316 SS Pressure Sensor Model 154N UltraStable
ISO – Standard
June 2001
PERFORMANCE SPECIFICATIONS
All parameters measured at 1.5mA drive and at 25°C a fter 10 second warm up, unless otherwise specified. Unless specificall y indicated, only those parameters
indicated as te sted are verified on ea ch part. Par ameters are specified for the compensated version only.
Notes
1. For amplified output circuits, 3.012V ± 1% interchangeability with gain
set resistor. See Application Note TN-003.
2. Measured at vacuum for absolute (A), ambient for gage (G).
3. Best fit straight line.
4. Over Temperature range: -20°C to +85°C Span with respect to +25°C.
5. 15 psi range sensors have a temperature error of ±0.75% (Max) of
Zero from -20°C to 85°C.
6. Guarantees output/input ratiometricity.
7. Load resistance to reduce measurement errors due to output loading.
8. Between case and sensing element.
9. Maximum temperature range for product with standard cable and
connector is –20°C to +105°C.
10. Weight depends upon configuration (cable, connector).
11. Standard gauge units are not recommended for vacuum applications.
For vacuum applications, contact factory.
12. For pressure ranges < 15 psi, consult factory on performance
specifications.
ORDERING INFORMATION
CONNECTIONS
APPLICATION SCHEMATIC
PARAMETERS MIN TYP MAX UNITS NOTES
Full Scale Output Span 75 100 150 mV 1, 2
Zero Offset -1 0 1 mV 2
Pressure Non-linearity -0.10 0.10 % Span 3
Pressure Hysteresis -0.05 0.02 0.05 % Span
Input Resistance 2,000 3,500 4,500 Ω
Output Resistance 4000 25000 Ω
Temperature Error - Span -0.75 0.75 % Span 4
Temperature Error - Zero -0.5 0.5 % Span 4, 5
Thermal Hysteresis - Span 0.05 % Span 4
Thermal Hysteresis - Zero 0.05 % Span 4
Long Term Stability - Offset 0.1 ±% Span/yr.
Long Term Stability - Span 0.1 ±% Span/yr.
Supply Current 0.5 1.5 2.0 mA 6
Output Load Resistance 5 MΩ7
Insulation Resistance (50 VDC) 50 MΩ8
Pressure Overload 3X Rated
Compensated Operating Temperature –20°C to +85°C 9
Operating Temperature Range –40°C to +125°C 9
Storage Temperature Range –50°C to +125°C 9
Media - Pressure Port Liquids and Gases compatible with 316L Stainless Steel
Media - Reference Port Compatible with Silicon, Pyrex, Gold, Fluorosilicone Rubber and
316L Stainless Steel
Weight 24 g 10
PAD NO. / CABLE FUNCTION
1+OUT
2-EX
3+EX
4-OUT
5GAIN
6GAIN
154N
(Note 12)
(Note 11)
LEADER Electronic Co., Ltd - 116, Jianghuali, Jianghua Road, Jiangmen City, GD, PRC
Tel: + 86 750 3101711, 3379183 Fax: + 86 750 3388669 Website: www.leadersensors.com - e-mail: leader@leadersensors.com