m4. VA 40W AC-DC Reliable Green Medical Adaptor G S M 4 0 B series MEAN WELL IC 4) CB As 1EC60950-1 ANSI/AAMI ES60601-1 a ee an54e. Pid onin Canbst. -P1J) (for 5-9v) EN60601-1 @ Features Applications * 2 pole AC inlet IEC320-C8, Class II power unit * Mobile clinical workstation * Medical safety approved (2 x MOPP) according to * Oral irrigator ANSI/AAMI ES60601-1 and IEC/EN60601-1 * Portable hemodialysis machine * Extremely low leakage current * No load power consumption<0.1W * Energy efficiency level VI and meet CoC Version 5 (Except 5~9V for Level V) * Comply with Korea K-MEPS(only GSM40B48-P1J) * -30~+70C wide range working temperature * Protections: Short circuit / Overload / Over voltage * LED indicator for power on * Lifetime > 90 K hours * 3 years warranty i Description GSM4O0B is a highly reliable, 40W desktop style single-output green medical adaptor series. This product is equipped with a 2-pin (no FG) standard IEC320-C8 power plug, adopting the input range from 80VAC to 264VAC. The entire series supplies different output voltages between 5VDC and 48VDC that can satisfy the demands for various kinds of medical electrical devices. The circuitry design meets the international medical standards (2*MOPP), having an ultra low leakage current (<50,A), fitting the medical devices in direct electrical contact with the patients. With the efficiency up to 91% and the extremely low no-load power consumption below 0.1W, GSM40B is compliant with USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP,and meet Code of Conduct (CoC) Version 5. The supreme feature allows the adaptor to save the energy when it is either under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical shock. GSM40B is approved with the international medical safety certificates. * Breath Machine * Medical computer monitor i Model Encoding DC plug type P1J: Standard model,2.1x5.5x11mm,C+,tuning fork type Other options available by customer requested (see Page 4~5) Output voltage IEC320-C8 AC inlet Rated wattage Series name File Name:GSM40B-SPEC 2018-11-214 W 40W AC-DC Reliable Green Medical Adaptor SPECIFICATION GSM40B series ORDER NO. GSM40B05-P1J| GSM40B07-P1J| GSM40B09-P1J| GSM40B12-P1J) GSM40B15-P1J} GSM40B18-P1J) GSM40B24-P1J| GSM40B48-P1J SAFETY MODEL NO. GSM40B05 | GSM40B07_ | GSM40B09 | GSM40B12 | GSM40B15 | GSM40B18 | GSM40B24 GSM40B48 DC VOLTAGE Note.2) 5V 7.5V ov 12V 15V 18V 24V 48Vv RATED CURRENT 5A 5.34A 4.45A 3.34A 2.67A 2.22A 1.67A 0.84A CURRENT RANGE 0~5A 0~5.34A 0~4.45A 0~3.34A 0~2.67A 0~2.22A 0~1.67A 0~0.84A RATED POWER (max.) 25W 40W 40wW 40W 40W 40W 40W 40wW OUTPUT | RIPPLE & NOISE (max.) Note.3) 80mVp-p 80mVp-p 100mVp-p 100mVp-p 100mVp-p 120mVp-p 150mVp-p 150mVp-p VOLTAGE TOLERANCE Note.4} 5.0% +5.0% +5.0% +3.0% +3.0% +3.0% +2.5% +2.5% LINE REGULATION Note.5| + 1.0% +1.0% +1.0% +1.0% +1.0% +1.0% +1.0% +1.0% LOAD REGULATION +5.0% +5.0% +5.0% +3.0% +3.0% +3.0% 42.5% +2.5% SETUP, RISE TIME Note.6| 1000ms, 30ms / 230VAC 1500ms, 30ms/115VAC at full load HOLD UP TIME (Typ.) 50ms / 230VAC 24ms / 115VAC at full load VOLTAGE RANGE Note.7/80~ 264VAC 113~370VDC FREQUENCY RANGE 47 ~ 63Hz INPUT EFFICIENCY (Typ.) 81% 85.5% 86% 88% 88.5% 89% 90% 91% AC CURRENT (Typ.) 1A/ 115VAC 0.5A/ 230VAC INRUSH CURRENT (Typ.) Cold start 30A/115VAC 60A/ 230VAC LEAKAGE CURRENT(max.) | Touch current < 50A/264VAC OVERLOAD 105 ~ 160% rated output power __ PROTECTION Protection type : Hiccup mode, recovers automatically after fault condition is removed 5.2~7,0V | 7.8~10.2V | 9.4~12.2V | 12.6~16.2V | 15.7~20.3V | 18.9~24.3V | 25.2~32.4V | 50.4~64.8V OVER VOLTAGE - Protection type : Shut down o/p voltage, re-power on to recover WORKING TEMP. -30 ~ +70C (Refer to "Derating Curve") WORKING HUMIDITY 20% ~ 90% RH non-condensing ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85C , 10 ~ 95% RH non-condensing TEMP. COEFFICIENT +0.03% /C (0~50C) VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes OPERATING ALTITUDE Note.8} 3000 meters sarery saoanos | EHH ENED ENE ANY ESET 1 ESSE. eon) PSE IBD, KONO ISOLATION LEVEL Primary-Secondary: 2xMOPP WITHSTAND VOLTAGE I/P-O/P:4KVAC ISOLATION RESISTANCE /P-O/P:100M Ohms / 500VDC / 25C/ 70% RH Parameter Standard Test Level / Note Conducted emission EN55011 (CISPR11), FCC PART 15/ CISPR22, Class B CAN ICES-3(B)/NMB-3(B), MSIP KN32 EMC EMISSION Radiated emission CAN ICES IB) NMB-3(6) MSIP KN Class B Harmonic current EN61000-3-2 Class A SAFETY & Voltage flicker EN61000-3-3 | mene EMC EN55024 , EN60601-1-2, EN61204-3 (Note 9) Parameter Standard Test Level / Note ESD EN61000-4-2 Level 4, 15KV air ; Level 4, 8KV contact RF field susceptibility EN61000-4-3 bevel 3, 10Viin( BoM Z~2.7GH2 Table 9, 9~28V/m( 385MHz~5.78GHz ) EFT bursts EN61000-4-4 Level 3, 2KV EMC IMMUNITY Surge susceptibility EN61000-4-5 Level 3, 1KV/Line-Line Conducted susceptibility EN61000-4-6 Level 3, 10V Magnetic field immunity EN61000-4-8 Level 4, 30A/m Voltage dip, interruption EN61000-4-11 toon inerotons 60 periods periods, MTBF 740K hrs min. MIL-HDBK-217F(25C) OTHERS | DIMENSION 125*50*31.5mm (L*W*H) PACKING 0.29Kg; 40pcs/12.6Kg/1.05CUFT PLUG See page 4~5 ; Other type available by customer requested CONNECTOR CABLE See page 4~5 ; Other type available by customer requested NOTE 1. All parameters are specified at 230VAC input, rated load, 25C 70% RH ambient. 2. DC voltage: The output voltage set at point measure by plug terminal & 50% load. 3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47f capacitor. 4. Tolerance: includes set up tolerance, line regulation, load regulation. 5. Line regulation is measured from low line to high line at rated load. 6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time. 7. Derating may be needed under low input voltages. Pleas check the derating curve for more details. 8. The ambient temperature derating of 3.5C/1000m with fanless models and of 5C/1000m with fan models for operating altitude higher than 2000m(6500ft). 9. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http:/Avwww.meanwell.com) File Name:GSM40B-SPEC 2018-11-214 "VA 40W AC-DC Reliable Green Medical Adaptor MEAN WELL GSM40B series Hi Derating Curve 100 60F LOAD (%) 40 80------- 50p--- >> ee ee -30 0 ro ND So wo So S oS a So 2 So x So (HORIZONTAL) AMBIENT TEMPERATURE (C) H Mechanical Specification 125 60 Power LED AC Inlet IEC320-C8 HB DC output plug Standard plug: P1J Hi Static Characteristics LOAD (%) 100 F 4 1 90 1 | 80 - 70 F 40 - 1 ! ! 1 I I 1 I sof | | 1 I I 1 1 1 ! ! 1 1 80 90 100 110 120 140 160 180 200 220 240 264 INPUT VOLTAGE (VAC) 60Hz Case No.GS60B _Unit:mm UL2464 16AWG 1000+50mm for 5 ~ 7.5V UL1185 16AWG 1000450mm for 9 ~ 12V UL1185 16AWG 1500+50mm for 15 ~ 48V 7010mm P1J Pin Assignment 55 "sosnm or Outside O-@ Inside File Name:GSM40B-SPEC 2018-11-21GSM40B series 4 "VA 40W AC-DC Reliable Green Medical Adaptor Optional DC plug: A B C Tuning Fork Style Type No. g y He OD ID L c P1l 5.5 2.1 9.5 Se P1L 5.5 2.5 9.5 A (Straight) P1M 5.5 2.5 11.0 a @, P1IR 55 21 | 9.5 7 P1JR 5.5 2.1 11.0 P1LR 5.5 2.5 9.5 (Right-angled) P1MR 55 25 11.0 A B C Barrel Style Type No. y ule )) ID L P2i 5.5 2.1 9.5 R= P2J 5.5 2.1 11.0 , a P2L 5.5 2.5 | 9.5 ~~ (Straight) P2M 5.5 2.5 | 11.0 . , P2IR 5.5 2.1 9.5 P2JR 5.5 2.1 11.0 P2LR 5.5 2.5 9.5 (Right-angled) | paMR 5.5 25 | 11.0 A B C Lock Style Type No. OD iD 1 A Locking P2S(S761K) 5.53 2.03 | 12.06 >. | ( (a P2K(761K) 5.53 2.54 | 12.06 & , | P2C(S760K) 5.53 2.03 | 9.52 SWITCHCRAFT original or equivalent P2D(760k) 5.53 2.54 9.52 A B C Type No. Min. Pin Style yp OD ID L C., P3A 2.35 0.7 11.0 Wi a : ee P3B 4.0 1.7 | 11.0 EIAJ equivalent P3C 4.75 1.7 11.0 A B C D Type No. Center Pin Style yp OD D L Icenterrin Cc P4A , ; . A -., 5.5 | 3.4 | 11.0 | 1.0 on Coma] | P4B 65 | 44 | 11.0) 14 EIAJ equivalent P4C 7.4 5.1 11.0 | 0.6 . oo, Pin Assignment Min. DIN 3 Pin with Lock (male Type No. ae oP PIN No. Output 5 1 +Vo We 2 Call Sly [2 R6B 2 Vo : g KYCON KPPX-3P equivalent 3 +Vo File Name:GSM40B-SPEC 2018-11-21m4. VA 40W AC-DC Reliable Green Medical Adaptor G S M 4 0 B series MEAN WELL Min. DIN 4 Pin with Lock (mal Type No rin Assignment in. in with Lock (male) y BIN No. Output 5 1 +Vo (oO o\\2 3 caf isl |e 2 -Vo : S853 Fs KYCON KPPX-4P equivalent 4 +Vo Min. DIN 4 Pin with Lock (femal Type No Pin Assignment in. in with Lock (female) y PIN No. Output 1 +Vo 2 -Vo OO) 23 23 (eam | R7BF 2 #e KYCON KPJX-CM-4S equivalent 4 +Vo . Pin Assignment DIN 5 Pin (male) Type No. | BIN No. Output { -Vo 3 IT | 2 Vo 9 8 Ye R1B 3 +Vo = 4 -Vo 5 +Vo . . Pin Assignment Stripped and tinned leads Type No. PIN No. Output L (red) a -_ ' *Vo LT 9 -} L1 (black) | by customer Length of Land L1 by request 2 -Vo (MW's standard length, L:_25 mm, L1:_5 mm) Hf Installation Manual Please refer to : http://www.meanwell.com/manual.html File Name:GSM40B-SPEC 2018-11-21