1587/1577 Insulation Multimeters Fluke Corporation 5
Model 1587
Output Voltage Display Range Resolution Test Current Resistance Accuracy
50 V 0.01 to 6.00 MΩ0.01 MΩ1 mA @ 50 kΩ± (3 % + 5 counts)
(0 % to + 20 %) 6.0 to 50.0 MΩ0.1 MΩ
100 V 0.01 to 6.00 MΩ0.01 MΩ
(0 % to + 20 %) 6.0 to 60.0 MΩ0.1 MΩ1 mA @ 100 kΩ± (3 % + 5 counts)
60 to 100 MΩ1 MΩ
250 V 0.1 to 60.0 MΩ0.1 MΩ1 mA @ 250 kΩ± (1.5 % + 5 counts)
(0 % to + 20 %) 60 to 250 MΩ1 MΩ
500 V 0.1 to 60.0 MΩ0.1 MΩ1 mA @ 500 kΩ± (1.5 % + 5 counts)
(0 % to + 20 %) 60 to 500 MΩ1 MΩ
1000 V 0.1 to 60.0 MΩ0.1 MΩ± (1.5 % + 5 counts)
(0 % to + 20 %) 60 to 600 MΩ1 MΩ1 mA @ 1 MΩ
0.6 to 2.0 GΩ100 MΩ± (10 % + 3 counts)
Model 1577
Output Voltage Display Range Resolution Test Current Resistance Accuracy
500 V 0.1 to 60.0 MΩ0.1 MΩ1 mA @ 500 kΩ± (2.0 % + 5 counts)
(0 % to + 20 %) 60 to 500 MΩ1 MΩ
1000 V 0.1 to 60.0 MΩ0.1 MΩ1 mA @ 1 MΩ±(2.0 % + 5 counts)
(0 % to + 20 %) 60 to 600 MΩ1 MΩ
Maximum voltage applied to any terminal: 1000 V ac rms
or dc
Storage temperature: -40 °C to 60 °C (-40 °F to 140 °F)
Operating temperature: -20 °C to 55 °C (-4 °F to 131 °F)
Temperature coefficient: 0.05 x (specified accuracy) per °C
for temperatures < 18 °C or > 28 °C (< 64 °F or > 82 °F)
Relative humidity, non-condensing: < °C
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Vibration: Random, 2 g, 5-500 Hz per MIL-PRF-28800F,
Class 2 instrument
Shock: 1 meter drop per IEC 61010-1 2nd Edition (1 meter
drop test, six sides, oak floor)
Electromagnetic compatibility:In an RF field of 3 V/M,
accuracy = specified accuracy except in temperature:
accuracy = specified accuracy ± 5 °C (9 °F). (EN 61326-1:1997)
Safety: Complies with ANSI/ISA 82.02.01 (61010-1) 2004,
CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd
Edition for measurement CAT III 1000 V and CAT IV 600 V
Certifications: CSA per standard CSA/CAN C22.2
No. 61010.1-04; TUV per standard EN 61010 Part 1-1002
Batteries: Four AA batteries (NEDA 15A or IEC LR6)
Battery life: Meter use 1000 hours; Insulation test use: Meter
can perform at least 1000 insulation tests with fresh alkaline
batteries at room temperature. These are standard tests of
1000 V into 1 MΩwith a duty cycle of 5 seconds on and 25
seconds off
Size: 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in
W x 8.00 in L)
Weight: 550 g (1.2 lb)
IP rating: IP40
Altitude (operating): 2000 m CAT III 1000 V, CAT IV 600 V;
3000 m CAT II 1000 V, CAT III 600 V
Storage: 12,000 m
Over-range capability: 1
10 % of range except for
capacitance which is 1 %
Compliance to EN 61557: IEC61557-1, IEC61557-2
1587/1577 General Specifications