60W AC-DC Reliable Green Medical Adaptor G S M 6 0 A series wD AWi.; CB FE CE FAI (@for5-9v) ANSI/AAMI ES60601-1 IEC60601-1 @ Features @ Applications * 3 pole AC inlet IEC320-C14, Class I power unit * Mobile clinical workstation * Medical safety approved (2 x MOPP) according to * Oral irrigator ANSI/AAMI ES60601-1 and IEC/EN60601-1 * Portable hemodialysis machine * Extremely low leakage current * Breath Machine * No load power consumption<0.1W * Energy efficiency level VI and meet CoC Version 5 (Except 5~9V for Level V) * -30~+70C wide range working temperature * Protections: Short circuit / Overload / Over voltage/ Over temperature * LED indicator for power on * Lifetime > 105 K hours * 3 years warranty * Medical computer monitor i Description GSME6O0A is a highly reliable, 60W desktop style single-output green medical adaptor series. This product is a class [ power unit(with FG),equipped with a standard IEC320-C14 AC inlet and adopting the input range from 80VAC to 264VAC. The entire series supplies different models with output voltages between 5VDC and 48VDC that can satisfy the demands for various types of medical electrical devices. The circuitry design meets the international medical standards (2*MOPP), having an ultra low leakage current (<100#A), fitting the medical devices in direct electrical contact with the patients. With the efficiency up to 91% and the extremely low no-load power consumption below 0. 1W, GSM60A is compliant with USA EISA 2007/DoE , Canada NRCan, Australia and New Zealand MEPS , EU ErP, and meet Code of Conduct(CoC) Version 5.The supreme feature allows the adaptor to save the energy when it is either under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case. GSM6O0A is certified for the international medical safety regulations. Mi Model Encoding P1J: Standard model,2.1x5.5x11mm,C+,tuning fork type DC plug type Other options available by customer requested (see Page 4~5) Output voltage IEC320-C14 AC inlet Rated wattage Series name File Name:GSM60A-SPEC 2018-11-21MW 60W AC-DC Reliable Green Medical Adaptor GSM60A series SPECIFICATION ORDER NO. GSM60A05-P1J | GSM60A07-P1J | GSM60A09-P1J | GSM60A12-P1J | GSM60A15-P1J | GSM60A18-P1J | GSM60A24-P1J | GSM60A48-P1J SAFETY MODEL NO. GSM60A05_ |GSM60A07 | GSM60A09_ |GSM60A12 |GSM60A15 |GSM60A18 | GSM60A24 GSM60A48 DC VOLTAGE Note.2) 5V 7.5V ov 12V 15V 18V 24V 48V RATED CURRENT 6A 6A 6A 5A 4A 3.33A 2.5A 1.25A CURRENT RANGE 0.1~6A 0.1~6A 0.1~6A 0.1~5A 0.1~4A 0.1~3.33A |0.1~2.5A 0.1 ~ 1.25A RATED POWER (max.) 30W 45W 54W 60W 60W 60W 60W 60W OUTPUT | RIPPLE & NOISE (max.) Note.3/ 80mVp-p 80mVp-p 100mVp-p 100mVp-p 100mVp-p 120mVp-p 150mVp-p 240mVp-p VOLTAGE TOLERANCE Note.4| +5.0% +5.0% +5.0% +3.0% +3.0% +3.0% +3.0% +2.5% LINE REGULATION Note.5| +1.0% +1.0% +1.0% +1.0% +1.0% +1.0% +1.0% +1.0% LOAD REGULATION +5.0% +5.0% +5.0% +3.0% +3.0% +3.0% +3.0% +2.5% SETUP, RISE TIME Note.6/ 1000ms, 30ms / 230VAC 1500ms, 30ms /115VAC at full load HOLD UP TIME (Typ.) 50ms / 230VAC 18ms/115VAC at full load VOLTAGE RANGE Note.7| 80~ 264VAC 113 ~370VDC FREQUENCY RANGE 47 ~ 63Hz INPUT EFFICIENCY (Typ.) 81.5% 86% 87.5% 88% 88.5% 89% 90.5% 91.5% AC CURRENT (Typ.) 1.4A/ 115VAC 1A/230VAC INRUSH CURRENT (Typ.) Cold start 30A/115VAC 60A/ 230VAC LEAKAGE CURRENT(max.) | Earth leakage current < 100 A/264VAC , Touch current < 100 uA/264VAC OVERLOAD 105 ~ 160% rated output power __ PROTECTION Protection type : Hiccup mode, recovers automatically after fault condition is removed 5.2~7.0V | 7.8 ~10.2V | 9.4~12.2V | 12.6~16.2V |15.7~20.3V |18.9~243V |25.2~32.4V |50.4~64.8V OVER VOLTAGE - Protection type : Shut down o/p voltage, re-power on to recover OVER TEMPERATURE Shut down o/p voltage, re-power on to recover WORKING TEMP. -30~+70C (Refer to "Derating Curve") WORKING HUMIDITY 20% ~ 90% RH non-condensing ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85C, 10 ~ 95% RH non-condensing TEMP. COEFFICIENT +0.03% /C (0~40C) VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes OPERATING ALTITUDE Note.8} 3000 meters SAFETY STANDARDS IEC60601-1, TUV EN60601-1, ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22.2 No. 60601-1:14 - Edition 3, EAC TP TC 004 approved ISOLATION LEVEL Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP WITHSTAND VOLTAGE I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:SHORT ISOLATION RESISTANCE /P-O/P, /P-FG:100M Ohms / 500VDC / 25C /70% RH Parameter Standard Test Level / Note . EN55011 (CISPR11), FCC PART 15/ CISPR22, Conducted emission CAN ICES-3(B)/NMB-3(B) Class B EMC EMISSION Radiated emission CAN ILS 3(B)NMB-3(8) PART 15 / CISPR22, Class B Harmonic current EN61000-3-2 Class A SAFETY & Voltage flicker EN61000-3-8 00 nee EMC EN55024 , EN60601-1-2, EN61204-3 (Note 9) Parameter Standard Test Level / Note ESD EN61000-4-2 Level 4, 15KV air ; Level 4, 8KV contact Level 3, 10V/m( 80MHz~2.7GHz ) RF field susceptibility EN61000-4-3 Table 9, 9~28V/m( 385MHz~5.78GHz ) EFT bursts EN61000-4-4 Level 3, 2KV EMC IMMUNITY Surge susceptibility EN61000-4-5 Level 3, 1KV/Line-Line , 2KV/Line-FG Conducted susceptibility EN61000-4-6 Level 3, 10V Magnetic field immunity EN61000-4-8 Level 4, 30A/m Voltage dip, interruption EN61000-4-11 toon iteruptions 350 erode Periods, MTBF 720K hrs min. MIL-HDBK-217F(25C_) OTHERS | DIMENSION 125*50*31.5mm (L*W*H) PACKING 0.32Kg; 40pces/ 13.8Kg/1.05CUFT PLUG See page 4~5 ; Other type available by customer requested CONNECTOR CABLE See page 4~5 ; Other type available by customer requested NOTE 1. All parameters are specified at 230VAC input, rated load, 25C 70% RH ambient. 2. DC voltage: The output voltage set at point measure by plug terminal & 50% load. 3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47f capacitor. 4. Tolerance: includes set up tolerance, line regulation, load regulation. 5. Line regulation is measured from low line to high line at rated load. 6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time. 7. Derating may be needed under low input voltages. Pleas check the derating curve for more details. 8. The ambient temperature derating of 3.5C/1000m with fanless models and of 5C/1000m with fan models for operating altitude higher than 2000m(6500ft), 9. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http:/Avww.meanwell.com) File Name:GSM60A-SPEC 2018-11-21MW 60W AC-DC Reliable Green Medical Adaptor GSM60A series Hi Derating Curve 100 70 LOAD (%) 80 |o-- nanan 50 AMBIENT TEMPERATURE (C) Hi Mechanical Specification 125 60 70 (HORIZONTAL) Power LED AC Inlet IEC320-C14 BH DC output plug Standard plug: P1J LOAD (%) Static Characteristics 80 70 5 60 50 40 80 90 UL2464 16AWG 1000+50mm for 5 ~ 15V UL1185 16AWG 1500+50mm for 18 ~ 48V 100 110 120 140 160 180 200 220 240 264 INPUT VOLTAGE (VAC) 60Hz CaseNo.GS60A_ Unit:mm 7010mm Pi Pin Assignment 5.5 1140.5mm -c# =, coy 2.1 Outside SG Inside AC FG -V connected to AC FG File Name:GSM60A-SPEC 2018-11-21MW 60W AC-DC Reliable Green Medical Adaptor G S M 6 0 A series Optional DC plug: Tuning Fork Style Type No. SD . ; Cc Pl 5.5 2.1 9.5 cyt P4L 5.5 25 | 9.5 A (Straight) P1M 5.5 2.5 11.0 iy @, C P1IR 5.5 21 | 9.5 P1JR 5.5 2.1 11.0 P1LR 5.5 2.5 9.5 (Right-angled) | P1MR 5.5 25 | 11.0 A B C Barrel Style Type No. y He OD ID L C P2l 5.5 2.1 9.5 P2J 5.5 2.1 11.0 A | . P2L 55 25 | 9.5 ba (Straight) P2M 5.5 25 | 11.0 ~ , Cc P2IR 5.5 2.1 9.5 P2JR 5.5 2.1 11.0 P2LR 5.5 2.5 9.5 (Right-angled) | B2MR 5.5 25 | 11.0 A B C Lock Style Type No. OD ID L A Locking P2S(S761K) 5.53 2.03 12.06 >. ra ies P2K(761K) 5.53 2.54 | 12.06 % Os P2C(S760K) | 5.53 | 2.03 | 9.52 SWITCHCRAFT original or equivalent P2D(760k) 5.53 2.54 9.52 A B C D Type No. Center Pin Style yp OD ID L Icomerrin A -&., P4A 5.5 | 3.4 | 11.0 | 1.0 6} @ coe Mi] | P4B 65 | 44] 11.0] 14 EIAJ equivalent P4C 7.4 5.1 | 11.0 | 0.6 . oo. Pin Assignment Min. DIN 3 Pin with Lock (male Type No. (male) uP PIN No. Output 5 1 +Vo Wr 12 ttt ll el [4 R6B 2 -Vo . 2 KYCON KPPX-3P equivalent 3 +Vo File Name:GSM60A-SPEC 2018-11-2160W AC-DC Reliable Green Medical Adaptor G S M 6 0 A series Min. DIN 4 Pin with Lock (mal Type No rin Assignment in. in with Lock (male) y PIN No. Output z 1 +Vo (oo\y23 mrt ld [4 2 Vo Wr; @) i ll R7B : we KYCON KPPX-4P equivalent 4 +Vo Min. DIN 4 Pin with Lock (female Type No See in nw ( ) "| PINNo. Output 1 +Vo 2 -Vo OO) 23 2 camel |] | eer rr KYCON KPJX-CM-4S equivalent 4 +Vo . Pin Assignment DIN 5 Pin (male) Type No. PIN No. Output 1 -Vo 3 TT R1B 2 a 4 428) LITE = 3 +Vo 4 -Vo 5 +Vo . . Pin Assignment Stripped and tinned leads Type No. PIN No. Output L___ (red) aa 1 1 +Vo CT 9 re ae ae L1 (black) | by customer Length of Land L1 by request 2 -Vo (MW's standard length, L:_25 mm, L1:_5 mm) Installation Manual Please refer to : http://www.meanwell.com/manual.html File Name:GSM60A-SPEC 2018-11-21