Electrical Specications
DC specications
Function Range Resolution
Accuracy ±
(% of reading + counts of
least signicant digit)
Test
current
Burden
voltage/
shunt
Input
impedance
U1231A/U1232A/
U1233A
Where
(applicable)
Where
(applicable)
Where
(applicable)
Voltage
600 mV10.1 mV 0.5% + 2 NA NA 11.18 MΩ
6 V 0.001 V 0.5% + 2 NA NA 11.18 MΩ
60 V 0.01 V 0.5% + 2 NA NA 10.1 MΩ
600 V 0.1 V 0.5% + 2 NA NA 10 MΩ
600 V (VZLOW)20.1 V 2% + 3 NA NA 3 kΩ
Resistance
600 Ω40.1 Ω 0.9% + 3 0.57 mA NA NA
6 kΩ40.001 kΩ 0.9% + 3 57 μA NA NA
60 kΩ 0.01 kΩ 0.9% + 3 5.7 μA NA NA
600 kΩ 0.1 kΩ 0.9% + 3 570 nA NA NA
6 MΩ50.001 MΩ 0.9% + 3 100 nA/10 MΩ NA NA
60 MΩ50.01 MΩ 1.5% + 3 100 nA/10 MΩ NA NA
Diode3
2 V 0.001 V 0.9% + 2 0.57 mA NA NA
Current4
60 μA10.01 μA 1.0% + 2 NA < 2.5 V/1 kΩ NA
600 μA10.1 μA 1.0% + 2 NA < 2.5 V/1 kΩ NA
6 A2,5 0.001 A 1.0% + 3 NA < 0.2 V/0.005 Ω NA
10 A2, 3 0.01 A 1.0% + 3 NA < 0.4 V/0.005 Ω NA
Notes for DC voltage specications:
1. The accuracy of the 600 mV range is specied after the Null function is used to subtract the thermal effect (by shorting the test leads).
2. For VZLOW (low input impedance) measurements, auto-ranging is disabled and the multimeter’s range is set to 600 V in the manual
ranging mode.
Notes for resistance specications:
1. Overload protection: 600 Vrms for short circuits with < 0.3 A current.
2. Maximum open voltage is < +3 V.
3. Built-in buzzer beeps when the resistance measured is less than 23 Ω ± 10 Ω. The multimeter can capture intermittent measurements longer
than 1 ms.
4. The accuracy of the 600 Ω to 6 kΩ range is specied after the Null function is used to subtract the test lead resistance and thermal effect (by
shorting the test leads).
5. For the ranges of 6 MΩ and 60 MΩ, the RH is specied for < 60%.
Notes for diode specications:
1. Overload protection: 600 Vrms for short circuits with < 0.3 A current.
2. Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased diode or semiconductor
junctions measured between 0.3 V and 0.8 V (0.3 V ≤ reading ≤ 0.8 V).
3. Open voltage for diode: < +3 V DC.
4. The maximum display for diode measurements is 2100 counts.
Notes for DC current specications:
1. Overload protection for 60 μA to 600 μA range: 600 Vrms for short circuits with < 0.3 A current.
2. Overload protection for 6 A to 10 A range: 11 A/1000 V; 10 × 38 mm fast-acting fuse.
3. Specication for 10 A range: 10 A continuous. Add 0.3% to the specied accuracy when measuring signals > 10 A to 20 A for 30 seconds
maximum. After
measuring currents > 10 A, cool down the multimeter for twice the duration of the measured time before proceeding with
low current measurements.
4. Only applicable for the U1232A/U1233A model
5. DC current range of 0.6 μA to 1 μA is not measureable on the U1232A and U1233A models.
05 | Keysight | U1230 Series Handheld Digital Multimeters (DMMs) – Data Sheet