LOW TO MEDIUM POWER AC/DC POWER SUPPLIES 80-110W AC/DC Universal Input Switch Mode Power Supplies 1
2 YEAR WARRANTY
NFS110 Medical Series
Single and quad output
7.0 x 4.25 x 1.8 inch package
Medical, dental and laboratory applications
Overvoltage and short circuit protection
110W with 20CFM
UL, VDE and CSA safety approvals
EN60601-1 and UL2601 medical approvals
The NFS110 medical series is a 80 to 110W universal input AC/DC power supply on a 7
x 4.25 inch card with a maximum component height of 1.8 inches for use in medical
applications. The NFS110 medical series has the same generic feature set as the
standard NFS110 series but have been designed with lower safety ground leakage and
higher isolation as required for medical safety approval. The NFS110 provides 80W of
output power with free air convection cooling which can be boosted to 110W with
20CFM of air. Standard features include overvoltage and short-circuit protection. The
series, with full medical safety approval to EN60601 and UL2601, meets conducted
emissions EN55022 level A. The NFS110 medical series is designed for use in low
power medical, dental and laboratory applications such as dialysis machines,
monitoring equipment, instrumentation and infusion pump controls.
All specifications are typical at nominal input, full load at 25°C unless otherwise stated
GENERAL SPECIFICATIONS
Hold-up time 110VAC @ 80W 35ms
110VAC @ 110W 17ms
230VAC @ 80W 140ms
230VAC @ 110W 100ms
Efficiency Multiple outputs 70% typ.
+5.1V single 70% typ.
12V and 15V singles 72% typ.
24V single 75% typ.
Isolation voltage Input/output 4000VAC
Input/chassis 1500VAC
Switching frequency At 100 Watts output 20 to 70kHz
At zero load 100 to 250kHz
Approvals and VDE0750, IEC60601
standards IEC1010, UL2601
(See Note 12) CSA C22.2 No. 125
Weight Singles 550g (19.4oz)
Multiple outputs 600g (21.2oz)
MTBF MIL-HDBK-217E 125,000 hours
ENVIRONMENTAL SPECIFICATIONS
Thermal performance Operating, see curve 0°C to +70°C
Non-operating –40°C to +85°C
0°C to +50°C, 80W
amb. convection cooled
+50°C to +70°C, Derate 2W/°C
amb. convection cooled
0°C to +50°C, 110W
20CFM forced air
+50°C to +70°C, Derate
20CFM forced air 2.75W/°C
Peak, 0°C to +50°C, 110W
max. 60 seconds
Relative humidity Non-condensing 5% to 95% RH
Altitude Operating 10,000 feet max.
Non-operating 40,000 feet max.
Vibration (See Note 11) 5Hz to 500Hz 2.4G approx.
OUTPUT SPECIFICATIONS
Voltage adjustability +5.1V output on multi’s ±3.0%
5.1V single ±3.0%
12V single 12V to 14V
15V single 15V to 18V
24V single 24V to 30V
Line regulation LL to HL, FL ±0.1% max.
All outputs on all units
Overshoot/undershoot At turn-on no load 0%
Temperature coefficient All outputs ±0.02%/°C
Overvoltage protection Multi output 5.1V only 6.25V±0.75V
5.1V single 6.25V±0.75V
12V single 15.75V±1.0V
15V single 22V±1.5V
24V single 33V±2.5V
Output power limit Primary power Pin max. 160W
limited Pout min. 110W
Short circuit protection Burst mode operation
INPUT SPECIFICATIONS
Input voltage range 90 to 253VAC
127 to 357VDC
Input frequency range 47Hz to 440Hz
Input surge current 110VAC, 50Hz 17A
230VAC, 50Hz 35A
Safety ground 132VAC 50µA
leakage current 264VAC 100µA
EMC CHARACTERISTICS
Conducted emissions EN55022, FCC part 15 Level A
Radiated emissions EN55022, FCC part 15 Level A
ESD air EN61000-4-2, level 3 Perf. criteria 1
ESD contact EN61000-4-2, level 4 Perf. criteria 1
Surge EN61000-4-5, level 3 Perf. criteria 1
Fast transients EN61000-4-4, level 3 Perf. criteria 1
Radiated immunity EN61000-4-3, level 3 Perf. criteria 2
Conducted immunity EN61000-4-6, level 3 Perf. criteria 2
SPECIFICATIONS
File Name: nfs110m.pdf Rev: 29 Aug 2005
For the most current data and application support visit www.artesyn.com/powergroup/products.htm
Notes
1Convection cooled, 80W maximum.
2Peak outputs lasting less than 60 seconds with duty cycle less than 10%.
Total peak power must not exceed 110W.
3Forced air, 20CFM at 1 atmosphere, 110W maximum.
4Figure is peak-to-peak. Output ripple is measured across a 50MHz
bandwidth using a 12 inch twisted pair terminated with a 47µF capacitor.
5Total regulation is defined at the static output regulation at 25°C, including
initial tolerance, line voltage within stated limits and output voltages
adjusted to their factory settings. Also for NFS110-7902P, for 24V output
stated regulation IA/ IB5. This output will maintain ±5.0% regulation if
IA5A, where IA= +5.1V output current and IB= +24V output current.
6Single output models have floating outputs which may be referenced as
either positive or negative. Higher voltage supplies, may be adjusted over
a wide output voltage range, as long as the total output power does not
exceed 80 Watts (natural convection) or 110 Watts (forced air).
7Power fail detect not available on single output models.
8Derating curve is application specific for ambient temperatures > 50°C, for
optimum reliability no part of the heatsink should exceed 90oC and no
semiconductor case temperature should exceed 100oC.
9Caution: Allow a minimum of 1 second after disconnecting the power when
making thermal measurements.
10 The user should read the PSU installation instructions in conjunction with
the relevant national safety regulations in order to ensure compliance.
11 Three orthogonal axes, random vibration, 10 minute test for each axis.
12 This product is only for inclusion by professional installers within other
equipment and must not be operated as a stand alone product.
13 NOTICE: Some models do not support all options. Please contact your
local Artesyn representative or use the on-line model number search tool at
http://www.artesyn.com/powergroup/products.htm to find a suitable
alternative.
NFS110 Medical Series
Single and quad output
LOW TO MEDIUM POWER AC/DC POWER SUPPLIES 80-110W AC/DC Universal Input Switch Mode Power Supplies 2
File Name: nfs110m.pdf Rev: 29 Aug 2005
OUTPUT OUTPUT CURRENTS RIPPLE (4) TOTAL MODEL NUMBERS (13)
VOLTAGE MAX (1) PEAK (2) FAN (3) REGULATION (5)
+5.1V 8.0A 20.0A 10.0A 50mV ±2.0% NFS110-7901P
+12.0V 4.5A 9.0A 5.0A 120mV ±3.0%
–12.0V 0.5A 1.5A 1.0A 120mV ±3.0%
–5.0V 0.5A 1.5A 1.0A 50mV ±3.0%
+5.1V (VA)8.0A 20.0A 10.0A 50mV ±2.0% NFS110-7902P
+24.0V (VB)3.5A 4.5A 4.5A 240mV +10/–5.0%
+12.0V 4.5A 9.0A 5.0A 120mV ±3.0%
–12V 0.5A 1.5A 1.0A 120mV ±3.0%
+5.1V 8.0A 20.0A 10.0A 50mV ±2.0% NFS110-7904P
+15.0V 4.0A 7.5A 5.0A 150mV ±3.0%
-15V 0.5A 1.5A 1.0A 150mV ±3.0%
-5V 0.5A 1.5A 1.0A 50mV ±3.0%
+5.1V (6) 16.0A 22.0A 20.0A 50mV ±2.0% NFS110-7905 (6,7)
12V (6) 7.0A 9.0A 9.0A 120mV ±2.0% NFS110-7912 (6,7)
15V (6) 5.0A 7.3A 7.3A 150mV ±2.0% NFS110-7915 (6,7)
24V (6) 3.5A 4.5A 4.5A 240mV ±2.0% NFS110-7924 (6,7)
TRANSIENT RESPONSE
NFS110-7901P +5.1V (7.5A to 10A) 150mV peak,
1ms recovery
+12V (2.5A to 5A) 100mV peak,
0.5ms recovery
-12V (0.5A to 1A) 100mV peak,
0.5ms recovery
-5V (0.5A to 1A) 100mV peak,
0.5ms recovery
NFS110-7902P +5.1V (7.5A to 10A) 150mV peak,
1ms recovery
+12V (2.5A to 5A) 100mV peak,
0.5ms recovery
-12V (0.5A to 1A) 100mV peak,
0.5ms recovery
24V (1.5A to 3A) 300mV peak,
1ms recovery
NFS110-7904P +5.1V (7.5A to 10A) 150mV peak,
1ms recovery
+15V (2.5A to 5A) 100mV peak,
0.5ms recovery
-15V (0.5A to 1A) 100mV peak,
0.5ms recovery
-5V (0.5A to 1A) 100mV peak,
0.5ms recovery
NFS110-7905 +5.1V (10A to 20A) 250mV peak,
1ms recovery
NFS110-7912 +12V (4.5A to 9A) 360mV peak,
1ms recovery
NFS110-7915 +15V (3.65A to 7.3A) 450mV peak,
1ms recovery
NFS110-7924 +24V (2.25A to 4.5A) 720mV peak,
1ms recovery
AC (J1) mating connector
Molex 09-50-3051 or Molex 09-91-0500 mating connector with 2478 or
equivalent crimp terminals.
DC (J2) mating connector
Molex 09-50-3131 or Molex 09-91-1300 mating connector with 2478 or
equivalent crimp terminals.
ALL DIMENSIONS IN INCHES (mm)
C
Optional
Power Fail
Detect Circuit
(Key = Pin 2)
Pin 1
6-32 UNC
(4PL)
PIN 1
1.00
±0.005
(25.4)
Voltage Adjust
J2
J1
F1
L1
C6
PIN 1
0.156
DIA. HOLE
(4PL)
5A, 250 VAC
0.475
±0.010
(12.07)
1.55 ±0.005
(39.37)
3.35 ±0.015 (85.09)
1.80 (45.72)
MAX
0.050
±0.020
(1.27)
4.250
±0.020
(107.95)
3.750
±0.005
(95.25)
0.250
±0.010
(6.35)
0.250
±0.010
(6.35)
6.50 ±0.005 (165.1)
7.00 ±0.020 (177.8)
T1
Power fail detect signal (See Note 7)
50ms ≤ T1 ≤ 200ms
T2 will vary with line and load
T3 ≥ 3ms
Pout: 110W
PFD output is an open collector
which will sink ≤ 40mA in the low state
HIGH
T1 T2
T3
4.75V
3.5V MIN
0-0.4 MAX
LOW
4.75V
5V OUTPUT
PFD SIGNAL
AC INPUT
OPTIONAL POWER FAIL DETECT
TIMING DIAGRAM
110W
80W
20 CFM FORCED AIR COOLING
40W
55W
0W
11W 11W MINIMUM LOAD REQUIRED AT 230VAC
NATURAL
CONVECTION
COOLING
0°C 10°C 20°C 30°C 40°C 50°C 60°C 70°C
DERATING CURVE (See Notes 8, 9)
Output Power (Watts)
N/C = no connection.
PIN CONNECTIONS
J1 –7901P –7902P –7904P SINGLES
Pin 1 AC Ground AC Ground AC Ground AC Ground
Pin 2 AC Neutral AC Neutral AC Neutral AC Neutral
Pin 3 AC Line AC Line AC Line AC Line
J2
Pin 1 +5.1V +5.1V +5.1V Vout
Pin 2 +5.1V +5.1V +5.1V Vout
Pin 3 +5.1V +5.1V +5.1V Vout
Pin 4 Return Return Return Return
Pin 5 Return Return Return Return
Pin 6 Return Return Return Return
Pin 7 Return Return Return Return
Pin 8 +12V +12V +15V Vout
Pin 9 +12V +12V +15V Vout
Pin 10 PFD PFD PFD N/C
Pin 11 –12V –12V –15V N/C
Pin 12 Removed for Key
Pin 13 –5V +24V –5V N/C
Datasheet © Artesyn Technologies
®
2005
The information and specifications contained in this datasheet are believed to be correct at time of publication. However, Artesyn Technologies accepts no responsibility for consequences arising
from printing errors or inaccuracies. Specifications are subject to change without notice. No rights under any patent accompany the sale of any such product(s) or information contained herein.
For the most current data and application support visit www.artesyn.com/powergroup/products.htm
NFS110 Medical Series
Single and quad output
LOW TO MEDIUM POWER AC/DC POWER SUPPLIES 80-110W AC/DC Universal Input Switch Mode Power Supplies 3
www.artesyn.com
File Name: nfs110m.pdf Rev: 29 Aug 2005
Mechanical Notes
AMetallic or non-metallic stand-offs (maximum diameter 5.4mm) can be
used in all four mounting holes without affecting safety approval.
BThe ground pad of the mounting hole near J1 allows system grounding
through a metal stand-off to the system chassis.
CThe heatsink is grounded, and allows system grounding by mechanical
connection to the system chassis.
DThe supply must be mechanically supported using the PCB mounting holes
and may be additionally supported by the heatsink mounting holes.
EIt is always advisable to attach the power supply heatsink to another
thermal dissipator (such as a chassis or finned heatsink etc). The resulting
decrease in heat sink mounted component temperatures will improve
power supply lifetime.
FA standard L-bracket and cover is available for mounting which contains all
screws connectors and necessary mounting hardware. Two different kits
are available, order part number ‘NFS110 COVER KIT’ or ‘NFS110C’.
International Safety Standard Approvals
VDE0750/EN60601-1/IEC601/IEC1010 File No. 10401-3336-1049
licence No. 2874
UL2601 File No. E147937
CSA C22.2 No. 125 File No. LR41062C