The UltraStable 600 from
Schaevitz Sensors, a division
of Measurement Specialties,
incorporates an ultra-rugged IP
67 rated housing. All wetted parts
are constructed of 316L stainless
steel, providing media isolation and
no internal “O” rings.
The UltraStable 600 also offers a wide
selection of standard pressure ranges
and electrical outputs. The unit's high
stability rating is provided by IC Sensors'
solid state UltraStable technology, which
provides stability over a wide
temperature range. The modular design
is adaptable to an infinite number of port
variations, and standard outputs include
.5V to 4.5V ratiometric, 0V to 5V, 1V to
5V, 0V to 10V, regulated and 4 to 20mA
current loop.
The UltraStable's small size vs. high
performance enables the miniaturization
of high accuracy pressure and vacuum
systems. The unit's new low-pressure
ranges add sensitivity that has
previously been unavailable in
transducers of this size. The competitive
price, coupled with its high accuracy,
may make the UltraStable 600 the best
choice for your application.
Field replacement can easily be
performed without sacrificing accuracy
due to the very small total error band of
.75% max.
A Division of
Applications
Refrigeration and HVAC controls
Process control
Marine systems
Tank level
Industrial machinery
Calibration equipment
High Stability Stainless
Steel Pressure Transducer
0.75% TEB (Total Error Band) with .5% TEB
available as an option
0.1% Accuracy - typically
-50 to 125º C Operating Temperature
100% stainless steel isolation. Small size
Wide variety of pressure ranges and
electrical outputs
Broad assortment of standard connectors
and ports
UltraStable sensing technology
CE versions available
Custom verisons available
Ranges starting from 15 psi FS
High field Interchangeability
Features
www.msiusa.com
UltraStable 600
UltraStable 600
UltraStable 600
04/02
0.75%
TEB
www.msiusa.com
A Division of
Schaevitz®Sensors
Division of Measurement Specialties
1000 Lucas Way
Hampton, VA 22666
1-800-745-8008
Phone: 757-766-1500
UltraStable 600
UltraStable 600
UltraStable 600
02/02
Units Amplified Ratiometric Amplified Amplified Notes
0.5 - 4.5 V out 0-5 V, 1-5 V, 0-10 V out 4 to 20 mA , two wire
Min. Typ. Max. Min. Typ. Max. Min. Typ. Max.
Supply voltage V 4.75 5 5.25 8 30 10 30 1
Supply Current mA 2.5 3 3.5 2.5 3 3.5 20
Min load impedance 100 K ohm 100 K ohm
Max load impedance RL=50* (Vsupply-10)
Accuracy %FS -0.1 +0.1 -0.1 +0.1 -0.1 +0.1 3, 4
Offset %FS -0.5 +0.5 -0.5 +0.5 -0.5 +0.5 2
Span %FS -0.5 +0.5 -0.5 +0.5 -0.5 +0.5 2
Long Term
stability %FS/year 0.1 0.1 0.1
Response time ms 1.0 1.0 1.0
Isolation
Resistance M ohm 50 50 50 5
Static Performance Specifications
Notes:
1) Supply voltage must be 12 V for units with 10 V out.
2) Room temperature calibration.
3) Combined BFSL, hysteresis and repeatability (Per ISA S37.2) .25% for 1Kpsi and above
4) Over temp range -20 to +85º C
5) @ 50 V DC
Environmental Performance
Specification Min. Typ. Max. Notes
Total error band -20 to +85º C,
-7 to 185º F -0.75 %FS 0.50 %FS +0.75 %FS 1, 3, 4
Operating temperature range -50ºC 125ºC
-58ºF 257º F 2,3
Storage temperature range -50ºC 125ºC
-58ºF 257º F 2
Notes:
1) With reference to 25 º C, (77 º F)
2) Standard 105º C max. for cable version 3) Lower ranges available - consult factory
4) 0.5% max availability - consult factory
www.msiusa.com
A Division of
Schaevitz®Sensors
Division of Measurement Specialties
1000 Lucas Way
Hampton, VA 22666
1-800-745-8008
Phone: 757-766-1500
Specification Value Notes
Proof pressure 3x rated pressure
Burst pressure 4x rated pressure or 20k psi 2
Vibration 20 G’s rms @2kHz
Shock 100G 11ms
Pressure cycles 1 million cycles 0 to full scale
Media compatibility All materials compatible with 316 stainless steel 1
Environmental protection IP 67 (Cable version)
Ordering Information
1)Other materials available - consult factory 2) Which ever is less, 10X burst available on
some pressure ranges (consult factory)
Output Connection Specials Port Pressure Range Pressure Units
3=.5 to 4.5V ratio (1)
4= 1 to 5V
6= 0 to 5V
7= 0 to 10V
8=4-20 mA
X=Special output
1) For ratiometric output, span changes with input voltage
(see supply voltage specifications)
2= 1/4-19 BSP
4= 7/16-20 UNF
5=1/4-18 NPT
6= 1/8-27 NPT
X= special port
PSI Ranges
015P = 0 to 15 psi
030P = 0 to 30 psi
050P = 0 to 50psi
100P =0 to 100psi
300P = 0 to 300psi
500P - 0 to 500psi
01KP = 0 to 1Kpsi
03KP = 0 to 3Kpsi
05KP = 0 to 5Kpsi
Bar Ranges
001B = 0 to 1 Bar
002B = 0 to 2 Bar
005B = 0 to 5 Bar
007B = 0 to 7 Bar
010B = 0 to 10 Bar
020B = 0 to 20 Bar
035B = 0 to 35 Bar
070B = 0 to 70 Bar
200B = 0 to 200 Bar
350B = 0 to 350 Bar
A = absolute
G = gauge
Example:
US664-000005-100PG is an UltraStable 600 with 0 to 5V out, Packard
Connector, nothing special, 1/4 NPT pressure port and 100 psi gage.
US681-C00002- 005PG is an UltraStable 600 with 4-20 mA, 2-wire, output,
2 feet of cable, CE certified, 1/4 BSP male pressure port and 5 psi gage.
Electrical
US6 Output Connection Specials Port type Pressure range Pressure unit
US6 X X - XXXXX X - XXXX X
1=Cable 2 feet
4 = Integral Packard
5 = PTIH-10-6P
(Bendix style)
6 = GSSR 300 (9.4
mm) (Square
Hirshman)
8 = PTIH-08-4P
(Bendix Style)
x = Special
Connection
For Non-Standard
the standard num-
ber of “00000”
changes to a higher
value .i.e 1XXXX
Special numbers
starting with a “C”
are CE rated
Mechanical Specifications
UltraStable 600
UltraStable 600
UltraStable 600
02/02
www.msiusa.com
A Division of
Schaevitz®Sensors
Division of Measurement Specialties
1000 Lucas Way
Hampton, VA 22666
1-800-745-8008
Phone: 757-766-1500
UltraStable 600
UltraStable 600
UltraStable 600
02/02
6-pin 4-pin
fl .875
fl .875
Note: Mating connector is available with 3 ft of cable PN (2001140).
fl .875 www.powerandsignal.com