T3DMM5-5 Data Sheet
5.5 Digit Digital Multimeter
Broad Measurement Range
DC: 1000 Volts
AC: 750 Volts
Current: 10A
Tools for Improved Debugging
Wide range of measurements – DC/AC voltage and
Current, Resistance, Capacitance, Frequency, Period,
Temperature, and more.
 More application coverage from a single Digital
multimeter.
True-RMS measurements – All AC Voltage and
Current ranges give True-RMS readings.
 Excellent accuracy regardless of the waveform shape.
Low level measurement, high sensitivity ranges
Voltage ranges as low as 200 mV full scale, DC Current
200 µA, AC Current 20 mA full scale.
High sensitivity ranges give greater accuracy of
small signal measurements.
Advanced measurement features – Min, Max,
Average, Standard Deviation dBm/dB, Pass/fail,
Histogram, Trend, Relative measurements.
Advanced features for today’s measurement needs.
Built-in cold terminal thermocouple compensation
4.3 inch (10.92 cm) color TFT-LCD 480 x 272 display.
Accurate Temperature measurements.
USB Device, USB Host and LAN support Remote control your measurements.
Key Specications
DC Voltage 200 mV to 1000 V
DC Current 200 µA to 10 A
True RMS AC Voltage 200 mV to 750 V
True RMS AC Current 20 mA to 10 A
2/4 Wire Resistance 200 Ohms to 100 MOhms
Connectivity USB Device, LAN
Remote Control SCPI, LabView Driver
PRODUCT OVERVIEW
2
Teledyne Test Tools T3DMM5-5 is a 5½ digit digital multimeter incorporating the latest 4.3 inch (10.92cm) dual-
display technology which can be congured to show data histograms, Data fluctuation Trends, Bar Graph, Statistics
or the traditional Number mode, all in an easy to use interface.
A great feature of the Teledyne Test Tools T3DMM5-5 is it’s ability to made highly accurate True RMS AC Voltage and
Current measurements, meaning no loss of accuracy even when measuring complex voltage and current waveforms.
The T3DMM5-5 is especially well suited for the needs of High precision multifunctional environments, as well as
supporting a full range of automatic measurements.
Main Functions
Basic Measurement Function
DC Voltage: 200 mV ~ 1000 V
DCCurrent:200μA~10A
AC Voltage: True-RMS, 200 mV ~ 750 V
AC Current: True-RMS, 20 mA ~ 10 A
2/4-WireResistance:200Ω~100MΩ
Capacitance:2nF~10000μF
ContinuityTest:Rangeisxedat2kΩ
Diode Test: Adjustable range is 0 ~ 4 V.
Frequency Measurement: 20 Hz ~ 1 MHz
PeriodMeasurement:1μs~0.05s
Temperature: Support for TC and RTD sensor
Max, Min, Average, Standard Deviation,
dBm/dB, Relative Measurement, Pass/Fail
Histogram,TrendChart
User-friendly Design
4.3” TFT-LCD, 480*272
Dual display, Chinese and English Menu
Built-in front panel accessible help system
File management (support for U-disc and local storage)
Application elds
Research Laboratory
Development Laboratory
Repair and Maintenance
Calibration Laboratory
Automatic Production Test
General bench-top use
Main Features
Real 5½ digit (240,000 count) readings resolution
Up to 150 rdgs/s measurement speed
True-RMS AC Voltage and AC Current measuring
1Gbflashmemoryformassstoragecongurationles
anddatales
Built-in cold terminal compensation for thermocouple
Standard interface: USB Device, USB Host, LAN
USB & LAN remote interfaces support common SCPI
command set. Compatible with other popular DMMs
onthemarket.
3
SPECIAL FEATURES
Dual Display
Statistics
Trend Chart
dBm Hold Measurement
Bar Chart
Histogram
Hold Measurement
Rear Panel Interfaces
4
SPECIFICATIONS
DC Characteristic Accuracy ± (% of Reading + % of Range)1)
Function Range2) Test current or
Load voltage
1 Year
23°C ± 5°C
Temperature coefcient
0°C ~ 18°C
28°C ~ 50°C
DC Voltage
200 mV 0.015 + 0.004 0.0015 + 0.0005
2 V 0.015 + 0.003 0.0010 + 0.0005
20 V 0.015 + 0.004 0.0020 + 0.0005
200 V 0.015 + 0.003 0.0015 + 0.0005
1000 V4) 0.015 + 0.003 0.0015 + 0.0005
DC Current
200μA < 8 mV 0.055 + 0.005 0.003 + 0.001
2 mA < 80 mV 0.055 + 0.005 0.002 + 0.001
20 mA < 0.05 V 0.095 + 0.020 0.008 + 0.001
200 mA < 0.5 V 0.070 + 0.008 0.005 + 0.001
2 A < 0.1 V 0.170 + 0.020 0.013 + 0.001
10 A5) < 0.3 V 0.250 + 0.010 0.008 + 0.001
Resistance3)
200Ω 1 mA 0.030 + 0.005 0.0030 + 0.0006
2KΩ 1 mA 0.020 + 0.003 0.0030 + 0.0005
20KΩ 100μA 0.020 + 0.003 0.0030 + 0.0005
200KΩ 10μA 0.020 + 0.010 0.0030 + 0.0005
2MΩ 1μA 0.040 + 0.004 0.0040 + 0.0005
10MΩ 200 nA 0.250 + 0.003 0.0100 + 0.0005
100MΩ 200nA||10MΩ 1.75 + 0.004 0.2000 + 0.0005
Diode Test 2.0 V6) 1 mA 0.05 + 0.01 0.0050 + 0.0005
4 V 100μA 0.05 + 0.01 0.0050 + 0.0005
Continuity Test 2000Ω 1 mA 0.05 + 0.01 0.0050 + 0.0005
Remarks:
1) Specicationsarefor0.5Hourwarm-up,“Slow”measurementrateandcalibrationtemperature18°C~28°C.
2) 20 %overrangeonallrangesexceptforDCV1000V,ACV750V,DCI10AandACI10A.
3) Specicationsarefor4-wiremeasurementor2-wiremeasurementunder“REF”operation.±0.2Ωofextraerrorwillbegeneratedifperforminga2-wiremeasurement
without“REF”operation.
4) Plus0.02mVoferrorper1Vaftertherst±500VDC.
5) 30 seconds OFF / 30 seconds ON is recommend for the continuous current greater than DC 7 A or AC RMS 7 A.
6) Accuracyspecicationsareonlyforvoltagemeasurementsattheinputterminals.Thetypicalvalueofcurrentinthemeasurementis1mA.Voltagedropatthediode
junction may vary with current supply.
5
SPECIFICATIONS
Function Range2) Frequency Range 1 Year
23°C ± 5°C
Temperature coefcient
0°C ~ 18°C
28°C ~ 50°C
True-RMS
AC Voltage3)
200 mV
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
2 V
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
20 V
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
200 V
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
750 V
20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 KHz 0.2 + 0.05 0.01 + 0.005
20 KHz – 50 KHz 1.0 + 0.05 0.01 + 0.005
50 KHz –100 KHz 3.0 + 0.05 0.05 + 0.010
True-RMS
AC Current4)
20 mA
20 Hz – 45 Hz 1.5 + 0.10 0.015 + 0.015
45 Hz – 2 KHz 0.50 + 0.10 0.015 + 0.006
2 KHz – 10 KHz 2.50 + 0.20 0.015 + 0.006
200 mA
20 Hz – 45 Hz 1.5 + 0.10 0.015 + 0.005
45 Hz – 2 KHz 0.50 + 0.10 0.015 + 0.005
2 KHz – 10 KHz 2.50 + 0.20 0.015 + 0.005
2 A
20 Hz – 45 Hz 1.5 + 0.20 0.015 + 0.005
45 Hz – 2 KHz 0.50 + 0.20 0.015 + 0.005
2 KHz – 10 KHz 2.50 + 0.20 0.015 + 0.005
10 A5)
20 Hz – 45 Hz 1.5 + 0.15 0.015 + 0.005
45 Hz – 2 KHz 0.50 + 0.15 0.015 + 0.005
2 KHz – 10 KHz 2.50 + 0.20 0.015 + 0.005
AC Characteristic Accuracy ± (% of Reading + % of Range)1)
Additional wave crest factor error (not Sine)6)
Wave crest coefcient Error (% Range)
1 – 2 0.05
2 – 3 0.2
Remarks:
1) Specicationsarefor0.5Hourwarm-up,“Slow”measurementrateandcalibrationtemperature18°C~28°C.
2) 20 %overrangeonallrangesexceptforDCV1000V,ACV750V,DCI10AandACI10A.
3) Specicationsareforamplitudeofsinewaveinput>5 %ofrange.Forinputsfrom1%to5 %ofrangeand<50kHz,addextra0.1%ofrangeerror.
For50kHzto100kHz,addextra0.1%ofrangeerror.
4) Specicationsareforsinewaveinput>5%ofrange.0.1%errorswillsbeaddedwhentherangeofinputsinewaveis1 %to5 %.
5) 30 seconds OFF / 30 seconds ON is recommend for the continuous current that higher than DC 7 A or AC RMS 7 A.
6) For frequency < 100 Hz
6
SPECIFICATIONS
Function Range Frequency Range 1 Year
23°C ± 5°C
Temperature coefcient
0°C ~ 18°C
28°C ~ 50°C
Frequency/Period 200 mV ~ 750 V2)
20 Hz – 2 KHz 0.01 + 0.003 0.002 + 0.001
2 KHz – 20 KHz 0.01 + 0.003 0.002 + 0.001
20 KHz – 200 KHz 0.01 + 0.003 0.002 + 0.001
200 KHz –1 MHz 0.01 + 0.006 0.002 + 0.002
Remarks:
1) Specicationsarefor0.5Hourwarm-up.
2) Exceptforspecialmarks,theACinputvoltageis15 %to120 %ofrangewhen<100kHzand30 %to120 %ofrangewhen>100kHz.750Vrangeislimitedto750Vrms.
The accuracy is 10 times % of reading when the measurement range of AC voltage is in 200 mV range.
Function Probe
Type
Probe
Model
Working
Temperature Range
Accuracy (one year;
23°C ± 5°C)
Temperature coefcient
0°C ~ 18°C
28°C ~ 50°C
Temperature
RTD 2) α=0.00385 -200°C~660 °C 0.16°C 0.08 + 0.002
TC3)
B0°C~1820°C 0.76°C 0.14°C
E-270°C~1000°C 0.5°C 0.02°C
J-210°C~1200°C 0.5°C 0.02°C
K-270°C~1372°C 0.5°C 0.03°C
N-270°C~1300°C 0.5°C 0.04°C
R-270°C~1768°C 0.5°C 0.09°C
S-270°C~1768°C 0.6°C 0.11°C
T-270°C~400°C 0.5°C 0.03°C
Remarks:
1) Specicationsarefor0.5Hourwarm-up,notincludeprobeerror.
2) Specicationsarefor4-wiremeasureor2-wiremeasureunder“REF”operation.
3) Built-incoldterminalcompensationforthermocouple,accuracyis±2°C.
Frequency and Period Characteristic Accuracy ± (% of Reading + % of Range)1)
Function Range2) Max Testing Current 1 Year
23°C ± 5°C
Temperature coefcient
0°C ~ 18°C
28°C ~ 50°C
Capacitance
2 nF 200 nA 3 + 1.0 0.08 + 0.002
20 nF 200 nA 1 + 0.5 0.02 + 0.001
200 nF 2μA 1 + 0.5 0.02 + 0.001
2μF 10μA 1 + 0.5 0.02 + 0.001
200μF 100μA 1 + 0.5 0.02 + 0.001
10000μF 1 mA 2 + 0.5 0.02 + 0.001
Remarks:
1) Specicationsarefor0.5Hourwarm-upand“REF”operation.Usingofnon-lmcapacitormaygenerateadditionalerrors.
2) Specicationsareforfrom1%to120 %on2nFrangeandrangesfrom10 %to120 %onotherranges.
Capacitance Characteristic Accuracy ± (% of Reading + % of Range)1)
Temperature Characteristic Accuracy ± (% of Reading + % of Range)1)
7
MEASURING METHOD AND OTHER CHARACTERISTICS
DC Voltage
Input Resistance 200mVand2VRange 10MΩor>10GΩselectable
20V,200Vand1000VRange 10MΩ±2 %
Input Bias Current <90pA,25°C
Input Protection 1000 V on all ranges
CMRR 120dB(Forthe1KΩunbalancedresistanceinLOlead,max±500VDC)
NMRR 60dBat“slow”measurementrate
20dBisaddedifthe“AC”lterisopen(disconnected).
Resistance
Testing Method 4-wire resistance or 2-wire resistance selectable
Input Protection 1000 V on all ranges
DC Current
Shunt Resistor
200μAsamplingvoltage<8mV
2 mA sampling voltage < 8 mV
1Ωfor20mAand200mA
0.01Ωfor2A,10A
Input Protection Rear panel : accessible 10 A, 250 V fast-melt fuse
Internal: 12 A, 250 V slow-melt fuse
Continuity/Diode Test
Measurement Method 1mA±5 %constant-currentsourceoropen-circuitvoltage
Beeper yes
Continuity Threshold Adjustable
Input Protection 1000 V
True-RMS AC Voltage
Measurement Method AC Coupled true RMS measure – up to 1000 V DC bias are permitted on every range.
Wave Crest Factor ≤3atfullscale
Input Impedance 1MΩ±2 %inparallelwith<100pFonallranges
AC Filter Bandwidth 20 Hz ~ 100 KHz
CMRR 60dB(Forthe1KΩimbalanceresistanceamongLoleadand<60Hz,Max±500VDC)
True-RMS AC Current
Measurement Method DC Coupled to the fuse and shunt; AC Coupled: True-RMS measurement
(measures the AC components only)
Wave Crest Factor ≤3atfullscale
Max Input <10 A (include DC component)
Shunt Resistor 1Ωfor20mA,200mA,0.01Ωfor2A,10A
Input Protection Rear panel: accessible 10 A, 250 V fast-melt fuse
Internal: 12 A, 250 V slow-melt fuse
Frequency/Period
Measurement Method Reciprocal-counting technique, AC Coupled input, AC voltage or AC current measurement
function
Additional Errors Percentage Error increases in all frequency counters when measuring low voltage or low
frequency signal.
Capacitance Measuring
Measurement Method Measure the rate of change of voltage generated during the current flow to the capacitor
Connection Type 2-wire
Input Protection 1000 V on all ranges
8
Temperature Measuring
Measurement Method Support for TC and RTD types of sensor
Trigger and Memory
Samples/Trigger 1 ~ 10000
Trigger Delay 6 ms ~ 10000 ms optional
External Trigger Input
Input Level TTL compatible (High level when left input terminal is
hanging in the air)
Trigger Condition Rising and Falling selectable
InputImpendance ≥20KΩ//400pF,DC-coupled
Min Pulse 500 us
VMC
Level TTL compatible
Output Polarity Positive and negative optional
OutputImpendance 200Ω,typical
History Records
Volatile Memory 10 K reading history records
Nonvolatile Memory 1GbNandFlash,Massstorage,congurationles,anddatales,SupportU-diskexternal
storage
Math Functions
Min/Max/Average, dBm, dB, Pass/Fail, Relative, Standard deviation, Hold, histogram, Trend chart, Bar chart
General Specications
Power Supply
AC 100 V ~ 120 V 50/60 Hz
AC 200 V ~ 240 V 50/60 Hz
Consumption 20 VA max
Mechanism
Dimension 282 mm × 260 mm × 105 mm
Weight 3.33 Kg
Other Characteristics
Display Screen 4.3” TFT-LCD, resolution 480*272
Environmental
OperatingTemperature:0 °Cto+40 °C
StorageTemperature:-10 °Cto+70 °C
Humidity:5 %to90 %relativehumidity(non-condensing)upto+30 °C.
Upperlimitderatesto50 %relativehumidity(non-condensing)at+40 °C
Operating Altitude: 2000 m (Max)
Indoor Use only
Pollution Degree 2 (per IEC61010-1:2010)
Safety Conforming to IEC61010-1:2001. Measure CAT I 1000 V/CAT II 600V
Class of pollution: 2
Remote Interface 10/100 Mbit LAN, USB2.0 Full Speed Device and Host
Programmer Language Standard SCPI
Warm Up Time 30 minutes
The T3DMM5-5 Digital Multimeter comes with a 3 year return to Teledyne LeCroy warranty.
MEASURING METHOD AND OTHER CHARACTERISTICS
9
Ordering information
Product Name Teledyne Test Tools T3DMM5-5 Digital Multimeter
Model T3DMM5-55.5 Digit Broad Measurement Range Digital Multimeter
Standard Accessories Two Test Leads, Two Alligator Clips
A USB Cable
AQuickStart
ACalibrationCerticate
Power Cord
© 2019 Teledyne Test Tools is a brand and trademark of Teledyne LeCroy Inc. All rights reserved. Speci cations, prices, availability and delivery subject to change without notice.
Product brand or brand names are trademarks or requested trademarks of their respective holders.
T3 stands for Teledyne Test Tools.
teledynelecroy.com
Company Pro le
Teledyne LeCroy is ableading provider of oscilloscopes, protocol
analyzers and related test and measurement solutions that
enable companies across abwide range of industries to design
and test electronic devices of all types. Since our founding
in 1964, we have focused on creating products that improve
productivity by helping engineers resolve design issues
faster and more effectively. Oscilloscopes are tools used by
designers and engineers to measure and analyze complex
electronic signals in order to develop high-performance
systems and to validate electronic designs in order to improve
time to market.
The Teledyne Test Tools brand extends the Teledyne LeCroy
product portfolio with a comprehensive range of test
equipment solutions. This new range of products delivers
a broad range of quality test solutions that enable engineers
to rapidly validate product and design and reduce time-to-
market. Designers, engineers and educators rely on Teledyne
Test Tools solutions to meet their most challenging needs for
testing, education and electronics validation.
Location and Facilities
Headquartered in Chestnut Ridge, New York, Teledyne
Test Tools and Teledyne LeCroy has sales, service and
development subsidiaries in the US and throughout
Europe and Asia. Teledyne Test Tools and Teledyne LeCroy
products are employed across abwide variety of industries,
including semiconductor, computer, consumer electronics,
education, military/aerospace, automotive/industrial, and
telecommunications.
Teledyne LeCroy
(US Headquarters)
700 Chestnut Ridge Road
Chestnut Ridge, NY. USA 10977-6499
Phone: 800-553-2769 or 845-425-2000
Fax Sales: 845-578-5985
Phone Support: 1-800-553-2769
Email Sales: contact.corp@teledynelecroy.com
Email Support: support@teledynelecroy.com
Web Site: http://teledynelecroy.com/
Teledyne LeCroy
(European Headquarters)
Teledyne LeCroy GmbH
Im Breitspiel 11c
D-69126 Heidelberg, Germany
Phone: +49 6221 82700
Fax: +49 6221 834655
Phone Service: +49 6221 8270 85
Phone Support: +49 6221 8270 28
Email Sales: contact.gmbh@teledynelecroy.com
Email Service: service.gmbh@teledynelecroy.com
Email Support: tlc.t3.appsupport.eu@teledyne.com
Web Site: http://teledynelecroy.com/germany
World wide support contacts can be found at:
https://teledynelecroy.com/support/contact/#
ABOUT TELEDYNE TEST TOOLS
Distributed by:
17apr19