RS-485 I/O Modules: ADAM-4000
RoHS
COMPLIANT
2002/95/EC
1-ch Thermocouple Input Module
1-ch Analog Input Module
1-ch RTD Input Module
Specifications
General
Power Consumption 0.7 W @ 24 VDC
Supported Protocols ASCII command
Analog Input
Channels 1
Input Connections 2 or 3-wire
Input Impedance 2 M
Input Type Pt or Ni RTD
RTD Types and Temperature Ranges
IEC RTD 100 ohms
Pt -100°C to +100°C a = 0.00385
Pt 0°C to +100°C a = 0.00385
Pt 0°C to +200°C a = 0.00385
Pt 0°C to +600°C a = 0.00385
JIS RTD 100 ohms
Pt -100°C to +100°C a = 0.003916
Pt 0°C to +100°C a = 0.003916
Pt 0°C to +200°C a = 0.003916
Pt 0°C to +600°C a = 0.003916
Ni RTD
Ni -80°C to +100°C
Ni 0°C to +100°C
Accuracy ±0.1% or better
Span Drift ±25 ppm/°C
Zero Drift ±3 µV/°C
Specifications
General
Power Consumption 1.2 W @ 24 VDC
Supported Protocols ASCII command
Analog Input
Channels 1
Input Impedance Voltage: 20 M
Current: 125 (Added
by user)
Input Type mV, V or mA
Input Range ±150 mV, ±500 mV, ±1 V,
±5 V, ±10 V and ±20 mA
Accuracy Voltage mode: ±0.1% or
better
Current mode: ±0.2% or
better
Span Drift ±25 ppm/°C
Zero Drift ±6 µV/°C
Digital Input
Channels 1
Logic level 0: 1 V max.
Logic level 1: 3.5 ~ 30 V
pull up current: 0.5 mA,
10 k resistor to 5 V
Event Counter Max. input freq.: 50 Hz
Min. input pulse width: 1
msec.
Digital Output
Channels 2, open collector to 30 V,
30 mA max. load
Power Dissipation 300 mW
ADAM-4011
ADAM-4012
ADAM-4013
ADAM-4012 ADAM-4013
Specifications
General
Power Consumption 1.4 W @ 24 VDC
Supported Protocols ASCII command
Analog Input
Channels 1
Input Impedance Voltage: 2 M
Current: 125 (Added
by user)
Input Type T/C, mV, V or mA
Input Range ±15 mV, ±50 mV, ±100 mV,
±500 mV, ±1 V,
±2.5 V, ±20 mA
Accuracy Voltage mode: ±0.1% or
better
Current mode: ±0.2% or
better
T/C Type and Temperature Range
J0 ~ 760°C R500 ~ 1,750°C
K0 ~ 1,370°C S500 ~ 1,750°C
T-100 ~ 400°C B500 ~ 1,800°C
E0 ~ 1,000°C
Span Drift ±25 ppm/°C
Zero Drift ±6 µV/°C
Digital Input
Channels 1
Logic level 0: 1 V max.
Logic level 1: 3.5 ~ 30 V
Pull up current: 0.5 mA,
10 k resistor to 5 V
Event Counter Max. input freq: 50 Hz
Digital Output
Channels 2, open collector to
30 V, 30 mA max. load
Power Dissipation 300 mW
Supports high/low alarms
ADAM-4011
Common Specifications
General
Power Input Unregulated 10 ~ 30 VDC
Connectors 1 x plug-in terminal block
(#14 ~ 22 AWG)
Watchdog Timer System (1.6 second)
Analog Input
Resolution 16-bit
Sampling Rate 10 sample/second
CMR @ 50/60 Hz 150 dB
NMR @ 50/60 Hz 100 dB
Isolation Voltage 3,000 VDC
Environment
Operating Humidity 5 ~ 95% RH
Operating -10 ~ 70°C (14 ~ 158°F)
Temperature
Storage Temperature -25 ~ 85°C (-13 ~ 185°F)
Ordering Information
ADAM-4011 1-ch Thermocouple
Input Module
ADAM-4012 1-ch Analog Input
Module
ADAM-4013 1-ch RTD Input Module
RoHS
COMPLIANT
2002/95/EC
RoHS
COMPLIANT
2002/95/EC
All product specifications are subject to change without notice Last updated : 19-Mar-2014
Mouser Electronics
Authorized Distributor
Click to View Pricing, Inventory, Delivery & Lifecycle Information:
Advantech:
ADAM-4011-D2E ADAM-4013-DE ADAM-4012-DE