DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
2018.08.23-A/7 Page 1 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
Selection Guide
Certification
Part No.
Input Voltage (VDC)
Output
Full Load
Efficiency (%)
Min./Typ.
Nominal
(Range)
Voltage(VDC)
Current (mA)
Max./Min.
UL/CE
G0505S-2WR2
5
(4.5-5.5)
±5
±200/±20
74/78
G0509S-2WR2
±9
±111/±12
74/78
G0512S-2WR2
±12
±83/±9
74/78
G0515S-2WR2
±15
±67/±7
76/80
H0505S-2WR2
5
400/40
73/77
H0512S-2WR2
12
167/17
75/79
H0515S-2WR2
15
133/14
75/79
G1205S-2WR2
12
10.8-13.2
±5
±200/±20
70/74
G1209S-2WR2
±9
±111/±12
76/80
G1212S-2WR2
±12
±83/±9
76/80
G1215S-2WR2
±15
±67/±7
73/77
H1205S-2WR2
5
400/40
72/76
H1212S-2WR2
12
167/17
75/79
H1215S-2WR2
15
133/14
77/81
--
G1505S-2WR2
15
(13.5-16.5)
±5
±200/±20
73/77
G1509S-2WR2
±9
±111/±12
76/80
G1515S-2WR2
±15
±67/±7
69/73
UL
H1505S-2WR2
5
400/40
73/77
--
H1515S-2WR2
15
133/14
78/82
UL/CE
G2405S-2WR2
24
21.6-26.4
±5
±200/±20
75/79
G2409S-2WR2
±9
±111/±12
77/81
G2412S-2WR2
±12
±83/±9
78/82
G2415S-2WR2
±15
±67/±7
77/81
H2405S-2WR2
5
400/40
75/79
H2412S-2WR2
12
167/17
78/82
H2415S-2WR2
15
133/14
80/84
* Note: The specified maximum capacitive load for positive and negative output is identical.
2W isolated 4.2kVAC/6kVDC, Fixed Input voltage,
unregulated Single or Dual Output
Patent Protected RoHS
FEATURES
High I/O isolation voltage of 4200VAC/6000VDC
Reinforced insulation
Patient leakage current A max.
High efficiency up to 84%
Operating ambient temperature -40to +85
Internal surface mount design
SIP package
Industry standard pin-out
EN60601-1, ANSI/AAMI ES60601-1 approval
(1xMOPP/2xMOOP)
G_S-2WR2 & H_S-2WR2 series meets reinforced insulation requirements. They are especially designed for applications where extremely high
isolation, low insulation capacitor with low leakage current in a compact package size is required. They are widely used in applications
such as medical, electrical grid, IGBT driver circuits and similar where:
1. The voltage of the input power supply is relatively stable with a variation of ±10%Vin or less;
2. An extremely high input to output isolation voltage of up to 4200VAC or 6000VDC is required;
3. The requirement for ripple & noise or a tight output regulation is not as strict.
DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
2018.08.23-A/7 Page 2 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
Input Specifications
Item
Operating Conditions
Min.
Typ.
Max.
Unit
Input Current (no-load/full load)
5V input
--
35/520
80/--
mA
12V input
--
15/217
40/--
15V input
--
18/171
40/--
24V input
--
10/106
25/--
Surge Voltage (1sec. max.)
5V input
-0.7
--
9
VDC
12V input
-0.7
--
18
15V input
-0.7
--
21
24V input
-0.7
--
30
Reflected Ripple Current*
--
0.2
--
A
Input Filter
Capacitor filter
Hot Plug
Unavailable
* Note: Please refer to DC-DC Converter Application Note for detailed description of Reflected ripple current testing method.
Output Specifications
Item
Operating Conditions
Min.
Typ.
Max.
Unit
Output Voltage Accuracy*
See Typical Performance Curves (Fig. 1)
Line Regulation
Input voltage change: ±1%
--
--
±1.2
--
Load Regulation
10%-100% load
5VDC output
--
--
20
%
9VDC output
--
--
15
12VDC output
--
--
15
15VDC output
--
--
15
Ripple & Noise**
20MHz bandwidth
--
100
150
mVp-p
Temperature Coefficient
100% full load
--
±0.02
--
%/
Output Short Circuit Protection***
--
--
3
s
Note: *Output voltage accuracy of G1515S-2WR2 with 10% load, Min. -5%.
**The parallel cable” method is used for Ripple and noise test, please refer to
DC-DC Converter Application Notes
for specific information.
*** At the end of the short circuit duration, the supply voltage must be disconnected from the modules.
General Specifications
Item
Operating Conditions
Min.
Typ.
Max.
Unit
Isolation Voltage
Input-output Electric strength test for 1 minute
4200
--
--
VAC
6000
--
--
VDC
Patient Leakage Current
250VAC50/60Hz
--
--
2
µA
Insulation Resistance
Input-output resistance at 500VDC
1000
--
--
MΩ
Insulation Capacitance
Input-output capacitance at 100kHz/0.1V
--
5
--
pF
Operating Temperature
-40
--
85
Storage Temperature
-55
--
125
Case Temperature Rise
Ta=25
--
25
--
Pin Soldering Resistance
Temperature
1.5mm away from the case for 10 seconds
--
--
300
Storage Humidity
Non-condensing
--
--
95
%RH
Switching Frequency
100% load, nominal input voltage
--
100
--
kHz
MTBF
MIL-HDBK-217F @ 25
3500
--
--
k hours
Transformer Creepage &
Clearance Distance
5
--
--
mm
PCB Creepage & Clearance
Distance
5.5
--
--
Note:1Patient leakage current and reinforced insulation is based on a 250 VAC, 50/60 Hz system input voltage.
2The UL certification (ANSI/AAMI ES60601-1, File No. E347375) of G_S-2WR2 & H_S-2WR2 series is approved, G_S-2WR2 & H_S-2WR2 series meets
1xMOPP/2xMOOP when system input voltage is 250VAC , 50/60Hz.
DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
2018.08.23-A/7 Page 3 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
Mechanical Specifications
Case Material
Black plastic; flame-retardant and heat-resistant (UL94 V-0)
Package Dimensions
19.50 x 9.80 x 12.50 mm
Weight
4.2g (Typ.)
Cooling Method
Free air convection
Typical Performance Curves
5VDC output
-10%
80%
20%
10% 40% 60%
-5%
0
+5%
+10%
+15%
100%
Typ
.
Min.
Max.
-7.5%
-2.5%
+2.5%
Tolerance Envelope Curve
Output Cur rent Percent
(Nominal Input Voltage)
Output Voltage Accuracy
Other output
-10%
80%
20%
10% 40% 60%
-5%
0
+5%
+10%
+15%
100%
Typ
.
Min.
Max .
-7.5%
-2.5%
+2.5%
Output Current Percent
(Nominal Input Voltage)
Output Voltage Accuracy
Tolerance Envelope Curve
Fig. 1
Operating Temperature( )
Temperature Curve
Output Power Percentage(%)
-40 0 40 85
80
60
100
120
40
20
0
Safe Operating Area
120
Fig. 2
Electromagnetic Compatibility (EMC)
EMI
CE
EN60601-1-2/CISPR 11 GROUP1 CLASS B (see Fig. 5 for recommended circuit)
RE
EN60601-1-2/CISPR 11 GROUP1 CLASS B (see Fig. 5 for recommended circuit)
EMS
ESD
EN60601-1-2/IEC/EN61000-4-2 Contact ±8kV performance Criteria B
DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
2018.08.23-A/7 Page 4 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
Design Reference
1. Typical application
Input and/or output ripple can be further reduced, by connecting a filter capacitor from the input and/or output terminals to ground as
shown in Fig.3.
Choosing suitable filter capacitor values is very important for a smooth operation of the modules, particularly to avoid start-up problems
caused by capacitor values that are too high. For recommended input and output capacitor values refer to Table 1.
For a tight output voltage regulation, including overvoltage, overcurrent and over temperature protection, we recommend the use of a
linear regulator that is connected in series to the input and/or output terminals as shown in Fig. 4.
S in g l e
V in
G N D
+Vo
0 V
D C
C in
D C
C ou t
D u a l
V in
G N D
+Vo
0 V
D C
C in
D C
C o ut
C ou t
-Vo
Fig. 3
Single
Vin
GND
0V
+Vo
DC DC
REG REG
Vin
GND
0V
+Vo
DC DC
-Vo
REG REG
REG
Dual
Fig. 4
Table 1: Recommended input and output capacitor values
Vin
(VDC)
Cin
(µF)
Single
Vout
(VDC)
Cout
(µF)
Dual
Vout
(VDC)
Cout
(µF)
5
10
5
10
±5
4.7
12/15
4.7
12
2.2
±9
2.2
24
2.2
15
1
±12/±15
1
2. EMC (CLASS B) compliance circuit
Vin
GND
+Vo
-Vo
( )0V
DC/DC LOAD
LDM
C1 C2 C3
Vin
GND
Fig. 5
Table 2: Recommended EMC filter values
Input voltage (V)
5/12/15
24
EMI
C1,C2
4.7µF /50V
C3
Refer to the Cout in Fig.3
LDM
6.8µH
15µH
Note: C1 and C2 of G1515S-2WR2 is 10µF/25V, LDM of G1515S-2WR2 is 22µH.
3. Output load requirements
For a reliable and efficient operation of the converter, the minimum load should never be less than 10% of the rated output load. If the total
required output power is below 10%, a parallel bleeding resistor is required on the output, ensuring that the sum of the power consumption is
always maintained at 10% minimum.
4. For additional information, please refer to DC-DC converter application notes on
www.mornsun-power.com
DC/DC Converter
G_S-2WR2 & H_S-2WR2 Series
2018.08.23-A/7 Page 5 of 5
MORNSUN Guangzhou Science & Technology Co., Ltd. reserves the copyright and right of final interpretation
Dimensions and Recommended Layout
Notes:
1. For additional information on Product Packaging please refer to www.mornsun-power.com. Packaging bag number: 58200013;
2. In order to guarantee product performance and datasheet compliance, the product must be operated within specifications and load
range requirement;
3. The specified maximum capacitive load is tested under full load condition and over the input voltage range;
4. All parameters in this datasheet were measured under following conditions: Ta=25, relative humidity <75%RH, nominal input voltage
and rated output load (unless otherwise specified);
5. All index testing methods in this datasheet are based on our Company’s corporate standards;
6. For special requirements and customization service, please contact your nearest MORNUSN sales representative or one of our
technicians;
7. Products are related to laws and regulations: see "Features" and "Electromagnetic Compatibility (EMC)";
8. Our products shall be handled according to ISO14001 and related environmental laws and regulations by qualified personnel only.
Mornsun Guangzhou Science & Technology Co., Ltd.
Address: No. 5, Kehui St. 1, Kehui Development Center, Science Ave., Guangzhou Science City, Luogang District, Guangzhou, P. R. China
Tel: 86-20-38601850 Fax: 86-20-38601272 E-mail: sales@mornsun.cn www.mornsun-power.com