20 MHz DIGITAL STORAGE/ANALOG
OSCILLOSCOPE
INSTRUCTION
MANUAL
Model 2522C
TEST INSTRUMENT SAFETY
WARNING
Normal use of test equipment exposes you to a certain amount of danger from electrical shock because testing
mu st of ten be p erf orme d whe re e xpo sed h igh vol tage is p res ent . An ele ctri cal shoc k ca usi ng 10 milliamps of current
to pass through the heart will stop most human heartbeats. Voltage as low as 35 volts dc or ac rms should be
considered dangerous and hazardous since it can produce a lethal current under certain conditions. Higher
voltage poses an even greater threat because such voltage can more easily produce a lethal current. Your normal
work habits should include all accepted practices that will prevent contact with exposed high voltage, and that will
steer current away from your heart in case of accidental contact with a high voltage. You will significantly reduce
the risk factor if you know and observe the following safety precautions:
1. Don’t expose high voltage needlessly in the equipment under test. Remove housings and covers only when necessary. Turn
off equipment while making test connections in high-voltage circuits. Discharge high-voltage capacitors after removing
power.
2. If possible, familiarize yourself with the equipment being tested and the location of its high voltage points. However,
remember that high voltage may appear at unexpected points in defective equipment.
3. Use an insulated floor material or a large, insulated floor mat to stand on, and an insulated work surface on which to place
equipment; make certain such surfaces are not damp or wet.
4. Use the time-proven “one hand in the pocket” technique while handling an instrument probe. Be particularly careful to avoid
contacting a nearby metal object that could provide a good ground return path.
5. When using a probe, touch only the insulated portion. Never touch the exposed tip portion.
6. When testing ac powered equipment, remember that ac line voltage is usually present on some power input circuits such as the
on-off switch, fuses, power transformer, etc. any time the equipment is connected to an ac outlet, even if the equipment is
turned off.
7. Some equipment with a two-wire ac power cord, including some with polarized power plugs, is the “hot chassis” type. This
includes most recent television receivers and audio equipment. A plastic or wooden cabinet insulates the chassis to protect the
customer. When the cabinet is removed for servicing, a serious shock hazard exists if the chassis is touched. Not only does this
present a dangerous shock hazard, but damage to test instruments or the equipment under test may result from connecting the
ground lead of most test instruments (including this oscilloscope) to a “hot chassis”. To make measurements in “hot chassis”
equipment, always connect an isolation transformer between the ac outlet and the equipment under test. The B&K Precision
Model TR-1 10 or 1604A Isolation Transformer, or Model 1653A or 1655A AC Power Supply is suitable for most applications.
To be on the safe side, treat all two wire ac powered equipment as “hot chassis” unless you are sure it has an isolated chassis or
an earth ground chassis.
8. Never work alone. Someone should be nearby to render aid if necessary. Training in CPR (cardio-pulmonary resuscitation)
first aid is highly recommended
3
TABLE OF CONTENTS
TEST INSTRUMENT SAFETY..........................................................................................................2
TABLE OF CONTENTS....................................................................................................................3
FEATURES.......................................................................................................................................4
SPECIFICATIONS ............................................................................................................................6
CONTROLS AND INDICATORS ......................................................................................................8
GENERAL FUNCTION CONTROLS ............................................................................................................8
VERTICAL CONTROLS ...............................................................................................................................8
HORIZONTAL CONTROLS........................................................................................................................10
TRIGGERING CONTROLS........................................................................................................................10
DIGITAL STORAGE CONTROLS ..............................................................................................................12
REAR PANEL CONTROLS........................................................................................................................12
OPERATING INSTRUCTIONS .......................................................................................................13
SAFETY PRECAUTIONS...........................................................................................................................13
EQUIPMENT PROTECTION PRECAUTIONS.........................................................................................13
OPERATING TIPS......................................................................................................................................14
INITIAL STARTING PROCEDURE.............................................................................................................14
SINGLE TRACE DISPLAY .........................................................................................................................14
DUAL TRACE DISPLAY.............................................................................................................................14
TRIGGERING.............................................................................................................................................15
MAGNIFIED SWEEP OPERATION............................................................................................................17
X Y OPERATION ......................................................................................................................................17
VIDEO SIGNAL OBSERVATION................................................................................................................17
DIGITAL STORAGE OPERATION .............................................................................................................17
MAINTENANCE..............................................................................................................................20
FUSE REPLACEMENT ..............................................................................................................................20
LINE VOLTAGE SELECTION.....................................................................................................................20
PERIODIC ADJUSTMENTS.......................................................................................................................20
CALIBRATION CHECK ..............................................................................................................................21
INSTRUMENT REPAIR SERVICE.............................................................................................................21
APPENDIX I....................................................................................................................................22
IMPORTANT CONSIDERATIONS FOR RISE TIME AND FALL TIME MEASUREMENTS.......................22
APPENDIX II...................................................................................................................................23
UNIQUE CHARACTERISTICS OF DIGITAL STORAGE OSCILLSCOPES...............................................23
ALIASING ...................................................................................................................................................23
EQUIVALENT TIME SAMPLING................................................................................................................23
SERVICE INFORMATION...............................................................................................................25
LIMITED ONE-YEAR WARRANTY ................................................................................................25
4
FEATURES
LOW COST, HIGH PERFORMANCE
B&K Precision’s Model 2522C is one of the lowest cost
digital storage oscilloscopes in the industry, yet it includes all
the basic features needed by most technicians and engineers.
Unlike other digital oscilloscopes, it has a full analog mode with
infinite resolution. Digital storage modes include refresh, roll,
and single sweep. Equivalent time storage techniques allow
storage of repetitive waveforms up to 20 MHz. This
oscilloscope is built by and backed by B&K Precision, a
company that has been selling reliable, durable, value priced
test instruments for over 50 years.
DIGITAL STORAGE FEATURES
High Resolution
2048 samples across the width of the screen (200 samples
per division) assures good resolution of stored waveforms.
1024 samples taken at sweep rates of 10 µs/div to 0.1
µs/div.
High Sampling Ra te
Sampling Rate up to 40 Ms/s (Megasamples per
second) for storing one-time events down to 50 nS.
Slow Event Display
X100 time base selections extend sweep time to 50
seconds/division. Excellent for displaying slow speed
events.
Equivalent Time Sampling
Allows repetitive waveforms up to 20 MHz to be stored.
Pretrigger Capture
Permits viewing activity before the trigger occurs.
Selectable 0%, 25%, 50%, or 75% pre-trigger view.
CRT FEAT URES
Rectangular CRT
Rectangular CRT with large 8 x 10 centimeter viewing
area.
Convenience
Trace rotation electrically adjustable from front panel. 0%,
10%, 90%, and 100% markers for rise time measurements.
DUAL TRACE FEATURES
Dual Trace
Model 2522C has two vertical input channels for dis-
playing two waveforms simultaneously. Selectable single
trace (either CH 1 or CH 2) or dual trace. Alternate or
chop sweep selectable at all sweep rates.
Sum and Difference Capability
Permits algebraic addition or subtraction of channel 1 and
channel 2 waveforms, displayed as a single trace. Useful
for differential voltage and distortion measurements.
HIGH FREQUENCY FEATURES
Wide Bandwidth
Conservatively-rated 3 dB bandwidth is dc to 20 MHz.
Fast Rise Time
Rise time is less than 18 nS.
Fast Sweep
Maximum sweep speed of 10 nS/div (with X10 MAG)
assures high frequencies and short-duration pulses are
displayed with high resolution.
VERTICAL FEATURES
High Sensitivity
5 mV/div sensitivity for full bandwidth. High-sensitivity
1 mV/div and 2 mV/div using PULL X5 gain control.
Calibrated Voltage Measurements
Accurate voltage measurements (±3%) on 10 calibrated
ranges from 5 mV/div to 5 V/div. Vertical gain fully
adjustable between calibrated ranges.
SWEEP FE ATURES
Calibrated Time Measurements
Accurate (±3%) time measurements. The main sweep has
21 calibrated ranges from 0.5 s/div to 0.1 S/div. Sweep
time is fully adjustable between calibrated ranges.
X10 Sweep Magnification
Allows closer examination of waveforms, increases
maximum sweep rate to 10 nS/div.
5
FEATU RES
TRIGGERING FEATURES
Two Trigger Modes
Selectable normal (triggered) or automatic sweep
modes.
Triggered Sweep
Sweep remains at rest unless adequate trigger signal
is applied. Fully adjustable trigger level and (+) or ()
slope.
AUTO Sweep
Selectable AUTO sweep provides sweep without
trigger input, automatically reverts to triggered
sweep operation when adequate trigger is applied.
Five Trigger Sources
Five trigger source selections, including CH 1, CH 2,
alternate, EXT, and LINE.
Video Sync
Frame (TV V) or Line (TV H) triggering selectable
for observing composite video waveforms. TV-H
position can also be used as low frequency reject
and TV-V position can be used as high frequency
reject.
Variable Holdoff
Trigger inhibit period after end of sweep adjustable. Permits
stable observation of complex pulse trains.
OTHER FEATURES
XY Operation
Channel 1 can be applied as horizontal deflection (X-axis)
while channel 2 provides vertical deflection (Y-axis).
Built-in Probe Adjust Square Wave
A 0.5 V p-p, 1 kHz square wave generator permits probe
compensation adjustment.
Channel 2 (Y) Output
A buffered 50 output of the channel 2 signal is available at
the rear panel for driving a frequency counter or other
instruments. The output is 50 mV/div (nominal) into 50 .
Supplied With Two Probes
6
SPECIFICATIONS
CRT
Type: 6-inch rectangular with integral graticule, P31
phosphor.
Display Area: 8 x 10 div (1 div = 1 cm).
Accelerating Voltage: 2 kV.
Phosphor: P31.
Trace Rotation: Electrical, front panel adjustable.
VERTICAL AMPLIFIERS (CH 1 and CH 2)
Sensitivity: 5 mV/div to 5 V/div, 1 mv/div to 1 V/div at X5
MAG.
Attenuator: 10 calibrated steps in 1-2-5 sequence. Vernier
control provides fully adjustable sensitivity between steps;
range 1/1 to at least 1/3.
Accuracy: 3%, 5 mV to 5 V/div; 5%, at X5 MAG.
Input Resistance: 1 M2%.
Input Capacitance: 25 pF ±10 pF.
Frequency Response:
5 mV/div to 5 V/div:
DC to 20 MHz ( - 3 dB).
X5 MAG:
DC to 10 MHz (- 3 dB).
Rise Time:
18 nS; 35 nS atX5 MAG.
Overshoot: Less than 5%.
Operating Modes:
CH 1:
CH 2:
DUAL:
ADD:
CH 1, single trace.
CH 2, single trace.
CH 1 and CH 2, dual trace.
Alternate or Chop selectable at
any sweep rate.
Algebraic sum of CH 1 + CH 2.
Chop Frequency: Approximately 500 kHz.
Polarity Reversal: CH 2 invert.
Maximum Input Voltage: 400 V (dc + ac peak).
HORIZONTAL AMPLIFIER
(Input through channel 1 input)
XY mode:
CH 1 = X axis.
CH2 = Y axis.
Sensitivity: Same as vertical channel 2.
Input Impedance: Same as vertical channel 2.
Frequency Response: DC to 1
MHz (-3 dB).
X-Y Phase Difference: 3° or less at 50 kHz. Maximum
Input Voltage: Same as vertical channel 1.
SWEEP SYSTEM
Time Base: 0.1 µS/div to 0.5 S/div in 1-2-5 sequence, 21
steps. Vernier control provides fully adjustable sweep
time between steps.
Accuracy: ±3%, except ±20% on 0.1 S/div.
Sweep Magnification: X10 ± 10%.
Holdoff: Continuously adjustable for main time base from
NORM to 5 times normal.
TRIGGERING
Trigger Modes:
AUTO (free run), NORM, TV-V, TV-H.
Trigger Source:
CH 1, CH 2, Alternate, EXT, LINE.
Slope:
(+) or (–).
Trigger Coupling:
AUTO: Sweep free-runs in absence of
suitable trigger signal.
NORM: Sweep triggered only by adequate
trigger signal.
TV-V: Video vertical sync pulses are selected.
Also usable for high frequency
reject.
TV-H: Video horizontal sync pulses are
selected. Also usable for low
frequency reject.
7
Trigger Sensitivity:
AC: 1.0 div (internal)
0.5 Vp-p (external)
30 Hz – 20 MHz
TV-V: 1.0 div (internal)
0.5 Vp-p (external)
20 Hz – 20 kHz
TV-H: 1.0 div (internal)
0.5 Vp-p (external)
3 kHz – 100 kHz
Maximum External Trigger Voltage: 300 V (dc +
ac peak).
DIGITAL STORAGE FACILITIES
Storage Word Size:
2048 x 8 bits/channel (2 k/channel with direct
sampling, 1 k/channel with equivalent time
sampling).
Vertical Resolution:
8 bit (1 in 256) approximately 25 steps/div.
Horizontal Resolution:
1 in 2048, approximately 200 samples/div.
Sampling Rate:
10 M samples/sec to 4 samples/sec, reduced in propor-
tion to time base. Direct sampling at time base settings
of 20 s/div and slower, equivalent time sampling at
time base settings of 10 µs/div and faster.
Time Base Expander:
For storage of slow time events, time base steps 10
ms/div and slower have selectable 1/1 or 1/100 rate.
1/100 rate expands time base from 1 sec/div to 50
sec/div in 1-2-5 sequence.
Equivalent Time Sampling Bandwidth:
20 MHz for repetitive waveforms.
Dot Joining:
Linear interpolation between samples.
Digital Display Modes:
Roll:
Stored data and display updated continually.
Refresh:
Stored data and display updated by triggered sweep.
Hold:
Freezes channel 1 and channel 2 data immediately.
Save CH2:
Freezes channel 2 data immediately.
Pretrigger Storage:
Available in single sweep mode, switchable to 0%,
25%, 50%, or 75%.
USB Host Interface
Save screen image to USB flash drive
OTHER SPECIFICATIONS
Cal/Probe Compensation Voltage: 0.5 V p-p ± 3% square wave, 1
kHz nominal.
CH 2 (Y) Output:
Output Voltage: 50 mV/div (nominal into 50 ohm load).
Output Impedance: Approximately 50 ohms.
Frequency Response: 20 Hz to 20 MHz, - 3 dB.
Power Requirements: 100–130 VAC or 200–260 VAC, 5 0/60 Hz,
55 watts.
Dimensions (H 3 W 3 D): 5.2” X 12.8” X 15.7”
(132 X 324 X 398 mm).
Weight: 18.7 lbs (8.5 kg).
Environment:
Within Specified Accuracy: +10° to +35° C, 10–80% relative
humidity.
Full Operation: 0° to +50° C, 10–80% relative humidity.
Storage: -30° to +70° C, 10–90% relative humidity.
ACCESSORIES SUPPLIED:
Two Switchable X1/X10 Probes.
Instruction Manual.
AC Line Cord.
8
CONTROLS AND INDICATORS
Fig. 1. Model 2522C Controls and Indicators.
GENERAL FUNCTION CONTROLS
X-Y:
used in conjunction with the
X-Y
control and
1. ON Indicator. Lights when oscilloscope is “on”. Trigger SOURCE switch to enable X-Y display
mode.
2. POWER Pushbutton. Turns oscilloscope “on” and
“off”. DUAL:
Displays the channel 1 and channel 2 signals simul-
3. INTENSITY Control. Adjusts brightness of trace. taneously. Dual-trace mode may be either alternate
or chopped sweep; see the description under
CHan
-
4. TRACE ROTATION Control. Adjusts to maintain
trace at a horizontal position. nel 1 POSition/PULL CHOP control.
ADD:
5. FOCUS Control. Adjusts trace focus. The inputs from channel 1 and channel 2 are
6. GND Terminal. Oscilloscope chassis ground
j
ack, and earth ground via three-wire ac power cord.
summed and displayed as a single signal. If the
Channel 2 POSition/PULL INVert control is
pulled out, the input from channel 2 is subtracted
7. CAL Terminal. Terminal provides 0.5 V p-p, 1 kHz
(nominal) square wave signal. This signal is useful for
checking probe compensation adjustment, as
well as
from channel 1 and the difference is displayed as a
single signal.
providing a rough check of vertical calibration. 9. CH1 AC-GND- D C Switch. Three-position lever
switch with the followin
g
p
ositions:
VERTICAL CONTROLS AC:
8. VERTical MODE Switch. Selects vertical display
mode. Four-position lever switch with the following
positions:
CH1:
Displays the channel 1 signal by itself.
CH2/X-Y:
CH2: displays the channel 2 signal by itself.
Channel 1 input signal is capacitively coupled; dc
component is blocked.
GND:
Opens signal path and grounds input to vertical
amplifier. This provides a zero-volt base line, the
position of which can be used as a reference when
performing dc measurements.
9
Fig. 2. Rear Panel Contr ol s and In dicators.
DC:
Direct coupling of channel 1 input signal; both ac
and dc components of signal produce vertical de-
flection.
10. CH1 ( X) Input Jack. Vertical input for channel 1.
X-axis input for X-Y operation.
11. CH1 (X) VOLTS/DIV Control. Vertical
attenuator for channel 1. Provides step adjustment of
vertical sensitivity. When channel 1 VARiable
control is set to CAL, vertical sensitivity is
calibrated in 10 steps from 5 mV/div to 5 V/div in a
1-2-5 sequence. When the X-Y mode of operation is
selected, this control provides step adjustment of
X-axis sensitivity.
12. CH1 VARia ble/PULL X5 MAG Con trol:
VARiable:
Rotation provides vernier adjustment of channel 1
vertical sensitivity. In the fully-clockwise (CAL)
position, the vertical attenuator is calibrated. Coun-
terclockwise rotation decreases gain sensitivity.
In X-Y operation, this control becomes the
vernier X-axis sensitivity control.
PULL X5 MAG:
When pulled out, increases vertical sensitivity by a
factor of five. Effectively provides two extra sensi-
tivity settings: 2 mV/div and 1 mV/div. In X-Y
mode, increases X-sensitivity by a factor of five.
13. CH1 POSition/PULL CHOP Control:
POSition:
Adjusts vertical position of channel 1 trace.
PULL CHOP:
When this switch is pulled out in the dual-trace mode, the
channel 1 and channel 2 sweeps are chopped and displayed
simultaneously (normally used at slower sweep speeds).
When it is pushed in, the two sweeps are alternately displayed,
one after the other (normally used at higher sweep speeds).
14. CH2 POSition/PULL INVert Control:
POSition:
Adjusts vertical position of channel 2 trace. In X-Y operation,
rotation adjusts vertical position of X-Y display.
PULL INVert:
When pushed in, the polarity of the channel 2 signal is normal.
When pulled out, the polarity of the channel 2 signal is
reversed, thus inverting the waveform.
15. CH2 VOLTS/DIV Control. Vertical attenuator for channel 2.
Provides step adjustment of vertical sensitivity. When channel
2 VARiable control is set to CAL, vertical sensitivity is
calibrated in 10 steps from 5 mV/div to 5 V/div in a 1-2-5
sequence. When the X-Y mode of operation is selected, this
control provides step adjustment of Y-axis sensitivity.
16. CH2 VARia ble/PULL X5 MAG Con trol:
VARiable:
Rotation provides vernier adjustment of channel 2 vertical
sensitivity. In the fully-clockwise (CAL) position, the
vertical attenuator is calibrated. Counterclockwise rotation
decreases gain sensitivity. In X-Y operation, this control
becomes the vernier Y-axis sensitivity control.
39 40
41 4243
17.
PULL X5 MAG:
When pulled out, increases vertical sensitivity by a
factor of five. Effectively provides two extra sensitivity
settings: 2 mV/div and 1 mV/div. In X-Y mode, increases
Y-sensitivity by a factor of five.
CH2 (Y) Input Jack. Vertical input for channel 2.
PULL NORM TRIG:
When pushed in, selects AUTOmatic triggering
mode. In this mode, the oscilloscope generates
sweep (free runs) in absence of an adequate trigger;
it automatically reverts to triggered sweep operation
when an adequate trigger signal is present. On the
18.
Y-axis input for X-Y operation.
CH2 AC-GND- DC Switch. Three-position lever Model 2522C, automatic triggering is applicable to
both the main sweep and delayed sweep.
switch with the following positions:
AC:
Channel 2 input signal is capacitively coupled; dc 24.
When pulled out, selects
NORM
al
triggered sweep
operation. A sweep is generated only when an
adequate trigger signal is present.
Trigger SOURCE Switch. Selects source of sweep
component is blocked.
GND:
Opens signal path and grounds input to vertical
amplifier. This provides a zero-volt base line, the
position of which can be used as a reference when
performing dc measurements.
DC:
Direct coupling of channel 2 input signal; both ac
and dc components of signal produce vertical deflection.
HORIZONTAL CONTROLS
trigger. Four-position lever switch with the following
positions:
CH1
Causes the channel 1 input signal to become the
sweep trigger, regardless of the
VERTical MODE
switch setting.
CH2:
The channel 2 signal becomes the sweep trigger,
regardless of the
VERTical MODE
switch setting.
ALTernate:
Selects alternate triggering, used in dual-trace
19.
Time Base TIME/DI V Co ntrol .
Provides step selec- mode, permits each waveform viewed to become its
20.
tion of sweep rate for the time base. When the VARi-
able Sweep control is set to CAL, sweep rate is
calibrated. This control has 21 steps, from 0.1 S/div
to 0.5 S/div, in a 1-2-5 sequence.
VARiable Sweep Control.
Rotation of control is ver-
own trigger source. Must be used with alternate
sweep, cannot be used with chop sweep.
EXT:
Signal from EXTernal TRIGger jack becomes
sweep trigger.
25. Trigger COUPLING Switch. Selects trigger cou-
21.
nier adjustment for sweep rate. In fully clockwise
(CAL) position, sweep rate is calibrated.
POSition/PULL X10 MAG Control.
pling. Four-position lever switch with the following
positions:
22.
POSition:
Horizontal (X) position control.
PULL X10 MAG:
Selects ten times sweep magnification when pulled
out, normal when pushed in. Increases maximum
sweep rate to 10 nS/div.
X-Y Switch. Used with the VERTical MODE switch
AC:
Trigger signal is capacitively coupled. Used for
most waveforms except video.
TV-V:
Used for triggering from television vertical sync
pulses. Also serves as lo-pass/dc (high frequency
reject) trigger coupling.
TV-H:
and Trigger SOURCE switch to select X-Y operating
mode. The channel 1 input becomes the X-axis and the
channel 2 input becomes the Y-axis. Trigger source
and coupling are disabled in this mode.
TRIGGERING CONTROLS
Used for triggering from television horizontal sync
pulses. Also serves as hi-pass (low frequency reject)
trigger coupling.
LINE:
Signal derived from input line voltage (5 0/60 Hz)
becomes trigger.
23. HOLDOFF/PULL NORM TRIG Control. 26. TRIGger LEVEL/PULL (–) SLOPE Control.
11
HOLDOFF:
Rotation adjusts holdoff time (trigger inhibit period
beyond sweep duration). When control is rotated
fully counterclockwise, the holdoff period is MIN-
inum (normal). The holdoff period increases pro-
gressively with clockwise rotation.
TRIGger LEVEL:
Trigger level adjustment; determines the point on
the triggering waveform where the sweep is trig-
gered. Rotation in the (–) direction (counterclock-
wise) selects more negative triggering point;
rotation in the (+) direction (clockwise) selects
more positive triggering point.
CONTROLS AND INDICATORS
12
PULL (—) SLOPE:
Two-position push-pull switch. The “in” position
selects a positive-going slope and the “out”
position selects a negative-going slope as
triggering point for main sweep.
27. EXTern al TRIGger Jack. External trigger input
for single- and dual-trace operation.
DIGITAL STORAGE CONTROLS
28. STORAGE/ANALOG Pushbutton. When
released, scope is in ANALOG mode. When
engaged, scope is in digital STORAGE mode. In
ANALOG mode, all digital mode controls (29–3 8)
are disabled. In STORAGE mode, all digital mode
controls (29–38) are enabled.
In STORAGE mode with all digital mode controls
(29–38) released, scope is in REFRESH mode of
storage, in which it updates the stored waveform
each time an adequate trigger signal is received (if
the previous update has been completed).
REFRESH mode is applicable at all TIME/DIV
settings.
29. SAVE CH2 Pushbutton. Engaging this switch
freezes and stores the channel 2 trace immediately.
The channel 2 display and memory cannot be
updated until this switch is released.
30. SAVE ALL Pushbutton. Engaging this switch
freezes and stores the channel 1 and channel 2
traces immediately. The display and memory cannot
be up-dated until this switch is released.
31. SLOW X100 Pushbutton. Engaging this switch
expands the time base settings of 10 ms/div to 0.5
S/div by a factor of 100 to become settings of 1
sec/div to 50 sec/div. Time base settings are normal
when this switch is released.
32. ROLL Pushbutton. Engaging this switch selects
the ROLL mode of operation, in which the trace
moves across the CRT from right to left like a strip
chart recorder (opposite of conventional oscilloscope
operation) and the display is continuously updated.
This update takes place even in the absence of any
trigger signal. The rolling speed equals the
TIME/DIV setting (and SLOW X100 if engaged).
ROLL mode is applicable only at time base settings
of 10 ms/div to 0.5 S/div.
33. SINGLE Pushbutton. Engaging this switch
enables the single sweep mode of digital storage
operation, in which the memory is filled in a single
sweep and continuously displayed. Single sweep
mode is applicable for all “real time” settings of the
TIME/DIV control, that is, 20 s/div and slower.
34. RESET Pushbutton. With the SINGLE switch
engaged, this switch readies the scope for single
sweep
operation. A suitable trigger signal arriving after
pressing the RESET switch initiates the single
sweep.
35. READY Indicator. Lights when the RESET switch readies
the scope for single sweep, and goes out when trigger signal
occurs.
36. PRE-TRIGGER Pushbuttons. Selects post-trigger or
pre-trigger storage conditions. In post-trigger condition (0%
pre-trigger), the entire stored waveform occurs after the trigger.
The trigger point is at the extreme left of the display, as in a
conventional analog oscilloscope display. In pre-trigger
conditions, a portion of the waveform occurring before the
trigger is displayed. Pre-trigger selections of 25%, 50%, and
75% are selectable. For 0% pre-trigger, release both
PRE-TRIGGER pushbuttons. For 25% pre-trigger, engage the
left pushbutton. For 50% pre-trigger, en-gage the right
pushbutton. For 75% pre-trigger, engage both pushbuttons.
Pre-trigger operation is applicable only to single sweep
operation, not to ROLL or REFRESH modes.
37. Data Out Sends data to USB BMP.
38. Transfer Indicator. When flashing, indicates waveform being
stored to USB drive.
REAR PANEL CONTROLS (see Fig. 2)
39. Fuse Holder/Line Voltage Selector. Contains fuse and selects
line voltage.
40. Power Cord Receptacle.
41. USB Host connector. Store screen data to USB flash drive.
42. Analog Output. Output terminal where sample of channel 2
analog signal is available. Amplitude of output is 50 millivolts
per division of vertical deflection seen on CRT when
terminated into 50 ohms. Output impedance is 50 ohms.
43. Factory use only Not used.
13
OPERATING INSTRUCTIONS
SAFETY PRECAUTIONS
WARNING
The following precauti o ns m ust be ob served
to help prevent electric shock.
1. When the oscilloscope is used to make measurements in
equipment that contains high voltage, there is al-ways a
certain amount of danger from electrical shock. The person
using the oscilloscope in such conditions should be a
qualified electronics technician or other-wise trained and
qualified to work in such circumstances. Observe the
TEST INSTRUMENT SAFETY recommendations listed
on the inside front cover of this manual.
2. Do not operate this oscilloscope with the case removed
unless you are a qualified service technician. High voltage
up to 2100 volts is present when the unit is operating with
the case removed.
3. The ground wire of the 3-wire ac power plug places the
chassis and housing of the oscilloscope at earth ground.
Use only a 3-wire outlet, and do not attempt to defeat the
ground wire connection or float the oscilloscope; to do so
may pose a great safety hazard.
4. Special precautions are required to measure or observe
line voltage waveforms with any oscilloscope. Use the
following procedure:
a. Do not connect the ground clip of the probe to either
side of the line. The clip is already at earth ground and
touching it to the hot side of the line may “weld” or
“disintegrate” the probe tip and cause possible injury,
plus possible damage to the scope or probe.
b. Insert the probe tip into one side of the line voltage
receptacle, then the other. One side of the receptacle
should be “hot” and produce the waveform. The other
side of the receptacle is the ac return and no waveform
should result.
EQUIPMENT PROTECTION PRECAUTIONS
The following precautions will help avoid
damage to the oscilloscope.
1. Never allow a small spot of high brilliance to remain
stationary on the screen for more than a few seconds. The
screen may become permanently burned. A spot will
occur when the scope is set up for X Y operation and no
signal is applied. Either reduce the intensity so the spot is
barely visible, apply signal, or switch back to normal
sweep operation. It is also advisable to use low intensity
with AUTO triggering and no signal applied for long
periods. A high intensity trace at the same position could
cause a line to become permanently burned onto the
screen.
2. Do not obstruct the ventilating holes in the case, as this
will increase the scope’s internal temperature.
3. Excessive voltage applied to the input jacks may
dam-age the oscilloscope. The maximum ratings of the
inputs are as follows:
CH 1 and CH2:
400 V dc + ac peak.
EXT TRIG:
300 V dc + ac peak.
Z-AXIS INPUT:
30 V ( dc and ac peak).
4. Always connect a cable from the ground
terminal of the oscilloscope to the chassis of the
equipment under test. Without this precaution, the entire
current for the equipment under test may be drawn
through the probe clip leads under certain circumstances.
Such conditions could also pose a safety hazard, which the
ground cable will prevent.
5. The probe ground clips are at oscilloscope and earth
ground and should be connected only to the earth ground
or isolated common of the equipment under test. To
measure with respect to any point other than the common,
use CH 2 – CH 1 subtract operation (ADD mode and
invert channel 2), with the channel 2 probe to the point of
measurement and the channel 1 probe to the point of
reference. Use this method even if the reference point is a
dc voltage with no signal.
OPERATING INSTRUCTIONS
14
OPERATING TIPS
The following recommendations will help obtain the best
performance from the oscilloscope.
1. Always use the probe ground clips for best results,
attached to a circuit ground point near the point of
measurement. Do not rely solely on an external
ground wire in lieu of the probe ground clips as
undesired signals may be introduced.
2. Avoid the following operating conditions:
a. Direct sunlight.
b. High temperature and humidity.
c. Mechanical vibration.
d. Electrical noise and strong magnetic fields, such
as near large motors, power supplies,
transformers, etc.
3. Occasionally check trace rotation, probe compensation,
and calibration accuracy of the oscilloscope using the
procedures found in the MAINTENANCE section of
this manual.
4. Terminate the output of a signal generator into its
characteristic impedance to minimize ringing, espe-
cially if the signal has fast edges such as square waves
or pulses. For example, the typical 50 output of a
square wave generator should be terminated into an
external 50 terminating load and connected to the
oscilloscope with 50 coaxial cable.
5. Probe compensation adjustment matches the probe to
the input of the scope. For best results, compensation
should be adjusted initially, then the same probe
al-ways used with the same channel. Probe
compensation should be readjusted when a probe
from a different oscilloscope is used.
INITIAL STARTING PROCEDURE
Until you familiarize yourself with the use of all controls,
the settings given here can be used as a reference point to
obtain a trace on the CRT in preparation for waveform
observation.
1. Set these controls as follows:
VERTical MODE to CH1.
CH1 AC/GND/DC to GND.
Select AUTO triggering (HOLD OFF pushed in)
Trigger COUPLING to AC. Trigger SOURCE
to CH1.
All POSition controls and INTENSITY control
centered (pointers facing up). Time Base control to 1
mS/div.
2. Press the red POWER pushbutton.
3. A trace should appear on the CRT. Adjust the trace brightness
with the INTENSITY control, and the trace sharpness with
the FOCUS control.
SINGLE TRACE DISPLAY
Either channel 1 or channel 2 may be used for single-trace operation.
To observe a waveform on channel 1:
1. Perform the steps of the “Initial Starting Procedure”.
2. Connect the probe to the CH 1 (X) input jack.
3. Connect the probe ground clip to the chassis or common of the
equipment under test. Connect the probe tip to the point of
measurement.
4. Move the CH1 AC/GND/DC switch out of the GND position
to either DC or AC.
5. If no waveforms appear, increase the sensitivity by turning the
CH 1 VOLTS/DIV control clockwise to a position that gives 2
to 6 divisions vertical deflection.
6. Position the waveform vertically as desired using the CH1
POSition control.
7. The display on the CRT may be unsynchronized. Refer to the
“Triggering” paragraphs in this section for procedures on
setting triggering and sweep time controls to obtain a stable
display showing the desired number of waveforms.
DUAL TRACE DISPLAY
In observing simultaneous waveforms on channel 1 and 2, the
waveforms are usually related in frequency, or one of the waveforms
is synchronized to the other, although the basic frequencies are
different. To observe two such related waveforms simultaneously,
perform the following:
1. Connect probes to both the CH 1 (X) and CH 2 (Y) input
jacks.
2. Connect the ground clips of the probes to the chassis or
common of the equipment under test. Connect the tips of the
probes to the two points in the circuit where waveforms are to be
measured.
3. To view both waveforms simultaneously, set the VERTical
MODE switch to DUAL and select either ALT (alternate) or
CHOP with the PULL CHOP switch.
4. In the ALT sweep mode (PULL CHOP switch pushed in),
one sweep displays the channel 1 signal and the next sweep
displays the channel 2 signal in an alternating sequence.
Alternate sweep is normally used for viewing high-frequency
or high-speed wave-forms at sweep times of 1 ms/div and faster,
but may be selected at any sweep time.
15
5. In the CHOP sweep mode (PULL CHOP switch pulled
out), the sweep is chopped (switched) between channel 1
and channel 2. Using CHOP, one channel does not have
to “wait” for a complete swept display of the other channel.
Therefore, portions ofboth channel’s waveforms are
displayed with the phase relation-ship between the two
waveforms unaltered. Chop sweep is normally used for
low-frequency or low-speed waveforms at sweep times of
1 ms/div and slower; or where the phase relationship
between channel 1 and channel 2 requires measurement.
If chop sweep is used at sweep times of 0.2 ms/div and
faster, the chop rate becomes a significant portion of the
sweep and may become visible in the displayed waveform.
However, you may select chop sweep at any sweep time
for special applications.
6. Adjust the channel 1 and 2 s t POSition controls to place
the channel 1 trace above the channel 2 trace.
7. Set the CH 1 and CH 2 VOLTS/DIV controls to a
position that gives 2 to 3 divisions of vertical deflection
for each trace. If the display on the screen is
unsynchronized, refer to the “Triggering” paragraphs in this
section of the manual for procedures for setting triggering
and sweep time controls to obtain a stable display
showing the desired number of waveforms.
8. When the VERTical MODE switch is set to ADD, the
algebraic sum of CH 1 + CH 2 is displayed as a single trace.
When the PULL INV switch is pulled out, the algebraic
difference of CH 1 – CH 2 is displayed.
9. If two waveforms have no phase or frequency
relation-ship, there is seldom reason to observe both
wave-forms simultaneously. However, these oscilloscopes
do permit the simultaneous viewing of two such unrelated
waveforms, using alternate triggering. Refer to the
paragraphs on “Triggering — Trigger SOURCE Switch”
for details on alternate triggering.
TRIGGERING
The Model 2522C Oscilloscope provides versatility in sync
triggering for ability to obtain a stable, jitter-free display in
single-trace, or dual-trace operation. The proper settings
depend upon the type of waveforms being observed and the type
of measurement desired. An explanation of the various controls
which affect synchronization is given to help you select the
proper setting over a wide range of conditions.
AUTO or NORM Triggering
1. In the AUTO mode (PULL NORM TRIG pushed in),
automatic sweep operation is selected. In automatic sweep
operation, the sweep generator free-runs to generate a
sweep without a trigger signal. However, it automatically
switches to triggered sweep operation if an acceptable
trigger source signal is present. The AUTO mode is handy
when first setting up the scope to observe a waveform; it
provides sweep for wave-
form observation until other controls can be properly set.
Once the controls are set, operation is often switched back
to the normal triggering mode, since it is more sensitive.
Automatic sweep must be used for dc measurements and
signals of such low amplitude that they will not trigger
the sweep.
2. In the NORM mode (PULL NORM TRIG pulled out),
normal triggered sweep operation is selected. The sweep
remains at rest until the selected trigger source signal
crosses the threshold level set by the TRIG LEVEL
control. The trigger causes one sweep to be generated,
after which the sweep again remains at rest until triggered.
In the normal triggering mode, there will be no trace
unless an adequate trigger signal is present. In the ALT
VERTICAL MODE of dual-trace operation with the
SOURCE switch also set to ALT, there will be no trace
unless both channel 1 and channel 2 signals are adequate
for triggering. Typically, signals that produce even one
division of vertical deflection are adequate for normal
triggered sweep operation.
Trigger COUPLING Switch
1. The AC position is used for most waveforms except
video. The trigger signal is capacitively coupled. Thus, it
blocks the dc component and references the average of the
“changing” portion of the waveform.
2. The TV H and TV V positions are primarily for viewing
composite video waveforms. Horizontal sync pulses are
selected as trigger when the trigger COUPLING switch is
set to the TV H position, and vertical sync pulses are
selected as trigger when the trigger COUPLING switch
is set to the TV V position. The TV H and TV V
positions may also be used as low frequency reject and
high frequency reject coupling, respectively. Additional
procedures for observing video waveforms are given later
in this section of the manual.
Trigger SOURCE Switch
The trigger SOURCE switch (CH 1, CH 2, etc.) selects the
signal to be used as the sync trigger.
1. If the SOURCE switch is set to CH 1 (or CH 2) the
channel 1 (or channel 2) signal becomes the trigger
source regardless of the VERTICAL MODE selection.
CH 1, or CH 2 are often used as the trigger source for
phase or timing comparison measurements.
2. By setting the SOURCE switch to ALT, alternating
triggering mode is activated. In this mode, the trigger
source alternates between CH 1 and CH 2 with each
sweep. This is convenient for checking amplitudes,
waveshape, or waveform period measurements, and even
permits simultaneous observation of two wave-forms
which are not related in frequency or period. However,
this setting is not suitable for phase or timing comparison
measurements. For such measurements, both traces must
be triggered by the same sync signal. Alternate triggering
can only be used in dual-trace
16
mode (VERT MODE set to DUAL), and with
alternate sweep only (PULL CHOP not engaged).
3. In the EXT position, the signal applied to the EXT
TRIG jack becomes the trigger source. This signal
must have a timing relationship to the displayed
wave-forms for a synchronized display.
4. In the LINE position of the COUPLING switch,
triggering is derived from the input line voltage (5
0/60 Hz) and the trigger SOURCE switch is disabled.
This is useful for measurements that are related to
line frequency.
TRIG LEVEL/PULL (–) SLOPE Control
(Refer to Fig. 3)
A sweep trigger is developed when the trigger source
signal crosses a preset threshold level. Rotation of the
TRIG LEVEL control varies the threshold level. In the +
direction (clockwise), the triggering threshold shifts to a
more positive value, and in the – direction
(counterclockwise), the triggering threshold shifts to a
more negative value. When the control is centered, the
threshold level is set at the approximate average of the
signal used as the triggering source. Proper adjustment of
this control usually synchronizes the display.
The TRIG LEVEL control adjusts the start of the sweep
to almost any desired point on a waveform. On sine wave
signals, the phase at which sweep begins is variable.
Note that if the TRIG LEVEL control is rotated toward
its extreme + or – setting, no sweep will be developed in
the normal trigger mode because the triggering threshold
exceeds the peak amplitude of the sync signal.
When the PULL (–) SLOPE control is set to the +
(“in”) position, the sweep is developed from the trigger
source waveform as it crosses a threshold level in a
positive-going direction. When the PULL (–) SLOPE
control is set to the
– (“out”) position, a sweep trigger is developed from the
trigger source waveform as it crosses the threshold level in
a negative-going direction.
TIME BASE Control
Set the Time Base TIME/DIV control to display the desired
number of cycles of the waveform. If there are too many cycles
displayed for good resolution, switch to a faster sweep time. If only a
line is displayed, try a slower sweep time. When the sweep time is
faster than the waveform being observed, only part of it will be
displayed, which may appear as a straight line for a square wave or
pulse waveform.
HOLDOFF Control
(Refer to Fig. 4)
A “holdoff” period occurs immediately after the completion of
each sweep, and is a period during which triggering of the next sweep
is inhibited. The normal holdoff period varies with sweep rate, but is
adequate to assure complete retrace and stabilization before the next
sweep trigger is permitted. The HOLDOFF control allows this
period to be extended by a variable amount if desired.
This control is usually set to the MIN position (fully
counterclockwise) because no additional holdoff period is necessary.
The HOLDOFF control is useful when a complex series of pulses
appear periodically such as in Fig. 4B. Improper sync may produce a
double image as in Fig. 4A. Such a display could be synchronized
with the VAR SWEEP control, but this is impractical because time
measurements are then uncalibrated. An alternate method of
synchronizing the display is with the HOLDOFF control. The
sweep speed remains the same, but the triggering of the next sweep is
“held off” for the duration selected by the HOLDOFF control. Turn
the HOLDOFF control clock-wise from the MIN position until the
sweep starts at the same point of the waveform each time.
Fig. 3. Function o f Slope and Level Controls. Fig. 4. Use of HOLDOFF Con trol.
OPERATING INSTRUCTIONS
17
MAGNIFIED SWEEP OPERATION
Since merely shortening the sweep time to magnify a portion
of an observed waveform can result in the desired portion
disappearing off the screen, magnified display should be
performed using magnified sweep.
Using the POSition control, move the desired portion of
waveform to the center of the CRT. Pull out the PULL X10 knob
to magnify the display ten times. For this type of display, the sweep
time is the Time Base TIME/DIV control setting
divided by 10. Rotation of the POSition control can then be
used to select the desired portion of the waveforms.
X Y OPERATION
XY operation permits the oscilloscope to perform many
measurements not possible with conventional sweep operation.
The CRT display becomes an electronic graph of two
instantaneous voltages. The display may be a direct comparison
of the two voltages such as stereoscope display of stereo signal
outputs. However, the XY mode can be used to graph almost
any dynamic characteristic if a transducer is used to change the
characteristic (frequency, temperature, velocity, etc.) into a
voltage. One common application is frequency response
measurements, where theY axis corresponds to signal amplitude
and the X axis corresponds to frequency.
1. Depress the XY switch and set the Trigger Source and
VERTical MODE switches to XY.
2. In this mode, channel 1 becomes the X axis input and
channel 2 becomes the Y axis input. The X and Y
positions are now adjusted using the POSition and
the channel 2 POSition controls, respectively.
3. Adjust the amount of vertical (Y axis) deflection with the
CH 2 VOLTS/DIV and VARIABLE controls.
4. Adjust the amount of horizontal (X axis) deflection with
the CH 1 VOLTS/DIV and VARIABLE controls.
VIDEO SIGNAL OBSERVATION
Setting the COUPLING switch to the TV-H or TV-V
position permits selection of horizontal or vertical sync pulses
for sweep triggering when viewing composite video waveforms.
When the TV-H mode is selected, horizontal sync pulses are
selected as triggers to permit viewing of horizontal lines of video.
A sweep time of about 10 µs/div is appropriate for displaying
lines of video. The VAR SWEEP control can be set to display
the exact number of waveforms desired.
When the TV-V mode is selected, vertical sync pulses are
selected as triggers to permit viewing of vertical fields and
frames of video. A sweep time of 2 ms/div is appropriate for
viewing fields of video and 5 ms/div for complete frames (two
interlaced fields) of video.
At most points of measurement, a composite video signal is
of the () polarity, that is, the sync pulses are negative and the
video is positive. In this case, use () SLOPE. If the waveform
is taken at a circuit point where the video wave-form is inverted,
the sync pulses are positive and the video is negative. In this
case, use (+) SLOPE.
DIGITAL STORAGE OPERATION
Digitizing Repetitive Waveforms
While the most powerful feature of a digital storage
oscilloscope (DSO) is its ability to capture one-time events, a
DSO can also digitize conventional repetitive waveforms, such
as those observed on a standard analog scope.
An analog/digital unit such as the Model 2522C allows the
user to set up sensitivity, sweep, and triggering in a familiar
analog setting, and then switch to the digital mode. This is a
good way to gain familiarity with the digital features of the
instrument. You may wish to pursue this method as described
below in the “Refresh Mode” paragraph.
Also, some repetitive signals can be viewed and measured
much more effectively in the digital store mode than in the
analog mode. One example is slow signals below 60 Hz.
Signals that appear as a flickering waveform or just a moving dot
on the screen in the analog mode are displayed as a bright,
non-flickering, easily viewed entire waveform in the digital
store mode.
The digital store mode also improves viewing and meas-
urement of repetitive signals with low trigger repetition rates
relative to the sweep rate. In the analog mode, such
wave-forms may be too faint for viewing. In the digital store
mode, the display is equivalent to a CRT with infinite
persistence. The waveform can be easily viewed.
Refresh Mode
In the refresh mode, the waveform is digitized and displayed
on the screen. After the complete waveform is digitized, the
next trigger causes the stored waveform to be replaced
(refreshed) by a new waveform.
1. Set up the oscilloscope to view a periodic waveform in
the ANALOG mode. Adjust the trigger, VOLTS/DIV,
sweep TIME/DIV, and vertical position controls as
desired.
2. Initially, disengage all of the digital storage control
switches (the group of switches beneath the CRT).
3. Select digital storage operation by setting the ANA-
LOG/STORAGE switch to STORAGE (engaged).
When all digital storage control switches are released, the
“refresh” mode is selected. The waveform should appear
on the display, relatively unchanged from the previously
displayed analog version. In the “refresh” mode, the
display is continually updated as long as a suitable trigger
signal remains present.
18
4. Once a waveform is digitized, it can be stored in
long-term memory by pressing the SAVE ALL or
SAVE CH2 switches. The SAVE CH2 switch
immediately stores the channel 2 waveform. The
SAVE ALL switch immediately freezes the display
and stores both waveforms. Once a waveform is
stored by engaging either of these switches, it will be
stored until the switch is disengaged or the power is
turned off. NOTES
Any of the operating modes previously dis-
cussed in analog operation (e.g. DUAL, ADD,
etc.) can be used in digital mode, except for
X–Y operation.
Time base settings of 10 µs/div and faster result
in a display that is acquired through “equivalent
time sampling”. This process develops the
digitized image over many cycles of the
repetitive signal; therefore, these higher sweep
speeds cannot be used for capture of one-time
events. Equivalent time sampling is discussed
in detail in
Appendix II, “Unique Characteristics of Digital
Storage Oscilloscopes”.
Digitizing One-Time Events
One of the most powerful features of a digital storage
oscilloscope (DSO) is its ability to capture one-time
events. To capture one-time events, single-sweep
operation is employed. This is done through the use of the
SINGLE switch. When pushed, this switch releases the
REFRESH or ROLL mode if previously engaged, and
readies the digital storage circuit to receive a trigger signal
— presumably the event to be captured or some other
time-related occurrence. When the event arrives, it is
stored in the memory and displayed. The procedure is as
follows:
1. Set the oscilloscope to run in analog mode (ANA-
LOG/DIGITAL switch to ANALOG). Select
NORM triggering, and adjust the TRIG LEVEL
control so that the unit triggers on the event to be
captured.
2. Set the oscilloscope to digital (STORAGE) mode by
engaging the ANALOG/STORAGE switch.
3. Initially, set the PRE TRIG switch to 50%. A more
detailed discussion of pre- and post-triggering is
given later in this section of the manual.
4. Press the RESET switch. The READY indicator
will momentarily light as the scope awaits the arrival
of the trigger signal. When that trigger occurs, the
READY indicator will go off, and the event being
monitored will roll to the center of the display and
stop. NOTES
Depending on the sweep TIME/DIV
setting, the anticipated event may roll to the
center of the screen very rapidly, or very slowly,
after the READY indicator goes out.
Single-sweep mode can be used in the presence of very rapidly
occurring events, even continuous waveforms if desired. In that
case, the period between pushing the RE-SET switch and the
arrival of the trigger may be very short or almost instantaneous.
As a result, the READY indicator may light for only a very
short time, perhaps not being visible at all.
Pushing the SINGLE switch automatically disengages
both the REFresh and ROLL modes, which are
continuous modes.
5. Once the waveform is captured, it can be stored in long-term
memory through the use of the SAVE CH2 or SAVE ALL
switches.
Pre- and Post-Trigger Capture
Another powerful feature of a DSO is ability to display
“pre-trigger” information, that is, events occurring before the arrival
of a trigger event. On the Model 2522C, pre-trigger operation is
available in single-sweep mode by setting the PRE TRIG switches.
As shown in the above section on “Digitizing One-Time Events”,
when 50% pre-trigger operation is selected, the event to be captured is
situated in the center of the memory (roughly the center of the display)
after storage. The waveform to the left of center represents activity
which occurred before that trigger arrived. Pre-Trigger selection of
25% and 75% are also available.
If the PRE-TRIG switches are released (0% pretrigger), then no
pre-trigger information is stored, and the trigger event rolls to the
extreme left of the display. In this case, all the information displayed
on the screen represents activity after the trigger event. You may
wish to use post-trigger operation first to observe where the trigger is
occurring on the waveform of interest, and then switch to pre-trigger
mode.
Roll Mode
In this mode of operation, the waveform rolls across the CRT from
right to left (as opposed to standard oscilloscopes, which have the trace
moving from left to right) in the same manner as most strip chart
recorders. It is most commonly used for viewing very slow events.
1. Set up the oscilloscope in analog mode so that the event to be
observed is properly positioned on the display. you may wish to
use AUTO triggering so that the scope continues to draw a
trace even if the event is especially slow.
2. Switch to digital mode (ANALOG/STORAGE switch to
STORAGE), and depress the ROLL switch. Select a
TIME/DIV setting that produces a roll at the desired speed. As
the sweep speed is decreased, the waveform will move across
the screen more slowly and the Roll feature will become more
apparent. It can also be slowed by a factor of 100 on some ranges;
see the discussion in the next section.
OPERATING INSTRUCTIONS
19
3. The rolling display can be frozen at any time by pressing the
SAVE CH2 or SAVE ALL switches (as discussed in the
section on “Digitizing Repetitive Waveforms”).
NOTE
ROLL mode cannot be used on sweep speeds
greater than 10 ms/div.
Expanded Sweep Settings—SLOW X100 Mode
In digital mode, the 10 ms/div to 0.5 s/div TIME/DIV ranges
can be expanded by a factor of 100 by depressing the SLOW
X100 switch. For example, the 10 ms/div setting becomes 1000
ms (1 sec)/div when this switch is engaged. This time base
expansion is extremely useful for observing very slow events.
With it, the scope is capable of recording an event up to 500
seconds in duration (.5 sec/div x100 becomes 50 sec/div).
Saving Screen images to a USB flash drive
The Model 2522C oscilloscope supports saving of screen data
to a USB flash drive. To save, simply press down the SAVE
ALL button first, then press down DATA OUT. The Transfer
light next to it will flash or lid up. This means data is being
transferred to USB drive connected in the rear USB port.
When transfer is done, light will turn off. Press back up the
DATA OUT button when finished.
.
20
MAINTENANCE
WARNING
The following instructions are fo r use by
qualified service personnel only. To
avoid electrical shock, do not perform
any servicing other than contained in
the operating instructions unless you
are qualified to do so.
High voltage up to 2000 V is present
when covers are removed and the unit is
operating. Remember that high voltage
may be retained indefinitely on high
voltage capacitors. Also remember that
ac line voltag e is present on line volta ge
input circuits any time the instrument is
plugged into an ac outlet, even if turned
of. Unplug the oscilloscope and dis-
charge high voltage capacitors before
performing servi ce proced ures.
FUSE REPLACEMENT
If the fuse blows, the “ON” indicator will not light and
the oscilloscope will not operate. The fuse should not
normally open unless a problem has developed in the unit.
Try to determine and correct the cause of the blown fuse,
then replace only with the correct value fuse. For 110/125
V line voltage operation, use an 800 mA, 250 V fuse. For
220/240 V line voltage operation, use a 600 mA, 250 V
fuse. The fuse is located on the rear panel adjacent to the
power cord receptacle.
Remove the fuseholder assembly as follows:
1. Unplug the power cord from rear of scope.
2. Insert a small screwdriver in fuseholder slot
(located between fuseholder and receptacle). Pry
fuseholder away from receptacle.
3. When reinstalling fuseholder, be sure that the fuse
is installed so that the correct line voltage is selected
(see LINE VOLTAGE SELECTION).
LINE VOLTAGE SELECTION
To select the desired line voltage, simply insert the fuse
and fuse holder so that the appropriate voltage is pointed to
by the arrow. Be sure to use the proper value fuse (see
label on rear panel).
PERIODIC ADJUSTMENTS
Probe compensation and trace rotation adjustments should be
checked periodically and adjusted if required. These procedures are
given below.
Probe Compensation
1. Connect probes to CH 1 and CH 2 input jacks. Per-form
procedure for each probe, one probe at a time.
2. Set the probe to X10 (compensation adjustment is not possible
in the X1 position).
3. Touch tip of probe to CAL terminal.
4. Adjust oscilloscope controls to display 3 or 4 cycles of CAL
square wave at 5 or 6 divisions amplitude.
5. Adjust compensation trimmer on probe for optimum square
wave (minimum overshoot, rounding off, and tilt). Refer to Fig.
5.
Fig. 5. Probe Compensation Adjustment.
Trace Rotation Adjustment
1. Set oscilloscope controls for a single trace display in CH 1
mode, and with the channel 1 AC-GND-DC switch set to
GND.
2. Use the channel 1 POSition control to position the trace over
the center horizontal line on the graticule scale. The trace
should be exactly parallel with the horizontal line.
3. Use the TRACE ROTATION adjustment on the front panel
to eliminate any trace tilt.
Correct
Compensation
Over
Compensation
Insufficient
Compensation
CALIBRATION CHECK
A general check of calibration accuracy may be made by
displaying the output of the CAL terminal on the screen.
This terminal provides a square wave of 0.5 V p-p. This
signal should produce a displayed waveform amplitude of
five divisions at 0.1 V/div sensitivity for both channel 1 and 2
(with probes set for direct). With probes set for X10, there
should be five divisions amplitude at 10 mV/div sensitivity.
The VARIABLE controls must be set to CAL during this
check.
INSTRUMENT REPAIR SERVICE
Because of the specialized skills and test equipment required
for instrument repair and calibration, many customers prefer
to rely upon B&K Precision for this service. To use this
service, even if the oscilloscope is no longer under warranty,
follow the instructions given in the SERVICE
INFORMATION portion of this manual. There is a flat rate
charge for instruments out of warranty.
NOTE
The CAL signal should be used only as a
general check of calibration accuracy, not
as a signal source for performing recali-
bration adjustments; a voltage
standarcalibrated at several steps and of
0.3% or better accuracy is required for
calibration adjustments.standarcalibrated
at several steps and of 0.3% or better
accuracy is required for calibration
adjustments.
MAINTENANCE
22
APPENDIX I
IMPORTANT CONSIDERATIONS FOR RISE TIME AND FALL TIME MEASUREMENTS
Error in Observed Measurement
The observed rise time (or fall time) as seen on the
CRT is actually the cascaded rise time of the pulse being
measured and the oscilloscope’s own risetime. The two
rise times are combined in square law addition as follows:
2 2
Tobserved = ( T pulse) + ( T scope)
The effect of the oscilloscope’s rise time is almost
negligible when its rise time is at least 3 times as fast as
that of the pulse being measured. Thus, slower rise times
may be measured directly from the CRT. However, for
faster rise time pulses, an error is introduced that
increases progressively as the pulse rise time approaches
that of the oscilloscope. Accurate measurements can still
be obtained by calculation as described below.
Direct Measurements
The Model 2522C has a rated rise time of 18 ns. Thus,
pulse rise times of about 54 ns or greater can be
measured directly. Most fast rise times are measured at
the fastest sweep speed and using X10 magnification. This
sweep rate is 10 ns/div. A rise time of less than about five
divisions at this sweep speed should be calculated.
Calculated Measurements
For observed rise times of less than 54 ns, the pulse rise time should
be calculated to eliminate the error introduced by the cascaded
oscilloscope rise time. Calculate pulse rise time as follows:
2 2
Tpulse = ( T observed) + ( T scope)
Limits of Measurement
Measurements of pulse rise times that are faster than the scope’s
rated rise time are not recommended because a very small reading
error introduces significant error into the calculation. This limit is
reached when the “observed” rise time is about 1.3 times greater than
the scope’s rated rise time, about 23 ns minimum for the Model
2522C.
Probe Considerations
For fast rise time measurements which approach the limits of
measurement, direct connection via 50 coaxial cable and 50
termination is recommended where possible. When a probe is used,
its rise time is also cascaded in square law addition. Thus the probe
rating should be considerably faster than the oscilloscope if it is to be
disregarded in the measurement.
23
APPENDIX II
UNIQUE CHARACTERISTICS OF DIGITAL STORAGE OSCILLSCOPES
Digital Storage Oscilloscopes (DSO’s) use a digital sampling
technique to convert analog input signals to a series of digital
words that can be stored in memory. Since digital sampling has
disadvantages as well as advantages, it is important to be aware
of these unique characteristics of DSO’s.
ALIASING
This DSO uses Real Time Sampling when sweep TIME/DIV
settings of 50 sec/div to 20 µs/div are selected. Real Time
Sampling simply means that samples of the input signals are
taken at equal spaces (e.g., every 0.25 ms when the 50 ms/div
range is selected). With Real Time Sampling, a phenomena
called “Aliasing” can occur when the input signal is not
sampled often enough. This causes the digitized signal to appear
to be of a lower frequency than that of the input signal. Unless
you have an idea what the input signal is supposed to look like,
you will usually be unaware that Aliasing is occurring.
Aliasing Example
To see what actually occurs when a Digital Storage Oscil-
loscope is Aliasing, perform the following example.
1. Apply a 10 kHz signal to the input jack and set the sweep
TIME/DIV control to 50 µs/div. You should see about 5
cycles of the waveform on the display. Since the DSO
samples the input waveform 200 times per division, each
cycle is sampled 400 times.
2. Now change the sweep TIME/DIV control to 2 ms/div.
The display should look crowded. Because the DSO takes
200 samples per division, the sample points are 10 µs
apart. Since the input signal is at a frequency of 10 kHz, it
is being sampled 10 times per cycle. The resulting display
is too crowded to be useful, however, it is not incorrect (it
is very similar to what you would see on a conventional
analog oscilloscope).
3. Change the setting of the sweep TIME/DIV control to 20
ms/div. Vary the frequency a slight amount (until the
display is readable) to obtain as few cycles as possible on
the CRT. If you were to calculate the frequency of the
signal from the display, you would come up with a much
lower frequency than that of the actual frequency of the
signal at the input jack. As an example, if three cycles are
displayed, the calculated frequency would be
approximately 15 Hz. This is
obviously incorrect. This occurs because the DSO is
taking one sample every 0.1 ms and a 10 kHz signal has
one cycle every 0.1 ms. What is actually happening is that
the frequency is off (not perfectly 10 kHz) by just enough
to cause the DSO to take one sample at a slightly
different place on each cycle of the waveform.
Avoiding Aliasing
Aliasing is not limited to the above example. This phe-
nomenon can occur anytime that at least two samples per cycle
are not taken (whenever the sweep TIME/DIV setting is much
too slow for the waveform being applied to the input).
Whenever the frequency of the signal is unknown, always
begin with the fastest real-time sweep speed (20 µs/div) or by
viewing the waveform in the analog mode of operation first.
NOTE
Viewing one-time events or glitches is not
possible when sweep TIME/DIV settings
higher than 20 s/div are selected. Viewing
one-time events poses no problem with Ali-
asing because Aliasing can occur only with
repetitive waveforms.
EQUIVALENT TIME SAMPLING
This oscilloscope uses a sampling method called Equivalent
Time Sampling when sweep speeds higher than 20 µs/div are
selected. This method permits viewing of repetitive waveforms
to 20 MHz, although the maximum digital sampling rate is 10
Msamples/sec. When the Equivalent Time Sampling mode is
activated, one sample is taken during each cycle. Of course if
one sample is taken during each cycle at the trigger point (the
same point on each cycle), only a flat trace would be produced.
Therefore, it is necessary to take each sample further (in time)
from the trigger point than the last sample. This incremental
delay is deter-mined by the sweep TIME/DIV control setting.
Because 1024 (1 k) samples are needed to fill the display, the
oscilloscope must sample 1024 cycles of the waveform.
Therefore, only repetitive waveforms should be observed in
this mode. Irregularities that are present on an otherwise
repetitive waveform are not likely to show up when the
Equivalent Time Sampling method is used. With only one
sample being taken during each cycle, it is very likely that
glitches and other irregularities will be skipped over.
Intentionally left blank
SERVICE INFORMATION
Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return the product in the
original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes,
connectors and accessories that you are using with the device.
Non-Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return the product
in the original packaging to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and
accessories that you are using with the device. Customers not on an open account must include payment in the form of a money order or credit
card. For the most current repair charges please refer to the service and support section on our website.
Return all merchandise to B&K Precision Corp. with pre-paid shipping. The flat-rate repair charge for Non-Warranty Service does not include
return shipping. Return shipping to locations in North America is included for Warranty Service. For overnight shipments and non-North
American shipping fees please contact B&K Precision Corp.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
714-921-9095
Include with the retur ne d instrument your complete return shipping addre ss, contact name, phone number and de scription of problem.
LIMITED ONE-YEAR WARRANTY
B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in
workmanship and materials for a period of one year from date of purchase.
B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be
accompanied by proof of the purchase date in the form of a sales receipt.
To obtain warranty coverage in the U.S.A., this product must be registered by completing a warranty registration form on our website
www.bkprecision.com within fifteen (15) days of purchase.
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthor i zed alter ations or
repairs. The warranty is void if the serial number is altered, defaced or removed.
B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting from loss of use. Some
states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion may not apply to you.
This warranty gives you specific rights and you may have other rights, which vary from state-to-state.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
714-921-9095
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
© 2009 B&K Precision Corp.
481-315-9-001 Printed in Taiwan