LG01-2000Axxx Liquid Flow Switch and Bubble Detector Detection of low liquid flow in the ml/min Range Totally non-invasive Ultra-fast response time of 100 ms Product Summary The LG01 Liquid Flow Switch enables detection of flow of liquid in the range of ml/min. The output signal is 5V when the flow rate is higher than the specified flow switch point and 0V below this limit or when air or bubbles are found inside the sensor's flow channel instead of liquid. 1 Excellent chemical resistance and bio-compatibility are ensured: The flow path of the LG01 Liquid Flow Switch is formed by a simple, straight glass capillary. For more information on communication, please refer to section 2 of this document. Sensing Performance Table 1: Model specific performance of LG01 (all data for medium H2O, 23C) Parameter H2O flow switching point H2O maximum flow rate Accuracya of switching point Switching hysteresis Mounting orientation sensitivityb Flow detection response time 63 Response time on power-up Operating temperature Ambient storage temperaturec Operating pressure Proof pressure LG01-2000A090 4.5 50 10 5 +2 LG01-2000A005 0.25 220 10 5 +14 50 120 +10...+50 -10...+60 3 5 Unit ml/min ml/min % of switching point % of switching point % of switching point ms ms C C bar bar aBetter available on request shift of switching point if mounted vertically cNon-condensing, flow path empty bMaximum 1.1 Operation with Other Liquids than Water Due to the measurement principle the LG01 flow switch characteristics is changing with the thermal properties of the liquid. For aqueous solutions the data for H2O as stated above is a good approximation. For hydrocarbon liquids such as oil, fuel, or organic solvents, the flow switch point shifts towards higher flow rates. For indicative purpose, data is provided for isopropyl alcohol (IPA) below. Table 2: Approximate switch levels of the LG01 liquid flow switch with hydrocarbon liquids Parameter Approximate flow switching point with IPA Maximum flow rate with IPA www.sensirion.com LG01-2000A090 LG01-2000A005 Unit 9 >250 1.4 >250 ml/min ml/min February 2015 V2 1/4 2 Communication with the Sensor An electrical voltage output indicates flow above (5 V) or below (0 V) the flow switch point and/or air or bubbles inside the sensors flow channel. The product comes fully calibrated for water - for volume applications flow calibration for other media is available on request. Table 3: DC Characteristics Power supply DC, VDD Operating current, VDD = 9 V, no load Voltage output high Voltage output low aUse 3 Min. Typ. Max. Units 5 7 12a V 6.8 mA 5 0 V V Connector Type: 5 pin Molex PCB Header 53261-0590 Mating connector: Molex 51021-0500. Table 4: Electrical Pinout Pin 1 2 3 4 5 2.1 Electrical Specifications Parameter 2.2 Electrical Connector and Sensor Pinout Do not connect Do not connect VDD GND Voltage output 1 2 3 4 5 9 V max. for best performance Fluidic Connection Table 5: Fluidic Specifications and Pressure Rating Parameter LG01-2000Axxx Wetted materials: Borosilicate Glass 3.3 PEEK FEP 1/4-28 flat-bottom for 1/16" or 1/8" OD plastic tubing (min. 2mm ID recommended)a 1 mbar 80 l Internal sensor tube material Fitting material Sealing material Fluid connector ports (Fittings) Pressure drop (at maximum flow rate, H2O, 23C) Total internal volume For fluid connection material use flat bottom 1/4-28 standard components e.g. IDEX parts P-200/P-300. Various adapters e.g. to Barb connectors are available from company Value Plastics as well. a For more information on the fluidic connection please find: "Application Note Sensor Ports and Tubing Connections" in the Download Center on our webpage (www.sensirion.com/liquidflow-download). 4 Connectivity Kit The optional LG01 connectivity kit contains material for electrical and fluidic connections Assembled flat ribbon cable, 30cm (1-100482-01) 2x barb adapter to connect ID 1.6 mm (1/16") tubing (Value Plastics ABR004-J1A-2) 2x barb adapter to connect ID 3.2 mm (1/8") tubing (Value Plastics ABR013-J1A-1) www.sensirion.com February 2015 V2 2/4 5 Mechanical Specifications Table 6: Mechanical Specifications Parameter LG01-2000Axxx Largest dimensions Total mass Inner diameter d 53 x 22 x 9 mm 6g 1.8 mm 531 All dimensions in mm Attention Fragile Mechanical shocks, forces induced by stiff fluid connection lines as well as connecting to the fittings without suitable tools lead to stress on the internal thin-walled glass capillary and can cause it to break. While tightening the fittings, fix the fluidic ports position with a wrench. Test for leakage after every time new connections are made. 6 ! Ordering Information For OEM applications the sensor can be purchased in larger quantities without any additional parts. Product LG01-2000A090 LG01-2000A005 LG01 Connectivity Kit www.sensirion.com Article Number 1-100796-01 1-100797-01 1-100814-01 February 2015 V2 3/4 Important Notices Warning, personal injury Do not use this product as safety or emergency stop devices or in any other application where failure of the product could result in personal injury (including death). Do not use this product for applications other than its intended and authorized use. Before installing, handling, using or servicing this product, please consult the datasheet and application notes. Failure to comply with these instructions could result in death or serious injury. If the Buyer shall purchase or use SENSIRION products for any unintended or unauthorized application, Buyer shall defend, indemnify and hold harmless SENSIRION and its officers, employees, subsidiaries, affiliates and distributors against all claims, costs, damages and expenses, and reasonable attorney fees arising out of, directly or indirectly, any claim of personal injury or death associated with such unintended or unauthorized use, even if SENSIRION shall be allegedly negligent with respect to the design or the manufacture of the product. ESD Precautions The inherent design of this component causes it to be sensitive to electrostatic discharge (ESD). To prevent ESD-induced damage and/or degradation, take customary and statutory ESD precautions when handling this product. Warranty This warranty does not apply to any equipment which has not been installed and used within the specifications recommended by SENSIRION for the intended and proper use of the equipment. EXCEPT FOR THE WARRANTIES EXPRESSLY SET FORTH HEREIN, SENSIRION MAKES NO WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH RESPECT TO THE PRODUCT. ANY AND ALL WARRANTIES, INCLUDING WITHOUT LIMITATION, WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY EXCLUDED AND DECLINED. SENSIRION is only liable for defects of this product arising under the conditions of operation provided for in the datasheet and proper use of the goods. SENSIRION explicitly disclaims all warranties, express or implied, for any period during which the goods are operated or stored not in accordance with the technical specifications. SENSIRION does not assume any liability arising out of any application or use of any product or circuit and specifically disclaims any and all liability, including without limitation consequential or incidental damages. All operating parameters, including without limitation recommended parameters, must be validated for each customer's applications by customer's technical experts. Recommended parameters can and do vary in different applications. SENSIRION reserves the right, without further notice, (i) to change the product specifications and/or the information in this document and (ii) to improve reliability, functions and design of this product. SENSIRION warrants solely to the original purchaser of this product for a period of 12 months (one year) from the date of delivery that this product shall be of the quality, material and workmanship defined in SENSIRION's published specifications of the product. Within such period, if proven to be defective, SENSIRION shall repair and/or replace this product, in SENSIRION's discretion, free of charge to the Buyer, provided that: Copyright (c) 2001-2015, SENSIRION. The LG01 flow switch complies with requirements of the following directives: notice in writing describing the defects shall be given to SENSIRION within fourteen (14) days after their appearance; such defects shall be found, to SENSIRION's reasonable satisfaction, to have arisen from SENSIRION's faulty design, material, or workmanship; the defective product shall be returned to SENSIRION's factory at the Buyer's expense; and the warranty period for any repaired or replaced product shall be limited to the unexpired portion of the original period. CMOSens(R) is a trademark of Sensirion All rights reserved REACH, RoHS and WEEE Statement EU Directive 1907/2006/EC concerning Registration, Evaluation, Authorization and Restriction of Chemicals (REACH) EU Directive 2002/96/EC on waste electrical and electronic equipment (WEEE), OJ13.02.2003; esp. its Article 6 (1) with Annex II. EU Directive 2002/65/EC on the restriction of certain hazardous substances in electric and electronic equipment (RoHS), OJ01.01.2011 Headquarters and Subsidiaries SENSIRION AG Laubisruetistr. 50 CH-8712 Staefa ZH Switzerland Sensirion Inc., USA phone: +1 805 409 4900 info_us@sensirion.com www.sensirion.com Sensirion Korea Co. Ltd. phone: +82 31 345 0031 3 info@sensirion.co.kr www.sensirion.co.kr phone: +41 44 306 40 00 fax: +41 44 306 40 30 info@sensirion.com www.sensirion.com Sensirion Japan Co. Ltd. phone: +81 3 3444 4940 info@sensirion.co.jp www.sensirion.co.jp Sensirion China Co. Ltd. phone: +86 755 8252 1501 info@sensirion.com.cn www.sensirion.com.cn Sensirion AG (Germany) phone: +41 44 927 11 66 info@sensirion.com www.sensirion.com To find your local representative, please visit www.sensirion.com/contact www.sensirion.com February 2015 V2 4/4 Mouser Electronics Authorized Distributor Click to View Pricing, Inventory, Delivery & Lifecycle Information: Sensirion: LG01 CONNECTIVITY KIT LG01-2000A005 LG01-2000A090