M5100
Heavy Industrial Pressure Transducer
M5100 www.meas-spec.com June 2012
1/7
CE Compliance
Wide Temperature Range
Compact
Variety of Pressure Ports and Electrical
Configurations
DESCRIPTION
The M5100 series pressure transducers from the Microfused™ line of MEAS, set a new price
performance standard for demanding commercial and heavy industrial applications. This series is suitable for
measurement of liquid or gas pressure, even for difficult media such as contaminated water, steam, and mildly
corrosive fluids.
The transducer pressure cavity is machined from a solid piece of 17-4 PH stainless steel. The standard
version includes a 1/4 NPT pipe thread allowing a leak-proof, all metal sealed system. There are no O-rings,
welds or organics exposed to the pressure media. The durability is excellent. The M5100 exceeds the latest
heavy industrial CE requirements including surge protection, and is over voltage protected to 16Vdc in both
positive and reverse polarity.
This product is geared to the OEM customer who uses medium to high volumes. The standard version
is suitable for many applications, but the dedicated design team at our Transducer Engineering Center stands
ready to provide a semi-custom design where the volume and application warrants.
FEATURES APPLICATIONS
Heavy Industrial CE Approval Advanced HVAC Systems
100 V/m EMI Protection Refrigeration Systems
Reverse Polarity Protection Automotive Test Stands
Extended Temperature Range Industrial Process Control
1% Total Error Band Pumps and Compressors
Compact Outline Hydraulic/Pneumatic Systems
-40°C to +125°C Operating Temperature Range
Agriculture Equipment
Energy Generation and Management
M5100
Heavy Industrial Pressure Transducer
M5100 www.meas-spec.com June 2012
2/7
STANDARD RANGES
Range psig Range Barg
0 to 50 0 to 3.5
0 to 100 0 to 7
0 to 200 0 to 14
0 to 300 0 to 20
0 to 500 0 to 35
0 to 1k 0 to 70
0 to 3k 0 to 200
0 to 5k 0 to 350
0 to 10k 0 to 700
DIN Range Barg
0 to 4
0 to 6
0 to 10
0 to 16
0 to 25
0 to 40
0 to 60
0 to 100
0 to 160
0 to 250
0 to 400
0 to 600
0 to 1000
*Intermediate ranges available upon request.
DIM A 1.71 [ 43.5 ] 1.92 [ 48.6 ]
DIM B 2.17 [ 55.2 ] 2.38 [ 60.5 ]
DIM A 1.71 [ 43.4 ] 1.92 [ 48.6 ]
DIM B 2.53 [ 64.3 ] 2.72 [ 69.1 ]
DIM A 1.74 [ 44.2 ] 1.93 [ 49.0 ]
DIM B 2.17 [ 55.0 ] 2.36 [ 59.8 ]
DIM A 1.67 [ 42.3 ] 1.86 [ 47.1 ]
DIM B 2.12 [ 53.8 ] 2.31 [ 58.6 ]
DIM A 2.12 [ 53.8 ] 2.12 [ 53.8 ]
DIM B 2.64 [ 67.1 ] 2.64 [ 67.1 ]
DIM A 1.71 [ 43.5 ] 1.92 [ 48.8 ]
DIM B 2.53 [ 64.3 ] 2.72 [ 69.1 ]
DIM A 1.69 [ 42.9 ] 1.90 [ 48.3 ]
DIM B 2.21 [ 56.1 ] 2.41 [ 61.2 ]
DIM A 1.71 [ 43.5 ] 1.92 [ 48.6 ]
DIM B 2.17 [ 55.2 ] 2.38 [ 60.5 ]
DIM A 1.71 [ 43.5 ] 1.92 [ 48.6 ]
DIM B 2.17 [ 55.2 ] 2.38 [ 60.5 ]
DIM A 1.71 [ 43.5 ] 1.92 [ 48.6 ]
DIM B 2.17 [ 55.2 ] 2.38 [ 60.5 ]
CONNECTION/ PRESSURE RANGE/ PRESSURE REF
PRESSURE RANGE
PG/BG PRESS REF
CODE CONNECTION
<500 PSI >500 PSI
<040BAR > 040BAR
1CABLE 2 FEET
ALPHA#3243,18AWG
4PACKARD CONNECTOR
SUPPLY ON PIN A
5BENDIX PTIH-10-6P
CONNECTOR
6
HIRSCHMANN
CONNECTOR
DIN 43650 FORM C
7
HIRSCHMANN
CONNECTOR
DIN 43650 FORM A
9PACKARD CONNECTOR
SUPPLY ON PIN B
D M12 CONNECTOR
RCABLE 10 METER
MANHATTAN#M39114,22AWG
MCABLE 1 METER
MANHATTAN#M39114,22AWG
PCABLE 5 METER
MANHATTAN#M39114,22AWG
M5100
Heavy Industrial Pressure Transducer
M5100 www.meas-spec.com June 2012
3/7
PERFORMANCE SPECIFICATIONS
Ambient Temperature: 25°C (unless otherwise specified)
PARAMETERS MIN TYP MAX UNITS NOTES
Accuracy (combined non linearity, hysteresis, and repeatability) -0.25 0.25 %F.S. 1
Isolation, Body to any Lead 1 MΩ @ 25Vdc
Pressure Cycles (Zero to Full Scale) 1.00E+7 FS Cycles
Proof Pressure 2X Rated
Burst Pressure 5X Rated
Long Term Stability (1 year) -0.25 0.25 %Span
Total Error Band -1.0 1.0 %F.S. 2
Compensated Temperature -20 +85 °C
Operating Temperature -40 +125 °C 3
Storage Temperature -40 +125 °C 3
Load Resistance (R
L
)
R
L
> 100k 4
Bandwidth DC to 1KHz (Typical)
Shock 50g, 11msec Half Sine Shock per MIL-STD-202G, Method 213B,
Condition A
Vibration ±20g, MIL-STD-810C, Procedure 514.2, Fig 514.2-2, Curve L
For custom configurations, consult factory.
Notes
1. Best fit straight line at 25°C.
2. TEB includes all accuracy errors, thermal errors, span and zero tolerances.
3. Cable 105°C max.
4. Voltage output.
CE Compliance
IEC 55022 Emissions Class A & B
IEC 61000-4-2 Electrostatic Discharge Immunity (6kV contact/8kV air)
IEC 61000-4-3 Electromagnetic Field Immunity (30V/m)
IEC 61000-4-4 Electrical Fast Transient Immunity (1kV)
IEC 61000-4-5 Surge Immunity (1kV)
IEC 61000-4-6 Conducted RF Immunity (10V)
IEC 61000-4-9 Magnetic Field Immunity (100A/m)
M5100
Heavy Industrial Pressure Transducer
M5100 www.meas-spec.com June 2012
4/7
DIMENSIONS
M5100
Heavy Industrial Pressure Transducer
M5100 www.meas-spec.com June 2012
5/7
4
2
1
3
M5100
Heavy Industrial Pressure Transducer
M5100 www.meas-spec.com June 2012
6/7
M5100
Heavy Industrial Pressure Transducer
M5100 www.meas-spec.com June 2012
7/7
OUTPUT OPTIONS
Code Output Supply (V)
3 0.5 – 4.5V (ratiometric) 5 ± 0.25 Ratiometric
4 1 – 5V 8 – 30
5 4 – 20mA 9 – 30
6 0 – 5V 8 – 30
7 0 – 10V 15 – 30
8 1 – 6V 8 – 30
ORDERING INFORMATION
M5131-000005-300PG
Type (G = Gage)
Pressure Port (See Pressure Port Options Table)
Specials (nnnnn = Custom Drawing)
Output (See Output Options Table)
Connection (See Electrical Connections Table)
Pressure Range (See Pressure Range Table)
Model
Units (P = psi, B = Bar)
NORTH AMERICA EUROPE ASIA
Measurement Specialties
45738 Northport Loop West
Fremont, CA 94538
Tel: 1-800-767-1888
Fax: 1-510-498-1578
Sales: pfg.cs.amer@meas-spec.com
Measurement Specialties
(Europe), Ltd.
26 Rue des Dames
78340 Les Clayes-sous-Bois, France
Tel: +33 (0) 130 79 33 00
Fax: +33 (0) 134 81 03 59
Sales: pfg.cs.emea@meas-spec.com
Measurement Specialties
(China), Ltd.
No. 26 Langshan Road
Shenzhen High-Tech Park (North)
Nanshan District, Shenzhen 518057
China
Tel: +86 755 3330 5088
Fax: +86 755 3330 5099
Sales: pfg.cs.asia@meas-spec.com
The information in this sheet has been carefully reviewed and is believed to be accurate; however, no responsibility is assumed for
inaccuracies. Furthermore, this information does not convey to the purchaser of such devices any license under the patent rights to the
manufacturer. Measurement Specialties, Inc. reserves the right to make changes without further notice to any product herein. Measurement
Specialties, Inc. makes no warranty, representation or guarantee regarding the suitability of its product for any particular purpose, nor does
Measurement Specialties, Inc. assume any liability arising out of the application or use of any product or circuit and specifically disclaims
any and all liability, including without limitation consequential or incidental damages. Typical parameters can and do vary in different
applications. All operating parameters must be validated for each customer application by customer’s technical experts. Measurement
Specialties, Inc. does not convey any license under its patent rights nor the rights of others.