Parameter Value Units Notes
General
Pressure Range 100 kPa ≈15 psi
200 kPa ≈30 psi
700 kPa ≈100 psi
Maximum Pressure 3x rated pressure
Electrical @ 77°F (25°C) unless otherwise stated
Excitation 3.0 V 10 VDC maximum
Input Impedance 5,000 ±20% Ω
Output Impedance 5,000 ±20% Ω
Environmental
Electrostatic Damage (ESD) Class 1
Operating Temperature Range –40°F to 257°F (–40°C to 125°C)
Mechanical (1)
Weight ≈0.0002 lb (0.10 g)
Media Compatibility Clean, dry air and non-corrosive gases
Parameter Units Minimum Type Maximum Notes
Performance Parameters (Note 2)
Offset mV/V ±10
Full Scale Output mV 60 ±20
Linearity %FSO ±0.20 3
Hysteresis and
Repeatability %FSO 0.1
Thermal Coefficient
of Zero %FSO/°C 0.04 4
Thermal Coefficient
of Resistance %/°C 0.3 4
Thermal Coefficient
of Sensitivity %FSO/°C –0.2 4
Thermal Hysteresis
of Zero %FSO 0.1 5
Long-Term
Stability of FSO %FSO 0.2 6
1. Standard IC industry bake operations should be used prior to surface mount
operations. Consult GE for further information.
2. Values measured at 3 VDC and 77°F (25°C), unless otherwise noted.
3. Best fit straight line.
4. Typical coefficients, between 32°F to 158°F (0° to 70°C).
5. 32°F to 158°F (0° to 70°C).
6. Typical value over one year.
NPP-301 Series
Specifications
GE
Sensing
Ordering Information
The code number to be ordered may be specified as follows:
NPP-
Code Description Shipping
301A-100A 15 psia (1.03 bar), non-ported IC tubes
301A-200A 30 psia (2.06 bar), non-ported IC tubes
301A-700A 100 psia (6.89 bar), non-ported IC tubes
301A-100AT 15 psia (1.03 bar), non-ported Tape and reel
301A-200AT 30 psia (2.06 bar), non-ported Tape and reel
301A-700AT 100 psia (6.89 bar), non-ported Tape and reel
301B-100A 15 psia (1.03 bar), ported IC tubes
301B-200A 30 psia (2.06 bar), ported IC tubes
301B-700A 100 psia (6.89 bar), ported IC tubes
301B-100AT 15 psia (1.03 bar), ported Tape and reel
301B-200AT 30 psia (2.06 bar), ported Tape and reel
301B-700AT 100 psia (6.89 bar), ported Tape and reel
NPP - ________ Typical model number