V18
1
TR60M SERIES
60W MEDICAL SWITCHING
ADAPTER
Features
* Universal Input Range 90~264VAC
* Meets EN60601-1 and EN55011 Class B
* Continuous Short Circuit Protection
* Over Voltage Protection
* Meet CEC Level
(Output Cable Length1800mm)
(TR60M Series meets CEC IV except
TR60M05 is Non-CEC Compliant)
(TR60M12
: Output Cable Length 1220mm 16AWG
)
* Efficiency & Standby Power Meet Level V (Option)
(Output Cable Length1800mm)
(TR60M12Output Cable Length720mm 16AWG)
(TR60M15Output Cable Length1220mm 16AWG)
(TR60M18,TR60M19Output Cable Length1500mm 18AWG)
* Meets 2 MOPP
Ordering
information
TR60MXX-
XX X XX
Model No. DC Plug Type OVP DC Cable Length and Type
E: With OVP 01: 720mm
02: 1220mm
03: 1800mm
11: 720mm with Ferrite Core
12:1220mm with Ferrite Core
13:1800mm with Ferrite Core
*18AWG/UL1185
MODEL OUTPUT
VOLTAGE
OUTPUT
CURRENT
RIPPLE &
NOISE
NOTE 2
VOLTAGE
ACCURACY
NOTE 1
LINE
REGULATION
NOTE 3
LOAD
REGULATION
NOTE 4
% EFF. (Typ.)
NOTE 5
TR60M05 5 V 6 A 50 mV ±4% ±1% ±6% 75%
TR60M12 12 V 5 A 120 mV ±2% ±1% ±5% 85%
TR60M15 15 V 4 A 150 mV ±2% ±1% ±3% 85%
TR60M18 18 V 3.33 A 180 mV ±2% ±1% ±2% 86%
TR60M19 19 V 3.15 A 190 mV ±2% ±1% ±2% 86%
TR60M24 24 V 2.5 A 240 mV ±2% ±1% ±2% 87%
TR60M36 36 V 1.66 A 360 mV ±2% ±1% ±2% 87%
TR60M48 48 V 1.25 A 480 mV ±2% ±1% ±2% 87%
Specifications are subject to change without notice.
V18
2
Specifications
TR60M Series Derating Curve
INPUT SPECIFICATIONS:
Voltage …....………....…..............…………………..……………… 90~264Vac
Frequency ………………………………………........……………..... 47 to 63Hz
Inrush Current ………………..……. Cold Start @25 80A max. @240Vac
Conducted EMI ……………………..…………….…… CISPR/FCC Class B
Leakage Current ……………………………………..……………… 0.1mA max.
OUTPUT SPECIFICATIONS:
Hold-up Time …………………………………………......… 8ms typ. @115Vac
Short Circuit Protection …………….…………………….….……Continuous
Over Voltage Protection ………………..................................................… Yes
Temperature Coefficient …………………….............................….…±0.05%/
GENERAL SPECIFICATIONS:
NOTE:
Isolation …………………………………………… Input to output =5,656VDC
1. Voltage accuracy at 60% full load.
Operating Temperature ................................... 0 ~ 60(see derating curve)
2. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for
Storage Temperature …………………………...….......…….......... -20 ~ 85
Ripple & Noise measurement @20MHz BW.
Humidity …………………..….......………........ 93% RH max. Non condensing
3. Line regulation is measured from 100Vac to 240Vac full load.
Cooling ………………………………….…………………Natural Convection
4. Load regulation is measured from 60% to 100% full load and from 60%
Switching Frequency ………………………………………… 100KHz Typical
to 20% full load (60% +/- 40% full load).
MTBF ……… MIL-HDBK-217F, GB, at 25/115VAC ………. 200Khrs min.
5. Typical efficiency at 230VAC and full load at 25.
Altitude ……………………………………………………………………… 3000m
6. ”Various TR Series adapters are PSE certified. PSE certification
Dimensions …….…. . 5.197x2.283x1.201 inches (132.00x58.00x30.50mm)
alone is not sufficient for importation into Japan. A valid PSE mark
Weight …………………... ………………………………… 345g(0.76 Pounds)
must contain the name of the importer as shown in the example below.
SAFETY AND EMC:
If PSE mark is required, the name of the registered importer must be
Emission and lmmunity (Ed. 4.0) .……. N55011, EN60601-1-2, EN61000-3-2
supplied to Cincon on order placement. Product labels will not contain
EN61000-3-3
PSE mark if importer name is not supplied. Consult factory or local
Safety (Ed. 3.1) … Class II, IEC60601-1:2005+A1:2012, EN60601-1:2006/A1:2013
representative for details”.
UL ANSI/AAMI ES60601-1:2005
Mechanical Specification
IEC320/C8
70.866[1800.00]±1.969[50.00]
LED
2.283[58.00]
5.197[132.00]
1.201[30.50]
DC Plug :Right Angle(ψ5.5/ψ2.1) L12mm
18AWG/1800mm
DC Plug type: V+ ─●)─ V-
Millimeters:X.XX±0.5
Tolerance:Inches:X.XXX±0.02
All Dimensions are in inches(mm)
Typical at 25˚C, nominal line and 75% load, unless otherwise Specified
0%
25%
50%
75%
100%
0 10 20 30 40 50 60
Ambient Temperature(
o
C)
Rated Load(%)
20LFM(Natural Convection)