www.sensirion.com February 2015 V2 1/4
LG01-2000Axxx
Liquid Flow Switch and Bubble Detector
Detection of low liquid flow in the ml/min Range
Totally non-invasive
Ultra-fast response time of 100 ms
Product Summary
The LG01 Liquid Flow Switch enables detection of
flow of liquid in the range of ml/min. The output signal
is 5V when the flow rate is higher than the specified
flow switch point and 0V below this limit or when air
or bubbles are found inside the sensors flow channel
instead of liquid.
Excellent chemical resistance and bio-compatibility
are ensured: The flow path of the LG01 Liquid Flow
Switch is formed by a simple, straight glass capillary.
For more information on communication, please refer
to section 2 of this document.
1 Sensing Performance
Table 1: Model specific performance of LG01 (all data for medium H2O, 23°C)
Parameter
LG01-2000A090
LG01-2000A005
Unit
H2O flow switching point
4.5
0.25
ml/min
H2O maximum flow rate
50
220
ml/min
Accuracya of switching point
±10
±10
% of switching point
Switching hysteresis
±5
±5
% of switching point
Mounting orientation sensitivityb
+2
+14
% of switching point
Flow detection response time 63
50
ms
Response time on power-up
120
ms
Operating temperature
+10…+50
°C
Ambient storage temperaturec
-10…+60
°C
Operating pressure
3
bar
Proof pressure
5
bar
aBetter available on request
bMaximum shift of switching point if mounted vertically
cNon-condensing, flow path empty
1.1 Operation with Other Liquids than Water
Due to the measurement principle the LG01 flow switch characteristics is changing with the thermal properties of the
liquid. For aqueous solutions the data for H2O as stated above is a good approximation.
For hydrocarbon liquids such as oil, fuel, or organic solvents, the flow switch point shifts towards higher flow rates.
For indicative purpose, data is provided for isopropyl alcohol (IPA) below.
Table 2: Approximate switch levels of the LG01 liquid flow switch with hydrocarbon liquids
Parameter
LG01-2000A090
LG01-2000A005
Unit
Approximate flow switching point with IPA
9
1.4
ml/min
Maximum flow rate with IPA
>250
>250
ml/min
www.sensirion.com February 2015 V2 2/4
2 Communication with the Sensor
An electrical voltage output indicates flow above (5 V)
or below (0 V) the flow switch point and/or air or
bubbles inside the sensors flow channel. The product
comes fully calibrated for water for volume
applications flow calibration for other media is
available on request.
2.1 Electrical Specifications
Table 3: DC Characteristics
Parameter
Typ.
Max.
Units
Power supply DC, VDD
7
12a
V
Operating current,
VDD = 9 V, no load
6.8
mA
Voltage output high
5
V
Voltage output low
0
V
aUse 9 V max. for best performance
2.2 Electrical Connector and Sensor Pinout
Connector Type: 5 pin Molex PCB Header
53261-0590 Mating connector: Molex 51021-0500.
Table 4: Electrical Pinout
Pin
1
Do not connect
2
Do not connect
3
VDD
4
GND
5
Voltage output
3 Fluidic Connection
Table 5: Fluidic Specifications and Pressure Rating
Parameter
LG01-2000Axxx
Wetted materials:
Internal sensor tube material
Borosilicate Glass 3.3
Fitting material
PEEK
Sealing material
FEP
Fluid connector ports (Fittings)
¼-28 flat-bottom for 1/16” or 1/8” OD plastic tubing
(min. 2mm ID recommended)a
Pressure drop (at maximum flow rate, H2O, 23°C)
1 mbar
Total internal volume
80 µl
a For fluid connection material use flat bottom ¼-28 standard components e.g. IDEX parts P-200/P-300. Various adapters e.g. to Barb
connectors are available from company Value Plastics as well.
For more information on the fluidic connection please find: “Application Note Sensor Ports and Tubing Connections in the
Download Center on our webpage (www.sensirion.com/liquidflow-download).
4 Connectivity Kit
The optional LG01 connectivity kit contains material for electrical and fluidic connections
Assembled flat ribbon cable, 30cm (1-100482-01)
2x barb adapter to connect ID 1.6 mm (1/16”) tubing (Value Plastics ABR004-J1A-2)
2x barb adapter to connect ID 3.2 mm (1/8”) tubing (Value Plastics ABR013-J1A-1)
1
2
3
4
5
www.sensirion.com February 2015 V2 3/4
5 Mechanical Specifications
Table 6: Mechanical Specifications
Parameter
LG01-2000Axxx
Largest dimensions
53 x 22 x 9 mm
Total mass
6 g
Inner diameter d
1.8 mm
All dimensions in mm
6 Ordering Information
For OEM applications the sensor can be purchased in larger quantities without any additional parts.
Product
Article Number
LG01-2000A090
1-100796-01
LG01-2000A005
1-100797-01
LG01 Connectivity Kit
1-100814-01
Attention Fragile
Mechanical shocks, forces induced by stiff fluid connection lines as well as connecting to the fittings without suitable
tools lead to stress on the internal thin-walled glass capillary and can cause it to break.
While tightening the fittings, fix the fluidic ports position with a wrench.
Test for leakage after every time new connections are made.
!
53±1
www.sensirion.com February 2015 V2 4/4
Important Notices
Warning, personal injury
Do not use this product as safety or emergency stop devices or in
any other application where failure of the product could result in
personal injury (including death). Do not use this product for
applications other than its intended and authorized use. Before
installing, handling, using or servicing this product, please
consult the datasheet and application notes. Failure to comply
with these instructions could result in death or serious injury.
If the Buyer shall purchase or use SENSIRION products for any
unintended or unauthorized application, Buyer shall defend, indemnify
and hold harmless SENSIRION and its officers, employees,
subsidiaries, affiliates and distributors against all claims, costs,
damages and expenses, and reasonable attorney fees arising out of,
directly or indirectly, any claim of personal injury or death associated
with such unintended or unauthorized use, even if SENSIRION shall be
allegedly negligent with respect to the design or the manufacture of the
product.
ESD Precautions
The inherent design of this component causes it to be sensitive to
electrostatic discharge (ESD). To prevent ESD-induced damage and/or
degradation, take customary and statutory ESD precautions when
handling this product.
Warranty
SENSIRION warrants solely to the original purchaser of this product for
a period of 12 months (one year) from the date of delivery that this
product shall be of the quality, material and workmanship defined in
SENSIRION’s published specifications of the product. Within such
period, if proven to be defective, SENSIRION shall repair and/or
replace this product, in SENSIRION’s discretion, free of charge to the
Buyer, provided that:
notice in writing describing the defects shall be given to
SENSIRION within fourteen (14) days after their appearance;
such defects shall be found, to SENSIRION’s reasonable
satisfaction, to have arisen from SENSIRION’s faulty design,
material, or workmanship;
the defective product shall be returned to SENSIRION’s factory at
the Buyer’s expense; and
the warranty period for any repaired or replaced product shall be
limited to the unexpired portion of the original period.
This warranty does not apply to any equipment which has not been
installed and used within the specifications recommended by
SENSIRION for the intended and proper use of the equipment.
EXCEPT FOR THE WARRANTIES EXPRESSLY SET FORTH
HEREIN, SENSIRION MAKES NO WARRANTIES, EITHER EXPRESS
OR IMPLIED, WITH RESPECT TO THE PRODUCT. ANY AND ALL
WARRANTIES, INCLUDING WITHOUT LIMITATION, WARRANTIES
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSLY EXCLUDED AND DECLINED.
SENSIRION is only liable for defects of this product arising under the
conditions of operation provided for in the datasheet and proper use of
the goods. SENSIRION explicitly disclaims all warranties, express or
implied, for any period during which the goods are operated or stored
not in accordance with the technical specifications.
SENSIRION does not assume any liability arising out of any application
or use of any product or circuit and specifically disclaims any and all
liability, including without limitation consequential or incidental
damages. All operating parameters, including without limitation
recommended parameters, must be validated for each customer’s
applications by customer’s technical experts. Recommended
parameters can and do vary in different applications.
SENSIRION reserves the right, without further notice, (i) to change the
product specifications and/or the information in this document and (ii) to
improve reliability, functions and design of this product.
Copyright © 2001-2015, SENSIRION.
CMOSens® is a trademark of Sensirion
All rights reserved
REACH, RoHS and WEEE Statement
The LG01 flow switch complies with requirements of the following
directives:
EU Directive 1907/2006/EC concerning Registration, Evaluation,
Authorization and Restriction of Chemicals (REACH)
EU Directive 2002/96/EC on waste electrical and electronic
equipment (WEEE), OJ13.02.2003; esp. its Article 6 (1) with Annex II.
EU Directive 2002/65/EC on the restriction of certain hazardous
substances in electric and electronic equipment (RoHS),
OJ01.01.2011
Headquarters and Subsidiaries
SENSIRION AG
Laubisruetistr. 50
CH-8712 Staefa ZH
Switzerland
phone: +41 44 306 40 00
fax: +41 44 306 40 30
info@sensirion.com
www.sensirion.com
Sensirion Inc., USA
phone: +1 805 409 4900
info_us@sensirion.com
www.sensirion.com
Sensirion Japan Co. Ltd.
phone: +81 3 3444 4940
info@sensirion.co.jp
www.sensirion.co.jp
Sensirion Korea Co. Ltd.
phone: +82 31 345 0031 3
info@sensirion.co.kr
www.sensirion.co.kr
Sensirion China Co. Ltd.
phone: +86 755 8252 1501
info@sensirion.com.cn
www.sensirion.com.cn
Sensirion AG (Germany)
phone: +41 44 927 11 66
info@sensirion.com
www.sensirion.com
To find your local representative, please visit www.sensirion.com/contact