NEVO+1200M MEDICAL AC/DC MODULAR CONFIGURABLE POWER SUPPLY DATA SHEET _______________________________________________________________________________________________ 6"x6"x1.61" 1200W 1.2kg SMALL POWERFUL LIGHT The NEVO+1200M is the smallest in its class and the ultimate power solution for medical applications where size, weight, low standby power and primary side inhibit are vital factors and delivers up to 1200 Watts from a 1.2kg 6" x 6" x 1.61" package. Each configured unit consists of an input module with up to eight output modules, where any combination of outputs can be fitted to create a power solution with up to sixteen isolated outputs. Standard features include intelligent fan control, wide output voltage adjust capability and primary side shutdown with standby power consumption of less than 3 Watts. A low noise fan option with virtually silent operation is also available, which allows you to use this innovative power supply in even the quietest of environments. The series carries full 3rd Edition 60601 safety approvals and complies with EN55022-B EMC Standards and features market leading specifications and design in application support. MAIN FEATURES * * * * * * Up to 1200 Watts of output power 6 x 6 x 1.61" footprint Efficiency up to 89% Primary side remote on/off function Standby power 3 Watts Field configurable * * * * UL60601 3rd Edition approved Industry leading power density (21W/in) Lightest modular design - only 1.2kg - 1000Watts/kg Parallel & series connection of modules * * * * * Remote current / voltage programming Accurate current sharing 2 x 5V 1A bias supply Low noise fan option 3 Year warranty NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016 Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com SPECIFICATIONS INPUT ELECTRICAL Details AC Input Voltage AC Input Frequency DC Input Voltage Power Rating Input Current Inrush Current Fusing Input Current Limit Efficiency Idle Power Idle Power Standby Power Power Factor Holdup UVLO Over temperature Reliability Leakage Current Output Bias voltage Output Bias current Power Good voltage Power Good current Inhibit voltage Inhibit current Global inhibit voltage Global inhibit current Nominal range is 100V to 240V Contact factory for 400Hz operation. Standard See graphs for de-rating 1200Watts output at 120Vrms input 265Vrms (cold start) 5x20 Fast acting S i g n a l s Parameter Min 85 47 120 See graphs All outputs fitted and enabled All outputs fitted and Disabled Latched off state, 120Vrms 1200Watts output at 120Vrms input Turn on only Internally monitored. Latching 40C 80% load Normal condition, 264V, 63Hz Two isolated Bias Outputs available Hiccup type current limit PNP open collector with internal 10k pull down resistor 10k ohm input impedance AC_OK voltage AC_OK current AC_OK warning Primary Bias voltage Primary Bias current Primary Remote On/Off 5k ohm input impedance High output Low output See user manual for exceptions Medically Isolated Hiccup type current limit Negative Edge Triggered, Refer to User Manual 17 78 115 4.8 0 8 0 2 0.2 3 0.6 4.7 0 -10 5 4.8 Typ 50/60 14 86 46 32 2.5 0.99 20 Max Units 264 63 370 1200 12 40 12.5 Vrms Hz Vdc Watts Amps Amps Amps Amps % Watts Watts Watts 89 0.99 21 84 125 2 190 5 mS Vrms C FPMH Amps V A V mA V mA V mA V V mA mS V A V 5.2 1 15 20 15 1.5 15 3 5.2 0.1 10 10 5 5.2 0.5 5 RELIABILITY Component Details Fan Mag Lev Std (2 Fans per unit) Input Output Min Max Units 3.8 FPMH Excluding FAN 2 FPMH See individual output datasheets 1 FPMH 3 Years Max Units 4000 1500 250 250 Vac Vac Vdc Vdc mm mm mm mm uA Warranty Parameter Isolation voltage Isolation clearance Isolation creepage Leakage current SAFETY Details Input to output (2 MOPP) Input to chassis (1 MOPP) Output to chassis Output to output Primary to secondary (reinforced) Primary to chassis (basic) Primary to secondary (reinforced) Primary to chassis (basic) Medical: 265Vac, 63Hz, 25C Min 7 2.5 12 4 300 MECHANICAL Parameter Details Size 154.5mm (L) x 152.4mm (W) x 41.0 1.0mm (H) Weight 720 gram + 60 gram per output module Mounting Bottom (see diagram for details) NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016 Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com Storage Parameter Temperature Humidity Altitude Air Pressure Relative, non-condensing Full power Derate input and outputs at 2.5%/C Relative, non-condensing (-200 to 2000m for UL60601-1) Temperature Operation ENVIRONMENTAL Details Humidity Altitude Air Pressure Noise level Measured 1m from fan intake Shock Vibration Min Max Units -40 5 -200 54 -20 50 5 -200 78 +85 95 5000 106 50 70 95 3000 106 42 61 C % m kPa C C % m kPa dBA dBA Unit at idle Unit at full power, 25C 3000 bumps at 10G (16ms) half sine wave 1.5G 10 to 200Hz sine wave, 20G for 15min in 3 axes random vibration EMC Emissions Parameter Standard Level Radiated electric field Conducted emissions Harmonic Distortion Flicker & Fluctuation EN55011, EN55022, FCC EN55011, EN55022, FCC EN61000-3-2 EN61000-3-3 EN61000-4-2 Electrostatic discharge (15kV air, 8kV contact) Radiated RFI EN61000-4-3 (10V/m) Fast Transient burst EN61000-4-4 (4kV) Input line surges EN61000-4-5 (1kV L-N, 2kV L-E) Conducted RFI EN61000-4-6 (10V) Power Freq. Magnetic Field EN61000-4-8 (10A/m) Voltage Dips EN61000-4-11 (EN55024) Note: To meet Class B radiated emissions the end user should add ferrites to I/P and O/P cables. Consult Vox Power for details. A (See Note) B Compliant Compliant Immunity 4 3 4 3 4 3 Compliant AGENCY APPROVALS Standard Details File IEC/EN60601-1 3rd Edition UL60601-1 3rd Edition CAN/CSA-C22.2 No. 60601-1 (2008) ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10) CE MARK LVD 2014/35/EU CB certificate and report available on request UL: E316486 Typical Line Efficiency (Pmax) 0,90 0,89 0,88 0,87 Efficiency Efficiency Typical Load Efficiency (220Vrms) 0,92 0,90 0,88 0,86 0,84 0,82 0,80 0,78 0,76 0,74 0,72 0,70 0,68 0,66 0,86 0,85 0,84 0,83 OP1 OP2 0,82 OP3 0,81 OP4 OP1 OP2 OP3 OP4 0,80 80 100 200 300 400 500 600 700 800 900 1000 1100 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260 1200 Input Voltage(Vrms) Output power (W) Temperature Derating 1400 1400 1200 1200 1000 Output Power (W) Output Power (W) Line Derating 1000 800 Derate at 10W per Volt below 120Vrms 600 800 Derate at 30W per degree celcius above 50 degree 600 400 No Derating for DC operation 400 200 200 0 -20 -10 0 10 20 30 40 50 60 70 0 80 100 120 140 160 180 200 220 240 260 Ambient Temperature (Celcius) Input Voltage (RMS) NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016 Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com MECHANICAL DIMENSIONS AND MOUNTING SCREWS SCREWS LOCATION DETAILS PENETRATION TIGHTENING MOUNTING M4 x 4 4mm max, including chassis 0.55 NM (1) OUTPUT MODULES M3 x 5, Countersink Posi, 16 Places Defined by screw 0.35 NM (1) CHASSIS LID AND FACEPLATE M3 x 5, Countersink Posi, 11 Places Defined by screw 0.35 NM (1) 1. Torque settings are for general reference only. The torque settings shown in the datasheet are the insert manufacturers recommended values. NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016 Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com CONNECTORS PINOUTS 6 1 J1 Circuit 1 2 3 Details Live Earth Neutral J6 J2a/b Details Circuit 1 2 3 4 5 6 7 8 9 10 11 12 Power Good Inhibit Power Good Inhibit Power Good Inhibit Power Good Inhibit Global Inhibit AC OK +5V 1A Bias Supply COM 1 2 3 4 5 6 Common +5V 500mA Bias Shut Down Reserved Reserved Reserved Slot A and E Slot B and F Slot C and G Slot D and H N E L J1 J6 J2a J2b 1 11 2 12 v REF J1 J2a/b J6 J1 DETAILS MANUFACTURER MAINS INPUT: 3 Pin, Barrier, 6-32 Steel Screws, 0.8 NM or 7IN LB Torque Cable 14-18AWG, 300V, 16A, 105C, use appropriately rated fork or ring terminal. GLOBAL SIGNALS: 12 Pin, 2mm, without Friction Lock, 24-30 AWG INPUT BIAS: OUTPUT SIGNALS: 6 Pin, 1.25mm, with Friction lock, 28-32 AWG HOUSING TERMINAL MOLEX MOLEX 511101251 503948051 MOLEX 510210600 500588000 Notes 1. Direct equivalents may be used for any connector parts. 2. All cables must be rated 105C min, equivalent to UL1015 PART NUMBERING SYSTEM NEVO Power Series Leakage Current Medical NEVO+1200 M - 1 1 2 2 3 3 4 4 - 0 0 0 Factory Use USE '0' for unused slots. Blanking plates will be inserted at factory. Slot A - Output # Slot H - Output # Slot B - Output # Slot G - Output # Slot C - Output # Slot F - Output # Slot D - Output # Slot E - Output # Our design team will assist with value add requirement if an application requires standard/non-standard accessories or non-nominal voltage settings. Once approved, the factory will issue a 3 or 4 digit code for your specific configuration which can be used for all future orders of the same configuration. When ordering an input unit with no outputs inserted, simply order NEVO+1200M. All specifications are believed to be correct at time of publishing. Vox Power Ltd reserves the right to make changes to any of its products and to change or improve any part of the specification, electrical or mechanical design or manufacturing process without notice. Vox Power Ltd does not assume any liability arising out of the use or application of any of its products and of any information to the maximum extent permitted by law. No license, express or implied, by estoppel or otherwise, to any intellectual property rights is granted by this document or by any products of Vox Power Ltd. VOX POWER LTD DISCLAIMS ALL WARRANTIES AND REPRESENTATIONS OF ANY KIND WHETHER EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, IMPLIED WARRANTIES OF SUITABILITY, FITNESS FOR PURPOSE, MERCHANTABILITY AND NONINFRINGEMENT. Please consult your local distributor or Vox Power directly to ensure that you have the latest revision before using the product and refer to the latest relevant user manual for further information relating to the use of the product. Vox Power Ltd products are not intended for use in connection with life support systems, human implantations, nuclear facilities or systems, aircraft, spacecraft, military or naval missile, ground support or control equipment used for the purpose of guidance navigation or direction of any aircraft, spacecraft or military or naval missile or any other application where product failure could lead to loss of life or catastrophic property damage. The user will hold Vox Power Ltd harmless from any loss, cost or damage resulting from its breach of these provisions. NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016 Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com