Fluke 700-IV
Current Shunt
Product Home
Description:
Conversion factor: 10 mV = 1 mA
Accuracy (% of input, 1 year): 0.025%
●
Input current: 0 to 55 mA● Input resistance: 250Ω, nominal● Output resistance: 10Ω, nominal
Accuracy specification applies from +18ºC
and 28-50ºC
●
Maximum input voltage: 30V dc●
Application: Use the Fluke 700-IV Current Shunt to calibrate and
troubleshoot mA/mA devices. On a 741 or 743 Documenting Process
Calibrator, Source mA and Measure V dc. The custom units function of the
calibrator automatically provides the correct conversion, and directly
displays/captures the current in mA.
Included: Shunt, instruction sheet.
Warranty: One year
© 1995 - 2000 Fluke Corporation