RIGOL
Users Guide
DSA1000&DSA1000A Series
Options&Accessories
Nov 2011
RIGOL Technologies, Inc.
RIGOL
DSA1000&DSA1000A Options&Accessories
1
Guaranty and Declaration
Copyright
© 2010 RIGOL Technologies, Inc. All Rights Reserved.
Trademark Information
RIGOL is a registered trademark of RIGOL Technologies, Inc.
Publication Number
UMD01104-1110
Notices
RIGOL products are protected by patent law in and outside of P.R.C.
RIGOL Technologies, Inc. reserves the right to modify or change parts of or all
the specifications and pricing policies at company’s sole decision.
Information in this publication replaces all previously corresponding material.
RIGOL shall not be liable for losses caused by either incidental or consequential
in connection with the furnishing, use or performance of this manual as well as
any information contained.
Any part of this document is forbidden to copy or photocopy or rearrange without
prior written approval of RIGOL.
Product Certification
RIGOL guarantees this product conforms to the national and industrial standards in
China. International standard conformance certification is in progress, e.g. ISO.
Contact Us
If you have any problem or requirement when using our products, please contact
RIGOL Technologies, Inc. or your local distributors, or visit: www.rigol.com
RIGOL
DSA1000&DSA1000A Options&Accessories
2
Content
Guaranty and Declaration .................................................................................. 1
Content ........................................................................................................... 2
DSA1000 Series Options&Accessories ................................................................. 3
DSA1000A Series Options&Accessories ............................................................... 4
Preamplifier ..................................................................................................... 5
Advanced Measurement Functions ...................................................................... 6
3 GHz Tracking Generator ................................................................................. 11
USB-GPIB Interface Converter .......................................................................... 13
Front Panel Cover ............................................................................................ 14
Desk Mount Instrument Arm ............................................................................. 15
Rack Mount Kit ................................................................................................ 16
Soft Carrying Bag ............................................................................................ 22
Other Accessories ............................................................................................ 24
RIGOL
DSA1000&DSA1000A Options&Accessories
3
DSA1000 Series Options&Accessories
Name
Descriptions
Options
Preamplifier
DSA1030-PA (For DSA1030 only)
Advanced Measurement Kit
DSA1000-AMK (For DSA1030 only)
3 GHz Tracking Generator
DSA1030-TG3 (For DSA1030 only)
PC software for DSA
Ultra Spectrum*
USB-GPIB Interface Converter for
Instrument
USB-GPIB
RF Demo Kit (Transmitter)
TX1000
Standard Accessories
Power Cord
1
USB Cable
1
Front Panel Cover
1
Quick Guide
1
CDROM
1 (Users Guide, Programming Guide)
Optional Accessories
Desk Mount Instrument Arm
ARM
DSA Accessories Package
DSA Utility Kit*
Front Panel Cover
DSA1000-FPCS
Soft Carrying Bag
DSA1000-SCBA
Rack Mount Kit
DSA1000-RMSA
*Note: please refer to the corresponding documents for more information.
RIGOL
DSA1000&DSA1000A Options&Accessories
4
DSA1000A Series Options&Accessories
Name
Description
Options
3 GHz Tracking Generator
DSA1030-TG3
PC software for PC
Ultra Spectrum*
USB to GPIB Interface Converter for
Instrument
USB-GPIB
RF Demo Kit (Transmitter)
TX1000*
Standard Accessories
Power Cord
1
USB Cable
1
Front Panel Cover
1
Quick Guide
1
CDROM
1 (Users Guide, Programming Guide)
Optional Accessories
Desk Mount Instrument Arm
ARM
RF Demo Kit
TX1000
DSA Accessories Package
DSA Utility Kit*
Front Panel Cover
DSA1000-FPCS
Soft Carrying Bag
DSA1000-SCBA
Rack Mount Kit
DSA1000-RMSA
*Note: please refer to the corresponding documents for more information.
RIGOL
DSA1000&DSA1000A Options&Accessories
5
Preamplifier
Turning on the preamplifier reduces the displayed average noise level in order to
distinguish small signals from the noise when working with small signals. For the
instruments supported option DSA1030-PA, pressing the front panel key AMPT, a
corresponding menu will appear on the right of the screen. Then, press RF Preamp
to turn on or off the preamplifier. An icon will be displayed on the left of the screen
when the preamplifier is on. The preamplifier will be disabled automatically when the
input attenuation is set higher than 30 dB.
The measurement results when the preamplifier is off and on are shown below,
respectively. As can be seen, small input signals can be displayed by the analyzer
when the preamplifier is on.
RIGOL
DSA1000&DSA1000A Options&Accessories
6
Advanced Measurement Functions
Option DSA1000-AMK provides plenty of advanced measurement functions including
T-Power (Time domain Power), ACP (Adjacent Channel Power), Chan Pwr (Channel
Power), OBW (Occupied Bandwidth), EBW (Emission Bandwidth), C/N Ratio, Harmo
Dist (Harmonic Distortion), TOI (Third Order Intermodulation) and Pass/Fail. The
measure mode can be Single or Continuous and you can control the measurement
status: Restart, Pause or Resume.
Pressing the front panel key Meas, the corresponding menu will appear on the right
of the screen. Press Meas Fctn and choose a measurement function. The screen will
be divided into two windows. The upper one is for basic measure, displaying sweep
trace, and the lower one shows the measurement results.
1. T-Power (Time domain Power)
Enables the Zero Span Mode and calculates the power within time domain. The
measurable power types are Peak, Average and RMS.
Measured Result Parameters
Measure Control
Start Line Stop Line
RIGOL
DSA1000&DSA1000A Options&Accessories
7
2. ACP (Adjacent Channel Power)
Measures the power of the main channel and the adjacent channels and
calculates the power difference between the main and each of the adjacent
channels. When enabled, both the span and resolution bandwidth of the analyzer
are adjusted to be smaller automatically.
3. Chan Pwr (Channel Power)
Measures the power and the power density within a specified channel bandwidth.
The span and bandwidth are automatically set to smaller values in this
measurement type.
Main CH BW
Adj CH BW
CH Spacing
Integ BW CH Pwr Span
RIGOL
DSA1000&DSA1000A Options&Accessories
8
4. OBW (Occupied Bandwidth)
Calculates the power within whole bandwidth by integral operation and works out
the occupied bandwidth by this value based on the specified power ratio. The
center frequency difference between the measured channel and the analyzer will
also be given in the measurement results.
5. EBW (Emission Bandwidth)
Measures the bandwidth of the two points at both sides of the max signal when
the amplitude of this max signal falls off X dB within the span range.
RIGOL
DSA1000&DSA1000A Options&Accessories
9
6. C/N Ratio
Measures the power of both the carrier and the noise within specified bandwidth
and calculates their ratio.
7. Harmo Dist (Harmonic Distortion)
Measures each order harmonic power and THD (Total Harmonic Distortion) of
carrier. The available range is up to 10 orders. And the fundamental wave
amplitude of the carrier must be higher than -50 dBm or else the measurement
will be invalid.
Noise BW
Offset Freq
Carrier BW
RIGOL
DSA1000&DSA1000A Options&Accessories
10
8. TOI (Third Order Intermodulation)
Measures the parameters of the TOI production generated by two signals which
have the same amplitude and similar frequency. The measured results include the
frequency and amplitude of the Base Lower, Base Upper, 3rd Order Lower and 3rd
Order Upper signals, as well as the Intercept of both the 3rd Order Lower and 3rd
Order Upper signals.
9. Pass/Fail
Compares the measured curve with the pre-edited, if the related rules are met,
the result will be Pass, or else is Fail.
RIGOL
DSA1000&DSA1000A Options&Accessories
11
3 GHz Tracking Generator
When the Tracking Generator (DSA1030-TG3 Option) is on, a signal with the same
frequency with the current sweep signal will be outputted from the GEN OUTPUT 50
Ω terminal on the front panel. The power of the signal could be set through the menu.
And two working modes, Power Sweep Output and Fixed Power Output, are available.
The characteristic of the DUT (Device Under Test) can be acquired by measuring the
response of the DUT stimulated by the Tracking Generator. For example, the
frequency response of a filter can be measured by using the Tracking Generator.
Connect the filter under measure to the RF INPUT 50Ω and GEN OUTPUT 50Ω
terminal of the analyzer.
Measurements Example
The frequency response of a 300 MHz low-pass filter is shown below.
Filter
RIGOL
DSA1000&DSA1000A Options&Accessories
12
Tracking Generator Specifications
Item
Condition
Value
Frequency Range
9 kHz to 3 GHz
Output Power
-20 dBm to 0 dBm, in 1 dB
steps
Output Flatness
10 MHz to 3 GHz,
Reference to 50 MHz
± 3 dB
RIGOL
DSA1000&DSA1000A Options&Accessories
13
USB-GPIB Interface Converter
By using RIGOL USB-GPIB interface converter, the spectrum analyzer could be
connected to a GPIB Bus Controller on PC. That is, the analyzer has been expanded
GPIB interface and could finish all sorts of work more easily through GPIB instructions.
And the main performance characters are shown below.
Connect the USB interface of the converter to the USB Host interface of the analyzer,
and connect the GPIB interface to the GPIB Bus Controller on PC.
Enable GPIB control of the analyzer through the USB
Host port.
GPIB primary address of the analyzer can be assigned
through the host instrument (PC).
Powered by USB Host, do not need external power
supply.
LED indicator provides the power status.
LED Indicator
Analyzer
PC
GPIB
USB
RIGOL
DSA1000&DSA1000A Options&Accessories
14
Front Panel Cover
The cover is provided to avoid damage to the front panel of the analyzer during
transportation.
Before putting your analyzer into action, remove the front panel cover by releasing
latches at both sides, and sliding in the direction of the arrows as shown in the figure
below.
RIGOL
DSA1000&DSA1000A Options&Accessories
15
Desk Mount Instrument Arm
Using an arm, the analyzer could be mounted on the workdesk to save your operation
space. The height and angle of the instrument could be adjusted freely to acquire a
supreme comfort and efficiency making measurement and observation conveniently.
To use the arm, please order the related option.
The effects figures of an analyzer installed to an arm are given below.
RIGOL
DSA1000&DSA1000A Options&Accessories
16
Rack Mount Kit
The analyzer could be installed in 19-inch standard machine cabinet, shown below.
1. Kit Parts List
No.
Name
Qty.
Part Number
Description
1-1
Front Panel
1
RM-DSA-1-01
1-2
Support Board
1
RM-DSA-1-02
1-3
Left Plate
1
RM-DSA-1-03
1-4
Right Plate
1
RM-DSA-1-04
1-5
Fixed Figure
2
RM-DSA-1-05
1-6
Rear Support Bar
1
RM-DSA-1-06
2-1
M4 Screw
20
RM-SCREW-01
M4*6 Phil-Slot Pan Head
Machine Screw Nail
2-2
M6 Screw
4
RM-SCREW-02
M6*16 Phil-Slot Pan Head
Machine Screw Nail
2-3
M6 Nut
4
RM-SCREW-03
M6*5 Square Machine Female
Screw Contain Lock Blade
2-4
M4 Screw
4
RM-SCREW-04
M4*8 Cross Recessed Pan
Head Screw
RIGOL
DSA1000&DSA1000A Options&Accessories
17
2-1 2-2 2-3 2-4
2. Installation Tool
PH2 Phillips Screwdriver (recommended)
RIGOL
DSA1000&DSA1000A Options&Accessories
18
3. Space Requirements for Installation
The following requirements must be fulfilled by the machine cabinet in which the
instrument is mounted.
Dimension of the machine cabinet must be standard 19-inch.
At least 6U (266.7mm) space should be provided by the machine cabinet.
The depth inside the machine cabinet should not be less than 300 mm.
The dimension of the instrument after being installed is shown below.
RIGOL
DSA1000&DSA1000A Options&Accessories
19
4. Installation Procedures
A PH2 Phillips Screwdriver is recommended. Note that the instrument will be
damaged or installed in rack incorrectly if the installation is not proper. Thus, only
authorized operators can execute the installation operation.
1) Install the right and left palate: aim the detents of right and left plate at the
openings on support board and insert respectively, then fix them using six M4
screws.
2) Place the instrument: Remove the battery* by pressing the button on top of it,
and then place the instrument into the confined position on the Support Board.
*Note: the battery is only supplied in China.
RIGOL
DSA1000&DSA1000A Options&Accessories
20
3) Fix the instrument: Fix a Rear Support Bar” and two “Fixed Figure” on the
instrument using four M4*8 Screws. Locate the instrument on proper location and
then fixed the Rear Support Bar” and “Fixed Figure on the left/right board and
the Support Board using eight M4*6 Screw.
4) Mount Front Panel: Let the screw holes fitting well together, and aiming the
instrument front panel at the opening of the Front Panel of the machine rack and
fix them with six M4 screws.
RIGOL
DSA1000&DSA1000A Options&Accessories
21
5) Load into machine cabinet: mount the rack with instrument-fixed into a standard
19-inch machine cabinet respectively using four M6 screws and square nuts.
6) Notice: The rack holds a height of 6U, and the holes those in compliance with the
arrow direction are the mounting holes for rack.
RIGOL
DSA1000&DSA1000A Options&Accessories
22
Soft Carrying Bag
Main features
Could be held by hand, one-shoulder or shoulders. In addition, one of the belts
could be adjusted as neck strap through the vertical strap loop, while another one
as waist strap. In this condition, the operator could execute the measurement
work, and it could reduce the load on shoulders.
Soft bag body. And the shoulder straps and belts are equipped by sponge to
increase the contact area of the case on human body. It is in conformity with
human engineering and provides greatly comforts.
A strap collection box at the bottom of the case is designed to collect the straps
while using belts and ensure the straps do not prevent users work.
Many pockets inside for organizing the instrument accessories; have strong
accommodating function.
The built-in water-proof bag can accommodate the case body entirely to avoid
damage from water or rain and ensure its normal use and service life; while not in
use, fold it up and place it in the hidden pocket outside of the case body and
without preventing normal applications.
The bag is designed by RIGOL for utility
model measurement instruments. It has
reasonable structure as adopting some
improvements according to the instrument
and practical application. It is in conformity
with human engineering and could help
you to carry instrument and its accessories
easily with comforts and conveniences.
RIGOL
DSA1000&DSA1000A Options&Accessories
23
Held by shoulders Held by Neck&Waist Carried by Hand
RIGOL
DSA1000&DSA1000A Options&Accessories
24
Other Accessories
Accessories along with the instrument are shown below. Note that the pictures here
are for your reference only. Some differences may exist.
Power Cord USB Cable
Quick Guide CDROM*
Front Panel Cover
*Note: The CD contains
User’s Guide
. If there are some modifications on it,
Please be according to the version on the RIGOL website.