Digital Multimeters
Tektronix DMM4020 Data Sheet
DMM4020
Features & Benets
Key Performance Specications
5.5 Digit Resolution
Basic VDC Accuracy of up to 0.015% (1 yr.)
200 mV to 1000 V Voltage Range, with up to 10 V Resolution
200 A to 10 A Current Range, with up to 1 nA Resolution
200 to 100 MOhm Range, with up to 1 mResolution
CAT I 1000 V, CAT II 600 V
Available Functions and Features
Volts, Ohms, and Amps Measurements
Diode and Continuity Testing
Frequency Measurements
2×4 Ohms 4-wire Measurement Technique
Dedicated DC Leakage Current Measurements
Six Dedicated Buttons for Fast Access to Instrument Setups
Limit Compare Mode for Pass/Fail Testing
Connectivity
Front 2×4 Measurement Inputs
RS-232 on Rear Panel for Quick PC Connectivity
Includes USB to RS-232 Interface Adapter Cable
Includes National Instrument’s LabVIEW SignalExpress™ TE Limited
Edition for Connecting Your Bench
3-year Warranty
Data Sheet
Measurements with the Push of a Button
As the circuits in embedded system designs become more sophisticated,
you must measure a multitude of different parameters to validate your
design. The Tektronix DMM4020 5.5 digit bench multimeter offers a broad
range of functions in one easy-to-use instrument.
Typical multimeter measurements volts, ohms, and amps are made with
a basic VDC accuracy of up to 0.015%, ensuring you have the performance
you need for your design. You can also use the DMM4020 to measure
frequency, and to perform continuity and diode tests. This allows you
to replace your counter, continuity tester, and DMM with one versatile
instrument, saving bench space and cost.
Measure Nanoampere Signals
Measuring standby current in today’s energy-efcient designs requires
looking at very low currents, often in the microamp or nanoamp range.
Using a traditional multimeter for this measurement can lead to inaccurate
results since they typically employ the shunt resistance technique for
measuring current.
The DMM4020 offers an improved method for measuring low currents. By
using a current-to-voltage op amp technique, the DMM4020 can make
current measurements with 1 nA of resolution and with minimal loading
affect on the circuit under test, giving results that reect real-world operation
of the device.
Designed to Make Your Work Easier
The DMM4020 multimeter is designed with the ease-of-use and familiar
operation you expect from Tektronix.
Intuitive Operation
Dedicated front-panel buttons provide fast access to frequently used
functions and parameters, reducing setup time. You no longer need to
search through software menus to nd the function you need.
Limit compare mode on the DMM4020.
Setup Buttons for Your Common Measurements
With six setup buttons on the front panel, you can save the settings for your
most common measurements. Simply set up the meter for a measurement,
then press shift followed by a setup button to save the measurement
settings. Now each time you perform that measurement, you simply press
the appropriate setup key.
Dual Display
With the unique dual display, you can measure two different parameters of
the same signal from one test connection.
Limit Compare
The limit compare mode provides pass/fail indicators to quickly show if a
test passes or fails to help eliminate mistakes, especially for results that
areclosetothelimit.
Simple and Accurate 4-wire Measurements
Patented split terminal jacks for the 2×4 ohms function allow you to perform
4-wire measurements using only two leads instead of four. Special test lead
accessories are available to enable you to establish the connection. You get
excellent resolution and accuracy plus the convenience and ease of using a
single pair of leads.
Easy Connectivity to Your PC
The RS-232 port on the back panel can be used to connect to your PC.
AUSBtoRS
-232 interface adapter cable is included standard with the
DMM4020 for connecting to your PC’s USB port.
2www.tektronix.com
Digital Multimeters Tektronix DMM4020
Signal Express acquiring data from Tektronix DMM4020 and DPO3052.
Connect Your Bench for Intelligent Debug
Easily capture, save, and analyze measurement results from your
multimeter with the special Tektronix Edition of National Instrument’s
LabVIEW SignalExpress™ software. Every DMM4020 multimeter ships
with a free copy of the Limited Edition version of SignalExpress for basic
instrument control, data logging, and analysis. The optional Professional
Edition offers over 200 built-in functions that provide additional signal
processing, advanced analysis, sweeping, limit testing, and user-dened
step capabilities.
SignalExpress supports the range of Tektronix bench instruments*1,
enabling you to connect your entire test bench. You can then access the
Signal Express used to export DMM4020 data into Excel.
feature-rich tools packed into each instrument from one intuitive software
interface. This allows you to automate complex measurements requiring
multiple instruments, log data for an extended period of time, time-correlate
data from multiple instruments, and easily capture and analyze your results,
all from your PC. Only Tektronix offers a connected test bench of intelligent
instruments to simplify and speed debug of your complex design.
Performance You Can Count On
In addition to industry-leading service and support, every DMM4020
multimeter comes backed with a three-year standard warranty.
*1NI LabVIEW Signal Express supports the following Tektronix instruments: MSO/DPO4000/3000/2000
Series oscilloscopes, TDS3000C/2000B/1000B Series oscilloscopes, AFG3000 Series arbitrary/function
generators, DMM4050/4040/4020 Series digital multimeters.
www.tektronix.com 3
Data Sheet
Characteristics
General Specications
Voltage
Characteristic Description
100V Setting 90 V to 110 V
120V Setting 108 V to 132 V
220V Setting 198 V to 242 V
240V Setting 216Vto264V
Frequency 47 Hz to 440 Hz
Power Consumption 15 VA peak (10 W average)
Dimensions
Dimension mm in.
Height 88 3.46
Width 217 8.56
Depth 297 11.7
Weight kg lb.
Net 2.1 4.6
Display
Vacuum Fluorescent Display, segment
Environment
Characteristic Description
Temperature
Operating 0 °C to 50 °C
Storage -40 °C to 70 °C
Warm Up ½ hour to full uncertainty specications
Relative Humidity (noncondensing)
Operating Uncontrolled (<10 °C)
<90%(1Cto2C)
<75%(2Cto4C)
<45%(4Cto5C)
Storage <95% (-40 °C to 70 °C)
Altitude
Operating 2,000 meters
Storage 12,000 meters
Vibration Complies with MIL-PRF-28800F Class 3
Safety Complies with IEC 61010-1:2001, ANSI/ISA 61010-1
(S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2
No. 61010.1:2004, CAT I 1000V / CAT II 600 V
EMC DesignedtocomplywithIEC
61326-1:1997+A1:1998+A2:2000
Triggering
Characteristic Description
Trigger Delay 400 ms
External Trigger Delay <2 ms
External Trigger Jitter <1 ms
Trigger Input TTL Levels
Trigger Output 5Vmax
Math Functions
Min/max, relative, hold, compare, and dB functions
Electrical
Characteristic Description
Input Protection 1000 V all ranges
Overrange 10% on the largest ranges of all functions except
continuity and diode test
Remote Interfaces
RS-232C (RS-232 to USB Adapter cable included)
Warranty
Three years
4www.tektronix.com
Digital Multimeters Tektronix DMM4020
Electrical Specications
DC Voltage Specications
Specications are valid for digit mode and after at least a half-hour warm-up.
Characteristic Description
Maximum Input 1000 V on any range
Common Mode
Rejection
120 dB at 50 or 60 Hz 0.1% (1 kunbalance)
Normal Mode Rejection 80 dB at Slow Rate
A/D Nonlinearity 15 ppm of range
Input Bias Current <30 pA at 25 °C
Settling Considerations Measurement settling times are affected by source
impedance, cable dielectric characteristics, and input
signal changes
Input Characteristics
ResolutionRange Full Scale
(5½ Digits) Slow Medium Fast
Input Impedance
200 mV 199.999 mV 1 V 10 V10V>10 G*2
2 V 1.99999 V 10 V100V100V>10 G*2
20 V 19.9999 V 100 V1000V 1000 V10 M±1%
200 V 199.999 V 1 mV 10 mV 10 mV 10 M±1%
1000 V 1000.00 V 10 mV 100 mV 100 mV 10 M±1%
*2At some dual-display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M.
Accuracy
Uncertainty*3
90 days 1 year
Range
23 °C ±5 °C 23 °C ±5 °C
Temperature Coefcient/°C
Outside 18 28 °C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005
2 V 0.01 + 0.002 0.015 + 0.003 0.001 + 0.0005
20 V 0.01 + 0.003 0.015 + 0.004 0.0020 + 0.0005
200 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
1000 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
*3Uncertainty given as ±(% of reading + % of range).
www.tektronix.com 5
Data Sheet
AC Voltage Specications
AC Voltage specications are for AC sinewave signals >5% of range. For inputs from
1% to 5% of range and <50 kHz, add an additional error of 0.1% of range, and for
50 kHz to 100 kHz, add 0.13% of range.
Characteristic Description
Maximum Input 750 VRMS or 1000 V peak or 8 × 107Volts-Hertz product
Measurement Method AC-coupled true RMS. Measures the AC component of
input with up to 1000 VDC bias on any range
AC Filter Bandwidth 20 Hz 100 kHz
Common Mode
Rejection
60 dB at 50 Hz or 60 Hz (1 kunbalance)
Maximum Crest Factor 3:1 at Full Scale
Additional Crest Factor
Errors (<100 Hz)
Crest Factor 1-2, 0.05% of full scale
CrestFactor2-3,0.2%offullscale
Only applies for non-sinusoid signals
Input Characteristics
ResolutionRange Full Scale
(5½ Digits) Slow Medium Fast
Input Impedance
200 mV 199.999 mV 1 V10V10V
2 V 1.99999 V 10 V100V100V
20 V 19.9999 V 100 V1000V 1000 V
200 V 199.999 V 1 mV 10 mV 10 mV
750 V 750.00 V 10 mV 100 mV 100 mV
1M±2% shunted by
<100 pf
Accuracy
Uncertainty*3
90 days1 year
Range Frequency
23 °C ±5 °C 23 °C ±5 °C
Temperature Coefcient/°C
Outside18–2C
20 Hz 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz–50kHz 0.3 + 0.05 0.35 +0.05 0.01 + 0.005
200 mV
50 kHz–100kHz 0.8 + 0.05 0.9 + 0.05 0.05 +0.01
20 Hz –45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz –20kHz 0.15 +0.05 0.2 + 0.05 0.01 + 0.005
20 kHz–50kHz 0.3 + 0.05 0.35 +0.05 0.01 + 0.005
2V
50 kHz 100 kHz 0.8 +0.05 0.9 +0.05 0.05+0.01
20 Hz–45Hz 0.8 +0.05 0.9 +0.05 0.01 + 0.005
45 Hz–20kHz 0.15 + 0.05 0.2 +0.05 0.01 + 0.005
20 kHz–50kHz 0.3 +0.05 0.35 + 0.05 0.01 + 0.005
20 V
50 kHz 100 kHz 0.8 +0.05 0.9 +0.05 0.05+0.01
20 Hz–45Hz 0.8 +0.05 0.9 +0.05 0.01 + 0.005
45 Hz 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
200 V
50 kHz 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
20 Hz–45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20kHz–50kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
750V
50kHz–100kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
*3Uncertainty given as ±(% of reading + % of range).
6www.tektronix.com
Digital Multimeters Tektronix DMM4020
Resistance
Specications are for 4-wire resistance function, or 2-wire resistance with REL. If
REL is not used, add 0.2 for 2-wire resistance plus lead resistance.
Characteristic Description
Measurement Method Current source referenced to LO input
Max Lead Resistance
(4-wire ohms)
10% of range per lead for 200 ,2kranges. 1 kper
lead on all other ranges.
Input Protection 1000 V on all ranges
Input Characteristics
ResolutionRange Full Scale
(5½ Digits) Slow Medium Fast
Current Source
200 199.999 0.001 0.01 0.01 0.8 mA
2k1.99999 k0.01 0.1 0.1 0.8 mA
20 k19.9999 k0.1 110.08 mA
200 k199.999 k110 10 0.008 mA
2M1.99999 M10 100 100 0.9 A
20 M19.9999 M100 1k1k0.16 A
100 M100.000 M1k10 k10 k0.16 A||10M
Accuracy
Uncertainty*3
90 days 1 year
Range
23 °C ±5 °C23 °C ±5 °C
Temperature Coefcient/°C
Outside 18–2C
200 0.02 + 0.004 0.03 + 0.004 0.003 + 0.0006
2k0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
20 k0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
200 k0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
2M0.03 + 0.003 0.04 + 0.004 0.004 + 0.0005
20 M0.2 + 0.003 0.25 + 0.003 0.01 + 0.0005
100 M1.5 + 0.004 1.75 + 0.004 0.2 + 0.0005
*3Uncertainty given as ±(% of reading + % of range).
www.tektronix.com 7
Data Sheet
DC Current
Characteristic Description
Input Protection Tool-accessible 11 A / 1000 V and 440 mA / 1000 V
fuses
Shunt Resistance 0.01 for 2 A and 10 A ranges
1for 20 mA and200mA
Burden voltage <5 mV for 200 A and 2 mA range
Input Characteristics
ResolutionRange Full Scale
(5½ Digits) Slow Medium Fast
Burden Voltage
200 A 199.999 A0.001A0.01A0.01A<5mV
2 mA 1999.99 A0.01A0.1A0.1A<5mV
20 mA 19.9999 mA 0.1 A1A1A <0.05 V
200 mA 19.9999 mA 1 A10A10A<0.5V
2 A 1.99999 A 10 A100A100A<0.1V
10 A 10.0000 A 100 A1mA 1mA <0.5V
Accuracy
Uncertainty*3
90 days 1 year
Range
23 °C ±5 °C 23 °C ±5 °C
Temperature Coefcient/°C
Outside 18 28 °C
200 A 0.02 + 0.005 0.03 + 0.005 0.003 + 0.001
2 mA 0.015 + 0.005 0.02 + 0.005 0.002 + 0.001
20 mA 0.03 + 0.02 0.04 + 0.02 0.005 + 0.001
200 mA 0.02 + 0.005 0.03 + 0.008 0.005 + 0.001
2 A 0.05 + 0.02 0.08 + 0.02 0.008 + 0.001
10 A 0.18 + 0.01 0.2 + 0.01 0.008 + 0.001
*3Uncertainty given as ±(% of reading + % of range).
8www.tektronix.com
Digital Multimeters Tektronix DMM4020
AC Current
The following AC current specications are for sinusoidal signals with amplitudes
greater than 5% of range. For inputs from 1% to 5% of range, add an additional
error of 0.1% of range.
Characteristic Description
Input Protection Tool-accessible 11 A / 1000 V and 440 mA / 1000 V
fuses
Measurement Method AC-coupled true RMS
Shunt Resistance 0.01 for 2 A and 10 A ranges
1for20mAand200mA
AC Filter Bandwidth 20 Hz 100 kHz
Maximum Crest Factor 3:1 at Full Scale
Additional Crest Factor
Errors (<100 Hz)
Crest Factor 1-2, 0.05% of full scale
CrestFactor2-3,0.2%offullscale
Only applies to non-sinusoid signals
Input Characteristics
ResolutionRange Full Scale
(5½ Digits) Slow Medium Fast
Burden Voltage
20 mA 19.9999 mA 0.1 A1A1A <0.05 V
200 mA 199.999 mA 1 A10A10A<0.5V
2 A 1.99999 A 10 A100A100A<0.1V
10 A 10.0000 A 100 A1mA 1mA <0.5V
Accuracy
Uncertainty*3
90 days 1 year
Range Frequency
23 °C ±5 °C 23 °C ±5 °C
Temperature Coefcient/°C
Outside18–2C
20 Hz - 45 Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.00520 mA
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
20 Hz - 45 Hz 0.8 + 0.05 1 + 0.06 0.015 + 0.005200 mA
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
20 Hz - 45 Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.0052A
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
20 Hz - 45 Hz 1 + 0.1 1.25 + 0.12 0.015 + 0.00510 A
45 Hz - 2 kHz 1 + 0.1 0.5 + 0.12 0.015 + 0.005
*3Uncertainty given as ±(% of reading + % of range).
www.tektronix.com 9
Data Sheet
Frequency
Characteristic Description
Gate Time 131 ms
Measurement
Method
AC-coupled input using the AC voltage
measurement function
Settling
Considerations
When measuring frequency after a DC offset
voltage change, errors may occur. For the most
accurate measurement, wait up to 1 second to allow
input-blocking RC time constant to settle
Measurement
Considerations
To minimize measurement errors, shield inputs
from external noise when measuring low-voltage,
low-frequency signals
Input Characteristics
Uncertainty
90 days 1 year
Range Frequency
23 °C ±5 °C 23 °C ±5 °C
Temperature Coefcient/°C
Outside18–2C
20 Hz 2 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
2 kHz 20 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
20 kHz 200 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
100mVto750V*
4, 5
200 kHz 1 MHz 0.01 + 0.004 0.01 + 0.006 0.002 + 0.002
*4Input >100 mV.
*5Limitedto8×10
7VHz.
Continuity
Characteristic Description
Continuity Threshold 20
Test Current 1mA
Response Time 100 S/s with audible tone
Rate Fast
Maximum Reading 199.99
Resolution 0.01
Diode Test
Characteristic Description
Response Time 100 S/s with audible tone
Rate Fast
Maximum Reading 1.9999 V
Resolution 0.1 mV
10 www.tektronix.com
Digital Multimeters Tektronix DMM4020
Ordering Information
Models
Model Description
DMM4020 5.5 Digit Multimeter
DMM4020 Includes: Meter, TL710 test leads, line cord, spare line fuse, statement
of cal practices, Warranty statement, Safety and Installation Guide, Connectivity
Installation Manual, CD-ROM with user manual (English, French, Italian, German,
Spanish, Simplied Chinese, Traditional Chinese, Korean, Russian, Japanese),
RS-232 to USB Adapter Cable, National Instruments LabVIEW SignalExpress™
Tektronix Edition, Limited Edition Software.
Please specify power plug when ordering.
Instrument Options
Power Plug Options
Option Description
Opt. A0 North America
Opt. A1 Universal Euro
Opt. A2 United Kingdom
Opt. A3 Australia
Opt. A5 Switzerland
Opt. A6 Japan
Opt. A10 China
Opt. A11 India
Opt. A99 No power cord
Opt. E1 Euro and UK power cords
Service Options*6
Option Description
Opt. CA1 Provides a single calibration event or coverage for the
designated calibration interval, whichever comes rst
Opt. C3 Calibration Service 3 Years
Opt. C5 Calibration Service 5 Years
Opt. D1 Calibration Data Report
Opt. R5 Repair Service 5 Years (including warranty)
*6Test Leads and accessories are not covered by the DMM warranty and Service Offerings. Refer to the
datasheet of each Test Lead and accessory model for its unique warranty and calibration terms.
Recommended Accessories and Software
Accessory Description
Calibration Manual 077-0365-xx
TL710 Premium Test Leads (196-3250-xx)
TL705 2×4 Wire Ohm 1000 V Precision Test Lead
TL725 2×4 Wire Ohm SMD Test Tweezers
AC4000 Soft Transit Case
HCTEK4321 Hard Carrying Case
Y8846S Single Rackmount Kit
Y8846D Dual Rackmount Kit
013-0369-xx Calibration Fixture 4-terminal short
SIGEXPTE NI LabVIEW SignalExpress Tektronix Edition Software
Full Version
Product(s) are manufactured in ISO registered facilities.
Product(s) complies with IEEE Standard 488.1-1987 and RS-232C.
www.tektronix.com 11
Data Sheet Contact Tektronix:
ASEAN / Australasia (65) 6356 3900
Austria +41 52 675 3777
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Belgium 07 81 60166
Brazil +55 (11) 3759-7627
Canada 1 (800) 661-5625
Central East Europe, Ukraine, and the Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Denmark +4580881401
Finland +41526
75 3777
France +33(0)169868181
Germany +49 (221) 94 77 400
Hong Kong (852) 2585-6688
India (91) 80-42922600
Italy +39 (02) 25086 1
Japan 81 (3) 6714-3010
Luxembourg +44 (0) 1344 392400
Mexico, Central/South America & Caribbean 52 (55) 54247900
Middle East, Asia, and North Africa +41 52 675 3777
The Netherlands 090 02 021797
Norway 800 16098
People’s Republic of China 86 (10) 6235 1230
Poland +41 52 675 3777
Portugal 80 08 12370
Republic of Korea 82 (2) 6917-5000
Russia & CIS +7 (495) 7484900
South Africa +27 11 206 8360
Spain (+34) 901 988 054
Sweden 020 08 80371
Switzerland +41 52 675 3777
Taiwan 886 (2) 2722-9622
United Kingdom & Ireland +44 (0) 1344 392400
USA 1 (800) 426-2200
For other areas contact Tektronix, Inc at: 1 (503) 627-7111
Updated 5 August 2009
For Further Information. Tektronix maintains a comprehensive, constantly expanding
collection of application notes, technical briefs and other resources to help engineers working
on the cutting edge of technology. Please visit www.tektronix.com
Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents,
issued and pending. Information in this publication supersedes that in all previously published material.
Specication and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of
Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks
of their respective companies.
08 Sep 2009 3MW-24431-0
www.tektronix.com