Keysight Technologies B2900A Series Precision Source/Measure Unit Data Sheet Introduction The Keysight Technologies, Inc. B2900A Series of Precision Source/Measure Units are compact and cost-effective bench-top Source/Measure Units (SMUs) with the capability to output and measure both voltage and current. An SMU combines the capabilities of a current source, a voltage source, a current meter and a voltage meter along with the capability to switch easily between these various functions into a single instrument. The Keysight B2900A series of SMUs provide best-in-class performance at a lower price than ever before. They have broad voltage (210 V) and current (3 A DC and 10.5 A pulsed) sourcing capability, excellent precision (minimum 10 fA/100 nV sourcing and measuring resolution ) and high measurement throughput. They also support an arbitrary waveform generation function. In addition, the Keysight B2900A series possess a superior graphical user interface with various viewing modes that dramatically improve test productivity, debug and characterization. The versatile integrated source and measurement capabilities of the Keysight B2900 series SMUs make them an ideal choice for testing semiconductors, active/passive components and a variety of other devices and materials. The B2900 series has a broad application range that spans from R&D and education uses to industrial development, test and manufacturing. Moreover, they work equally well as either standalone or system components. The Keysight B2900A series consists of four models, the B2901A, B2902A, B2911A and B2912A, differentiated through their available features (number of digits displayed, measurement resolution, minimum timing interval, supported viewing modes, etc.) and by the number of SMU channels (one or two) they contain. This makes it easy to select the exact price/performance point to meet your testing needs. 03 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet - Integrated voltage/current 4-quadrant precision source and measurement capabilities for easy and accurate I/V measurement - Wide coverage up to 210 V, 3 A DC/10.5 A pulse - 10 fA/100 nV minimum measurement resolution (61/2 digits) - 10 fA/100 nV minimum sourcing resolution (61/2 digits) - The 4.3" front panel color display supports both graphical and numerical view modes - High resolution arbitrary waveform generation (AWG) and list sweep functions (10 s minimum interval) - High speed digitizing capability (maximum 100000 points/s sample rate) - Free application software to facilitate PC -based instrument control - IVI-COM drivers, and SCPI supporting conventional SMU command set for basic compatibility - LXI Core Conformant, USB2.0, GPIB, LAN and digital I/O interface Cost-effective source/ measurement solutions offer superior performance and a best-in-class graphical user interface. Key Speciications of B2900A Series B2901A B2902A B2911A B2912A 1 2 1 2 210 V 210 V 210 V 210 V DC 3.03 A 3.03 A 3.03 A 3.03 A Pulsed 10.5 A 10.5 A 10.5 A 10.5 A 31.8 W 31.8 W 31.8 W 31.8 W Number of channels Max output Voltage Current Power Source Measurement Max digits Digits 51/2 51/2 61/2 61/2 Min resolution Voltage 1 V 1 V 100 nV 100 nV Current 1 pA 1 pA 10 fA 10 fA Max digits Digits 61/2 61/2 61/2 61/2 Min resolution Voltage 100 nV 100 nV 100 nV 100 nV Current 100 fA 100 fA 10 fA 10 fA Min programmable interval for List sweep/AWG waveform 20 s 20 s 10 s 10 s Min trigger interval for digitizing (Max sample rate) 20 s (50000 pts/s) 20 s (50000 pts/s) 10 s (100000 pts/s) 10 s (100000 pts/s) View modes Single view Dual view Graph view Roll view 04 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Speciications Speciication conditions The source and measurement accuracy are speciied under the conditions listed below. 1. Temperature: 23 C 5 C 2. Humidity: 30 % to 80 % RH 3. After 60 minutes warm-up 4. Ambient temperature change less than 3 C after self-calibration execution 5. Calibration period: 1 year 6. Measurement speed: 1 PLC (Power Line Cycle) Maximum voltage and current DC or pulsed1 Pulsed only1 Max voltage Max current 210 V 0.105 A 21 V 1.515 A 2 6V 3.03 A 2 200 V 1.515 A 6V 10.5 A Current [A] 10.5 Current [A] 3.03 1.515 0.1 -210 -21 -6 21 -0.1 6 -1.515 1.515 Voltage [V] Voltage [V] 210 -200 -6 6 -1.515 200 -3.03 DC or pulsed1 -10.5 Pulsed only1 1. See "Maximum pulse width and duty cycle" in Pulse Source Supplemental Characteristics for applicable maximum voltage and current. 2. Max current limitation: For 21 V/1.515 A and 6 V/3.03 A ranges, total max current is limited by the table below for using 2 channels. Max current is not limited for using 1 channel only. Maximum current limitation Ch1 voltage Ch2 voltage Max total current limitation of Ch1 and Ch2 (0 V < V 6 V) (0 V < V 6 V) Ch1 current + Ch2 current 4 A (0 V < V 6 V) (6 V < V 21 V) Ch1 current + Ch2 current x 1.6 4 A (6 V < V 21 V) (0 V < V 6 V) Ch1 current + Ch2 current x 0.625 2.5 A (6 V < V 21 V) (6 V < V 21 V) Ch1 current + Ch2 current 2.5 A 05 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Source speciications Voltage source speciications Range B2901A/B2902A Programming resolution B2911A/B2912A Accuracy (% reading + offset) Noise (peak to peak) 0.1 Hz to 10 Hz1 Max voltage (over range) 200 mV 1 V 100 nV (0.015 % + 225 V) 10 V 210 mV 2 V 10 V 1 V (0.02 % + 350 V) 20 V 2.1 V 20 V 100 V 10 V (0.015 % + 5 mV) 200 V 21 V 200 V 1 mV 100 V (0.015 % + 50 mV) 2 mV 210 V Noise (peak to peak) 0.1 Hz to 10 Hz 1 Max current (over range) 1. Supplemental characteristics Current source speciications Range Programming resolution B2901A/B2902A B2911A/B2912A Accuracy (% reading + offset) 10 nA 2 -- 10 fA (0.10 % + 50 pA) 1 pA 10.5 nA 100 nA 1 pA 100 fA (0.06 % + 100 pA) 2 pA 105 nA 1 A 10 pA 1 pA (0.025 % + 500 pA) 25 pA 1.05 A 10 A 100 pA 10 pA (0.025 % + 1.5 nA) 60 pA 10.5 A 100 A 1 nA 100 pA (0.02 % + 25 nA) 2 nA 105 A 1 mA 10 nA 1 nA (0.02 % + 200 nA) 6 nA 1.05 mA 10 mA 100 nA 10 nA (0.02 % + 2.5 A) 200 nA 10.5 mA 100 mA 1 A 100 nA (0.02 % + 20 A) 600 nA 105 mA 1 A 10 A 1 A (0.03 % + 1.5 mA) 70 A 1.05 A 1.5 A 10 A 1 A (0.05 % + 3.5 mA) 100 A 1.515 A 3 A 100 A 10 A (0.4 % + 7 mA) 120 A 3.03 A 10 A3 100 A 10 A 1. Supplemental characteristics 2. 10 nA range is not available for B2901A/B2902A. 3. 10 A range is available only for pulse mode, not available for DC mode. 4. Measurement speed: 0.01 PLC (0.4 % + 25 mA)4 10.5 A 06 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Source supplemental characteristics Temperature coeficient (0 C to 18 C and 28 C to 50 C): (0.1 x accuracy speciication)/C. Max output power and source/sink limits: 31.8 W 6 V @ 3.03 A, 21 V @ 1.515 A, 210 V @ 105 mA, four quadrant source or sink operation. Current limit/compliance: Accuracy is same as current source. Minimum value is 1 % of range, or 1 nA in 10 nA range. Voltage limit/compliance: Accuracy is same as voltage source. Minimum value is 1 % of range, or 20 mV in 200 mV range Over range : 101 % of source range for 1.5 A and 3 A ranges. 105 % of source range other than 1.5 A and 3 A ranges. No over range for 200 V range with current exceeding 105 mA pulse only condition. Over temperature protection: Output turns off then resets at over temperature sensed internally. Voltage output settling time: Time required to reach within 0.1 % of inal value at open load condition. Step is 10 % to 90 % range 200 mV, 2 V ranges: < 50 s 20 V range: < 110 s 200 V range: < 700 s Slew rate: 0.36 V/s, 20 V and 10 mA ranges, 10 M load resistance Current output settling time: Time required to reach within 0.1 % (0.3 % for 3 A range) of inal value at short condition. Step is 10 % to 90 % range 10 nA, 100 nA ranges: < 10 ms 1 A range: < 500 s 10 A, 100A ranges: < 250 s 1 mA to 3 A ranges: < 80 s Noise 10 Hz to 20 MHz (V source): 3 mVrms, 20 V range. V source overshoot: < (0.1 % + 10 mV). Step is 10 % to 90 % range, resistive load. I source overshoot: < 0.1 % (< 0.3 % for 3 A range). Step is 10 % to 90 % range, resistive load. Voltage source range change overshoot: 250 mV. 100 k load, 20 MHz bandwidth Current source range change overshoot: 250 mV/R load, 20 MHz bandwidth 07 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Pulse source supplemental characteristics Minimum programmable pulse width: 50 s Pulse width programming resolution: 1 s Pulse width deinition: The time from 10 % leading to 90 % trailing edge as follows. Pulse width Peak level 90% Base level 10% Maximum pulse width and duty cycle: Pulsed DC or pulsed Pulsed only 1. DC Max voltage Max peak current Max base current Pulse width Max duty cycle 210 V 0.105 A 0.105 A 50 s to 99999.9 s 99.9999 % 210 V 0.105 A 21 V 1.515 A1 1.515 A1 50 s to 99999.9 s 99.9999 % 21 V 1.515 A1 6V 3.03 A1 3.03 A1 50 s to 99999.9 s 99.9999 % 6V 3.03 A1 200 V 1.515 A 50 mA 50 s to 2.5 ms 2.5 % 180 V 1.05 A 50 mA 50 s to 10 ms 2.5 % 6V 10.5 A 0.5 A 50 s to 1 ms 2.5 % Max voltage Max current Max current limitation: For 21 V/1.515 A and 6 V / 3.03 A ranges, total max current is limited by the table in page 3 for using 2 channels. Max current is not limited for using 1 channel only. Minimum pulse width at the given voltage, current and settling conditions: Source value Limit value Load Source settling (% of range) Min pulse width 200 V 1.5 A 200 0.1 % 1 ms 6V 10.5 A 0.6 0.1 % 0.2 ms 1.5 A 200 V 65 0.1 % 2.5 ms 10.5 A 6V 0.5 0.1 % 0.2 ms 08 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Measurement speciications Voltage measurement speciications Range B2901A/B2902A B2911A/B2912A Accuracy (% reading + offset) 200 mV 100 nV 100 nV (0.015 % + 225 V) 2 V 1 V 1 V (0.02 % + 350 V) 20 V 10 V 10 V (0.015 % + 5 mV) 200 V 100 V 100 V (0.015 % + 50 mV) 1. Measurement resolution Supplemental characteristics Current measurement speciications Range 10 nA1 Measurement resolution B2901A/B2902A B2911A/B2912A Accuracy (% reading + offset) -- 10 fA (0.10 % + 50 pA) 100 nA 100 fA 100 fA (0.06 % + 100 pA) 1 A 1 pA 1 pA (0.025 % + 500 pA) 10 A 10 pA 10 pA (0.025 % + 1.5 nA) 100 A 100 pA 100 pA (0.02 % + 25 nA) 1 mA 1 nA 1 nA (0.02 % + 200 nA) 10 mA 10 nA 10 nA (0.02 % + 2.5 A) 100 mA 100 nA 100 nA (0.02 % + 20 A) 1 A 1 A 1 A (0.03 % + 1.5 mA) 1.5 A 1 A 1 A (0.05 % + 3.5 mA) 3 A 10 A 10 A (0.4 % + 7 mA) 10 A 10 A (0.4 % + 25 mA)3 10 A2 1. 10 nA range is not available for B2901A/B2902A. 2. 10 A range is available only for pulse mode, not available for DC mode. 3. Measurement speed: 0.01 PLC 09 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Measurement supplemental characteristics Temperature coeficient (0 C to 18 C and 28 C to 50 C): (0.1 x accuracy speciication)/C. Over range : 102 % of measurement range for 1.5 A and 3 A ranges. 106 % of measurement range other than 1.5 A and 3 A ranges. Voltage measurement range change overshoot: < 250 mV. 100 k load, 20 MHz bandwidth Current measurement range change overshoot: < 250 mV/R load, 20 MHz bandwidth Derating accuracy for measurement speed less than 1 PLC : Add % of range using the following table for measurement with PLC < 1. Derating accuracy with PLC setting< 1 PLC 0.1 PLC Voltage range Current range 0.2 V 2 V to 200 V 10 nA 100 nA 1 A to 100 mA 1 A to 3 A 0.01 % 0.01 % 0.1 % 0.01 % 0.01 % 0.01 % 0.01 PLC 0.05 % 0.02 % 1% 0.1 % 0.05 % 0.02 % 0.001 PLC 0.5 % 0.2 % 5% 1% 0.5 % 0.2 % Triggering and timing speciications Triggering Digital I/O trigger in to trigger out1: 5 s Digital I/O trigger in to source change1: 5 s LXI trigger in to source change1: Minimum 100 s, Typical 200 s, Maximum Unknown LXI trigger in to measurement1: Minimum 100 s, Typical 200 s, Maximum Unknown Internal event to external LXI trigger output1: Minimum 100 s, Typical 200 s, Maximum Unknown LAN[0-7] Event Receiving Latency1: Unknown LAN[0-7] Event Sending Latency1: Unknown Minimum trigger interval: 10 s 1. Supplemental characteristics Timer Trigger timing resolution: Timer accuracy: Arm/trigger delay: Arm/trigger interval: Arm/trigger event: Time stamp: 1 s to 100 ms 50 ppm 0 to 100000 s B2901A/B2902A: 20 s to 100000 s B2911A/B2912A: 10 s to 100000 s 1 to 100000 TIMER value automatically saved when each measurement is triggered. 10 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Other supplemental characteristics Output characteristics Sensing Modes: 2-wire or 4-wire (Remote-sensing) connections Low terminal connection: Chassis grounded or loating Output connectors: Banana jack. Triaxial connections are recommended for sourcing and measuring less than 1 nA. A banana jack to triaxial adapter is available for low current measurement. Output location: Channel 1 at front, and channel 2 at rear. Maximum load: Normal mode: 0.01 F High Capacitance mode: 50 F DC loating voltage: Max 250 V DC between low force and chassis ground. Guard offset voltage (V source): < 4 mV Remote sense operation range: Max voltage between High Force and High Sense = 3 V Max voltage between Low Force and Low Sense = 3 V Common mode isolation: > 1 G, < 4500 pF Maximum sense lead resistance: 1 k for rated accuracy Sense input impedance: > 10 G High capacitance mode The high capacitance mode permits the measurement of devices with capacitances greater than the normal mode maximum load value of 0.01 F. In high capacitance mode the maximum allowed value of the load capacitance is 50 F. Voltage output settling time: Time required to reach within 0.1 % of inal value with 4.7 F capacitive load on a ixed range at speciied current range and limit value. 200 mV, 2 V ranges: 600 s, at 1 A limit 20 V range: 1.5 ms, at 1 A limit 200 V range: 20 ms, at 100 mA limit Current measurement settling time: Time required to reach within 0.1 % of inal value after voltage source is stabilized on ixed range. Vout is 5 V unless noted. 1 A range: 230 ms 10 A, 100 A ranges: 23 ms 1 mA, 10 mA ranges: 0.23 ms 100 mA to 3 A ranges: 100 s Mode change delay: Delay into high cap mode: 1 A range: 230 ms 10 A, 100 A ranges: 23 ms 1 mA to 3 A ranges: 1 ms Delay out of high cap mode: All ranges: 10 ms Noise 10 Hz to 20 MHz (20 V range): 4.5 mVrms Voltage source range change overshoot (20 V range or below): < 250 mV, 20 MHz bandwidth High Capacitance mode working conditions: V/I mode: Voltage source mode only Range: Current measurement range is limited to ixed range only. 10 nA and 100 nA ranges are not available. Current limit: 1 A 11 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Resistance measurement Resistance measurement can be performed under either auto or manual test conditions. Auto resistance measurement is performed in current source and voltage measurement mode. The total auto resistance measurement error is shown in the table below. The total error of a manual resistance measurement can be calculated using the voltage and current accuracy information as shown below. Source I mode, manual ohm measurement (4-wire): Total error = Vmeas/Isrc = R reading x (gain error % of V range + gain error % of I range + offset error of I range/Isrc value %) + (offset error of V range/Isrc value) Source V mode, manual ohm measurement (4-wire): Total error = Vsrc/Imeas = 1/ [1/R reading x (gain error % of I range + gain error % of V range + offset error of V range/Vsrc value %) + (offset error of I range/Vsrc value)] Measurement speed = 1 PLC Applicable for temperature: 23 C 5 C Example of total error calculation: I source value = 1 mA at 1 mA range V measure range = 2 V range Total error (% reading + offset) = (0.02 % + 0.02 % + 200 nA/1 mA) + (350 V/1 mA) = 0.06 % + 0.35 Typical performance of auto resistance measurement (4-wire), 2 V range Range Resolution Test current Current range Total error (% reading + offset) 2 1 1A 1A 0.2 % + 0.00035 20 10 100 mA 100 mA 0.06 % + 0.0035 200 100 10 mA 10 mA 0.065 % + 0.035 2 k 1 m 1 mA 1 mA 0.06 % + 0.35 20 k 10 m 100 A 100 A 0.065 % + 3.5 200 k 100 m 10 A 10 A 0.06 % + 35 2 M 1 1 A 1 A 0.095 % + 350 20 M 10 100 nA 100 nA 0.18 % + 3.5 k 200 M 100 10 nA 100 nA 1.08 % + 35 k System speeds Maximum sweep operation reading rates (reading/second) for 50 Hz Measure speed Measure to memory Measure to GPIB Source measure to memory Source measure to GPIB < 0.001 PLC 20000 12500 19500 12500 0.01 PLC 4500 3950 4500 3950 0.1 PLC 500 490 500 490 1 PLC 49 49 49 49 Operation reading rate varies by number of sweep steps. Number of sweep steps is speciied 12 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet General Speciications Environment For use in indoor facilities Temperature range Operating: 0 C to +55 C Storage: -30 C to +70 C Humidity range Operating: 30 % to 80 % RH, non-condensing Storage: 10 % to 90 % RH, non-condensing Altitude Operating: 0 m to 2000 m Storage: 0 m to 4600 m Power supply AC voltage: 90 V to 264 V Line frequency: 47 Hz to 63 Hz Maximum volt-amps: 250 VA Cooling Forced air. Side intake and rear exhaust. EMC IEC61326-1/EN61326-1 AS/NZS CISPR 11 KC: RRA Notiication amending Radio Waves Act Article 58-2 Safety IEC61010-1/EN61010-1 CAN/CSA-C22.2 No. 61010-1-04, C/US Certiications CE, cCSAus, C-Tick, KC Dimensions Without handle & feet: 88 mm H x 213 mm W x 450 mm D Bench coniguration (with handle & feet): 180 mm H x 260 mm W x 480 mm D Weight B2901A/B2911A: 5.0 kg B2902A/B2912A: 6.4 kg Operation and Functions Front panel interface 4.3" Color TFT WQVGA (480x272, with LED backlight) provides a graphical user interface (GUI) with the following features. - 4-view modes (Single, Dual, Graph, Roll view) - Easy-to-use key operations - Easy access to basic and advanced source/measurement capabilities Single view Single view provides basic and advanced settings and display capabilities for the selected channel from the front panel of the instrument. No additional controller or software is required. Dual view Dual view provides the basic settings and display capabilities for both channels 1 and 2. Up to 61/2 digits can be displayed in dual view mode. This mode is available only for the B2902A/B2912A. 13 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Operation and Functions, continued Graph view Graph View displays measurement results on XY graphs (such as I-V and I-t/V-t curves) on up to 2 channels. This is useful for quick evaluation of device characteristics, especially those obtained from sweep measurements. Roll view Roll view draws I-t or V-t curves similar to the curves drawn by a strip chart recorder. Up to 1000 acquired data points can be displayed and updated while the measurement is still in progress. Roll view's continuous measurement capability is especially useful for monitoring low frequency phenomena. Roll view is available only for the B2911A/B2912A. Easy-to-use key operations - 10-key, assist key and rotary knob for easy operation. - 10-key enables direct input of alphanumeric values. - Assist key guides operation on the front panel. Easy access to source/measurement capabilities Settings: - V/I source mode - Source value - Limit (compliance) value - Sweep parameters - Pulse parameters - Measure speed - Range setting - Trigger setting Result displays: - Numeric results fully up to 61/2 digits both in Single and Dual view modes - Characterization in Graph View and Roll View - Data list Advanced capabilities: - Limit testing by user deinable limits working with component handlers - Trace memory for accumulating measurement (or math) result and collecting their statistics File operations (USB memory): - Save - System coniguration - Measurement/math result - Limit test result - Trace data - Graph screen dump - Load - System coniguration - List sweep data 14 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Source/measurement capabilities Sweep measurement Number of steps: 1 to 2500 Sweep mode: linear, logarithmic (log) or list Sweep direction: single or double Type: DC, or pulse Min programmable value to create list sweep waveform: B2901A/B2902A: Min 20 s with 1 s resolution B2911A/B2912A: Min 10 s with 1 s resolution Digitizing/sampling measurement Digitizing/sampling measurement can be performed by appropriate setting of measure speed and trigger interval. Min trigger interval (Max speed of measurement): B2901A/B2902A: 20 s (50000 points/s) B2911A/B2912A: 10 s (100000 points/s) Data buffers Data buffers are array of measurement elements. Each elements hold the following values. - Measured voltage - Measured current - Measured resistance - Time stamp - Measurement status - Source setting Max buffer size: 100000 points / channel Program and interface capabilities Programming language The Keysight B2900A series supports SCPI (Standard Commands for Programmable Instruments). Default command set: Supports all of the advanced features of Keysight B2900A series. Conventional command set: Supports industry standard conventional SCPI command set for basic compatibility Program memory Program memory allows you to store long strings of SCPI command lines once into the B2900A series' volatile memory and then recall those strings multiple times while the program is executing using a single SCPI command. By storing the command strings in memory, the time that would have been spent sending those same commands over a communication bus is eliminated. For tests that utilize lots of repeated code (such as subroutines), program memory can dramatically reduce test times. Max number of characters of program name: 32, with alphabets, numbers, hyphens and underscores. Max memory size: 100 kB (2500 lines typical) 15 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Program and interface capabilities, continued LXI LXI Core Conformant. The Keysight B2901A/B2902A/B2911A/B2912A follows speciied LAN protocols and adhere to LXI requirements such as a built-in Web control server and IVI-COM driver. Ethernet: USB2.0: GPIB: 10/100Base-T USB-TMC488 protocol (Rear x 1) IEEE-488.2 compliant U USB ile system USB 2.0 high-speed mass storage (MSC) class device (Front x 1) Digital I/O interface Connector: Input/output pins: Absolute max input voltage: Absolute min input voltage: Max logic low input voltage: Min logic high input voltage: Max source current: Max sink current: 5 V power supply pin: Safety interlock pin: 25-pin female D. 14 open drain I/O bits 5.25 V -0.25 V 0.8 V 2.0 V 1 mA, Vout = 0 V 50 mA, Vout = 5 V Limited to 500 mA, solid state fuse protected. One active high pin and one active low pin. Activation of both pin enables output voltage > 42 V. Maximum number of simultaneously triggered units (using Digital I/O)1: 8 1. Supplemental characteristic 16 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Furnished Accessories Power cable USB cable Quick Reference (English) Product Reference CD-ROM (including PDF manuals, Keysight B2900A Quick I/V Measurement Software and drivers) Keysight I/O Library Suite Software and drivers Keysight B2900A Quick I/V Measurement Software The Keysight B2900A series includes PC-based Keysight B2900A Quick I/V Measurement Software. This powerful software makes it easy to quickly setup and perform I/V measurements and to display the measurement data in a table or graph without the need to perform any programming. The software allows you to control up to four SMU channels over a GPIB or LAN connection, or one B2900A series unit via a USB connection. Operating environment: Operating system: Windows 7 (64bit/32bit), XP SP3 (32-bit). Other requirements: Microsoft .NET framework 4.0 or greater, and IO Libraries 16.0 or greater. Keysight B2900A Graphical Web Interface The Keysight B2900A Graphical Web Interface is a web browser based instrument control panel. It enables you to set up and perform a measurement easily and quickly from a web browser using the Keysight B2900A's built-in web server. This allows you to control one Keysight B2900A series unit over a LAN connection. IVI-C or IVI-COM drivers Compatible with Windows 7 (64bit/32bit), XP SP2 (32-bit) IO Libraries 16.0 or greater. Supports Keysight VEE, Microsoft Visual Studio (Visual Basic, Visual C++, Visual C#), National Instruments LabWindows and LabVIEW. LabVIEW driver (VI) National Instruments LabVIEW 7.0 or greater. LabView drivers are available at NI.COM. 17 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet Ordering Information Model number B2901A Precision Source/Measure Unit, 1ch, 100 fA, 210 V, 3 A DC/10.5 A pulse B2902A Precision Source/Measure Unit, 2ch, 100 fA, 210 V, 3 A DC/10.5 A pulse B2911A Precision Source/Measure Unit, 1ch, 10 fA, 210 V, 3 A DC/10.5 A pulse B2912A Precision Source/Measure Unit, 2ch, 10 fA, 210 V, 3 A DC/10.5 A pulse If you need more source capabilities or lower noise sourcing, the best choice is Keysight Power Source products. Options Printed manual (User's Guide) ABA English ABJ Japanese AB0 Traditional Chinese AB2 Simpliied Chinese Keysight B2961A/B2962A 6.5 Digit Low Noise Power Source www.keysight.com/find/precisionSOURCE Calibration A6J ANSI Z540 compliant calibration UK6 Commercial calibration certiicate with test data Rack mount kit 1CM Rack mount kit Accessories Banana to triaxial adapter N1294A-001 Banana - Triax adapter for 2-wire (non Kelvin) connection N1294A-002 Banana - Triax adapter for 4-wire (Kelvin) connection Interlock cable for 16442B N1294A-011 Interlock cable for 16442B (1.5 m) N1294A-012 Interlock cable for 16442B (3.0 m) Keysight B1500A Semiconductor Device Analyzer www.keysight.com/find/B1500A BNC trigger adapter N1294A-031 GPIO-BNC trigger adapter Test ixture N1295A Device/component test ixture For details of differences, refer to the comparison table in the page 2. Keysight B1505A Power Device Analyzer/ Curve Tracer (1500 A/10 kV) www.keysight.com/find/B1505A 18 | Keysight | B2900A Series Precision Source/Measure Unit - Data Sheet myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Keysight is a founding member of the LXI consortium. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight's commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight's measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/b2900a For more information on Keysight Technologies' products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada Brazil Mexico United States (877) 894 4414 55 11 3351 7010 001 800 254 2440 (800) 829 4444 Asia Paciic Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Other AP Countries 1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 (65) 6375 8100 Europe & Middle East Austria Belgium Finland France Germany Ireland Israel Italy Luxembourg Netherlands Russia Spain Sweden Switzerland United Kingdom 0800 001122 0800 58580 0800 523252 0805 980333 0800 6270999 1800 832700 1 809 343051 800 599100 +32 800 58580 0800 0233200 8800 5009286 800 000154 0200 882255 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-09-23-14) This information is subject to change without notice. (c) Keysight Technologies, 2013 - 2014 Published in USA, August 3, 2014 5990-7009EN www.keysight.com