by-side. Other Accessories 34131A Hard transit case 34162A Accessory pouch E5810A LAN/GPIB gateway 6 www.agilent.com www.agilent.com/find/DMM myAgilent www.agilent.com/find/myagilent A personalized view into the information most relevant to you. www.axiestandard.org AdvancedTCA(R) Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Agilent is a founding member of the AXIe consortium. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium. www.pxisa.org PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. Agilent Channel Partners www.agilent.com/find/channelpartners Get the best of both worlds: Agilent's measurement expertise and product breadth, combined with channel partner convenience. For more information on Agilent Technologies' products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/ThreeYearWarranty www.agilent.com/find/contactus Agilent's combination of product reliability and three-year warranty coverage is another Americas way we help you achieve your business goals: Canada (877) 894 4414 increased confidence in uptime, reduced cost Brazil (11) 4197 3600 of ownership and greater convenience. Three-Year Warranty Agilent Advantage Services www.agilent.com/find/AdvantageServices Accurate measurements throughout the life of your instruments. www.agilent.com/quality Mexico United States 01800 5064 800 (800) 829 4444 Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Other AP Countries 1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 (65) 375 8100 Europe & Middle East Belgium 32 (0) 2 404 93 40 Denmark 45 45 80 12 15 Finland 358 (0) 10 855 2100 France 0825 010 700* *0.125 /minute 49 (0) 7031 464 6333 Germany Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 United Kingdom 44 (0) 118 927 6201 For other unlisted countries: www.agilent.com/find/contactus (BP-3-1-13) Product specifications and descriptions in this document subject to change without notice. (c) Agilent Technologies, Inc. 2013 Published in USA, April 2, 2013 5989-3738EN