1 Pt100 KN 1510
The KN Series Ceramic Wire Wound PRTDs are suitable for
general applications requiring temperature stability.
Applications: Industrial resistance thermometers, especially
in chemical, power generation plants and analytical
equipment.
Construction: A platinum coil is sealed inside a high purity
aluminum oxide ceramic body. Lead wires are shear force
resistant and assure proper connection to extension leads
and cables.
Models
1Pt100 KN 1510
32.206.913
32.206.914
32.206.915
W0.3
W0.15
W0.1
0.14
15+2
- 0 10.0±0.5
1.0±0.15 0.20±0.01
Temperature stability:
Excellent long-term stability
Tolerance class:
Also available:
Temperature range:
Temperature
coefficient: Tc = 3850 ppm/K
Leads: Palladium-gold alloy
Measuring current: 1 mA
Insulation resistance
after assembly: > 100 MOhm @ 25 °C
Nominal resistance:
100 Ohm @ 0 °C
W0.3 (Class B) = -196 to +660 °C
W0.15 (Class A) = -196 to +600 °C
)
W0.1 (Class 1/3 B) = -100 to +350 °C
(Heraeus exceeds IEC 60751: -100 to +450 °C
- According to IEC 60751:2008
- Other standards and narrower
tolerances are available on request
- Platinum-gold alloy
- Different temperature coefficients
(3916 ppm/K - old JIS)
- Extension leads
- Two separated coils can be embedded
in one ceramic body
The measuring point is located at 8 mm from the end of the sensor body
Models
Technical Specification
0.2
0.3 9.0
3.0
Status 01/2009. Technical modifications reserved.
Self Heating
0°C (K/mW) Water current
V=0.4m/s
Air stream
V=3m/s
t
0.5 t
0.9 t 0.9
t
0.5
Description Tolerance
IEC 60751 Order No. Dimensions
mm
L D d l
Response time
Status: HSTUSA 10/09
1901 Route 130
North Brunswick, NJ 08902
Phone 732-940-4400 Fax 732-940-4445
Email info.hst-us@heraeus.com
www.hst-us.com
Heraeus Sensor Technology USA